Effects of Skin Pigmentation on Pulse Oximeter Accuracy at Low Saturation

Abstract

Background: It is uncertain whether skin pigmentation affects pulse oximeter accuracy at low HbO2 saturation. Methods: The accuracy of finger pulse oximeters during stable, plateau levels of arterial oxygen saturation (Sao2) between 60 and 100% were evaluated in 11 subjects with darkly pigmented skin and in 10 with light skin pigmentation. Oximeters tested were the Nellcor N-595 with the OxiMax-A probe (Nellcor Inc., Pleasanton, CA), the Novametrix 513 (Novametrix Inc., Wallingford, CT), and the Nonin Onyx (Nonin Inc., Plymouth, MN). Semisupine subjects breathed air-nitrogen-carbon dioxide mixtures through a mouthpiece. A computer used end-tidal oxygen and carbon dioxide concentrations determined by mass spectrometry to estimate breath-by-breath Sao2, from which an operator adjusted inspired gas to rapidly achieve 2- to 3-min stable plateaus of desaturation. Comparisons of oxygen saturation measured by pulse oximetry (Spo2) with Sao2 (by Radiometer OSM3) were used in a multivariate model to determine the interrelation between saturation, skin pigmentation, and oximeter bias (Spo2 - Sao2). Results: At 60-70% Sao2, Spo2 (mean of three oximeters) overestimated Sao2 (bias +/- SD) by 3.56 +/- 2.45% (n = 29) in darkly pigmented subjects, compared with 0.37 +/- 3.20% (n = 58) in lightly pigmented subjects (P < 0.0001). The SD of bias was not greater with dark than light skin. The dark-light skin differences at 60-70% Sao2 were 2.35% (Nonin), 3.38% (Novametrix), and 4.30% (Nellcor). Skin pigment-related differences were significant with Nonin below 70% Sao2, with Novametrix below 90%, and with Nellcor at all ranges. Pigment-related bias increased approximately in proportion to desaturation. Conclusions: The three tested pulse oximeters overestimated arterial oxygen saturation during hypoxia in dark-skinned individuals.

Overview

Integrity signals

Registry

Crossref · 10.1097/00000542-200504000-00004

Cites

16 of 16 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Verification

Claims the paper makes

This paper’s full text isn’t openly available, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

3 more claims about this paper

Create a free account to see every claim other papers make about this one.

See all claims — free account

Citations

Cites

Show all 16 references

Cited by

See all 434

Metadata

Authors
Bickler, P. E., Feiner, J. R., & Severinghaus, J. W.
Affiliations
  • Professor.
  • Associate Professor.
  • Professor Emeritus.
Venue
Anesthesiology
Published
2005-04-01
Volume
102
Issue
4
Pages
715-719
ISSN
0003-3022 (ISSN-L)
OpenAlex
W2142600751
Cited by
427
References
16
Citations by year
2026: 35 · 2025: 46 · 2024: 49 · 2023: 54 · 2022: 86 · 2021: 36 · 2020: 20 · 2019: 12 · 2018: 7 · 2017: 5 · 2016: 10 · 2015: 6 · 2014: 4 · 2013: 10 · 2012: 10
FWCI
4.93
Type
journal article
Language
en
Open access
closed
Locations
  • Anesthesiology (publishedVersion)
  • PubMed (publishedVersion)
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
  • Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
Keywords
pulse oximetry, skin pigmentation, dark skin, arterial oxygen saturation, low oxygen saturation, hypoxia, pulse oximeter accuracy, desaturation, end-tidal oxygen
MeSH terms
Adult, Black or African American, Female, Humans, Male, Middle Aged, Oximetry, Oxygen, Oxyhemoglobins, Skin Pigmentation, Reproducibility of Results

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Effects of Skin Pigmentation on Pulse Oximeter Accuracy at Low Saturation. https://app.cochat.ai/papers/10.1097/00000542-200504000-00004
The paper
Bickler, P. E., Feiner, J. R., & Severinghaus, J. W. (2005). Effects of Skin Pigmentation on Pulse Oximeter Accuracy at Low Saturation. Anesthesiology. https://doi.org/10.1097/00000542-200504000-00004

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.