Effects of Skin Pigmentation on Pulse Oximeter Accuracy at Low Saturation
Abstract
Background: It is uncertain whether skin pigmentation affects pulse oximeter accuracy at low HbO2 saturation. Methods: The accuracy of finger pulse oximeters during stable, plateau levels of arterial oxygen saturation (Sao2) between 60 and 100% were evaluated in 11 subjects with darkly pigmented skin and in 10 with light skin pigmentation. Oximeters tested were the Nellcor N-595 with the OxiMax-A probe (Nellcor Inc., Pleasanton, CA), the Novametrix 513 (Novametrix Inc., Wallingford, CT), and the Nonin Onyx (Nonin Inc., Plymouth, MN). Semisupine subjects breathed air-nitrogen-carbon dioxide mixtures through a mouthpiece. A computer used end-tidal oxygen and carbon dioxide concentrations determined by mass spectrometry to estimate breath-by-breath Sao2, from which an operator adjusted inspired gas to rapidly achieve 2- to 3-min stable plateaus of desaturation. Comparisons of oxygen saturation measured by pulse oximetry (Spo2) with Sao2 (by Radiometer OSM3) were used in a multivariate model to determine the interrelation between saturation, skin pigmentation, and oximeter bias (Spo2 - Sao2). Results: At 60-70% Sao2, Spo2 (mean of three oximeters) overestimated Sao2 (bias +/- SD) by 3.56 +/- 2.45% (n = 29) in darkly pigmented subjects, compared with 0.37 +/- 3.20% (n = 58) in lightly pigmented subjects (P < 0.0001). The SD of bias was not greater with dark than light skin. The dark-light skin differences at 60-70% Sao2 were 2.35% (Nonin), 3.38% (Novametrix), and 4.30% (Nellcor). Skin pigment-related differences were significant with Nonin below 70% Sao2, with Novametrix below 90%, and with Nellcor at all ranges. Pigment-related bias increased approximately in proportion to desaturation. Conclusions: The three tested pulse oximeters overestimated arterial oxygen saturation during hypoxia in dark-skinned individuals.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Related work on pulse oximetry, which relies on the same optical principles, has also documented skin-pigmentation-linked measurement errors [14,15]”
From PPG-Based Heart Rate Accuracy in Hispanic Adults with Fitzpatrick III–V Skin Tones: An Evaluation of Body Composition and Skin-Tone Effects (2026) · cites this paper as [14]
In the citing paper
“Notably, darker skin pigmentation may influence PPG-based SpO2 and PR readings [7-10].”
From Impact of skin pigmentation on pulse oximetry blood oxygenation and wearable pulse rate accuracy: systematic review and meta-analysis (2024) · cites this paper as [10]
In the citing paper
“Measurements of the pulse oximeter are known to be affected by the patient’s skin color, as the device systematically overestimates oxygen saturation levels in nonwhite patients.”
From Addressing bias in big data and AI for health care: A call for open science (2021) · cites this paper as [2]
In the citing paper
“Accuracy errors in pulse oximeter readings have been associated with SpO2 overestimations, especially at low saturation conditions [23,24,73,93].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [23]
In the citing paper
“The effect of skin pigmentation in pulse oximeter readings under hypoxia conditions was tested in 11 dark-skinned male subjects and 10 light-skinned males, who were made to breathe an air mixture (nitrogen-carbon monoxide). The three pulse oximeters used showed overestimation in dark-skinned individuals that increases linearly as SpO2 levels decreases [23].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [23]
3 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Recent Developments in Pulse Oximetry
Biebuyck, J. F., Severinghaus, J. W., & Kelleher, J. F. (1992) · Anesthesiology · doi:10.1097/00000542-199206000-00024 · Cited by 331
Self-citationJubran, A., & Tobin, M. J. (1990) · Chest · doi:10.1378/chest.97.6.1420 · Cited by 296
Errors in 14 pulse oximeters during profound hypoxia
Severinghaus, J. W., Naifeh, K. H., & Koh, S. O. (1989) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf01617877 · Cited by 209
Self-citationRalston, A., Webb, R. K., & Runciman, W. B. (1991) · Anaesthesia · doi:10.1111/j.1365-2044.1991.tb11501.x · Cited by 195
Effects of Intravenously Administered Dyes on Pulse Oximetry Readings
Scheller, M. S., UNGER, R. J., & Kelner, M. J. (1986) · Anesthesiology · doi:10.1097/00000542-198611000-00023 · Cited by 191
The Effect of Nail Polish on Pulse Oximetry
COT, C. J., Goldstein, E. A., Fuchsman, W. H., & Hoaglin, D. C. (1988) · Anesthesia & Analgesia · doi:10.1213/00000539-198807000-00013 · Cited by 171
Ries, A. L., Prewitt, L. M., & Johnson, J. J. (1989) · Chest · doi:10.1378/chest.96.2.287 · Cited by 145
Reliability of noninvasive oximetry in black subjects during exercise and hypoxia
Zeballos, R. J., & Weisman, I. M. (1991) · American Review of Respiratory Disease · doi:10.1164/ajrccm/144.6.1240 · Cited by 108
Effect of skin pigmentation on pulse oximetry accuracy in the emergency department
Adler, J. N., Hughes, L. A., Vtvilecchia, R., & Jr., C. A. C. (1998) · Academic Emergency Medicine · doi:10.1111/j.1553-2712.1998.tb02772.x · Cited by 102
Accuracy of pulse oximetry in pigmented patients.
Bothma, P. A., Joynt, G. M., Lipman, J., Hon, H., Mathala, B., Scribante, J., & Kromberg, J. G. R. (1996) · PubMed · Cited by 50
Cited by
Racial bias in pulse oximetry measurement
Sjoding, M. W., Dickson, R. P., Iwashyna, T. J., Gay, S. E., & Valley, T. S. (2020) · New England Journal of Medicine · doi:10.1056/nejmc2029240 · Cited by 1,114
Addressing bias in big data and AI for health care: A call for open science
Norori, N., Hu, Q., Aellen, F. M., Faraci, F. D., & Tzovara, A. (2021) · Patterns · doi:10.1016/j.patter.2021.100347 · Cited by 903
BTS guideline for emergency oxygen use in adult patients
O’Driscoll, B. R., Howard, L., Davison, A. G., & on behalf of the British Thoracic Society (2008) · Thorax · doi:10.1136/thx.2008.102947 · Cited by 740
BTS guideline for oxygen use in adults in healthcare and emergency settings
O’Driscoll, B. R., Howard, L., Earis, J. E., & Mak, V. H. F. (2017) · Thorax · doi:10.1136/thoraxjnl-2016-209729 · Cited by 731
Investigating sources of inaccuracy in wearable optical heart rate sensors
Bent, B., Goldstein, B. A., Kibbe, W. A., & Dunn, J. P. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0226-6 · Cited by 673
Metadata
- Authors
- Bickler, P. E., Feiner, J. R., & Severinghaus, J. W.
- Affiliations
- Professor.
- Associate Professor.
- Professor Emeritus.
- ORCID
- Bickler, Philip E.
- Venue
- Anesthesiology
- Published
- 2005-04-01
- Volume
- 102
- Issue
- 4
- Pages
- 715-719
- ISSN
- 0003-3022 (ISSN-L)
- OpenAlex
- W2142600751
- Cited by
- 427
- References
- 16
- Citations by year
- 2026: 35 · 2025: 46 · 2024: 49 · 2023: 54 · 2022: 86 · 2021: 36 · 2020: 20 · 2019: 12 · 2018: 7 · 2017: 5 · 2016: 10 · 2015: 6 · 2014: 4 · 2013: 10 · 2012: 10
- FWCI
- 4.93
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Anesthesiology (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
- Keywords
- pulse oximetry, skin pigmentation, dark skin, arterial oxygen saturation, low oxygen saturation, hypoxia, pulse oximeter accuracy, desaturation, end-tidal oxygen
- MeSH terms
- Adult, Black or African American, Female, Humans, Male, Middle Aged, Oximetry, Oxygen, Oxyhemoglobins, Skin Pigmentation, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Effects of Skin Pigmentation on Pulse Oximeter Accuracy at Low Saturation. https://app.cochat.ai/papers/10.1097/00000542-200504000-00004
- The paper
- Bickler, P. E., Feiner, J. R., & Severinghaus, J. W. (2005). Effects of Skin Pigmentation on Pulse Oximeter Accuracy at Low Saturation. Anesthesiology. https://doi.org/10.1097/00000542-200504000-00004
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.