Biological reactions to wear debris in total joint replacement
Abstract
The vast majority of total hip prostheses currently implanted consist of a hard metal or ceramic femoral head articulating against an ultra-high molecular weight polyethylene (UHMWPE) acetabular cup. Over the last 10 years, evidence has accumulated to show that these prostheses are prone to failure due to late aseptic loosening and few survive beyond 25 years. With an increasing need to implant hip prostheses in the younger, more active patient the need to understand the mechanisms of failure and to develop artificial hip joints using alternative materials have become major issues in the orthopaedic community. This review focuses initially on our current understanding of the biological reactions to UHMWPE prosthetic wear debris in vivo and in vitro since this is believed to be the main cause of late aseptic loosening. While the precise mechanisms of osteolysis induced by UHMWPE wear debris have not been elucidated, the major message to emerge is that it is not the wear volume that determines the biological response to the debris, but the concentration of the wear volume that is within the critical size range (0.2-0.8 μm) for macrophage activation. The review then considers whether the problem of wear-debris-induced osteolysis may be overcome with the use of new generation metal-on-metal or ceramic-on-ceramic prostheses. For metal-on-metal prostheses, the prospects for increasing the osteolysis free life of the implant are good but additional biological problems associated with the nanometre size and reactivity of the wear particles in vivo may emerge. For the ceramic-on-ceramic prostheses, although initial prospects are encouraging, more data are needed on the characteristics of the wear particles generated in vivo before predictions can be made. It is concluded that the pre-clinical testing of any new materials for joint replacement must include an analysis of the wear particle characteristics and their biological reactivity in addition to the usual assessment of wear.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Bertolini, D. R., Nedwin, G. E., Bringman, T. S., Smith, D. D., & Mundy, G. R. (1986) · Nature · doi:10.1038/319516a0 · Cited by 1,434
Effect of femoral head size on wear of the polyethylene acetabular component.
Livermore, J., Ilstrup, D., & Morrey, B. (1990) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-199072040-00008 · Cited by 1,024
An interleukin 1 like factor stimulates bone resorption in vitro
Gowen, M., Wood, D. D., Ihrie, E. J., McGuire, M. K. B., & Russell, R. G. G. (1983) · Nature · doi:10.1038/306378a0 · Cited by 1,012
Schmalzried, T. P., Jasty, M., & Harris, W. H. (1992) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-199274060-00006 · Cited by 989
Goldring, S. R., Schiller, A. L., Roelke, M., Rourke, C. M., OʼNeil, D. A., & Harris, W. H. (1983) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-198365050-00001 · Cited by 849
Reactions of the articular capsule to wear products of artificial joint prostheses
Willert, H. ‐., & Semlitsch, M. (1977) · Journal of Biomedical Materials Research · doi:10.1002/jbm.820110202 · Cited by 822
Mechanism and Clinical Significance of Wear Debris-Induced Osteolysis
AMSTUTZ, H. C., CAMPBELL, P., KOSSOVSKY, N., & CLARKE, I. C. (1992) · Clinical Orthopaedics and Related Research · doi:10.1097/00003086-199203000-00003 · Cited by 670
Roentgen stereophotogrammetric analysis as a predictor of mechanical loosening of knee prostheses
Ryd, L., Albrektsson, B., Carlsson, L., Dansgard, F., Herberts, P., Lindstrand, A., Regner, L., & Toksvig-Larsen, S. (1995) · The Journal of Bone and Joint Surgery. British volume · doi:10.1302/0301-620x.77b3.7744919 · Cited by 603
Green, T. (1998) · Biomaterials · doi:10.1016/s0142-9612(98)00140-9 · Cited by 569
Widespread dissemination of metal debris from implants
Case, C., Langkamer, V., James, C., Palmer, M., Kemp, A., Heap, P., & Solomon, L. (1994) · The Journal of Bone and Joint Surgery. British volume · doi:10.1302/0301-620x.76b5.8083255 · Cited by 444
Cited by
The role of macrophages in osteolysis of total joint replacement
Ingham, E., & Fisher, J. (2005) · Biomaterials · doi:10.1016/j.biomaterials.2004.04.035 · Cited by 667
Self-citationThe role of polymethylmethacrylate bone cement in modern orthopaedic surgery
Webb, J. C. J., & Spencer, R. F. (2007) · The Journal of Bone and Joint Surgery. British volume · doi:10.1302/0301-620x.89b7.19148 · Cited by 386
Metadata
- Authors
- Ingham, E., & Fisher, J.
- Venue
- Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Publisher
- SAGE Publications
- Published
- 2000-01-01
- Volume
- 214
- Issue
- 1
- Pages
- 21-37
- ISSN
- 0954-4119 (ISSN-L)
- 2041-3033
- OpenAlex
- W2068645319
- Cited by
- 708
- References
- 83
- FWCI
- 26.29
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Biological reactions to wear debris in total joint replacement. https://app.cochat.ai/papers/10.1243/0954411001535219
- The paper
- Ingham, E., & Fisher, J. (2000). Biological reactions to wear debris in total joint replacement. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. https://doi.org/10.1243/0954411001535219
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.