Comparison of blood glucose meters in a New Zealand diabetes centre
The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
Background: To assess the accuracy and precision of the Roche Performa and Medisense Optium Xceed (5 and 10 s reading) blood glucose meters. Methods: Capillary blood samples were taken from 100 patients attending a diabetes centre and blood glucose measured on Roche Performa ( n = 4) and Medisense Optium Xceed 5 s ( n = 2) and 10 s reading ( n = 2) meters. Venous plasma glucose from samples taken simultaneously was measured by the laboratory hexokinase method as reference standard. Imprecision was determined on the meters by replicate analysis ( n = 20) of control solutions provided by the manufacturers and also patient venous whole-blood samples. Accuracy was assessed relative to the reference method by Bland–Altman plots, Passing and Bablok regression analysis, and both Clarke and consensus error grid analysis. Coefficients of variation (CVs) were calculated to determine imprecision. Results: Bland–Altman and Passing–Bablok analysis confirmed significant systematic bias for all meters, with relative under-reading of higher glucose concentrations. Error grid analysis showed that <5% readings exceeded ±20% (or ±0.83 mmol/L for readings <4 mmol/L) deviation from the reference method (1%, 2% and 4% for the Roche, Optium 5 and 10 s meters, respectively). CVs were all <4% for the control solutions and <6% for patient samples. Conclusions: Both Roche Performa and Medisense Optium glucose meters (5 and 10 s readings) perform satisfactorily and are acceptable for operational use.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Kirkman, M. S., Lernmark, A., Metzger, B. E., & Nathan, D. M. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2010.161596 · Cited by 1,360
Parkes, J. L., Slatin, S. L., Pardo, S., & Ginsberg, B. H. (2000) · Diabetes Care · doi:10.2337/diacare.23.8.1143 · Cited by 620
Farmer, A., Wade, A., Goyder, E., Yudkin, P., French, D., Craven, A., Holman, R., Kinmonth, A.-L., & Neil, A. (2007) · BMJ · doi:10.1136/bmj.39247.447431.be · Cited by 407
O’Kane, M. J., Bunting, B., Copeland, M., & Coates, V. E. (2008) · BMJ · doi:10.1136/bmj.39534.571644.be · Cited by 326
Approved IFCC recommendation on reporting results for blood glucose (abbreviated)
D'Orazio, P. (2005) · Clinical Chemistry · doi:10.1373/clinchem.2005.051979 · Cited by 208
Sell-Monitoring of Blood Glucose
(1996) · Diabetes Care · doi:10.2337/diacare.19.1.s62 · Cited by 89
Cited by
The Accuracy of Point-of-Care Glucose Measurements
Rebel, A., Rice, M. A., & Fahy, B. G. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600228 · Cited by 144
McTavish, L., & Wiltshire, E. (2011) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2010.00725.x · Cited by 58
“The glucometers used satisfy the ISO15197 standard and Clarke error grid analysis for accuracy and precision (23).”
MethodGrzych, G., Defauwes, I., de Tullio, P., Pekar, J. D., Brousseau, T., Lippi, G., & Cavalier, E. (2025) · Critical Reviews in Clinical Laboratory Sciences · doi:10.1080/10408363.2025.2533855 · Cited by 4
Kim, Y. J., Choi, J. Y., & Kim, K. R. (2016) · International Journal of Bio-Science and Bio-Technology · doi:10.14257/ijbsbt.2016.8.2.38 · Cited by 1
Metadata
- Authors
- Florkowski, C., Budgen, C., Kendall, D., Lunt, H., & Moore, M. P.
- Venue
- Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
- Publisher
- SAGE Publications
- Published
- 2009-05-19
- Volume
- 46
- Issue
- 4
- Pages
- 302-305
- ISSN
- 0004-5632 (ISSN-L)
- 1758-1001
- OpenAlex
- W2003105358
- Cited by
- 18
- References
- 7
- FWCI
- 1.57
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.