Glucose meters: evaluation of the new formulation measuring strips from Roche (Accu-Chek) and Abbott (MediSense)

Open access
Dimeski, G, Jones, B W, Tilley, V, Greenslade, M N, Russell, A W · 2010 · Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
Cited by 31

The DOI matches its Crossref record; 5 of the 5 references it cites were found in a registry or an index.

Abstract

Background: Both Roche and Abbott have released new glucose meter strips. They supply the entire Australian hospital market. The present study compared the performance of the new strips utilizing various specimen types (capillary, venous lithium heparin whole blood, venous lithium heparin plasma and serum) and evaluated how well they comply with the International Standards Organization (ISO) 15197 criteria. Methods: The study included imprecision, patient comparison and interference studies. Participants with and without diabetes were recruited to evaluate the performance of various specimen types against the Beckman DxC800 glucose method. The strips were tested for different interferences: galactose, maltose, lactose, Icodextrin, Intragam, paracetamol, sodium, ascorbic acid, variable strip storage temperature, haematocrit, haemolysis and lipaemia. Results: The imprecision of the two strips was ∼5% or less, except for the Abbott strip at very low values (1.4 mmol/L), ∼7%. In total, 78% and 84%, respectively, of the results from the finger prick capillary specimens with the Roche (Accu-Chek Performa meter) and Abbott (Optium Xceed meter) strips, not 95% or greater as recommended by the ISO guideline, were within the recommended limits compared with reference plasma estimation on laboratory analysers. Galactose, ascorbic acid, haematocrit and sodium on the Roche and ascorbic acid and haematocrit on the Abbott strip continue to interfere to a variable degree with the glucose measurement. Conclusion: Analytically small differences exist between the glucose meter strips. The most significant analytical difference with the strips was at low glucose levels when compared with laboratory analyses and this may be of clinical importance. The impact of some of the interferences is variable between the two strips. Individuals, health-care professionals and health-care institutions should consider these data when selecting glucose meters for the management of people with diabetes mellitus.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1258/acb.2010.009291

Cites

5 of 5 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 5 references

Cited by

See all 31

Metadata

Authors
Dimeski, G., Jones, B. W., Tilley, V, Greenslade, M. N., & Russell, A. W.
Venue
Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
Published
2010-06-30
Volume
47
Issue
4
Pages
358-365
ISSN
  • 0004-5632 (ISSN-L)
  • 1758-1001
OpenAlex
W2164620477
Cited by
31
References
5
FWCI
1.52
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Glucose meters: evaluation of the new formulation measuring strips from Roche (Accu-Chek) and Abbott (MediSense). https://app.cochat.ai/papers/10.1258/acb.2010.009291
The paper
Dimeski, G., Jones, B. W., Tilley, V, Greenslade, M. N., & Russell, A. W. (2010). Glucose meters: evaluation of the new formulation measuring strips from Roche (Accu-Chek) and Abbott (MediSense). Annals of Clinical Biochemistry: International Journal of Laboratory Medicine. https://doi.org/10.1258/acb.2010.009291

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.