Two site evaluation of the performance of a new generation point-of-care glucose meter for use in a neonatal intensive care unit

The DOI matches its Crossref record; the references it cites are not checked yet.

Abstract

BACKGROUND: Monitoring of blood glucose in neonatal intensive care unit (NICU) patients is important in maintaining normoglycaemia and reducing the risk of hypoglycaemia. Point-of-care testing (POCT) glucose meters provide short turnaround times but some have been reported to be affected by haematocrit interference and other biochemical or biological substances in their accuracy and performance. The aim of this study was to assess the performance of a new POCT glucose meter in a challenging preterm neonatal population. METHODS: The new Nova Biomedical StatStrip™ (Nova Biomedical) was tested on 159 heparinised whole blood samples from NICU patients obtained for blood gas analysis. Accuracy (bias) of the meter and analytical interferences were evaluated by comparing the results of the meter with the results of the blood gas analyser routinely used for glucose measurements in this NICU setting. RESULTS: The results of the StatStrip glucose meter correlated very well with the reference routine method across a wide glucose concentration range (13-389 mg/dL) and were not affected by the level of haematocrit, by sample pH or by medication. CONCLUSIONS: The StatStrip meter showed good clinical accuracy and performance for measuring and monitoring glucose levels in NICU patients, with special respect to preterm infants, and therefore can act as a perfect alternative to a blood gas analyser for measuring blood glucose in NICU patients.

Overview

Integrity signals

Registry

Crossref · 10.1515/cclm-2012-0864

Cites

Statistics

Recomputing p-values needs the paper’s full text.

Verification

Claims the paper makes

This paper’s full text isn’t openly available, so its claims and self-check weren’t run. The registry and citation checks still apply.

Citations

The citation lists are queued

They are fetched later and appear here then.

References 20 · Cited by 18

Research works printed but not linked: 36

  1. Jackson, 2006, Special report : false positive finger stick blood glucose readings after high - dose intravenous vitamin Available from http www orthomolecular org library jom pdf pdf Accessed on November, J Orthomol Med, 21, 188
  2. August, 2003, for Standardization ISO In vitro diagnostic test systems Requirements for blood - glucose monitoring systems for self - testing in managing diabetes mellitus Available from : http www iso org iso iso catalogue catalogue tc catalogue detail htm csnumber, International Organization, 15197
  3. Newman, 2002, Monitoring neonatal hypoglycemia with the Accu - chek Advantage II Glucose Meter : the cautionary tale of galactosemia, Clin Chem, 2071, 10.1093/clinchem/48.11.2071 · DOI · Paper page
  4. Sander, 2010, Christoph Galactose interference on POCT Glucose Point of Care, Analysis
  5. Jackson, 2006, Special report : false positive finger stick blood glucose readings after high - dose intravenous vitamin Available from http www orthomolecular org library jom pdf pdf Accessed on November, J Orthomol Med, 21, 188
  6. Ho, 2003, Evaluation of point of care devices in the measurement of low blood glucose in neonatal practice Neonatal, Young Arch Dis Child Fetal, 89
  7. Roth, 2010, Evaluation of different POCT devices for glucose measurement in a clinical neonatal setting, Eur J Pediatr, 169
  8. Orazio, 2005, Approved IFCC recommendation on reporting results for blood glucose abbreviated, Clin Chem
  9. Burrin, 1990, KG What is blood glucose can it be measured, Med, 199
  10. Ho, 2003, Evaluation of point of care devices in the measurement of low blood glucose in neonatal practice Neonatal, Young Arch Dis Child Fetal, 89
  11. Avery, 2005, GB MG MM MD pathophysiology management of the newborn th ed Philadelphia, neonatology
  12. Aynsley, 1991, Glucose a for, fuel thought Paediatr Health, 27, 21
  13. Sander, 2010, Christoph Galactose interference on POCT Glucose Point of Care, Analysis
  14. Tang, 2001, Oxygen effects on glucose meter measurements with glucose dehydrogenase - and oxidase - based test strips for point - of - care testing, Crit Care Med, 29, 10.1097/00003246-200105000-00038 · DOI · Paper page
  15. August, 2003, for Standardization ISO In vitro diagnostic test systems Requirements for blood - glucose monitoring systems for self - testing in managing diabetes mellitus Available from : http www iso org iso iso catalogue catalogue tc catalogue detail htm csnumber, International Organization, 15197
  16. Tang, 2000, Effects of pH on glucose measurements with handheld glucose meters and a portable glucose analyzer for point - of - care testing, Arch Pathol Lab Med, 124
  17. Balion, 2006, Screening for hypoglycemia at the bedside in the neonatal intensive care unit with the glucose meter, BMC Pediatrics, 6, 10.1186/1471-2431-6-28 · DOI · Paper page
  18. Tang, 2000, Effects of different hematocrit levels on glucose measurements with handheld meters for point - of - care testing, Arch Pathol Lab Med, 124
  19. Burrin, 1990, KG What is blood glucose can it be measured, Med, 199
  20. Bellini, 2007, Reliability assessment of glucose measurement by HemoCue analyser in a neonatal intensive care unit, Clin Chem Lab Med, 10.1515/CCLM.2007.302 · DOI
  21. Rainey, 1995, Monitoring blood glucose meters, Am J Clin Pathol, 103
  22. Louis, 2002, St An evaluation of three glucose meter systems and their performance in relation to criteria of acceptability for neonatal specimens, Clin Chim Acta, 322
  23. Louis, 2002, St An evaluation of three glucose meter systems and their performance in relation to criteria of acceptability for neonatal specimens, Clin Chim Acta, 322
  24. Watkinson, 2012, The effects of precision haematocrit pH and oxygen tension on point - of - care glucose measurement in critically ill patients : a prospective study, Ann Clin Biochem, 49
  25. Avery, 2005, GB MG MM MD pathophysiology management of the newborn th ed Philadelphia, neonatology
  26. Tang, 2000, Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer, Am J Clin Pathol, 113, 75, 10.1309/QAW1-X5XW-BVRQ-5LKQ · DOI · Paper page
  27. Aynsley, 1991, Glucose a for, fuel thought Paediatr Health, 27, 21
  28. Beardsall, 2010, of glucose levels in the newborn, Measurement Early Hum Dev, 263, 10.1016/j.earlhumdev.2010.05.005 · DOI
  29. Makaya, 2012, Point - of - care glucose monitoring on the neonatal unit, Paediatr Health, 342
  30. Kost, 1998, Multicentre study of oxygen - insensitive handheld glucose point - of - care testing in critical care / hospital / ambulatory patients in the United States and, Crit Care Med, 26, 581, 10.1097/00003246-199803000-00036 · DOI · Paper page
  31. Pidcoke, 2010, Anemia causes hypoglycemia in intensive care unit patients due to error in single - channel glucometers : methods of reducing patient risk, Crit Care Med, 471, 10.1097/CCM.0b013e3181bc826f · DOI · Paper page
  32. Dimeski, 2010, Glucose meters : evaluation of the new formulation measuring strips from, Russell Ann Clin Biochem, 358, 10.1258/acb.2010.009291 · DOI · Paper page
  33. Cornblath, 2000, Controversies regarding definition of neonatal hypoglycemia : suggested operational thresholds, Pediatrics, 105
  34. Hellman, 2012, Glucose meter inaccuracy and the impact on the care of patients, Diabetes Metab Res Rev, 28, 207, 10.1002/dmrr.2271 · DOI · Paper page
  35. Suppl, 2010, Standards of medical care in diabetes, American Diabetes Association Diabetes Care, 1
  36. Dungan, 2007, Glucose measurement : confounding issues in setting targets for inpatient management, Diabetes Care, 30, 403, 10.2337/dc06-1679 · DOI · Paper page

Metadata

Authors
Tendl, K. A., Christoph, J., Bohn, A., Herkner, K. R., Pollak, A., & Prusa, A.-R.
Affiliations
  • Medical University of Vienna (AT)
  • Kinderkrankenhaus auf der Bult (DE)
Corresponding author
Tendl, Kristina A.
Venue
Clinical Chemistry and Laboratory Medicine
Published
2013-01-01
Volume
51
Issue
9
Pages
1747-54
ISSN
  • 1434-6621 (ISSN-L)
  • 1437-4331
OpenAlex
W1981753413
Cited by
18
References
20
Citations by year
2025: 2 · 2024: 1 · 2023: 2 · 2019: 3 · 2018: 1 · 2016: 1 · 2015: 4 · 2014: 3 · 2013: 1
FWCI
2.08
Type
journal article
Language
en
Open access
closed
Locations
  • Clinical Chemistry and Laboratory Medicine (CCLM) (publishedVersion)
  • PubMed (publishedVersion)
Topics
  • Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
  • Health Sciences › Medicine › Pediatrics, Perinatology and Child Health › Neonatal Health and Biochemistry
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
Keywords
point-of-care glucose meters, neonatal intensive care, preterm infants, blood glucose monitoring, hypoglycemia prevention, analytical interference, blood gas analyser, method comparison
MeSH terms
Blood Glucose, Humans, Infant, Newborn, Intensive Care Units, Neonatal, Reagent Strips, Blood Glucose Self-Monitoring, Point-of-Care Systems

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Two site evaluation of the performance of a new generation point-of-care glucose meter for use in a neonatal intensive care unit. https://app.cochat.ai/papers/10.1515/cclm-2012-0864
The paper
Tendl, K. A., Christoph, J., Bohn, A., Herkner, K. R., Pollak, A., & Prusa, A.-R. (2013). Two site evaluation of the performance of a new generation point-of-care glucose meter for use in a neonatal intensive care unit. Clinical Chemistry and Laboratory Medicine. https://doi.org/10.1515/cclm-2012-0864

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.