Two site evaluation of the performance of a new generation point-of-care glucose meter for use in a neonatal intensive care unit
The DOI matches its Crossref record; the references it cites are not checked yet.
Abstract
BACKGROUND: Monitoring of blood glucose in neonatal intensive care unit (NICU) patients is important in maintaining normoglycaemia and reducing the risk of hypoglycaemia. Point-of-care testing (POCT) glucose meters provide short turnaround times but some have been reported to be affected by haematocrit interference and other biochemical or biological substances in their accuracy and performance. The aim of this study was to assess the performance of a new POCT glucose meter in a challenging preterm neonatal population. METHODS: The new Nova Biomedical StatStrip™ (Nova Biomedical) was tested on 159 heparinised whole blood samples from NICU patients obtained for blood gas analysis. Accuracy (bias) of the meter and analytical interferences were evaluated by comparing the results of the meter with the results of the blood gas analyser routinely used for glucose measurements in this NICU setting. RESULTS: The results of the StatStrip glucose meter correlated very well with the reference routine method across a wide glucose concentration range (13-389 mg/dL) and were not affected by the level of haematocrit, by sample pH or by medication. CONCLUSIONS: The StatStrip meter showed good clinical accuracy and performance for measuring and monitoring glucose levels in NICU patients, with special respect to preterm infants, and therefore can act as a perfect alternative to a blood gas analyser for measuring blood glucose in NICU patients.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Citations
The citation lists are queued
They are fetched later and appear here then.
References 20 · Cited by 18
Research works printed but not linked: 36
- Jackson, 2006, Special report : false positive finger stick blood glucose readings after high - dose intravenous vitamin Available from http www orthomolecular org library jom pdf pdf Accessed on November, J Orthomol Med, 21, 188
- August, 2003, for Standardization ISO In vitro diagnostic test systems Requirements for blood - glucose monitoring systems for self - testing in managing diabetes mellitus Available from : http www iso org iso iso catalogue catalogue tc catalogue detail htm csnumber, International Organization, 15197
- Newman, 2002, Monitoring neonatal hypoglycemia with the Accu - chek Advantage II Glucose Meter : the cautionary tale of galactosemia, Clin Chem, 2071, 10.1093/clinchem/48.11.2071 · DOI · Paper page
- Sander, 2010, Christoph Galactose interference on POCT Glucose Point of Care, Analysis
- Jackson, 2006, Special report : false positive finger stick blood glucose readings after high - dose intravenous vitamin Available from http www orthomolecular org library jom pdf pdf Accessed on November, J Orthomol Med, 21, 188
- Ho, 2003, Evaluation of point of care devices in the measurement of low blood glucose in neonatal practice Neonatal, Young Arch Dis Child Fetal, 89
- Roth, 2010, Evaluation of different POCT devices for glucose measurement in a clinical neonatal setting, Eur J Pediatr, 169
- Orazio, 2005, Approved IFCC recommendation on reporting results for blood glucose abbreviated, Clin Chem
- Burrin, 1990, KG What is blood glucose can it be measured, Med, 199
- Ho, 2003, Evaluation of point of care devices in the measurement of low blood glucose in neonatal practice Neonatal, Young Arch Dis Child Fetal, 89
- Avery, 2005, GB MG MM MD pathophysiology management of the newborn th ed Philadelphia, neonatology
- Aynsley, 1991, Glucose a for, fuel thought Paediatr Health, 27, 21
- Sander, 2010, Christoph Galactose interference on POCT Glucose Point of Care, Analysis
- Tang, 2001, Oxygen effects on glucose meter measurements with glucose dehydrogenase - and oxidase - based test strips for point - of - care testing, Crit Care Med, 29, 10.1097/00003246-200105000-00038 · DOI · Paper page
- August, 2003, for Standardization ISO In vitro diagnostic test systems Requirements for blood - glucose monitoring systems for self - testing in managing diabetes mellitus Available from : http www iso org iso iso catalogue catalogue tc catalogue detail htm csnumber, International Organization, 15197
- Tang, 2000, Effects of pH on glucose measurements with handheld glucose meters and a portable glucose analyzer for point - of - care testing, Arch Pathol Lab Med, 124
- Balion, 2006, Screening for hypoglycemia at the bedside in the neonatal intensive care unit with the glucose meter, BMC Pediatrics, 6, 10.1186/1471-2431-6-28 · DOI · Paper page
- Tang, 2000, Effects of different hematocrit levels on glucose measurements with handheld meters for point - of - care testing, Arch Pathol Lab Med, 124
- Burrin, 1990, KG What is blood glucose can it be measured, Med, 199
- Bellini, 2007, Reliability assessment of glucose measurement by HemoCue analyser in a neonatal intensive care unit, Clin Chem Lab Med, 10.1515/CCLM.2007.302 · DOI
- Rainey, 1995, Monitoring blood glucose meters, Am J Clin Pathol, 103
- Louis, 2002, St An evaluation of three glucose meter systems and their performance in relation to criteria of acceptability for neonatal specimens, Clin Chim Acta, 322
- Louis, 2002, St An evaluation of three glucose meter systems and their performance in relation to criteria of acceptability for neonatal specimens, Clin Chim Acta, 322
- Watkinson, 2012, The effects of precision haematocrit pH and oxygen tension on point - of - care glucose measurement in critically ill patients : a prospective study, Ann Clin Biochem, 49
- Avery, 2005, GB MG MM MD pathophysiology management of the newborn th ed Philadelphia, neonatology
- Tang, 2000, Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer, Am J Clin Pathol, 113, 75, 10.1309/QAW1-X5XW-BVRQ-5LKQ · DOI · Paper page
- Aynsley, 1991, Glucose a for, fuel thought Paediatr Health, 27, 21
- Beardsall, 2010, of glucose levels in the newborn, Measurement Early Hum Dev, 263, 10.1016/j.earlhumdev.2010.05.005 · DOI
- Makaya, 2012, Point - of - care glucose monitoring on the neonatal unit, Paediatr Health, 342
- Kost, 1998, Multicentre study of oxygen - insensitive handheld glucose point - of - care testing in critical care / hospital / ambulatory patients in the United States and, Crit Care Med, 26, 581, 10.1097/00003246-199803000-00036 · DOI · Paper page
- Pidcoke, 2010, Anemia causes hypoglycemia in intensive care unit patients due to error in single - channel glucometers : methods of reducing patient risk, Crit Care Med, 471, 10.1097/CCM.0b013e3181bc826f · DOI · Paper page
- Dimeski, 2010, Glucose meters : evaluation of the new formulation measuring strips from, Russell Ann Clin Biochem, 358, 10.1258/acb.2010.009291 · DOI · Paper page
- Cornblath, 2000, Controversies regarding definition of neonatal hypoglycemia : suggested operational thresholds, Pediatrics, 105
- Hellman, 2012, Glucose meter inaccuracy and the impact on the care of patients, Diabetes Metab Res Rev, 28, 207, 10.1002/dmrr.2271 · DOI · Paper page
- Suppl, 2010, Standards of medical care in diabetes, American Diabetes Association Diabetes Care, 1
- Dungan, 2007, Glucose measurement : confounding issues in setting targets for inpatient management, Diabetes Care, 30, 403, 10.2337/dc06-1679 · DOI · Paper page
Metadata
- Authors
- Tendl, K. A., Christoph, J., Bohn, A., Herkner, K. R., Pollak, A., & Prusa, A.-R.
- Affiliations
- Medical University of Vienna (AT)
- Kinderkrankenhaus auf der Bult (DE)
- Corresponding author
- Tendl, Kristina A.
- Venue
- Clinical Chemistry and Laboratory Medicine
- Publisher
- Walter de Gruyter GmbH
- Published
- 2013-01-01
- Volume
- 51
- Issue
- 9
- Pages
- 1747-54
- ISSN
- 1434-6621 (ISSN-L)
- 1437-4331
- OpenAlex
- W1981753413
- Cited by
- 18
- References
- 20
- Citations by year
- 2025: 2 · 2024: 1 · 2023: 2 · 2019: 3 · 2018: 1 · 2016: 1 · 2015: 4 · 2014: 3 · 2013: 1
- FWCI
- 2.08
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Clinical Chemistry and Laboratory Medicine (CCLM) (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Pediatrics, Perinatology and Child Health › Neonatal Health and Biochemistry
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Keywords
- point-of-care glucose meters, neonatal intensive care, preterm infants, blood glucose monitoring, hypoglycemia prevention, analytical interference, blood gas analyser, method comparison
- MeSH terms
- Blood Glucose, Humans, Infant, Newborn, Intensive Care Units, Neonatal, Reagent Strips, Blood Glucose Self-Monitoring, Point-of-Care Systems
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Two site evaluation of the performance of a new generation point-of-care glucose meter for use in a neonatal intensive care unit. https://app.cochat.ai/papers/10.1515/cclm-2012-0864
- The paper
- Tendl, K. A., Christoph, J., Bohn, A., Herkner, K. R., Pollak, A., & Prusa, A.-R. (2013). Two site evaluation of the performance of a new generation point-of-care glucose meter for use in a neonatal intensive care unit. Clinical Chemistry and Laboratory Medicine. https://doi.org/10.1515/cclm-2012-0864
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.