Comparisons between a pencil beam and two fan beam dual energy X-ray absorptiometers used for measuring total body bone and soft tissue

Tothill, P, Hannan, W J, Wilkinson, S·2001·The British Journal of Radiology

Abstract

A pencil beam Hologic QDR 1000W scanner (1000), a fan beam QDR 4500A scanner (4500) and a fan beam Lunar Expert scanner (Expert) were compared for bone mineral and body composition measurement accuracy. Phantoms were scanned with each instrument to assess magnification effects and to compare calibrations for bone mineral and fat proportion. 41 volunteers were scanned with both the 1000 and the 4500, and 21 patients with both the 4500 and the Expert. The height of a bone within the body affected the measured bone mineral content (BMC) and, to a lesser extent, the bone mineral density (BMD). There were differences in calibration against recognized standards for fat proportion between the three instruments. The 1000 underestimated low fat proportions and the 4500 underestimated high fat proportions. Fat results for the Expert were closer to nominal values. Comparisons on volunteers showed that measured mean total body BMD was 4% higher and BMC was 7% higher with the 1000 compared with the 4500; some regional differences were greater. Mean values of per cent fat were equal, but the total and regional regression coefficients were well above unity. Mean BMD was 3% higher and mean BMC was 10% higher with the Expert compared with the 4500, but most regression coefficients for these comparisons were less than unity. Mean values of per cent fat were equal, but regression coefficients were above unity. Errors due to magnification are acceptable. Differences between the instruments are appreciable, but can be accommodated by cross-calibration.

Integrity signals

Registry

Crossref · 10.1259/bjr.74.878.740166

Cites

23 of 23 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

This paper’s full text isn’t openly available, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Tothill, P., Hannan, W. J., & Wilkinson, S.
Affiliations
  • Western General Hospital (GB)
  • Woolmanhill Hospital (GB)
Venue
The British Journal of Radiology
Published
2001-02
Volume
74
Issue
878
Pages
166-176
ISSN
  • 0007-1285 (ISSN-L)
  • 1748-880X
OpenAlex
W2161104497
Cited by
71
References
23
Citations by year
2025: 2 · 2023: 1 · 2021: 3 · 2019: 1 · 2017: 2 · 2016: 1 · 2014: 1 · 2013: 7 · 2012: 8
FWCI
2.65
Type
journal article
Language
en
Open access
closed
Locations
  • British Journal of Radiology (publishedVersion)
  • PubMed (publishedVersion)
Topics
  • Health Sciences › Medicine › Orthopedics and Sports Medicine › Bone health and osteoporosis research
  • Health Sciences › Medicine › Physiology › Body Composition Measurement Techniques
  • Physical Sciences › Engineering › Biomedical Engineering › Advanced X-ray and CT Imaging
Keywords
dual-energy X-ray absorptiometry, bone mineral content, bone mineral density, body composition, instrument comparison, phantom calibration, magnification effects, cross-calibration, soft tissue
MeSH terms
Adipose Tissue, Adult, Body Composition, Connective Tissue, Female, Humans, Male, Middle Aged, Regression Analysis, Absorptiometry, Photon, Bone Density, Phantoms, Imaging

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review