CoChat

Metal wear and tissue response in failed titanium alloy total hip replacements

Witt, JD, Swann, M · 1991 · The Journal of Bone and Joint Surgery. British volume
Cited by 161

The DOI matches its Crossref record.

Abstract

Thirteen total hip replacements with titanium alloy femoral components required revision for loosening at an average of two years after implantation. At revision the soft tissues around the implant were darkly stained and a proliferative membrane had invaded the cement-bone interface. The femoral components showed polishing of parts of their shot-blasted surfaces. Histology showed a fibroblastic reaction with abundant titanium lying free and within histiocytes, and a scanty foreign-body giant-cell reaction. Surface analysis of the removed femoral components and chemical analysis of the excised tissues is described. Tissue reaction in response to the metal-wear debris may have contributed to the early failure of these implants.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1302/0301-620x.73b4.2071635

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 161

Metadata

Authors
Witt, J., & Swann, M.
Venue
The Journal of Bone and Joint Surgery. British volume
Published
1991-07
Volume
73-B
Issue
4
Pages
559-563
ISSN
  • 0301-620X (ISSN-L)
  • 2044-5377
OpenAlex
W2415377797
Cited by
161
References
0
FWCI
10.57
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Metal wear and tissue response in failed titanium alloy total hip replacements. https://app.cochat.ai/papers/10.1302/0301-620x.73b4.2071635
The paper
Witt, J., & Swann, M. (1991). Metal wear and tissue response in failed titanium alloy total hip replacements. The Journal of Bone and Joint Surgery. British volume. https://doi.org/10.1302/0301-620x.73b4.2071635

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.