Comparison of Standing Posture Bioelectrical Impedance Analysis with DXA for Body Composition in a Large, Healthy Chinese Population
Open accessAbstract
Bioelectrical impedance analysis (BIA) is a common method for assessing body composition in research and clinical trials. BIA is convenient but when compared with other reference methods, the results have been inconclusive. The level of obesity degree in subjects is considered to be an important factor affecting the accuracy of the measurements. A total of 711 participants were recruited in Taiwan and were sub-grouped by gender and levels of adiposity. Regression analysis and Bland-Altman analysis were used to evaluate the agreement of the measured body fat percentage (BF%) between BIA and DXA. The BF% measured by the DXA and BIA methods (Tanita BC-418) were expressed as BF%DXA and BF%BIA8, respectively. A one-way ANOVA was used to test the differences in BF% measurements by gender and levels of adiposity. The estimated BF%BIA8 and BF%DXA in the all subjects, male and female groups were all highly correlated (r = 0.934, 0.901, 0.916, all P< 0.001). The average estimated BF%BIA8 (22.54 ± 9.48%) was significantly lower than the average BF%DXA (26.26 ± 11.18%). The BF%BIA8 was overestimated in the male subgroup (BF%DXA< 15%), compared to BF%DXA by 0.45%, respectively. In the other subgroups, the BF%BIA8 values were all underestimated. Standing BIA estimating body fat percentage in Chinese participants have a high correlation, but underestimated on normal and high obesity degree in both male and female subjects.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Single-frequency devices and segmental measures demonstrate the largest differences when compared to DXA, with the inaccuracy increasing in conjunction with higher levels of BMI.
“However, single-frequency devices and segmental measures demonstrate the largest differences when compared to DXA, with the inaccuracy increasing in conjunction with higher levels of BMI [55,56,57,58].”
From The Utility of Body Composition Assessment in Nutrition and Clinical Practice: An Overview of Current Methodology (2021) · cites this paper as [57]
The Results (‘BF%BIA8 bias on the basis of BF%DXA’; page unavailable) show that the eight-electrode BIA device’s underestimation of body-fat percentage increased across lean, normal and obese groups. The paper groups participants by DXA-measured body-fat percentage, not BMI, and does not establish that single-frequency devices or segmental measures have the largest differences from DXA compared with other devices or measures. (LLM, conf=0.88)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
1 retracted
Changes in energy balance and body composition at menopause: a controlled longitudinal study.
Poehlman, E. T., Toth, M. J., & Gardner, A. W. (1995) · Annals of Internal Medicine · doi:10.7326/0003-4819-123-9-199511010-00005 · Cited by 520 · cited 1× · see citing works
“Body composition information can be applied in clinical trials and other medicine-related fields [27, 28].”
BackgroundRetractedSTATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965 · see citing works
Bioelectrical impedance underestimates total and truncal fatness in abdominally obese women
Neovius, M., Hemmingsson, E., Freyschuss, B., & Uddén, J. (2006) · Obesity · doi:10.1038/oby.2006.199 · Cited by 119 · see citing works
Validity and accuracy of regional bioelectrical impedance devices to determine whole-body fatness
Lukaski, H. C., & Siders, W. A. (2003) · Nutrition · doi:10.1016/s0899-9007(03)00166-7 · Cited by 73 · cited 1× · see citing works
“The manufacturers of the currently available BIA measuring equipment have not released their built-in prediction equations or identified a suitable testing population, which limits the reference and application value of BIA [14].”
Background
Flagged first, then by citations
Research works printed but not linked: 26
- [1]Finucane MM, Stevens GA, Cowan MJ, Danaei G, Lin JK, Paciorek CJ, et al National, regional, and global trends in body-mass index since 1980: systematic analysis of health examination surveys and epidemiological studies with 960 country-years and 9·1 million participants. Lancet 2011; 377: 557– 67. 10.1016/S0140-6736(10)62037-5 21295846 PMC4472365 · DOI · Paper page
- [2]Hall ME, do Carmo JM, da Silva AA, Juncos LA, Wang Z, Hall JE. Obesity, hypertension, and chronic kidney disease. Int J Nephrol Renovasc Dis 2014; 7: 75– 88. 10.2147/IJNRD.S39739 24600241 PMC3933708 · DOI · Paper page
- [3]Katch FI. Practice curves and errors of measurement in estimating underwater weight by hydrostatic weighing. Med Sci Sports 1969; 1: 212– 6.
- [4]Johansson AG, Forslund A, Sjödin A, Mallmin H, Hambraeus L, Ljunghall S. Determination of body composition–a comparison of dual-energy x-ray absorptiometry and hydrodensitometry. Am J Clin Nutr 1993; 57: 323– 6.. 8438765
- [5]Temple D, Denis R, Walsh MC, Dicker P, Byrne AT. Comparison of anthropometric-based equations for estimation of body fat percentage is a normal-weight and overweight female cohort: validation via air-displacement plethysmography. Public Health Nutr 2015; 18: 446– 52. 10.1017/S1368980014000597 24735997 PMC10271096 · DOI
- [6]Durnin JV, Womersley J. Body fat assessed from total body density and its estimation from skinfold thickness: measurements on 481 men and women aged from 16 to 72 years. Br J Nutr 1974; 32: 77– 97.. 4843734
- [7]Tan YX, Nuñez C, Sun Y, Zhang K, Wang Z, Heymsfield SB. New electrode system for rapid whole-body and segmental bioimpedance assessment. Med Sci Sports Exerc 1997; 29: 1269– 73.. 9309641
- [8]Bosaeus M, Karlsson T, Holmäng A, Ellegård L. Accuracy of quantitative magnetic resonance and eight-electrode bioelectrical impedance analysis in normal weight and obese women. Clin Nutr 2014; 33: 417– 7. 10.1016/j.clnu.2013.06.017 23871192 · DOI
- [9]Sun G, French CR, Martin GR, Younghusband B, Green RC, Xie YG, et al Comparison of multifrequency bioelectrical impedance analysis with dual-energy X-ray absorptiometry for assessment of percentage body fat in large, healthy population. Am J Clin Nutr 2005; 81: 74– 8.. 15640463
- [10]Kim H, Kim CH, Kim DW, Park M, Park HS, Min SS, et al External cross-validation of bioelectrical impedance analysis for the assessment of body composition in Korean adults. Ntr Res Pract 2011; 5: 246– 52. 10.4162/nrp.2011.5.3.246 PMC3133758 21779529 · DOI
- [11]Böhm A, Heitmann BL. The use of bioelectrical impedance analysis for body composition in epidemiological studies. Eur J Clin Nutr 2013; 67 Suppl 1: S79– S85. 10.1038/ejcn.2012.168 23299875 · DOI
- [12]Niqam P, Misra A, Colles SL. Comparison of DEXA-derived body fat measurement to two race-specific bioelectrical impedance equations in healthy Indians. Diabetes Metab Syndr 2013; 7: 72– 7. 10.1016/j.dsx.2013.02.031 23680244 · DOI
- [13]Wang JG, Zhang Y, Chen HE, Li Y, Cheng XG, Xu L, et al Comparison of two bioelectrical impedance analysis devices with dual energy X-ray absorptiometry and magnetic resonance imaging in the estimation of body composition. J Strength Cond Res 2013; 27: 236– 43. 10.1519/JSC.0b013e31824f2040 22344056 · DOI
- [15]Wing L, Hui SS, Wong SH. Validity of bioelectrical impedance measurement in predicting fat-free mass of Chinese children and adolescents. Med Sci Monit 2014; 20: 2298– 310. 10.12659/MSM.890696 25398209 PMC4245105 · DOI
- [17]Erceg DN, Dieli-Conwright CM, Rossuello AE, Jensky NE, Sun S, Schroeder ET. The Stay healthy bioelectrical impedance analyzer predicts body fat in children and adults. Nutr Res 2010; 30: 297– 304. 10.1016/j.nutres.2010.04.009 20579521 · DOI
- [18]Pietrobelli A, Rubiano F, St-Onge MP, Heymsfield SB. New bioimpedance analysis system: improved phenotyping with whole-body analysis. Eur J Clin Nutr 2004; 58: 1479– 84.. 15138459
- [19]Lin LI. A concordance correlation coefficient to evaluate reproducibility. Biometrics 1989; 45: 255– 68.. 2720055
- [20]Ludbrook J. Statistical techniques for comparing measurers and methods of measurement: a critical review. Clin Exp Pharmacol Physiol 2002; 29: 527– 36.. 12060093
- [24]Hemmingsson E, Udden J, Neovius M. No apparent progress in bioelectrical impedance accuracy: validation against metabolic risk and DXA. Obesity 2009; 17: 183– 7. 10.1038/oby.2008.474 18997678 · DOI
- [25]Mally K, Trentmann, Heller M, Dittmar M. Reliability and accuracy of segmental bioelectrical impedance analysis for assessing muscle and fat mass in older Europeans: a comparison with dual-energy X-ray absorptiometry. Eur J Appl Physiol 2011; 111: 1879– 87. 10.1007/s00421-010-1795-x 21234595 · DOI · Paper page
- [26]Lee LC, Hsieh KC, Wu CS, Chen YJ, Chiang J, Chen YY. Validity of standing posture eight-electrode bioelectrical impedance to estimate body composition in Taiwanese elderly. Int J Gerontology 2014; 8: 137– 42. 10.1016/j.ijge.2013.08.010. · DOI
- [27]Sentongo TA, Semeao EJ, Piccoli DA, Stallings VA, Zemel BS. Growth, body composition, and nutritional status in children and adolescents with Crohn's disease. J Pediatr Gastroenterol Nutr 2000; 31: 33– 40.. 10896068
- [29]Oshima Y, Shiga T. Within-day variability of whole-body and segmental bioelectrical impedance in a standing position. Eur J Clin Nutr 2006; 60: 938– 41.. 16482076
- [30]Gupta N, Balasekaran G, Victor Govindaswamy V, Hwa CY, Shun LM. Comparison of body composition with bioelectric impedance (BIA) and dual energy X-ray absorptiometry (DEXA) among Singapore Chinese. J Sci Med Sport 2011; 14: 33– 5. 10.1016/j.jsams.2010.04.005 20619734 · DOI
- [31]Tothill P, Hannan WJ, Wilkinson S. Comparisons between a pencil beam and two fan beam dual energy X-ray absorptiometers used for measuring total body bone and soft tissue. Br J Radiol 2001; 74: 166– 76.. 11718390
- [32]Tothill P, Hannan WJ. Comparisons between Hologic QDR 1000W, QDR 4500 A, and Lunar Expert dual energy X-ray absorptiometry scanners used for measuring total body bone and soft tissue. Ann N Y Acad Sci 2002; 904: 63– 71.. 10865711
Cited by
Sorted by citations
Marra, M., Sammarco, R., De Lorenzo, A., Iellamo, F., Siervo, M., Pietrobelli, A., Donini, L. M., Santarpia, L., Cataldi, M., Pasanisi, F., et al. (2019) · Contrast Media & Molecular Imaging · doi:10.1155/2019/3548284 · Cited by 401 · see citing works
Holmes, C. J., & Racette, S. B. (2021) · Nutrients · doi:10.3390/nu13082493 · Cited by 402 · see citing works
Son, J. W., Han, B., Bennett, J. P., Heymsfield, S., & Lim, S. (2025) · Obesity Reviews · doi:10.1111/obr.13844 · Cited by 74 · see citing works
- Authors
- Chen, K.-T., Chen, Y.-Y., Wang, C.-W., Chuang, C.-L., Chiang, L.-M., Lai, C.-L., Lu, H.-K., Dwyer, G. B., Chao, S.-P., Shih, M.-K., & Hsieh, K.-C.
- Affiliations
- National Chung Hsing University (TW)
- National Taiwan University of Sport (TW)
- Jen-Ai Hospital (TW)
- East Stroudsburg University (US)
- Ministry of Health and Welfare (TW)
- National Kaohsiung University of Hospitality and Tourism (TW)
- Charder Electronic Co., Ltd. (Taiwan) (TW)
- Corresponding author
- Hsieh, Kuen-Chang
- Venue
- PLOS ONE
- Publisher
- Public Library of Science (PLoS)
- Published
- 2016-07-28
- Volume
- 11
- Issue
- 7
- Pages
- e0160105
- ISSN
- 1932-6203 (ISSN-L)
- PMC
- PMC4965215
- OpenAlex
- W2505736967
- Cited by
- 43
- References
- 4
- Citations by year
- 2026: 4 · 2025: 3 · 2024: 7 · 2023: 2 · 2022: 7 · 2021: 8 · 2020: 7 · 2019: 5
- FWCI
- 1.34
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- public-domain
- creativecommons.org/publicdomain/zero/1.0/ (vor)
- Locations
- PLoS ONE (publishedVersion, public-domain)
- Europe PMC (PubMed Central) (publishedVersion, public-domain)
- PubMed Central (publishedVersion, public-domain)
- PubMed (publishedVersion)
- RePEc: Research Papers in Economics (submittedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion)
- Funders
- the Ministry of Health and Welfare of the Republic of China (PG-10417)
- the National Science Council of the republic of China (NSC 100-2410-H-H-028-MY3)
- Topics
- Health Sciences › Medicine › Physiology › Body Composition Measurement Techniques
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrical and Bioimpedance Tomography
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Obesity, Physical Activity, Diet
- Keywords
- bioelectrical impedance analysis, dual-energy X-ray absorptiometry, body composition, body fat percentage, standing posture, obesity, measurement agreement, Chinese population, sex differences
- MeSH terms
- Body Composition, Female, Humans, Male, Taiwan, Absorptiometry, Photon, Electric Impedance
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review