Analytical Validation of a High-Sensitivity Cardiac Troponin T Assay
Open accessAbstract
Background: We report the development of a novel high-sensitivity cardiac troponin T (hs-cTnT) assay, a modification of the Roche fourth-generation cTnT assay, and validation of the analytical performance of this assay. Methods: Validation included testing of analytical sensitivity, specificity, interferences, and precision. We established the 99th percentile cutoff from healthy reference populations (n = 616). In addition, we studied differences in time to a positive result when using serial measurements of hs-cTnT vs cTnT in patients with a confirmed diagnosis of non-ST elevation myocardial infarction (non-STEMI). Results: The hs-cTnT assay had an analytical range from 3 to 10 000 ng/L. At the 99th percentile value of 13.5 ng/L, the CV was 9% using the Elecsys® 2010 analyzer. The assay was specific for cTnT without interferences from human cTnI or cTnC, skeletal muscle TnT, or hemoglobin concentrations up to 1000 mg/L, above which falsely lower values would be expected. When the assay was evaluated clinically, a hs-cTnT higher than the 99th percentile concentration identified a significantly higher number of patients with non-STEMI on presentation (45 vs 20 patients, P = 0.0004) compared with cTnT, and a final diagnosis of non-STEMI was made in 9 additional patients (55 vs 46 patients, P = 0.23) after serial sampling. Time to diagnosis was significantly shorter using hs-cTnT compared with cTnT [mean 71.5 (SD 108.7) min vs 246.9 (82.0) min, respectively; P < 0.01]. Conclusions: The analytical performance of hs-cTnT complies with the ESC-ACCF-AHA-WHF Global Task Force recommendations for use in the diagnosis of MI.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Antman, E., Bassand, J.-P., Klein, W., Ohman, M., Lopez Sendon, J. L., Rydén, L., Simoons, M., & Tendera, M. (2000) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(00)00804-4 · Cited by 4,821
Thygesen, K. A. (2007) · Circulation · Cited by 3,126
Task Force Members, Thygesen, K., Alpert, J. S., White, H. D., Biomarker Group, Jaffe, A. S., Apple, F. S., Galvani, M., Katus, H. A., Newby, L. K., et al. (2007) · European Heart Journal · doi:10.1093/eurheartj/ehm355 · Cited by 2,541
Self-citationEarly Diagnosis of Myocardial Infarction with Sensitive Cardiac Troponin Assays
Reichlin, T., Hochholzer, W., Bassetti, S., Steuer, S., Stelzig, C., Hartwiger, S., Biedert, S., Schaub, N., Buerge, C., Potocki, M., et al. (2009) · New England Journal of Medicine · doi:10.1056/nejmoa0900428 · Cited by 1,687
Januzzi, J. L., van Kimmenade, R., Lainchbury, J., Bayes-Genis, A., Ordonez-Llanos, J., Santalo-Bel, M., Pinto, Y. M., & Richards, M. (2006) · European Heart Journal · doi:10.1093/eurheartj/ehi631 · Cited by 1,169
Sensitive Troponin I Assay in Early Diagnosis of Acute Myocardial Infarction
Keller, T., Zeller, T., Peetz, D., Tzikas, S., Roth, A., Czyz, E., Bickel, C., Baldus, S., Warnholtz, A., Fröhlich, M., et al. (2009) · New England Journal of Medicine · doi:10.1056/nejmoa0903515 · Cited by 1,150
Cardiac Troponin and Outcome in Acute Heart Failure
Peacock, W. F., De Marco, T., Fonarow, G. C., Diercks, D., Wynne, J., Apple, F. S., & Wu, A. H. B. (2008) · New England Journal of Medicine · doi:10.1056/nejmoa0706824 · Cited by 730
Latini, R., Masson, S., Anand, I. S., Missov, E., Carlson, M., Vago, T., Angelici, L., Barlera, S., Parrinello, G., Maggioni, A. P., et al. (2007) · Circulation · doi:10.1161/circulationaha.106.655076 · Cited by 709
NACB WRITING GROUP MEMBERS, Morrow, D. A., Cannon, C. P., Jesse, R. L., Newby, L. K., Ravkilde, J., Storrow, A. B., Wu, A. H. B., & Christenson, R. H. (2007) · Circulation · doi:10.1161/circulationaha.107.182882 · Cited by 610
Cited by
Third Universal Definition of Myocardial Infarction
Thygesen, K., Alpert, J. S., Jaffe, A. S., Simoons, M. L., Chaitman, B. R., & White, H. D. (2012) · Circulation · doi:10.1161/cir.0b013e31826e1058 · Cited by 2,985
Self-citationAnalytical Characteristics of High-Sensitivity Cardiac Troponin Assays
Apple, F. S., Collinson, P. O., & for the IFCC Task Force on Clinical Applications of Cardiac Biomarkers (2012) · Clinical Chemistry · doi:10.1373/clinchem.2011.165795 · Cited by 861
Third Universal Definition of Myocardial Infarction
Thygesen, K., Alpert, J. S., Jaffe, A. S., Simoons, M. L., Chaitman, B. R., White, H. D., & For the Universal Definition of MI, W. G. J. E. (2012) · Global Heart · doi:10.1016/j.gheart.2012.08.001 · Cited by 544
Self-citationBiomarkers in atrial fibrillation: a clinical review
Hijazi, Z., Oldgren, J., Siegbahn, A., Granger, C. B., & Wallentin, L. (2013) · European Heart Journal · doi:10.1093/eurheartj/eht024 · Cited by 298
Clinical Use of High-Sensitivity Cardiac Troponin in Patients With Suspected Myocardial Infarction
Twerenbold, R., Boeddinghaus, J., Nestelberger, T., Wildi, K., Rubini Gimenez, M., Badertscher, P., & Mueller, C. (2017) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2017.07.718 · Cited by 255
Metadata
- Authors
- Giannitsis, E., Kurz, K., Hallermayer, K., Jarausch, J., Jaffe, A. S., & Katus, H. A.
- Venue
- Clinical Chemistry
- Publisher
- Oxford University Press (OUP)
- Published
- 2010-02-01
- Volume
- 56
- Issue
- 2
- Pages
- 254-261
- ISSN
- 0009-9147 (ISSN-L)
- 1530-8561
- OpenAlex
- W2098105324
- Cited by
- 1,084
- References
- 24
- FWCI
- 35.39
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Analytical Validation of a High-Sensitivity Cardiac Troponin T Assay. https://app.cochat.ai/papers/10.1373/clinchem.2009.132654
- The paper
- Giannitsis, E., Kurz, K., Hallermayer, K., Jarausch, J., Jaffe, A. S., & Katus, H. A. (2010). Analytical Validation of a High-Sensitivity Cardiac Troponin T Assay. Clinical Chemistry. https://doi.org/10.1373/clinchem.2009.132654
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.