Isotope-Dilution Liquid Chromatography–Tandem Mass Spectrometry Candidate Reference Method for Total Testosterone in Human Serum
Open accessThe DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
BACKGROUND: We developed and evaluated a candidate reference measurement procedure (RMP) to standardize testosterone measurements, provide highly accurate and precise value assignments for the CDC Hormone Standardization Program, and ensure accurate and comparable results across testing systems and laboratories. METHODS: After 2 liquid/liquid extractions of serum with a combination of ethyl acetate and hexane, we quantified testosterone by isotope-dilution liquid chromatography–tandem mass spectrometry with electrospray ionization in the positive ion mode monitoring 289→97 m/z (testosterone) and 292→112 m/z (3C13 testosterone). We used calibrator bracketing and gravimetric measurements to give higher specificity and accuracy to serum value assignments. The candidate RMP was evaluated for accuracy by use of NIST-certified reference material SRM971 and validated by split-sample comparison to established RMPs. We evaluated intraassay and interassay imprecision, measurement uncertainty, potential interferences, and matrix effects. RESULTS: A weighted Deming regression comparison of the candidate RMP to established RMPs showed agreement with no statistical difference (slope 0.99, 95% CI 0.98–1.00, intercept 0.54, 95% CI −1.24 to 2.32) and a bias of ≤0.3% for NIST SRM971. The candidate RMP gave maximum intraassay, interassay, and total percent CVs of 1.5%, 1.4%, and 1.7% across the concentrations of testosterone typically found in healthy men and women. We tested structural analogs of testosterone and 125 serum samples and found no interferences with the measurement. CONCLUSIONS: This RMP for testosterone can serve as a higher-order standard for measurement traceability and can be used to provide an accuracy base to which routine methods can be compared in the CDC Hormone Standardization Program.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Bhasin, S., Cunningham, G. R., Hayes, F. J., Matsumoto, A. M., Snyder, P. J., Swerdloff, R. S., & Montori, V. M. (2006) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2005-2847 · Cited by 965
Low Serum Testosterone and Mortality in Older Men
Laughlin, G. A., Barrett-Connor, E., & Bergstrom, J. (2008) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2007-1792 · Cited by 798
Wang, C., Catlin, D. H., Demers, L. M., Starcevic, B., & Swerdloff, R. S. (2004) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2003-031287 · Cited by 580
Androgen therapy in women: An Endocrine Society clinical practice guideline
Wierman, M. E., Basson, R., Davis, S. R., Khosla, S., Miller, K. K., Rosner, W., & Santoro, N. (2006) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2006-1121 · Cited by 301
Introduction: Strategies to set global quality specifications in laboratory medicine
C. G. Fraser, A. K. (1999) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365519950185184 · Cited by 172
Vesper, H. W., Bhasin, S., Wang, C., Tai, S. S., Dodge, L. A., Singh, R. J., Nelson, J., Ohorodnik, S., Clarke, N. J., Salameh, W. A., et al. (2009) · Steroids · doi:10.1016/j.steroids.2009.01.004 · Cited by 152
Self-citationStandardization of testosterone measurements in humans
Vesper, H. W., & Botelho, J. C. (2010) · The Journal of Steroid Biochemistry and Molecular Biology · doi:10.1016/j.jsbmb.2010.03.032 · Cited by 113
Self-citation
Cited by
Therapeutic Drug Monitoring of Tacrolimus-Personalized Therapy: Second Consensus Report
Brunet, M., van Gelder, T., Åsberg, A., Haufroid, V., Hesselink, D. A., Langman, L., Lemaitre, F., Marquet, P., Seger, C., Shipkova, M., et al. (2019) · Therapeutic Drug Monitoring · doi:10.1097/ftd.0000000000000640 · Cited by 707
Nilsson, M. E., Vandenput, L., Tivesten, Å., Norlén, A.-K., Lagerquist, M. K., Windahl, S. H., Börjesson, A. E., Farman, H. H., Poutanen, M., Benrick, A., et al. (2015) · Endocrinology · doi:10.1210/en.2014-1890 · Cited by 336
Mass spectrometry and immunoassay: how to measure steroid hormones today and tomorrow
Taylor, A. E., Keevil, B., & Huhtaniemi, I. T. (2015) · European Journal of Endocrinology · doi:10.1530/eje-15-0338 · Cited by 320
LC–MS/MS analysis of steroids in the clinical laboratory
Keevil, B. G. (2016) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2016.04.009 · Cited by 149
Zhou, H., Wang, Y., Gatcombe, M., Farris, J., Botelho, J. C., Caudill, S. P., & Vesper, H. W. (2017) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-017-0529-x · Cited by 48
“In addition, measurement bias was evaluated following CLSI document EP9-A2 (25) using 40 single-donor serum materials with reference values assigned by higher-order reference methods listed in the Joint Committee for Traceability in Laboratory Medicine (JCTLM) database (NIST (26), University of Ghent (27, 28), and CDC…”
Self-citation
Metadata
- Authors
- Botelho, J. C., Shacklady, C., Cooper, H. C., Tai, S. S.-C., Uytfanghe, K. V., Thienpont, L. M., & Vesper, H. W.
- Venue
- Clinical Chemistry
- Publisher
- Oxford University Press (OUP)
- Published
- 2013-02-01
- Volume
- 59
- Issue
- 2
- Pages
- 372-380
- ISSN
- 0009-9147 (ISSN-L)
- 1530-8561
- OpenAlex
- W2106587026
- Cited by
- 95
- References
- 7
- FWCI
- 3.40
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Isotope-Dilution Liquid Chromatography–Tandem Mass Spectrometry Candidate Reference Method for Total Testosterone in Human Serum. https://app.cochat.ai/papers/10.1373/clinchem.2012.190934
- The paper
- Botelho, J. C., Shacklady, C., Cooper, H. C., Tai, S. S.-C., Uytfanghe, K. V., Thienpont, L. M., & Vesper, H. W. (2013). Isotope-Dilution Liquid Chromatography–Tandem Mass Spectrometry Candidate Reference Method for Total Testosterone in Human Serum. Clinical Chemistry. https://doi.org/10.1373/clinchem.2012.190934
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.