Reliability of Pulse Oximetry in Titrating Supplemental Oxygen Therapy in Ventilator-Dependent Patients
Abstract
Pulse oximetry is widely used in the critical care setting, but few studies have examined its usefulness in clinical decision making. One area where pulse oximetry might be useful is in the titration of fractional inspired O2 concentration (FIO2) in ventilator-dependent patients. Unfortunately, documented guidelines for this use do not exist, and in a survey of directors of intensive care units, we found that they employed a wide range of target O2 saturation (SpO2) values. Consequently, we undertook a study to determine if SpO2 could be reliably substituted for measurements of arterial O2 tension (PaO2) when adjusting FIO2 in ventilator-dependent patients. We examined a number of SpO2 target values in 54 critically ill patients aiming for a PaO2 of greater than or equal to 60 mm Hg, while minimizing the risk of O2 toxicity. In white patients, we found that a SpO2 target of 92 percent was reliable in predicting a satisfactory level of oxygenation. However, in black patients, such a SpO2 reading was commonly associated with significant hypoxemia (PaO2 as low as 49 mm Hg), and a higher SpO2 target, 95 percent, was required. In addition, inaccurate oximetry readings (ie, greater than 4 percent difference between SpO2 and direct SaO2 measurements) were more common in black (27 percent) than in white patients (11 percent, p less than 0.05). In conclusion, a SpO2 target of 92 percent was reliable when titrating supplemental O2 in white patients receiving mechanical ventilation; however, in black patients, such a SpO2 reading was commonly associated with significant hypoxemia, and a higher SpO2 target, 95 percent, was required to ensure a satisfactory level of oxygenation.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“A study about applying a pulse oximeter device in the titration of fractional inspired O2 concentration in ventilator-dependent patients revealed that a SpO2 reading of 92% was reliable when titrating supplemental O2 in light-skinned patients receiving mechanical ventilation. However, in dark-skinned patients, the same SpO2 value was commonly related to significant hypoxemia, and a higher SpO2”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [88]
In the citing paper
“A study performed in 1990 in ventilator-dependent patients when titrating supplemental O2 report inaccurate measurements in dark-skinned patients [88].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [88]
Citations
Cites
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Continuous or Nocturnal Oxygen Therapy in Hypoxemic Chronic Obstructive Lung Disease
NOCTURNAL OXYGEN THERAPY TRIAL GROUP (1980) · Annals of Internal Medicine · doi:10.7326/0003-4819-93-3-391 · Cited by 2,202
Harrison's Principles of Internal Medicine
Bigby, J. (1988) · Archives of Dermatology · doi:10.1001/archderm.1988.01670020093028 · Cited by 1,249
Yelderman, M., & New, W. (1983) · Anesthesiology · doi:10.1097/00000542-198310000-00015 · Cited by 488
Tremper, K. K., & Barker, S. J. (1989) · Anesthesiology · doi:10.1097/00000542-198901000-00019 · Cited by 453
Complications of Percutaneous Radial-artery Cannulation: An Objective Prospective Study in Man
Bedford, R. F., & Wollman, H. (1973) · Anesthesiology · doi:10.1097/00000542-197303000-00006 · Cited by 293
Respiratory Monitoring in the Intensive Care Unit
Tobin, M. J. (1988) · American Review of Respiratory Disease · doi:10.1164/ajrccm/138.6.1625 · Cited by 200
Self-citationADULT RESPIRATORY DISTRESS SYNDROME
Kollef, M. H. (1996) · Southern Medical Journal · doi:10.1097/00007611-199604000-00024 · Cited by 199
History of blood gas analysis.
Jw, S., & Astrup, P. (1987) · PubMed · Cited by 187
The Effect of Nail Polish on Pulse Oximetry
COT, C. J., Goldstein, E. A., Fuchsman, W. H., & Hoaglin, D. C. (1988) · Anesthesia & Analgesia · doi:10.1213/00000539-198807000-00013 · Cited by 171
History of blood gas analysis. VI. Oximetry
Severinghaus, J. W., & Astrup, P. B. (1986) · The Journal of Clinical Monitoring · doi:10.1007/bf02851177 · Cited by 153
Ries, A. L., Prewitt, L. M., & Johnson, J. J. (1989) · Chest · doi:10.1378/chest.96.2.287 · Cited by 145
Noninvasive Detection of Profound Arterial Desaturations Using a Pulse Oximetry Device
Mihm, F. G., & Halperin, B. D. (1985) · Anesthesiology · doi:10.1097/00000542-198501000-00020 · Cited by 142
The Accuracy and Response Characteristics of a Simplified Ear Oximeter
Rebuck, A. S., Chapman, K. R., & D’Urzo, A. D. (1983) · CHEST Journal · doi:10.1378/chest.83.6.860 · Cited by 127
Bias and Precision of Pulse Oximeters and Arterial Oximeters
Nickerson, B. G., Sarkisian, C. A., & Tremper, K. K. (1988) · CHEST Journal · doi:10.1378/chest.93.3.515 · Cited by 124
Pulse Oximetry: Technical Aspects of Machine Design
Pologe, J. A. (1987) · International Anesthesiology Clinics · doi:10.1097/00004311-198702530-00009 · Cited by 110
Accuracy of two ear oximeters at rest and during exercise in pulmonary patients.
Ries, A. L., Farrow, J. T., & Clausen, J. L. (1985) · PubMed · doi:10.1164/arrd.1985.132.3.685 · Cited by 106
Ear oximetry: accuracy and practicability in the assessment of arterial oxygenation.
Saunders, N. A., Powles, A. C. P., & Rebuck, A. S. (1976) · PubMed · doi:10.1164/arrd.1976.113.6.745 · Cited by 105
Efficacy of pulse oximetry and capnometry in postoperative ventilatory weaning
Niehoff, J., DelGuercio, C., LaMorte, W. W., HUGHES-GRASBERGER, S. L., Heard, S. O., Dennis, R. C., & Yeston, N. S. (1988) · Critical Care Medicine · doi:10.1097/00003246-198807000-00011 · Cited by 85
A clinical evaluation of the accuracy of the Nellcor N-100 and Ohmeda 3700 pulse oximeters
Cecil, W. T., Thorpe, K. J., Fibuch, E. E., & Tuohy, G. F. (1988) · Journal of Clinical Monitoring · doi:10.1007/bf01618105 · Cited by 51
ACCURACY OF A PULSE OXIMETER IN THE CRITICALLY ILL ADULT
Tremper, K. K., Hufstedler, S. M., Barker, S. J., Adams, A. L., Wong, D. H. W., Zaccari, J., Benik, K., & Lemons, V. (1985) · Anesthesiology · doi:10.1097/00000542-198509001-00175 · Cited by 37
The Effect of Fingerprinting Ink on Pulse Oximetry
Battito, M. F. (1989) · Anesthesia & Analgesia · doi:10.1213/00000539-198908000-00031 · Cited by 11
Research works printed but not linked: 1
- Berlin, 1988, Pulse oximetry: a technology that needs direction, Respir Care, 33, 243
Cited by
Sorted by citations
Rice, T. W., Wheeler, A. P., Bernard, G. R., Hayden, D. L., Schoenfeld, D., & Ware, L. B. (2007) · CHEST Journal · doi:10.1378/chest.07-0617 · Cited by 847
BTS guideline for emergency oxygen use in adult patients
O’Driscoll, B. R., Howard, L., Davison, A. G., & on behalf of the British Thoracic Society (2008) · Thorax · doi:10.1136/thx.2008.102947 · Cited by 727
BTS guideline for oxygen use in adults in healthcare and emergency settings
O’Driscoll, B. R., Howard, L., Earis, J. E., & Mak, V. H. F. (2017) · Thorax · doi:10.1136/thoraxjnl-2016-209729 · Cited by 710
Why COVID-19 Silent Hypoxemia Is Baffling to Physicians
Tobin, M. J., Laghi, F. A., & Jubran, A. (2020) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.202006-2157cp · Cited by 601
Self-citationJubran, A. (2015) · Critical Care · doi:10.1186/s13054-015-0984-8 · Cited by 511
Self-citation
Metadata
- Authors
- Jubran, A., & Tobin, M. J.
- Affiliations
- The University of Texas Health Science Center at Houston (US)
- Venue
- Chest
- Publisher
- Elsevier BV
- Published
- 1990-06
- Volume
- 97
- Issue
- 6
- Pages
- 1420-1425
- ISSN
- 0012-3692 (ISSN-L)
- OpenAlex
- W1996518979
- Cited by
- 296
- References
- 22
- Citations by year
- 2026: 16 · 2025: 18 · 2024: 25 · 2023: 29 · 2022: 54 · 2021: 11 · 2020: 8 · 2019: 4 · 2018: 4 · 2017: 3 · 2016: 5 · 2015: 5 · 2014: 13 · 2013: 4 · 2012: 6
- FWCI
- 7.56
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- CHEST Journal (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Respiratory Support and Mechanisms
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- pulse oximetry, supplemental oxygen, mechanical ventilation, arterial oxygen tension, hypoxemia, racial differences, arterial oxygen saturation, oxygen toxicity, critically ill patients
- MeSH terms
- Algorithms, Black or African American, Female, Humans, Intensive Care Units, Male, Middle Aged, Oximetry, Oxygen, Oxygen Inhalation Therapy, Partial Pressure, Ventilators, Mechanical, White People
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review