A clinical evaluation of the accuracy of the Nellcor N-100 and Ohmeda 3700 pulse oximeters
Abstract
Two pulse oximeters, the Nellcor N-100 and the Ohmeda 3700, were compared with arterial blood values and with each other in a clinical evaluation of performance. Three hundred twenty-nine simultaneously sampled blood/oximeter data pairs from use of both makes of pulse oximeters on each of 152 test subjects were included in the comparison analysis for each oximeter. Among the patients, disease type and severity and hospital location varied widely. Basic descriptive statistics and linear regression analysis were employed to facilitate comparison. Both oximeters displayed a statistically significant but clinically insignificant bias when compared with arterial blood oxyhemoglobin: -0.31 (P = 0.023) and 0.59 (P = 0.001) for the Ohmeda 3700 pulse oximeter and the Nellcor N-100 pulse oximeter, respectively. Relative to arterial blood oxyhemoglobin, the 95% tolerance intervals were +4.84 to -5.45 (10.3) for the Ohmeda 3700 and +6.94 to -5.76 (12.7) for the Nellcor N-100. Regressed against [oxyhemoglobin + carboxyhemoglobin + methemoglobin] as x, the Nellcor N-100 read y = 0.85(x) + 12.5, r = 0.83, P less than 0.0001, and the Ohmeda 3700 read y = 1.02(x) - 5.3, r = 0.86, P less than 0.0001.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
A pioneering study comparing two pulse oximeters on 152 patients showed a statistically significant loss of accuracy in SpO2 readings for dark-skinned patients for one device.
“A pioneering study to compare two pulse oximeters on 152 patients showed a statistically significant loss of accuracy in SpO2 readings for dark-skinned patients for one device [43].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [43]
The skin pigmentation effect on pulse oximetry accuracy has long been studied, reporting some contradictory conclusions.
“The skin pigmentation effect on pulse oximetry accuracy has long been studied reporting some contradictory conclusions.”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [43]
Other subjective scales were employed in 13 studies (35.14%).
“and other subjective scales were employed in 13 studies (35.14%) [30,42,43,44,45,50,52,56,72,78,80,87,88].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [43]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Yelderman, M., & New, W. (1983) · Anesthesiology · doi:10.1097/00000542-198310000-00015 · Cited by 488 · see citing works
Handbook of Tables for Probability and Statistics.
Jones, D. A., & Beyer, W. E. (1968) · Journal of the Royal Statistical Society Series C (Applied Statistics) · doi:10.2307/2985692 · Cited by 422 · see citing works
History of blood gas analysis.
Jw, S., & Astrup, P. (1987) · PubMed · Cited by 187 · see citing works
History of blood gas analysis. VII. Pulse oximetry
Severinghaus, J. W., & Honda, Y. (1987) · The Journal of Clinical Monitoring · doi:10.1007/bf00858362 · Cited by 180 · see citing works
Fanconi, S., Doherty, P., EDMONDS, J. F., Barker, G. A., & BOHN, D. J. (1985) · The Journal of Pediatrics · doi:10.1016/s0022-3476(85)80507-2 · Cited by 158 · see citing works
Statistics and ethics in medical research: V--Analysing data.
Altman, D. G. (1980) · BMJ · doi:10.1136/bmj.281.6253.1473 · Cited by 61 · see citing works
Comparison of a pulse oximeter with an ear oximeter and an in-vitro oximeter
Mackenzie, N. (1985) · The Journal of Clinical Monitoring · doi:10.1007/bf02832255 · Cited by 37 · see citing works
Yelderman, M., New, W., & Chalon, J. (1984) · Survey of Anesthesiology · doi:10.1097/00132586-198406000-00031 · Cited by 32 · see citing works
Flagged first, then by citations
1 of the 9 references couldn’t be matched to a known paper.
Research works printed but not linked: 4
- Shippy MB, Petterson MT, Whitman RA, Shivers CR. A clinical evaluation of the BTI Biox II ear oximeter. Respir Care 1984;29:730-735
- Cecil WT, Petterson MT, Lamoonpun S, Rudolph CD. Clinical evaluation of the Ohmeda Biox IIA ear oximeter in the critical care environment. Respir Care 1985;30:179- 183
- Cecil WT, Morrison LS, Lamoonpun S. Clinical evalua- tion of the Ohmeda Biox III pulse oximeter: a comparison of finger and ear cuvettes. Respir Care 1985;30:840-845
- Severinghaus JW. Discussion II. In: Payne JP, Severing- haus JW, eds. Pulse oximetry. London: Springer-Verlag 1986:45-53
Cited by
Sorted by citations
Pulse Oximetry: Theory and Applications for Noninvasive Monitoring
Mendelson, Y. (1992) · Clinical Chemistry · doi:10.1093/clinchem/38.9.1601 · Cited by 320 · see citing works
Jubran, A., & Tobin, M. J. (1990) · Chest · doi:10.1378/chest.97.6.1420 · Cited by 296 · see citing works
Oximetry of retinal vessels by dual-wavelength imaging: calibration and influence of pigmentation
Beach, J. M., Schwenzer, K. J., Srinivas, S. P., Kim, D., & Tiedeman, J. S. (1999) · Journal of Applied Physiology · doi:10.1152/jappl.1999.86.2.748 · Cited by 263 · see citing works
Meta-analysis of arterial oxygen saturation monitoring by pulse oximetry in adults
Jensen, L. A., Onyskiw, J. E., & Prasad, N. G. N. (1998) · Heart & Lung · doi:10.1016/s0147-9563(98)90086-3 · Cited by 243 · see citing works
Kelleher, J. F. (1989) · The Journal of Clinical Monitoring · doi:10.1007/bf01618369 · Cited by 205 · see citing works
Ralston, A., Webb, R. K., & Runciman, W. B. (1991) · Anaesthesia · doi:10.1111/j.1365-2044.1991.tb11501.x · Cited by 190 · see citing works
Neonatal and Pediatric Pulse Oximetry
Salyer, J. W. (2003) · Respiratory Care · doi:10.4187/respcare.03480386 · Cited by 183 · see citing works
Valbuena, V. S. M., Seelye, S., Sjoding, M. W., Valley, T. S., Dickson, R. P., Gay, S. E., Claar, D., Prescott, H. C., & Iwashyna, T. J. (2022) · BMJ · doi:10.1136/bmj-2021-069775 · Cited by 127 · see citing works
Cabanas, A. M., Fuentes-Guajardo, M., Latorre, K., León, D., & Martín-Escudero, P. (2022) · Sensors · doi:10.3390/s22093402 · Cited by 124 · see citing works
A review of the effect of skin pigmentation on pulse oximeter accuracy
Al-Halawani, R., Charlton, P. H., Qassem, M., & Kyriacou, P. A. (2023) · Physiological Measurement · doi:10.1088/1361-6579/acd51a · Cited by 119 · see citing works
Validity of pulse oximetry during exercise in elite endurance athletes
Martin, D., Powers, S. K., Cicale, M. J., Collop, N. A., Huang, D. T.-C., & Criswell, D. S. (1992) · Journal of Applied Physiology · doi:10.1152/jappl.1992.72.2.455 · Cited by 84 · see citing works
Martín, D., Johns, C., Sorrell, L., Healy, E., Phull, M., Olusanya, O., Peters, M. J., & Fabes, J. (2024) · British Journal of Anaesthesia · doi:10.1016/j.bja.2024.01.023 · Cited by 80 · see citing works
Effects of acute hypoxia on theVO2max of trained and untrained subjects
Martin, D., & OʼKroy, J. A. (1993) · Journal of Sports Sciences · doi:10.1080/02640419308729961 · Cited by 74 · see citing works
Continuous Noninvasive Hemoglobin Monitoring
Barker, S. J., Shander, A., & Ramsay, M. A. E. (2015) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000000605 · Cited by 72 · see citing works
Pulse Oximetry Is Accurate in Acute Anemia From Hemorrhage
Jay, G. D., Hughes, L., & Renzi, F. P. (1994) · Annals of Emergency Medicine · doi:10.1016/s0196-0644(94)70158-x · Cited by 59 · see citing works
Materialized Oppression in Medical Tools and Technologies
Liao, S., & Carbonell, V. (2022) · The American Journal of Bioethics · doi:10.1080/15265161.2022.2044543 · Cited by 57 · see citing works
Pulse Oximetry: The Working Principle, Signal Formation, and Applications
Leppänen, T., Kainulainen, S., Korkalainen, H., Sillanmäki, S., Kulkas, A., Töyräs, J., & Nikkonen, S. (2022) · Advances in experimental medicine and biology · doi:10.1007/978-3-031-06413-5_12 · Cited by 40 · see citing works
A Novel Hemoglobin, Bonn, Causes Falsely Decreased Oxygen Saturation Measurements in Pulse Oximetry
Zur, B., Hornung, A., Breuer, J., Doll, U., Bernhardt, C., Ludwig, M. Z., & Stoffel‐Wagner, B. (2008) · Clinical Chemistry · doi:10.1373/clinchem.2007.095158 · Cited by 39 · see citing works
Reliability of Six Pulse Oximeters in Chronic Obstructive Pulmonary Disease
Hannhart, B., Michalski, H., Delorme, N., Chapparo, G., & Polu, J. M. (1991) · CHEST Journal · doi:10.1378/chest.99.4.842 · Cited by 39 · see citing works
Version Reporting and Assessment Approaches for New and Updated Activity and Heart Rate Monitors
Collins, T., Woolley, S. I., Oniani, S., Pires, I. M., García, N. M., Ledger, S., & Pandyan, A. D. (2019) · Sensors · doi:10.3390/s19071705 · Cited by 38 · see citing works
An evaluation of pulse oximetry in prehospital care
Aughey, K., HESS, D. R., Eitel, D. R., Bleecher, K., Cooley, M., Ogden, C. S., & Sabulsky, N. K. (1991) · Annals of Emergency Medicine · doi:10.1016/s0196-0644(05)81432-5 · Cited by 33 · see citing works
A prospective study of intraoperative pulse oximetry failure
Freund, P. R., Overand, P. T., Cooper, J., Jacobson, L., Bosse, S., Walker, B., Posner, K. L., & Cheney, F. W. (1991) · The Journal of Clinical Monitoring · doi:10.1007/bf01619270 · Cited by 32 · see citing works
Ninety years of pulse oximetry: history, current status, and outlook
Quaresima, V., Ferrari, M., & Scholkmann, F. (2024) · Journal of Biomedical Optics · doi:10.1117/1.jbo.29.s3.s33307 · Cited by 28 · see citing works
Noninvasive Monitoring of Oxygen and Carbon Dioxide
Wiedemann, H. P., & McCarthy, K. J. (1989) · Clinics in Chest Medicine · doi:10.1016/s0272-5231(21)00625-0 · Cited by 26 · see citing works
Pulse oximetry and finger blood pressure measurement during open-heart surgery
Kurki, T. S. O., Smith, N. T., Sanford, T. J., & HEAD, N. (1989) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf01618251 · Cited by 20 · see citing works
Pälve, H., & Vuori, A. (1991) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1016/1053-0770(91)90154-l · Cited by 19 · see citing works
Reliability Assessment of New and Updated Consumer-Grade Activity and Heart Rate Monitors
Oniani, S., Woolley, S. I., Pires, I. M., García, N. M., Collins, T., Ledger, S., & Pandyan, A. D. (2018) · Keele Research Repository (Keele University) · doi:10.13140/rg.2.2.35628.72328 · Cited by 18 · see citing works
A clinical evaluation of pulse oximetry during thoracic surgery
Desiderio, D. P., Wong, G. Y., Shah, N. K., Liu, J., Loughlin, C. J., & Bedford, R. F. (1990) · Journal of Cardiothoracic Anesthesia · doi:10.1016/0888-6296(90)90443-j · Cited by 15 · see citing works
Monitorage continu de la saturation du sang veineux mêlé en oxygène
Journois, D., & Safran, D. (1993) · Annales Françaises d Anesthésie et de Réanimation · doi:10.1016/s0750-7658(05)80107-8 · Cited by 10 · see citing works
Oxygenation during local anaesthesia for cataract surgery
Pälve, H., & Ali‐Melkkilä, T. (1991) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.1991.tb03269.x · Cited by 10 · see citing works
Martin, D., Day, J. A., Ward, G., Carter, E. P., & Chesrown, S. E. (1994) · Pediatric Pulmonology · doi:10.1002/ppul.1950180403 · Cited by 8 · see citing works
Plaisime, M. V. (2023) · American Journal of Epidemiology · doi:10.1093/aje/kwad019 · Cited by 8 · see citing works
Pulse Oximetry: Instrumentation and Clinical Applications
Welch, J. P., DeCesare, R., & HESS, D. R. (1990) · Respiratory Care · doi:10.1177/194336549003500608 · Cited by 7 · see citing works
Case–control studies of novel hemoglobin anomalies as differential diagnosis in sleep apnea syndrome
Zur, B., Ludwig, M. Z., & Stoffel‐Wagner, B. (2012) · Sleep And Breathing · doi:10.1007/s11325-012-0720-4 · Cited by 6 · see citing works
Basing intubation of acutely hypoxemic patients on physiologic principles
Laghi, F. A., Shaikh, H., & Caccani, N. (2024) · Annals of Intensive Care · doi:10.1186/s13613-024-01327-w · Cited by 5 · see citing works
Failure rate of pulse oximetry in the postanesthesia care unit
Gillies, B. S., Posner, K. L., Freund, P. R., & Cheney, F. W. (1993) · The Journal of Clinical Monitoring · doi:10.1007/bf01618673 · Cited by 5 · see citing works
Clinician behavior when skin tone affects test results
Alsan, M. M., Liran Einav, B., Finkelstein, A. N., & Zhang, J. (2026) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.2530585123 · Cited by 2 · see citing works
Pulse oximetry in patients with pigmented skin: What I should know
Brookman, S., Mukadam, T., Owasil, S., Thachettu, A., Urquhart, D. S., Dhawan, A. K., & Gupta, A. K. (2024) · Paediatric Respiratory Reviews · doi:10.1016/j.prrv.2024.06.001 · Cited by 2 · see citing works
Design and Implementation of Pulse Oximeter Synchronized with Respirators
Baloglu, S., Saritas, I. E., & Gölcük, A. (2022) · International Scientific Conference on Information, Communication and Energy Systems and Technologies (ICEST) · doi:10.1109/icest55168.2022.9828719 · Cited by 1 · see citing works
Duench, S. A. (2009) · UWSpace (University of Waterloo) · Cited by 1 · see citing works
Low pulse oximeter-measured hemoglobin oxygen saturations with hemoglobin Cheverly
Hohl, R. J., Sherburne, A. R., Feeley, J. E., Huisman, T. H. J., & Burns, C. P. (1998) · American Journal of Hematology · doi:10.1002/(sici)1096-8652(199811)59:3
Juvé, J. I., Roca, J. V., & Puigdevall, J. M. R. (2004) · Elsevier eBooks · doi:10.1016/b978-84-8174-708-9.50017-7
Chan, K.-C., Aitken, A., & Leung, M. P. (1991) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1016/1053-0770(91)90018-o
Cecil, W. T. (1988) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf01621825
Self-citationAOYAGI, T. (1998) · JAPANES JOURNAL OF MEDICAL INSTRUMENTATION · doi:10.4286/ikakikaigaku.68.8_315
Barker, S. J. (2013) · Elsevier eBooks · doi:10.1016/b978-0-323-11237-6.00011-x
Hemkumar, A., Hermann, J., Wilson, B., & Rajbhandary, P. L. (2025) · Design and Quality for Biomedical Technologies · doi:10.1117/12.3042294
Pilot Study: Using Reflectance as a Proxy for Skin Pigmentation to Assess Pulse Oximeter Accuracy
Branstad, A., Buttar, S., Kabir, R., Byrapuneni, S., Eid, M., Sandozi, A., BELTRAND, E., Evans, M. D., Joy, B. F., Goertzen, L., et al. (2026) · Hospital Pediatrics · doi:10.1542/hpeds.2025-008471
Effects of Skin Pigmentation on Pulse Oximetry: A Pilot Study
Cotton, S. A., Dahdoul, T., Best, N. C., Guo, Y., Lee, J.-A., Malhotra, A., & McGuire, W. C. (2026) · American Journal of Critical Care · doi:10.4037/ajcc2026253
- Authors
- Cecil, W. T., Thorpe, K. J., Fibuch, E. E., & Tuohy, G. F.
- Affiliations
- Saint Luke's Hospital (US)
- ORCID
- Cecil, William T.
- Venue
- Journal of Clinical Monitoring
- Published
- 1988-01
- Volume
- 4
- Issue
- 1
- Pages
- 31-36
- ISSN
- 1387-1307 (ISSN-L)
- 0748-1977
- 1573-2614
- OpenAlex
- W2006936900
- Cited by
- 51
- References
- 9
- Citations by year
- 2026: 3 · 2025: 1 · 2024: 4 · 2023: 2 · 2022: 5 · 2019: 1 · 2018: 1 · 2015: 1 · 2013: 1 · 2012: 1
- FWCI
- 6.75
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.springer.com/tdm (tdm)
- Locations
- Journal of Clinical Monitoring and Computing (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
- Keywords
- pulse oximetry, arterial oxygen saturation, oxyhemoglobin, carboxyhemoglobin, methemoglobin, measurement bias, tolerance intervals, device comparison
- MeSH terms
- Female, Humans, Male, Middle Aged, Monitoring, Physiologic, Oximetry, Regression Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review