Skin color and ear oximetry
Abstract
Measurements of arterial blood oxygen saturation from two ear oximeters were compared with 655 simultaneously drawn arterial blood samples in 187 patients grouped by skin color quantified by the Munsell color system. Technical problems including warning lights and messages with the two ear oximeters were recorded. There were significantly more technical problems in patients with the darkest skin color associated with inability to obtain a reading or warning message indicating poor tissue penetration of the signal (18 and 15 percent vs 1 percent). When readings could be obtained, the ear oximetry readings were found to be slightly less accurate in the darker patient groups. These findings suggest that dark skin color may affect the performance and accuracy of ear oximeters, including the newer type of pulse oximeters.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Inaccurate PPG heart rate measurements occur up to 15% more frequently in dark skin than in light skin.
“previous research demonstrated that inaccurate PPG HR measurements occur up to 15% more frequently in dark skin as compared to light skin”
From Investigating sources of inaccuracy in wearable optical heart rate sensors (2020) · cites this paper as [27]
Darker skin contains more melanin, which absorbs more green light than lighter skin.
“likely because darker skin contains more melanin which absorbs more green light than lighter skin.”
From Investigating sources of inaccuracy in wearable optical heart rate sensors (2020) · cites this paper as [27]
Darker skin pigmentation may influence PPG-based SpO2 and PR readings.
“Notably, darker skin pigmentation may influence PPG-based SpO2 and PR readings [7-10].”
From Impact of skin pigmentation on pulse oximetry blood oxygenation and wearable pulse rate accuracy: systematic review and meta-analysis (2024) · cites this paper as [7]
Objective scales such as the Fitzpatrick Skin Pigmentation (FSP) scale or the Munsell scale were used by 6/37 studies (16.21%) and 4/37 studies (10.81%), respectively.
“We find that objective scales such as the Fitzpatrick Skin Pigmentation (FSP) scale or the Munsell scale were used by 6/37 studies (16.21%) [61,69,70,71,76,77] and 4/37 studies (10.81%) [47,48,74,89], respectively.”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [89]
Another study in 187 pulmonary patients undergoing a stress test reported technical problems in patients with slightly less accurate readings.
“Besides, another study in 187 pulmonary patients undergoing a stress test reported technical problems in patients with slightly less accurate readings [89].”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [89]
Cites
The Accuracy and Response Characteristics of a Simplified Ear Oximeter
Rebuck, A. S., Chapman, K. R., & D’Urzo, A. D. (1983) · CHEST Journal · doi:10.1378/chest.83.6.860 · Cited by 127 · see citing works
Evaluation of the Ohmeda 3700 Pulse Oximeter
Kagle, D. M., Alexander, C. M., Berko, R. S., et al. (1987) · Anesthesiology · doi:10.1097/00000542-198703000-00018 · Cited by 109 · see citing works
Accuracy of two ear oximeters at rest and during exercise in pulmonary patients.
Ries, A. L., Farrow, J. T., & Clausen, J. L. (1985) · PubMed · doi:10.1164/arrd.1985.132.3.685 · Cited by 106 · see citing works
Self-citationEar oximetry: accuracy and practicability in the assessment of arterial oxygenation.
Saunders, N. A., Powles, A. C. P., & Rebuck, A. S. (1976) · PubMed · doi:10.1164/arrd.1976.113.6.745 · Cited by 105 · see citing works
Ear oximetry during combined hypoxia and exercise
Smyth, R. J., D'Urzo, A. D., Slutsky, A. S., Galko, B. M., & Rebuck, A. S. (1986) · Journal of Applied Physiology · doi:10.1152/jappl.1986.60.2.716 · Cited by 80 · see citing works
Validity of Ear Oximetry in Clinical Exercise Testing
Hansen, J. E., & Casaburi, R. H. (1987) · CHEST Journal · doi:10.1378/chest.91.3.333 · Cited by 74 · see citing works
Douglas, N. J., Brash, H. M., WRAITH, P. K., et al. (1979) · PubMed · doi:10.1164/arrd.1979.119.2.311 · Cited by 69 · see citing works
Ear oximetry in clinical practice.
Chaudhary, B. A., & Burki, N. K. (1978) · PubMed · doi:10.1164/arrd.1978.117.1.173 · Cited by 52 · see citing works
Evaluation of Biox IIA ear oximeter.
Tweeddale, P. M., & Douglas, N. J. (1985) · Thorax · doi:10.1136/thx.40.11.825 · Cited by 33 · see citing works
Noninvasive oximetry in pigmented patients.
Wang, Y. T., & Poh, S. C. (1985) · PubMed · Cited by 10 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
ATS/ACCP statement on cardiopulmonary exercise testing
(2003) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.167.2.211 · Cited by 3,026 · see citing works
Investigating sources of inaccuracy in wearable optical heart rate sensors
Bent, B., Goldstein, B. A., Kibbe, W. A., & Dunn, J. P. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0226-6 · Cited by 673 · see citing works
Effects of skin pigmentation on pulse oximeter accuracy at low saturation
Bickler, P. E., Feiner, J. R., & Severinghaus, J. W. (2005) · Anesthesiology · doi:10.1097/00000542-200504000-00004 · Cited by 427 · see citing works
Jubran, A., & Tobin, M. J. (1990) · Chest · doi:10.1378/chest.97.6.1420 · Cited by 296 · see citing works
Meta-analysis of arterial oxygen saturation monitoring by pulse oximetry in adults
Jensen, L. A., Onyskiw, J. E., & Prasad, N. G. N. (1998) · Heart & Lung · doi:10.1016/s0147-9563(98)90086-3 · Cited by 243 · see citing works
Shi, C., Goodall, M., Dumville, J., Hill, J., Norman, G., Hamer, O., Clegg, A., Watkins, C. L., Georgiou, G., Hodkinson, A., et al. (2022) · doi:10.1101/2022.02.16.22271062 · Cited by 158 · see citing works
A Review of Wearable Multi-Wavelength Photoplethysmography
Ray, D., Collins, T., Woolley, S., & Ponnapalli, P. (2023) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2021.3121476 · Cited by 155 · see citing works
Valbuena, V. S. M., Seelye, S., Sjoding, M. W., Valley, T. S., Dickson, R. P., Gay, S. E., Claar, D., Prescott, H. C., & Iwashyna, T. J. (2022) · BMJ · doi:10.1136/bmj-2021-069775 · Cited by 127 · see citing works
Ensuring that biomedical AI benefits diverse populations
Zou, J., & Schiebinger, L. (2021) · EBioMedicine · doi:10.1016/j.ebiom.2021.103358 · Cited by 124 · see citing works
Cabanas, A. M., Fuentes-Guajardo, M., Latorre, K., León, D., & Martín-Escudero, P. (2022) · Sensors · doi:10.3390/s22093402 · Cited by 124 · see citing works
Racial discrepancy in pulse oximeter accuracy in preterm infants
Vesoulis, Z., Tims, A., Lodhi, H., Lalos, N., & Whitehead, H. (2022) · Journal of Perinatology · doi:10.1038/s41372-021-01230-3 · Cited by 120 · see citing works
Effect of skin pigmentation on pulse oximetry accuracy in the emergency department
Adler, J. N., Hughes, L. A., Vtvilecchia, R., & Jr., C. A. C. (1998) · Academic Emergency Medicine · doi:10.1111/j.1553-2712.1998.tb02772.x · Cited by 102 · see citing works
The Effect of Skin Pigmentation on the Accuracy of Pulse Oximetry in Infants with Hypoxemia
Foglia, E. E., Whyte, R. K., Chaudhary, A., Mott, A., Chen, J., Propert, K. J., & Schmidt, B. (2017) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2016.11.043 · Cited by 96 · see citing works
Ajmal, Boonya-Ananta, T., Rodriguez, A. J., Du Le, V. N., & Ramella-Roman, J. C. (2021) · Biomedical Optics Express · doi:10.1364/boe.439893 · Cited by 79 · see citing works
Singh, S., Bennett, M. R., Chen, C., Shin, S., Ghanbari, H., & Nelson, B. W. (2024) · Journal of Medical Internet Research · doi:10.2196/62769 · Cited by 36 · see citing works
Gaskin, L., & Thomas, J. (1995) · Physiotherapy · doi:10.1016/s0031-9406(05)66819-8 · Cited by 14 · see citing works
Haxha, S., Nwibor, C., Ali, M., Sakel, M., Saunders, K., Dyo, V., & Nabakooza, S. (2024) · Sensors · doi:10.3390/s24113301 · Cited by 12 · see citing works
- Authors
- Ries, A. L., Prewitt, L. M., & Johnson, J. J.
- Affiliations
- University of California San Diego (US)
- Corresponding author
- Ries, Andrew L.
- ORCID
- Ries, Andrew L.
- Venue
- Chest
- Publisher
- Elsevier BV
- Published
- 1989-08
- Volume
- 96
- Issue
- 2
- Pages
- 287-290
- ISSN
- 0012-3692 (ISSN-L)
- OpenAlex
- W1966473319
- Cited by
- 145
- References
- 10
- Citations by year
- 2026: 6 · 2025: 8 · 2024: 13 · 2023: 7 · 2022: 37 · 2021: 7 · 2020: 6 · 2019: 1 · 2018: 2 · 2017: 1 · 2016: 3 · 2015: 1 · 2014: 1 · 2013: 1 · 2012: 1
- FWCI
- 6.00
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- CHEST Journal (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Center for Research Resources (RR00827)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
- Keywords
- ear oximetry, skin color, dark skin pigmentation, arterial oxygen saturation, pulse oximetry, Munsell color system, measurement accuracy, technical errors
- MeSH terms
- Blood Gas Monitoring, Transcutaneous, Ear, External, Exercise Test, Humans, Lung Diseases, Prospective Studies, Skin Pigmentation
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review