Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis
Open accessAbstract
Nowadays, pulse oximetry has become the standard in primary and intensive care units, especially as a triage tool during the current COVID-19 pandemic. Hence, a deeper understanding of the measurement errors that can affect precise readings is a key element in clinical decision-making. Several factors may influence the accuracy of pulse oximetry, such as skin color, body temperature, altitude, or patient movement. The skin pigmentation effect on pulse oximetry accuracy has long been studied reporting some contradictory conclusions. Recent studies have shown a positive bias in oxygen saturation measurements in patients with darkly pigmented skin, particularly under low saturation conditions. This review aims to study the literature that assesses the influence of skin pigmentation on the accuracy of these devices. We employed the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) statement to conduct a systematic review retrospectively since February 2022 using WOS, PubMed, and Scopus databases. We found 99 unique references, of which only 41 satisfied the established inclusion criteria. A bibliometric and scientometrics approach was performed to examine the outcomes of an exhaustive survey of the thematic content and trending topics.
AI snapshot
Written from the paper’s full text.
TL;DR
This PRISMA systematic review and bibliometric analysis of 41 studies (from 99 unique references) finds that pulse oximetry often overestimates oxygen saturation in dark-skinned patients, especially at low saturation, increasing occult hypoxemia risk. It also documents heterogeneous skin-pigmentation classification and risk-of-bias concerns across the literature.
Key findings
- The review included 41 studies from 31 sources published between 1976 and 2022; among 11/35 studies at low risk of bias, 8/11 (72.73%) reported inaccuracies related to skin pigmentation, while 3/11 (27.27%) reported no demonstrable loss of accuracy.
- Overestimations in dark-skinned patients were reported especially at low saturation: Feiner et al. found bias up to 8% at lower saturation, Bickler et al. found overestimation increasing linearly as SpO2 decreased, and Sjoding et al. found dark-skinned patients with SpO2 92–96% had arterial oxygen saturation <88% 11.7% of the time versus 3.6% in light-skinned patients (3 times higher).
- Henry et al. reported SpO2 overestimating arterial blood gas-derived oxygen saturation by 1.57% on average, and Wiles et al. reported biases of 0.28%, −0.33%, and −0.75% for “White,” “Asian,” and “Black” patients, respectively.
- Skin pigmentation was classified by Fitzpatrick scale in 6/37 studies (16.21%), Munsell scale in 4/37 (10.81%), EEL reflectance spectrophotometer in 1 (2.70%), ancestry descriptors in 11 (29.73%), and other subjective scales in 13 (35.14%).
Methods
- PRISMA-based retrospective systematic review of PubMed, Web of Science, and Scopus from January 1975 to February 2022, using PICO-based research questions; 99 unique references were found and 41 satisfied inclusion criteria.
- Bibliometric and scientometric analysis used Bibliometrix R-package/Biblioshiny and VOSviewer for keyword co-occurrence and cluster analysis; extracted data included author, year, journal, citations, author keywords, and research outputs.
- Risk of bias was assessed with QUADAS-2 for 35/41 studies, excluding reviews, letters, and studies lacking required clinical/experimental procedure or data; robvis was used for visualization.
Limitations
- Stated: there is no standard risk-of-bias tool for methods-comparison studies, so QUADAS-2 was used; many included studies classified skin pigmentation by unstandardized or qualitative judgments such as “dark,” “black,” “light,” or “white,” and some older negative studies may be misleading due to small dark-skinned samples or weak pigmentation correlation.
- Stated: most pulse oximeters have been calibrated using light-skinned individuals, not all studies reported the selected characteristics, and more studies with clinical/patient-reported data and standardized skin pigmentation scales are required.
- Evident: as a systematic review and bibliometric analysis, the paper does not pool effect sizes or perform a meta-analysis of accuracy across studies; its conclusions rely on qualitative synthesis of heterogeneous designs, populations, and oximeter models.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 154 cited sources checked · 0 supported · 0 flagged
- Findingown data
9/35 studies (25.71%) were at high risk of bias for one domain.
“We found 9/35 studies (25.71%) to be at high risk of bias for one domain [17,21,44,45,51,79,82,87,88].”
Quote in the paperNot a substantial claim - Findingreasoning
6 studies were in both groups: at high risk of bias for one domain and unclear risk of bias for at least two domains.
“We also found that 6 of these studies were in both groups (at high risk of bias for one domain and unclear risk of bias for at least two domains) [17,44,79,82,87,88].”
Quote in the paperNot a substantial claim - Findingown data
The review identified 41 studies that fulfilled the inclusion criteria, from 31 different sources between 1976 and 2022.
“We identified 41 studies that fulfilled the inclusion criteria from 31 different sources between 1976 and 2022.”
- Also stated in: § Abstract · plain
“of which only 41 satisfied the established inclusion criteria”
Quote in the paperNot a substantial claim - Also stated in: § Abstract · plain
- Findingcitation
Among the 11/35 low risk studies, 3/11 (27.27%) reported no demonstrable loss of accuracy caused by skin pigmentation.
“3/11 (27.27%) reported no demonstrable loss of accuracy caused by skin pigmentation [46,48,70]”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [46] Bothma, 1996, Accuracy of pulse oximetry in pigmented patients, S. Afr. Med. J.
Not checked in the free run.
- cites [48] Foglia, 2017, The Effect of Skin Pigmentation on the Accuracy of Pulse Oximetry in Infants with Hypoxemia, J. Pediatr. source
Not checked in the free run.
- cites [70] Allado, 2021, Innovative measurement of routine physiological variables (heart rate, respiratory rate and oxygen saturation) using a remote photoplethysmography imaging system: A prospective comparative trial protocol, BMJ Open source
Not checked in the free run.
- Findingown data
11/35 (31.43%) studies were at low risk of bias for at least three domains.
“Finally, 11/35 (31.43%) studies were found to be at low risk of bias for at least three domains [23,38,46,48,69,70,71,77,78,84,89].”
Quote in the paperNot a substantial claim
- Findingown data
The main reason for high risk of bias was selection of patients, where skin pigmentation levels were classified by an unstandardized or qualitative judgment such as "dark", "black", "light" or "white".
“The main reason for this bias is attributed to the selection of patients, where the skin pigmentation levels were classified by an unstandardized or qualitative judgment such as “dark”, “black”, “light” or “white”.”
Quote in the paperSubstantial claim - Findingown data
21/35 studies (60.00%) were at unclear risk of bias for at least two domains.
“Additionally, 21/35 studies (60.00%) were classified to be at unclear risk of bias for at least two domains [17,24,30,41,42,43,44,47,50,53,56,72,75,76,79,80,81,82,87,88,93].”
Quote in the paperNot a substantial claim - Findingown data
Other subjective scales were employed in 13 studies (35.14%).
“and other subjective scales were employed in 13 studies (35.14%) [30,42,43,44,45,50,52,56,72,78,80,87,88].”
Quote in the paperNot a substantial claim0 of 10 sources checked- cites [30] Al-Naji, A., Khalid, G.A., Mahdi, J.F., and Chahl, J. (2021). Non-Contact SpO2 Prediction System Based on a Digital Camera. Appl. Sci., 11. source
Not checked in the free run.
- cites [42] Emery, 1987, Skin pigmentation as an influence on the accuracy of pulse oximetry, J. Perinatol. Off. J. Calif. Perinat. Assoc.
Not checked in the free run.
- cites [43] Cecil, 1988, A clinical evaluation of the accuracy of the Nellcor N-100 and Ohmeda 3700 pulse oximeters, J. Clin. Monit. source
Not checked in the free run.
- cites [44] Wang, 1985, Noninvasive oximetry in pigmented patients, Ann. Acad. Med.
Not checked in the free run.
- cites [45] Gabrielczyk, 1988, Pulse oximetry and postoperative hypothermia, Anaesthesia source
Not checked in the free run.
- cites [50] Sjoding, 2020, Racial Bias in Pulse Oximetry Measurement, N. Engl. J. Med. source
Not checked in the free run.
- cites [52] Okunlola, 2022, Pulse Oximeter Performance, Racial Inequity, and the Work Ahead, Respir. Care source
Not checked in the free run.
- cites [56] Avant, 1997, G; Lowe N.; Torres Jr., A. Comparison of accuracy and signal consistency of two reusable pulse oximeter probes in critically ill children, Respir. Care
Not checked in the free run.
- cites [72] Verkruysse, 2021, Pulse oximetry based on photoplethysmography imaging with red and green light: Calibratability and challenges, J. Clin. Monit. Comput. source
Not checked in the free run.
- cites [78] Kumar, 2015, DistancePPG: Robust non-contact vital signs monitoring using a camera, Biomed. Opt. Express source
Not checked in the free run.
- Findingown data
Ancestry descriptors were used in 11 studies (29.73%).
“On the other hand, ancestry descriptors were used in 11 studies (29.73%) [17,21,23,24,41,51,53,61,75,82,93]”
Quote in the paperNot a substantial claim0 of 10 sources checked- cites [17] Wiles, 2021, The effect of patient ethnicity on the accuracy of peripheral pulse oximetry in patients with COVID-19 pneumonitis: A single-centre, retrospective analysis, Anaesthesia source
Not checked in the free run.
- cites [21] Mendelson, 1988, Noninvasive pulse oximetry utilizing skin reflectance photoplethysmography, IEEE Trans. Biomed. Eng. source
Not checked in the free run.
- cites [23] Bickler, 2005, Effects of skin pigmentation on pulse oximeter accuracy at low saturation, Anesthesiology source
Not checked in the free run.
- cites [24] Feiner, 2007, Dark skin decreases the accuracy of pulse oximeters at low oxygen saturation: The effects of oximeter probe type and gender, Anesth. Analg. source
Not checked in the free run.
- cites [41] Saunders, 1976, Ear oximetry: Accuracy and tracticability in the assessment of arterial oxygenation, Am. Rev. Respir. Dis.
Not checked in the free run.
- cites [51] Vesoulis, 2021, Racial discrepancy in pulse oximeter accuracy in preterm infants, J. Perinatol. source
Not checked in the free run.
- cites [53] Henry, 2022, Disparities in Hypoxemia Detection by Pulse Oximetry Across Self-Identified Racial Groups and Associations with Clinical Outcomes, Crit. Care Med. source
Not checked in the free run.
- cites [61] Shi, C., Goodall, M., Dumville, J., Hill, J.E., Norman, G., Hamer, O., Clegg, A., Watkins, C.L., Georgiou, G., and Alexander Hodkinson, A. (2022). Article The effects of skin pigmentation on the accuracy of pulse oximetry in measuring oxygen saturation: A systematic review and meta-analysis The effects of skin pigmentation on the accuracy of pulse and meta-analysis. PLoS Med. source
Not checked in the free run.
- cites [75] Baek, 2018, The Effect of Optical Crosstalk on Accuracy of Reflectance-Type Pulse Oximeter for Mobile Healthcare, J. Healthc. Eng. source
Not checked in the free run.
- cites [82] Wouters, 2002, Accuracy of pulse oximeters: The European multi-center trial, Anesth. Analg.
Not checked in the free run.
- Premisecitation
In low saturation conditions, a 10:1 variation in LED intensity can lead to an error of 2.5%.
“Another study to determine the effect of varying LED intensity on pulse oximeter accuracy showed that in low saturation conditions, a 10:1 variation in LED intensity can lead to an error of 2.5% [103].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [103] Reynolds, 1991, The effect of varying LED intensity on pulse oximeter accuracy, J. Med. Eng. Technol. source
Not checked in the free run.
- Premisecitation
Melanin attenuates the wavelength of incident light and limits penetration to subcutaneous tissue because it is highly light-absorbing.
“Melanin attenuates the wavelength of the incident light and limits the penetration to the subcutaneous tissue because is high light-absorbing [80,101].”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [80] Fallow, 2013, Influence of skin type and wavelength on light wave reflectance, J. Clin. Monit. Comput. source
Not checked in the free run.
- cites [101] Fine, J., Branan, K.L., Rodriguez, A.J., Boonya-ananta, T., Ramella-Roman, J.C., McShane, M.J., and Coté, G.L. (2021). Sources of Inaccuracy in Photoplethysmography for Continuous Cardiovascular Monitoring. Biosensors, 11. source
Not checked in the free run.
- Rests on an assumption: Dark skin contains more melanin.
- Premisecitation
In M.W. Sjoding et al., the probability of dark-skinned patients with SpO2 reading between 92% and 96% having arterial oxygen saturation less than 88% was 3 times higher (11.7%) than light-skinned patients (3.6%) with the same SpO2 measurement.
“An important finding of this study is that the probability of dark-skinned patients with SpO2 reading between 92% and 96% having an arterial oxygen saturation of less than 88% was 3 times higher (11.7%) than light-skinned patients (3.6%) with the same SpO2 measurement.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [50] Sjoding, 2020, Racial Bias in Pulse Oximetry Measurement, N. Engl. J. Med. source
Not checked in the free run.
- Premisecitation
Accuracy errors in pulse oximeter readings have been associated with SpO2 overestimations, especially at low saturation conditions.
“Accuracy errors in pulse oximeter readings have been associated with SpO2 overestimations, especially at low saturation conditions [23,24,73,93].”
Quote in the paperSubstantial claim0 of 4 sources checked- cites [23] Bickler, 2005, Effects of skin pigmentation on pulse oximeter accuracy at low saturation, Anesthesiology source
Not checked in the free run.
- cites [24] Feiner, 2007, Dark skin decreases the accuracy of pulse oximeters at low oxygen saturation: The effects of oximeter probe type and gender, Anesth. Analg. source
Not checked in the free run.
- cites [73] Philip, 2021, Racial bias in pulse oximetry: More statistical detail may help tackle the problem, BMJ source
Not checked in the free run.
- cites [93] Lee, 1990, Potential application of photoplethysmography technique in evaluating microcirculatory status of STAMP patients: Preliminary report, J. Rehabil. Res. Dev. source
Not checked in the free run.
- Premisecitation
A diagnostic accuracy study evaluated 10 of the most purchased pulse oximeters with arterial blood gas measurement SaO2 as reference standard and found poorer SpO2 performance in 5 of the 10 pulse oximeters.
“A diagnostic accuracy study under the Standards for Reporting of Diagnostic Accuracy Studies [104] evaluated 10 of the most purchased pulse oximeters with arterial blood gas measurement SaO2 as the reference standard and found a poorer SpO2 performance in 5 of the 10 pulse oximeters.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [71] Harskamp, 2021, Performance of popular pulse oximeters compared with simultaneous arterial oxygen saturation or clinical-grade pulse oximetry: A cross-sectional validation study in intensive care patients, BMJ Open Respir. Res.
Not checked in the free run.
- Premisecitation
A pioneering study comparing two pulse oximeters on 152 patients showed a statistically significant loss of accuracy in SpO2 readings for dark-skinned patients for one device.
“A pioneering study to compare two pulse oximeters on 152 patients showed a statistically significant loss of accuracy in SpO2 readings for dark-skinned patients for one device [43].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [43] Cecil, 1988, A clinical evaluation of the accuracy of the Nellcor N-100 and Ohmeda 3700 pulse oximeters, J. Clin. Monit. source
Not checked in the free run.
- Premisecitation
One of the earliest studies on this topic in 1976 reported reading errors in dark-skinned patients, reflecting lower blood oxygen saturation values.
“One of the earliest studies on this topic in 1976 already reported reading errors in dark-skinned patients, reflecting lower blood oxygen saturation values [41].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [41] Saunders, 1976, Ear oximetry: Accuracy and tracticability in the assessment of arterial oxygenation, Am. Rev. Respir. Dis.
Not checked in the free run.
- Premisecitation
Low-certainty evidence about overestimations in dark-skinned patients was reported.
“Although not all the reported studies assess specifically the influence of skin pigmentation, low-certainty evidence about overestimations in dark-skinned patients was reported.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [61] Shi, C., Goodall, M., Dumville, J., Hill, J.E., Norman, G., Hamer, O., Clegg, A., Watkins, C.L., Georgiou, G., and Alexander Hodkinson, A. (2022). Article The effects of skin pigmentation on the accuracy of pulse oximetry in measuring oxygen saturation: A systematic review and meta-analysis The effects of skin pigmentation on the accuracy of pulse and meta-analysis. PLoS Med. source
Not checked in the free run.
- Premisecitation
A recent retrospective cohort study of high-risk patients with COVID-19 pneumonia found that mortality was 48% lower in patients who used a pulse oximeter to monitor SpO2 than those without one.
“Indeed, a recent retrospective cohort study of high-risk patients with COVID-19 pneumonia found that mortality was 48% lower in patients who used a pulse oximeter to monitor SpO2 than those without one [19].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [19] Nematswerani, 2021, The impact of routine pulse oximetry use on outcomes in COVID-19-infected patients at increased risk of severe disease: A retrospective cohort analysis, S. Afr. Med. J. Suid-Afr. Tydskr. Vir Geneeskd.
Not checked in the free run.
- Premisecitation
Recent studies have shown a positive bias in oxygen saturation measurements in patients with darkly pigmented skin, particularly under low saturation conditions.
“Recent studies have shown a positive bias in oxygen saturation measurements in patients with darkly pigmented skin, particularly under low saturation conditions.”
- Also stated in: § 1. Introduction · plain
“More recent studies have also revealed that dark skin pigmentation influences the accuracy and performance of pulse oximeters devices, resulting in overestimations [17,49,50,51,52,53] with increased incidence in the risk for occult hypoxemia (SO2<88% despite normal SpO2>92%) [54,55].”
Quote in the paperSubstantial claim0 of 8 sources checked- cites [17] Wiles, 2021, The effect of patient ethnicity on the accuracy of peripheral pulse oximetry in patients with COVID-19 pneumonitis: A single-centre, retrospective analysis, Anaesthesia source
Not checked in the free run.
- cites [49] Bent, 2020, Investigating sources of inaccuracy in wearable optical heart rate sensors, NPJ Digit. Med. source
Not checked in the free run.
- cites [50] Sjoding, 2020, Racial Bias in Pulse Oximetry Measurement, N. Engl. J. Med. source
Not checked in the free run.
- cites [51] Vesoulis, 2021, Racial discrepancy in pulse oximeter accuracy in preterm infants, J. Perinatol. source
Not checked in the free run.
- cites [52] Okunlola, 2022, Pulse Oximeter Performance, Racial Inequity, and the Work Ahead, Respir. Care source
Not checked in the free run.
- cites [53] Henry, 2022, Disparities in Hypoxemia Detection by Pulse Oximetry Across Self-Identified Racial Groups and Associations with Clinical Outcomes, Crit. Care Med. source
Not checked in the free run.
- cites [54] Wong, 2021, Analysis of Discrepancies between Pulse Oximetry and Arterial Oxygen Saturation Measurements by Race and Ethnicity and Association with Organ Dysfunction and Mortality, JAMA Netw. Open source
Not checked in the free run.
- cites [55] Valbuena, 2021, Racial Bias in Pulse Oximetry Measurement Among Patients About to Undergo Extracorporeal Membrane Oxygenation in 2019–2020: A Retrospective Cohort Study, Chest
Not checked in the free run.
- Also stated in: § 1. Introduction · plain
- Premisecitation
Using multi-wavelength devices improves accuracy because they are less sensitive than classical devices to variations in hemoglobin saturation.
“Using multi-wavelength devices improves accuracy because they are less sensitive than the classical devices to the variations in the hemoglobin saturation [38].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [38] Raghavachari, R. (2018). An applicable approach for extracting human heart rate and oxygen saturation during physical movements using a multi-wavelength illumination optoelectronic sensor system. Progress in Biomedical Optics and Imaging—Proceedings of SPIE, SPIE.
Not checked in the free run.
- Premisecitation
Overestimations during exercise were attributed to high levels of carboxyhaemoglobin.
“overestimations were attributed to high levels of carboxyhaemoglobin [83].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [83] Gaskin, 1995, Pulse Oximetry and Exercise, Physiotherapy source
Not checked in the free run.
- Premisecitation
A study of two ear oximeters under normoxic and hypoxic conditions in 33 healthy male subjects (mostly with dark skin pigmentation) showed unacceptable readings for SpO2 values of less than 85% for one device and less than 90% for the other.
“A study of the reliability of two ear oximeters under normoxic and hypoxic conditions in 33 healthy male subjects (mostly with dark skin pigmentation) showed unacceptable readings for SpO2 values of less than 85% for one device and less than 90% for the other [87].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [87] Zeballos, 1991, Reliability of noninvasive oximetry in black subjects during exercise and hypoxia, Am. Rev. Respir. Dis. source
Not checked in the free run.
- Premisecitation
A recent study on 46 subjects at center wavelengths of 840 nm, 675 nm, and 580 nm found significant bias at lower temperatures when calibrating a camera-based SpO2 oximeter using red and green light under normoxic and hypoxic conditions.
“A recent study was performed on 46 subjects at center wavelengths of 840 nm (near-infrared), 675 nm (red), and 580 nm (green) to evaluate the feasibility of calibrating a camera-based SpO2 oximeter using red and green light on data under normoxic and hypoxic conditions found significant bias at lower temperatures [72].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [72] Verkruysse, 2021, Pulse oximetry based on photoplethysmography imaging with red and green light: Calibratability and challenges, J. Clin. Monit. Comput. source
Not checked in the free run.
- Interpretationnone given
There is growing evidence that pulse oximeters are less accurate in dark-skinned individuals at lower saturation (<80%) resulting in overestimations.
“However, there is growing evidence that pulse oximeters are less accurate in dark-skinned individuals at lower saturation (<80%) resulting in overestimations.”
Quote in the paperNot a substantial claim - Premisecitation
A small but statistically significant difference between SaO2 and SpO2 in light and dark skin phototypes was found in 404 ICU patients classified by FSP.
“A small but statistically significant difference between SaO2 and SpO2 in light and dark skin phototypes [76] was found.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [76] Ebmeier, 2018, A Two Centre Observational Study of Simultaneous Pulse Oximetry and Arterial Oxygen Saturation Recordings in Intensive Care Unit Patients, Anaesth. Intensive Care source
Not checked in the free run.
- Premisecitation
Zeballos and Weisman found overestimations in the two compared oximeters in darkly pigmented subjects.
“Although they did not refer to a specific scale of skin pigmentation degree, they also found overestimations in the two compared oximeters in darkly pigmented subjects.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [87] Zeballos, 1991, Reliability of noninvasive oximetry in black subjects during exercise and hypoxia, Am. Rev. Respir. Dis. source
Not checked in the free run.
- Premisecitation
In the study of the 10 most purchased pulse oximeters, 5 of 35 patients (14.3%) had a dark phototype (IV-VI) showing less accurate SpO2 measurements.
“In the study of the 10 most purchased pulse oximeters, 5 of 35 patients (14.3%) had a dark phototype (IV–VI) showing less accurate SpO2 measurements [104].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [104] Bossuyt, 2015, STARD 2015: An updated list of essential items for reporting diagnostic accuracy studies, BMJ source
Not checked in the free run.
- Premisecitation
A study in patients with acute COVID-19 pneumonia found a small bias between arterial blood gas analysis SaO2 and SpO2 measurements of 0.28%, −0.33% and −0.75% for 194 patients of "White" (135), "Asian" (34) and "Black" (19) ethnic origin, respectively.
“Besides, another study was performed in patients with acute COVID-19 pneumonia to find out a relationship between patient ancestry and the accuracy of pulse oximetry found a small bias between arterial blood gas analysis SaO2 and SpO2 measurements of 0.28%, −0.33% and −0.75% for 194 patients of “White” (135), “Asian” (34) and “Black” (19) ethnic origin, respectively, [17].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [17] Wiles, 2021, The effect of patient ethnicity on the accuracy of peripheral pulse oximetry in patients with COVID-19 pneumonitis: A single-centre, retrospective analysis, Anaesthesia source
Not checked in the free run.
- Premisecitation
Black henna produces more errors in SpO2 readings than red henna.
“They reported that black henna produces more errors in SpO2 readings than red henna [84], which is expectable since black henna produces a darker skin.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [84] Harakati, 1994, The effect of henna paste on oxygen saturation reading obtained by pulse oximetry, Trop. Geogr. Med.
Not checked in the free run.
- Premisecitation
Stell et al. reported that skin pigmentation had a significant effect on measurement bias with substantial discrepancies in two models.
“They reported that skin pigmentation had a significant effect on measurement bias with substantial discrepancies in two models [69].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [69] Stell, 2022, Exploring the impact of pulse oximeter selection within the COVID-19 home-use pulse oximetry pathways, BMJ Open Respir. Res.
Not checked in the free run.
- Interpretationreasoning
These studies may be misleading because of a weak correlation with skin pigmentation by using uncertain descriptors of ethnicity to classify the population as "Black" or "White".
“These studies may be misleading because of a weak correlation with skin pigmentation by using uncertain descriptors of ethnicity to classify the population as “Black” or “White”.”
Quote in the paperSubstantial claim - Premisecitation
A study in 1990 in ventilator-dependent patients when titrating supplemental O2 reported inaccurate measurements in dark-skinned patients.
“A study performed in 1990 in ventilator-dependent patients when titrating supplemental O2 report inaccurate measurements in dark-skinned patients [88].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [88] Jubran, 1990, Reliability of pulse oximetry in titrating supplemental oxygen therapy in ventilator-dependent patients, Chest source
Not checked in the free run.
- Premisecitation
Another study in 187 pulmonary patients undergoing a stress test reported technical problems in patients with slightly less accurate readings.
“Besides, another study in 187 pulmonary patients undergoing a stress test reported technical problems in patients with slightly less accurate readings [89].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [89] Ries, 1989, Skin color and ear oximetry, Chest source
Not checked in the free run.
- Premisecitation
A multi-ethnic population study found that overestimations were more pronounced in hypoxic conditions, jaundice and darker-skinned patients.
“A multi-ethnic population study composed of 22 Chinese, 6 Malay and 5 Indian patients was also studied, finding that overestimations were more pronounced in hypoxic conditions, jaundice and darker-skinned patients [85].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [85] Lee, 1993, Factors influencing pulse oximetry as compared to functional arterial saturation in multi-ethnic Singapore, Singap. Med. J.
Not checked in the free run.
- May combine two claims. 39
- Classified as probably not a substantial claim. 28
May combine two claims.
Findingown data6 of the 41 references studied in this review examined the performance of pulse oximetry during the COVID-19 pandemic, which is 50% of the references published since 2019.
“We found that 6 of the 41 references studied in this review examined the performance of pulse oximetry during the COVID-19 pandemic [17,30,50,51,69,71], that is the 50% of the references published since 2019.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataThe three most influential papers were M. Kumar et al. with 236 TC and an average of 29.50 cites per year, Y. Mendelson et al. with 181 TC and an average of 5.17 cites per year, and A.C. Ralston et al. with 149 TC and an average of 4.66 cites per year.
“The three most influential papers were M. Kumar et al. [78] with 236 TC and an average of 29.50 cites per year, Y. Mendelson et al. [21] with 181 TC and an average of 5.17 cites per year, and A.C. Ralston et al. [86] with 149 TC and an average of 4.66 cites per year.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataTwo prominent peaks in 2015 and 2020 correspond to M. Kumar et al. and M.W. Sjoding et al. with 29.50 and 37.00 TC per year, respectively.
“Two prominent peaks in 2015 and 2020 are observed, corresponding to the references of M. Kumar et al. [78] and M.W. Sjoding et al. [50] with 29.50 and 37.00 TC¯ per year, respectively.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown data35 of 41 studies were considered for risk of bias assessment.
“We have considered 35/41 studies to perform risk of bias assessment of the individual studies shown in Figure 2.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingown dataThe principal reasons for exclusion were studies not focused on humans (5), skin pigmentation influence not evaluated (4), and oxygen saturation not evaluated (4).
“The principal reasons for exclusion were: studies that did not focus on humans (5); skin pigmentation influence not evaluated (4); oxygen saturation not evaluated (4).”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataMajor issues for unclear risk domains were possible bias in interpretation or conduction of the reference standard, and flow and timing in terms of patient flow and all patients receiving the same reference standard.
“The major issues about these domains were related to possible bias about the interpretation or conduction of the reference standard, and about flow and timing in terms of patient flow and all patients receiving the same reference standard.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataThe most cited reference within the selection was P.E. Bickler et al. with 10 LC and a LC/TC ratio of 7.63%.
“However, the most cited reference within our selection was P.E. Bickler et al. [23] with 10 LC and a LC/TC ratio of 7.63%.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataAfter title and abstract examination, 54 references were potentially appropriate for full-text reading.
“After titles and abstract examination, 54 references were potentially appropriate for full-text reading.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingcitation1/12 (8.33%) assessed that further studies with other clinical and patient-reported data are required to understand the effects of skin pigmentation on accuracy.
“1/12 (8.33%) assessed that further studies with other clinical and patient-reported data are required to understand the effects of skin pigmentation on the accuracy [77]”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [77] Sanyal, 2018, Algorithms for Monitoring Heart Rate and Respiratory Rate From the Video of a User’s Face, IEEE J. Transl. Eng. Health Med. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataThe search identified 99 unique references.
“We found 99 unique references, of which only 41 satisfied the established inclusion criteria.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataM.W. Sjoding et al. has recently attracted much attention with 111 TC in the last two years.
“Let us remark that the reference by M.W. Sjoding et al. has recently attracted much attention with 111 TC in the last two years.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataObjective scales such as the Fitzpatrick Skin Pigmentation (FSP) scale or the Munsell scale were used by 6/37 studies (16.21%) and 4/37 studies (10.81%), respectively.
“We find that objective scales such as the Fitzpatrick Skin Pigmentation (FSP) scale or the Munsell scale were used by 6/37 studies (16.21%) [61,69,70,71,76,77] and 4/37 studies (10.81%) [47,48,74,89], respectively.”
Quote in the paperNot a substantial claim0 of 10 sources checked- cites [61] Shi, C., Goodall, M., Dumville, J., Hill, J.E., Norman, G., Hamer, O., Clegg, A., Watkins, C.L., Georgiou, G., and Alexander Hodkinson, A. (2022). Article The effects of skin pigmentation on the accuracy of pulse oximetry in measuring oxygen saturation: A systematic review and meta-analysis The effects of skin pigmentation on the accuracy of pulse and meta-analysis. PLoS Med. source
Not checked in the free run.
- cites [69] Stell, 2022, Exploring the impact of pulse oximeter selection within the COVID-19 home-use pulse oximetry pathways, BMJ Open Respir. Res.
Not checked in the free run.
- cites [70] Allado, 2021, Innovative measurement of routine physiological variables (heart rate, respiratory rate and oxygen saturation) using a remote photoplethysmography imaging system: A prospective comparative trial protocol, BMJ Open source
Not checked in the free run.
- cites [71] Harskamp, 2021, Performance of popular pulse oximeters compared with simultaneous arterial oxygen saturation or clinical-grade pulse oximetry: A cross-sectional validation study in intensive care patients, BMJ Open Respir. Res.
Not checked in the free run.
- cites [76] Ebmeier, 2018, A Two Centre Observational Study of Simultaneous Pulse Oximetry and Arterial Oxygen Saturation Recordings in Intensive Care Unit Patients, Anaesth. Intensive Care source
Not checked in the free run.
- cites [77] Sanyal, 2018, Algorithms for Monitoring Heart Rate and Respiratory Rate From the Video of a User’s Face, IEEE J. Transl. Eng. Health Med. source
Not checked in the free run.
- cites [47] Adler, 1998, Effect of skin pigmentation on pulse oximetry accuracy in the emergency department, Acad. Emerg. Med. Off. J. Soc. Acad. Emerg. Med. source
Not checked in the free run.
- cites [48] Foglia, 2017, The Effect of Skin Pigmentation on the Accuracy of Pulse Oximetry in Infants with Hypoxemia, J. Pediatr. source
Not checked in the free run.
- cites [74] Poets, 2019, Noninvasive Monitoring and Assessment of Oxygenation in Infants, Clin. Perinatol. source
Not checked in the free run.
- cites [89] Ries, 1989, Skin color and ear oximetry, Chest source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataThe keyword "pulse oximetry" has the strongest strength in the middle of the chart connected with the rest of the keywords as "oxygen saturation" and "patient monitoring".
“The keyword “pulse oximetry” has the strongest strength in the middle of the chart connected with the rest of the keywords as “oxygen saturation”, and “patient monitoring”.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
FindingcitationA portable EEL reflectance spectrophotometer to measure natural pigmentation (not darkened by sun-tanning) was employed in 1 study (2.70%).
“Besides, a portable EEL reflectance spectrophotometer to measure natural pigmentation (not darkened by sun-tanning) was employed in 1 study ( 2.70%) [46].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [46] Bothma, 1996, Accuracy of pulse oximetry in pigmented patients, S. Afr. Med. J.
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataThe keyword "COVID-19" appears in 2020 near to the keywords "hypoxemia", "oxygen-therapy", "intensive-care", "diagnostic accuracy", and "dark-skin population".
“The keyword “COVID-19” appears in 2020 near to the keywords, “hypoxemia”, “oxygen-therapy”, “intensive-care”, “diagnostic accuracy”, and “dark-skin population” among others.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingnone givenTo date, the effects of skin pigment on pulse oximeter performance have been studied for only a relatively small number of pulse oximeter models, and most of them have been calibrated using light-skinned individuals.
“It is also worthy to remark that, to date, the effects of skin pigment on pulse oximeter performance have been studied for only a relatively small number of pulse oximeter models, and most of them have been calibrated using light-skinned individuals.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationA prevalence of occult hypoxemia in dark-skinned subjects has been recently confirmed in three studies, showing that dark-skinned patients may have over three times the risk of experiencing occult hypoxemia during hospitalization compared to light-skinned patients.
“A prevalence of occult hypoxemia in dark-skinned subjects has been recently confirmed in three studies, showing that dark-skinned patients may have over three times the risk of experiencing occult hypoxemia during hospitalization compared to light-skinned patients [53,54,55].”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [53] Henry, 2022, Disparities in Hypoxemia Detection by Pulse Oximetry Across Self-Identified Racial Groups and Associations with Clinical Outcomes, Crit. Care Med. source
Not checked in the free run.
- cites [54] Wong, 2021, Analysis of Discrepancies between Pulse Oximetry and Arterial Oxygen Saturation Measurements by Race and Ethnicity and Association with Organ Dysfunction and Mortality, JAMA Netw. Open source
Not checked in the free run.
- cites [55] Valbuena, 2021, Racial Bias in Pulse Oximetry Measurement Among Patients About to Undergo Extracorporeal Membrane Oxygenation in 2019–2020: A Retrospective Cohort Study, Chest
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataTwo authors independently performed the bibliographic search.
“Two authors (AMC and MFG) independently performed the bibliographic search, while the search algorithm was reviewed and discussed with the rest of the authors (PM, KL or DL).”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationM.W. Sjoding et al. found overestimations in dark-skinned patients with an increased risk of occult hypoxemia.
“Besides, they found overestimations in dark-skinned patients with an increased risk of occult hypoxemia [50].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [50] Sjoding, 2020, Racial Bias in Pulse Oximetry Measurement, N. Engl. J. Med. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe five authors performed the quality and eligibility assessment of relevant references.
“The five authors performed the quality and eligibility assessment of relevant references.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationBickler et al. tested the effect of skin pigmentation under hypoxia in 11 dark-skinned male subjects and 10 light-skinned males breathing a nitrogen-carbon monoxide mixture; the three pulse oximeters used showed overestimation in dark-skinned individuals that increases linearly as SpO2 levels decrease.
“The effect of skin pigmentation in pulse oximeter readings under hypoxia conditions was tested in 11 dark-skinned male subjects and 10 light-skinned males, who were made to breathe an air mixture (nitrogen-carbon monoxide). The three pulse oximeters used showed overestimation in dark-skinned individuals that increases linearly as SpO2 levels decreases [23].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [23] Bickler, 2005, Effects of skin pigmentation on pulse oximeter accuracy at low saturation, Anesthesiology source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductcitationRisk of bias was assessed using the QUADAS-2 protocol based on four domains: patient selection, index test, reference standard, and flow and timing.
“Although there is no a standard risk of bias tool for methods-comparison studies, we follow the QUADAS-2 protocol based on four domains: patient selection, index test, reference standard, and flow and timing of patients [90].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [90] Whiting, 2011, QUADAS-2: A revised tool for the quality assessment of diagnostic accuracy studies, Ann. Intern. Med. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationA review of 13 models of pulse oximeters suggested that a high percentage of carboxyhemoglobin or methaemoglobin are related to failures in saturation readings and can hide hypoxemia.
“A review of 13 models of pulse oximeters suggested that a high percentage of carboxyhemoglobin or methaemoglobin are related to failures in saturation readings and can hide hypoxemia [86].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [86] Ralston, 1991, Potential errors in pulse oximetry. I. Pulse oximeter evaluation, Anaesthesia source
Not checked in the free run.
May combine two claims.
PremisecitationFallow et al. found a better resolution using a green-light wavelength at rest and a green or blue-light wavelength in motion across all skin types.
“Indeed, Fallow et al. found a better resolution using a green-light wavelength at rest and a green or blue-light wavelength in motion across all skin types [80].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [80] Fallow, 2013, Influence of skin type and wavelength on light wave reflectance, J. Clin. Monit. Comput. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductown dataTitles, abstracts, and author keywords were first independently analyzed by two authors, and full texts were examined by five authors to check inclusion criteria.
“Titles, abstracts, and author’s keywords were first independently analyzed to get enough information about the eligibility of two authors (AMC and MFG, or PM). Second, we carefully examined the full text to check whether the references satisfy the inclusion criteria by the five authors.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim. May combine two claims.
PremisecitationFeiner et al. conclude that pulse oximeter inaccuracy is reasonably small at SpO2>80% because of dark skin pigmentation, but a bias of up to 8% was detected at lower saturation conditions in dark-skinned individuals.
“Feiner et al. conclude that pulse oximeter inaccuracy is reasonably small at SpO2>80% because of dark skin pigmentation [24]. However, a bias of up to 8% was detected at lower saturation conditions in dark-skinned individuals.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [24] Feiner, 2007, Dark skin decreases the accuracy of pulse oximeters at low oxygen saturation: The effects of oximeter probe type and gender, Anesth. Analg. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataThe review applied the PRISMA statement to conduct the systematic review.
“The Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) statement was applied to pre-identified the selected studies indexed in PubMed, Web of Science (WOS), and Scopus databases.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationLee et al. found that estimation of arterial oxygen pressure, SaO2, varied significantly (p < 0.05) in multi-ethnic individuals (22 Chinese, 6 Malay and 5 Indian), and the difference between SpO2 and SaO2 appeared to increase with darker skin pigmentation, with overestimations more pronounced under conditions of hypoxia and jaundice.
“Lee et al. also observed this bias in an older study, finding that the estimation of arterial oxygen pressure, SaO2, varied significantly (p < 0.05) in multi-ethnic individuals (22 Chinese, 6 Malay and 5 Indian). The results suggested that the difference between SpO2 and SaO2 appeared to increase with darker skin pigmentation, with overestimations more pronounced under conditions of hypoxia and ja”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [85] Lee, 1993, Factors influencing pulse oximetry as compared to functional arterial saturation in multi-ethnic Singapore, Singap. Med. J.
Not checked in the free run.
May combine two claims.
PremisecitationM.W. Sjoding et al. conducted a study in two large cohorts at the University of Michigan Hospital comparing SpO2 and arterial blood gas measurements in over 10,000 people with 14-20% dark-skinned patients in both cohorts.
“They conducted this study in two large cohorts at the University of Michigan Hospital comparing measurements of SpO2 and arterial blood gas of a sample of over 10,000 people with 14–20% of dark-skinned patients in both cohorts.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [50] Sjoding, 2020, Racial Bias in Pulse Oximetry Measurement, N. Engl. J. Med. source
Not checked in the free run.
Classified as probably not a substantial claim. May combine two claims.
PremisecitationBecause dark skin contains more melanin, several studies assert that pulse oximetry may not be reliable in dark-skinned patients.
“Because dark skin contains more melanin, several studies assert that pulse oximetry may not be reliable in dark-skinned patients [21,23,24,50,51,53,69,73,75,76,78,80,87,88,89,93].”
Quote in the paperSubstantial claim- Rests on an assumption: Dark skin contains more melanin.
May combine two claims.
PremisecitationStudies on healthy adults with different skin pigmentation showed that red and green wavelengths could estimate oxygen saturation with good agreement and lower error ratio compared to the traditional pulse oximeter.
“Studies on healthy adults with different skin pigmentation showed that red and green wavelengths could estimate oxygen saturation with good agreement and lower error ratio compared to the traditional pulse oximeter [30,80,102].”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [30] Al-Naji, A., Khalid, G.A., Mahdi, J.F., and Chahl, J. (2021). Non-Contact SpO2 Prediction System Based on a Digital Camera. Appl. Sci., 11. source
Not checked in the free run.
- cites [80] Fallow, 2013, Influence of skin type and wavelength on light wave reflectance, J. Clin. Monit. Comput. source
Not checked in the free run.
- cites [102] Maeda, 2011, The advantages of wearable green reflected photoplethysmography, J. Med. Syst. source
Not checked in the free run.
May combine two claims.
PremisecitationBaek et al. performed desaturation experiments at 60% to 100% on healthy adults with different skin pigmentation; conventional sensors showed a large error in dark-skinned subjects, while the crosstalk-free sensor did not present SpO2 measurement errors according to skin color.
“Accuracy improvements were found with a crosstalk-free sensor designed by Baek et al. who performed desaturation experiments at 60% to 100% on healthy adults with different skin pigmentation [75]. The conventional sensors showed a large error in dark-skinned subjects, while the sensor which prevented optical crosstalk did not present SpO2 measurement errors according to skin color.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [75] Baek, 2018, The Effect of Optical Crosstalk on Accuracy of Reflectance-Type Pulse Oximeter for Mobile Healthcare, J. Healthc. Eng. source
Not checked in the free run.
Classified as probably not a substantial claim. May combine two claims.
PremisecitationDarker skin pigmentation affects the reliability of pulse oximeters, and when used by a patient for home monitoring, confirmation of a medical-grade oximeter is required.
“They advise that darker skin pigmentation affects the reliability of these pulse oximeters, and when used by a patient for home monitoring, confirmation of a medical-grade oximeter is required [71].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [71] Harskamp, 2021, Performance of popular pulse oximeters compared with simultaneous arterial oxygen saturation or clinical-grade pulse oximetry: A cross-sectional validation study in intensive care patients, BMJ Open Respir. Res.
Not checked in the free run.
May combine two claims.
PremisecitationPatients with hidden hypoxemia have higher rates of in-hospital mortality and organ dysfunction.
“This finding is of particular importance since patients with hidden hypoxemia have higher rates of in-hospital mortality and organ dysfunction [55].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [55] Valbuena, 2021, Racial Bias in Pulse Oximetry Measurement Among Patients About to Undergo Extracorporeal Membrane Oxygenation in 2019–2020: A Retrospective Cohort Study, Chest
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationSome older studies did not identify a considerable bias related to skin pigmentation, suggesting that dark skin pigmentation does not affect the quality of the signal.
“Conversely, some older studies did not identify a considerable bias related to skin pigmentation [44,45,46,56] suggesting that dark skin pigmentation does not affect the quality of the signal [47].”
Quote in the paperNot a substantial claim0 of 5 sources checked- cites [44] Wang, 1985, Noninvasive oximetry in pigmented patients, Ann. Acad. Med.
Not checked in the free run.
- cites [45] Gabrielczyk, 1988, Pulse oximetry and postoperative hypothermia, Anaesthesia source
Not checked in the free run.
- cites [46] Bothma, 1996, Accuracy of pulse oximetry in pigmented patients, S. Afr. Med. J.
Not checked in the free run.
- cites [56] Avant, 1997, G; Lowe N.; Torres Jr., A. Comparison of accuracy and signal consistency of two reusable pulse oximeter probes in critically ill children, Respir. Care
Not checked in the free run.
- cites [47] Adler, 1998, Effect of skin pigmentation on pulse oximetry accuracy in the emergency department, Acad. Emerg. Med. Off. J. Soc. Acad. Emerg. Med. source
Not checked in the free run.
May combine two claims.
PremisecitationA study about applying a pulse oximeter device in titration of fractional inspired O2 concentration in ventilator-dependent patients revealed that a SpO2 reading of 92% was reliable when titrating supplemental O2 in light-skinned patients receiving mechanical ventilation, but in dark-skinned patients the same SpO2 value was commonly related to significant hypoxemia, and higher SpO2 measurements of
“A study about applying a pulse oximeter device in the titration of fractional inspired O2 concentration in ventilator-dependent patients revealed that a SpO2 reading of 92% was reliable when titrating supplemental O2 in light-skinned patients receiving mechanical ventilation. However, in dark-skinned patients, the same SpO2 value was commonly related to significant hypoxemia, and a higher SpO2”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [88] Jubran, 1990, Reliability of pulse oximetry in titrating supplemental oxygen therapy in ventilator-dependent patients, Chest source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationGreen-wavelength pulse oximeters display the greatest modulation depth with pulsatile blood absorption, i.e., a greater absorbance for both deoxyhemoglobin and oxyhemoglobin than infrared light.
“Green-wavelength pulse oximeters display the greatest modulation depth with pulsatile blood absorption, i.e., a greater absorbance for both deoxyhemoglobin and oxyhemoglobin than infrared light [40].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [40] Tamura, 2014, Wearable photoplethysmographic sensors—past and present, Electronics source
Not checked in the free run.
May combine two claims.
InterpretationcitationThe main reason for the upsurge of publications in 2021 may be attributed to the increase of pulse oximetry for triage in patients because of the COVID-19 pandemic and recent concern about accuracy across all skin types.
“The main reason for this surge may be attributed to the increase of this technique for triage in patients because of the current COVID-19 pandemic [12,13,14,15,16,17], and also because of the recent concern about the accuracy of these devices across all skin types [20,57,58,59,60,61].”
Quote in the paperNot a substantial claim0 of 10 sources checked- cites [12] Scherrenberg, M., Wilhelm, M., Hansen, D., Völler, H., Cornelissen, V., Frederix, I., Kemps, H., and Dendale, P. (2020). The future is now: A call for action for cardiac telerehabilitation in the COVID-19 pandemic from the secondary prevention and rehabilitation section of the European Association of Preventive Cardiology. Eur. J. Prev. Cardiol., 1–21. source
Not checked in the free run.
- cites [13] Channa, A., Popescu, N., Skibinska, J., and Burget, R. (2021). The Rise of Wearable Devices during the COVID-19 Pandemic: A Systematic Review. Sensors, 21. source
Not checked in the free run.
- cites [14] Philip, 2020, Working accuracy of pulse oximetry in COVID-19 patients stepping down from intensive care: A clinical evaluation, BMJ Open Respir. Res.
Not checked in the free run.
- cites [15] Michard, 2021, COVID-19: Pulse oximeters in the spotlight, J. Clin. Monit. Comput. source
Not checked in the free run.
- cites [16] England NHS (2020, March 26). Specialty Guides for Patient Management during the Coronavirus Pandemic. Guidance for the Role and Use of Non-Invasive Respiratory Support in Adult Patients with COVID-19 (Confirmed or Suspected) 6 April 2020, Version 2. Available online: https://amhp.org.uk/app/uploads/2020/03/Guidance-Respiratory-Support.pdf.
Not checked in the free run.
- cites [17] Wiles, 2021, The effect of patient ethnicity on the accuracy of peripheral pulse oximetry in patients with COVID-19 pneumonitis: A single-centre, retrospective analysis, Anaesthesia source
Not checked in the free run.
- cites [20] World Health Organization (2021). COVID-19 Clinical Management: Living Guidance, 25 January 2021, World Health Organization. Technical Documents.
Not checked in the free run.
- cites [57] Warren, R., and Wyden, C.B. (2021). 2021.01.25 Letter to FDA re Bias in Pulse Oximetery Measurements, Technical Report; U.S. Food and Drug Administration.
Not checked in the free run.
- cites [58] NHS Race & Health Observatory (2021). Pulse Oximetry and Racial Bias: Recommendations for National Healthcare, Regulatory and Research Bodies, National Health Service England. Technical Report March.
Not checked in the free run.
- cites [59] Hunasikatti, 2021, Racial bias in accuracy of pulse oximetry and its impact on assessments of hypopnea and T90 in clinical studies, J. Clin. Sleep Med. source
Not checked in the free run.
May combine two claims.
PremisecitationMore recent studies have revealed that dark skin pigmentation influences the accuracy and performance of pulse oximeter devices, resulting in overestimations, with increased incidence in the risk for occult hypoxemia (SO2<88% despite normal SpO2>92%).
“More recent studies have also revealed that dark skin pigmentation influences the accuracy and performance of pulse oximeters devices, resulting in overestimations [17,49,50,51,52,53] with increased incidence in the risk for occult hypoxemia (SO2<88% despite normal SpO2>92%) [54,55].”
Quote in the paperSubstantial claim0 of 8 sources checked- cites [17] Wiles, 2021, The effect of patient ethnicity on the accuracy of peripheral pulse oximetry in patients with COVID-19 pneumonitis: A single-centre, retrospective analysis, Anaesthesia source
Not checked in the free run.
- cites [49] Bent, 2020, Investigating sources of inaccuracy in wearable optical heart rate sensors, NPJ Digit. Med. source
Not checked in the free run.
- cites [50] Sjoding, 2020, Racial Bias in Pulse Oximetry Measurement, N. Engl. J. Med. source
Not checked in the free run.
- cites [51] Vesoulis, 2021, Racial discrepancy in pulse oximeter accuracy in preterm infants, J. Perinatol. source
Not checked in the free run.
- cites [52] Okunlola, 2022, Pulse Oximeter Performance, Racial Inequity, and the Work Ahead, Respir. Care source
Not checked in the free run.
- cites [53] Henry, 2022, Disparities in Hypoxemia Detection by Pulse Oximetry Across Self-Identified Racial Groups and Associations with Clinical Outcomes, Crit. Care Med. source
Not checked in the free run.
- cites [54] Wong, 2021, Analysis of Discrepancies between Pulse Oximetry and Arterial Oxygen Saturation Measurements by Race and Ethnicity and Association with Organ Dysfunction and Mortality, JAMA Netw. Open source
Not checked in the free run.
- cites [55] Valbuena, 2021, Racial Bias in Pulse Oximetry Measurement Among Patients About to Undergo Extracorporeal Membrane Oxygenation in 2019–2020: A Retrospective Cohort Study, Chest
Not checked in the free run.
May combine two claims.
InterpretationreasoningThe older findings of no considerable bias may be misleading because of a small sample size of the dark-skinned population or a weak correlation with skin pigmentation by using uncertain descriptors of ancestry to classify the population.
“However, these findings may be misleading because of a small sample size of the dark-skinned population or a weak correlation with skin pigmentation by using uncertain descriptors of ancestry to classify the population.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationConventional red and infrared pulse oximeters are reliable at rest and in normoxic or hyperoxic conditions, but have presented worsening accuracy in hypoxic conditions.
“Conventional red and infrared pulse oximeters are reliable at rest and in normoxic or hyperoxic conditions, but have presented worsening accuracy in hypoxic conditions [23].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [23] Bickler, 2005, Effects of skin pigmentation on pulse oximeter accuracy at low saturation, Anesthesiology source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationA recent comprehensive review comparing SpO2 measurements with arterial blood gas analysis SaO2 in different patients and conditions reported accurate but imprecise measurements across all levels of skin pigmentation.
“A recent comprehensive review of different studies comparing SpO2 measurements with arterial blood gas analysis SaO2 in different patients and conditions has reported accurate but imprecise measurements across all levels of skin pigmentation [61].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [61] Shi, C., Goodall, M., Dumville, J., Hill, J.E., Norman, G., Hamer, O., Clegg, A., Watkins, C.L., Georgiou, G., and Alexander Hodkinson, A. (2022). Article The effects of skin pigmentation on the accuracy of pulse oximetry in measuring oxygen saturation: A systematic review and meta-analysis The effects of skin pigmentation on the accuracy of pulse and meta-analysis. PLoS Med. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThe current World Health Organization COVID-19 management guideline recommends that symptomatic people with COVID-19 use a home pulse oximeter for self SpO2 monitoring "as part of a package of care".
“Therefore, the current World Health Organization COVID-19 management guideline recommends that symptomatic people with COVID-19 to use a home pulse oximeter for self SpO2 monitoring “as part of a package of care” [20].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [20] World Health Organization (2021). COVID-19 Clinical Management: Living Guidance, 25 January 2021, World Health Organization. Technical Documents.
Not checked in the free run.
Classified as probably not a substantial claim.
PremisecitationSeveral health organizations and healthcare professionals have raised growing concerns about the clinical accuracy of pulse oximetry in dark-skinned patients, particularly in detecting occult hypoxemia.
“Due to this increasing evidence, several health organizations and healthcare professionals have raised growing concerns about the clinical accuracy of pulse oximetry in dark-skinned patients [20,57,58,59,60,61], particularly in detecting occult hypoxemia [54,55].”
Quote in the paperSubstantial claim0 of 8 sources checked- cites [20] World Health Organization (2021). COVID-19 Clinical Management: Living Guidance, 25 January 2021, World Health Organization. Technical Documents.
Not checked in the free run.
- cites [57] Warren, R., and Wyden, C.B. (2021). 2021.01.25 Letter to FDA re Bias in Pulse Oximetery Measurements, Technical Report; U.S. Food and Drug Administration.
Not checked in the free run.
- cites [58] NHS Race & Health Observatory (2021). Pulse Oximetry and Racial Bias: Recommendations for National Healthcare, Regulatory and Research Bodies, National Health Service England. Technical Report March.
Not checked in the free run.
- cites [59] Hunasikatti, 2021, Racial bias in accuracy of pulse oximetry and its impact on assessments of hypopnea and T90 in clinical studies, J. Clin. Sleep Med. source
Not checked in the free run.
- cites [60] Holder, 2022, The Big Consequences of Small Discrepancies: Why Racial Differences in Pulse Oximetry Errors Matter, Crit. Care Med. source
Not checked in the free run.
- cites [61] Shi, C., Goodall, M., Dumville, J., Hill, J.E., Norman, G., Hamer, O., Clegg, A., Watkins, C.L., Georgiou, G., and Alexander Hodkinson, A. (2022). Article The effects of skin pigmentation on the accuracy of pulse oximetry in measuring oxygen saturation: A systematic review and meta-analysis The effects of skin pigmentation on the accuracy of pulse and meta-analysis. PLoS Med. source
Not checked in the free run.
- cites [54] Wong, 2021, Analysis of Discrepancies between Pulse Oximetry and Arterial Oxygen Saturation Measurements by Race and Ethnicity and Association with Organ Dysfunction and Mortality, JAMA Netw. Open source
Not checked in the free run.
- cites [55] Valbuena, 2021, Racial Bias in Pulse Oximetry Measurement Among Patients About to Undergo Extracorporeal Membrane Oxygenation in 2019–2020: A Retrospective Cohort Study, Chest
Not checked in the free run.
May combine two claims.
PremisecitationThe skin pigmentation effect on pulse oximetry accuracy has long been studied, reporting some contradictory conclusions.
“The skin pigmentation effect on pulse oximetry accuracy has long been studied reporting some contradictory conclusions.”
- Also stated in: § 1. Introduction · plain
“Skin pigmentation impact on the accuracy of SpO2 monitoring has long been studied [24,41,42,43,44,45,46,47,48,49,50,51].”
Quote in the paperSubstantial claim0 of 10 sources checked- cites [24] Feiner, 2007, Dark skin decreases the accuracy of pulse oximeters at low oxygen saturation: The effects of oximeter probe type and gender, Anesth. Analg. source
Not checked in the free run.
- cites [41] Saunders, 1976, Ear oximetry: Accuracy and tracticability in the assessment of arterial oxygenation, Am. Rev. Respir. Dis.
Not checked in the free run.
- cites [42] Emery, 1987, Skin pigmentation as an influence on the accuracy of pulse oximetry, J. Perinatol. Off. J. Calif. Perinat. Assoc.
Not checked in the free run.
- cites [43] Cecil, 1988, A clinical evaluation of the accuracy of the Nellcor N-100 and Ohmeda 3700 pulse oximeters, J. Clin. Monit. source
Not checked in the free run.
- cites [44] Wang, 1985, Noninvasive oximetry in pigmented patients, Ann. Acad. Med.
Not checked in the free run.
- cites [45] Gabrielczyk, 1988, Pulse oximetry and postoperative hypothermia, Anaesthesia source
Not checked in the free run.
- cites [46] Bothma, 1996, Accuracy of pulse oximetry in pigmented patients, S. Afr. Med. J.
Not checked in the free run.
- cites [47] Adler, 1998, Effect of skin pigmentation on pulse oximetry accuracy in the emergency department, Acad. Emerg. Med. Off. J. Soc. Acad. Emerg. Med. source
Not checked in the free run.
- cites [48] Foglia, 2017, The Effect of Skin Pigmentation on the Accuracy of Pulse Oximetry in Infants with Hypoxemia, J. Pediatr. source
Not checked in the free run.
- cites [49] Bent, 2020, Investigating sources of inaccuracy in wearable optical heart rate sensors, NPJ Digit. Med. source
Not checked in the free run.
- Also stated in: § 1. Introduction · plain
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationDuring exercise, the main potential errors are due to motion artifact, which complicates the use of the pulse oximeter and the accurate interpretation of the data.
“Although during exercise, the main potential errors are due to motion artifact, which complicates the use of the pulse oximeter and the accurate interpretation of the data [29,30,31]”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [29] Tobin, 2002, A characterization of motion affecting pulse oximetry in 350 patients, Anesth. Analg.
Not checked in the free run.
- cites [30] Al-Naji, A., Khalid, G.A., Mahdi, J.F., and Chahl, J. (2021). Non-Contact SpO2 Prediction System Based on a Digital Camera. Appl. Sci., 11. source
Not checked in the free run.
- cites [31] 2014, Improved motion robustness of remote-PPG by using the blood volume pulse signature, Physiol. Meas. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThe FSP scale is commonly used to describe skin pigmentation according to the skin tanning response to UVR.
“The FSP scale is commonly used to describe skin pigmentation according to the skin tanning response to UVR [105,106,107].”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [105] Fitzpatrick, 1963, The epidermal melanin unit system, Dermatol. Wochenschr.
Not checked in the free run.
- cites [106] Fitzpatrick, 1988, The validity and practicality of sun-reactive skin types I through VI, Arch. Dermatol. source
Not checked in the free run.
- cites [107] Moreiras, 2021, Visible light and human skin pigmentation: The importance of skin phototype, Exp. Dermatol. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationFallow et al. found that phototype-V (dark brown) skin type has a significant signal-to-noise ratio than all other skin types.
“Fallow et al. classified 23 healthy subjects (11 males and 12 females, 20–59 years old) finding that phototype-V (dark brown) skin type has a significant signal-to-noise ratio than all other skin types [80].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [80] Fallow, 2013, Influence of skin type and wavelength on light wave reflectance, J. Clin. Monit. Comput. source
Not checked in the free run.
May combine two claims.
InterpretationcitationPulse oximetry performance in the dark-skinned population may lead to overestimations with increased incidence of occult hypoxemia and risk of negative health effects from diseases like COVID-19.
“To sum up, we conclude that pulse oximetry performance in the dark-skinned population may lead to overestimations with increased incidence of occult hypoxemia and risk of negative health effects from diseases like COVID-19 [51,53,55].”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [51] Vesoulis, 2021, Racial discrepancy in pulse oximeter accuracy in preterm infants, J. Perinatol. source
Not checked in the free run.
- cites [53] Henry, 2022, Disparities in Hypoxemia Detection by Pulse Oximetry Across Self-Identified Racial Groups and Associations with Clinical Outcomes, Crit. Care Med. source
Not checked in the free run.
- cites [55] Valbuena, 2021, Racial Bias in Pulse Oximetry Measurement Among Patients About to Undergo Extracorporeal Membrane Oxygenation in 2019–2020: A Retrospective Cohort Study, Chest
Not checked in the free run.
May combine two claims.
InterpretationreasoningThe recent camera-based SpO2 study did not stratify results according to level of skin pigmentation, which can introduce considerable errors.
“However, they did not stratify their results according to the level of skin pigmentation, which can introduce considerable errors.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationA study of 26,603 patients across four self-identified racial groups found that pulse oximetry accuracy for detection of occult hypoxemia is not consistent across these groups, with SpO2 measurements on average overestimating arterial blood gas-derived oxygen saturation by 1.57%.
“also found that the pulse oximetry accuracy for the detection of occult hypoxemia is not consistent across these self-identified ethnic groups with SpO2 measurements on average overestimating arterial blood gas-derived oxygen saturation by 1.57% [53].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [53] Henry, 2022, Disparities in Hypoxemia Detection by Pulse Oximetry Across Self-Identified Racial Groups and Associations with Clinical Outcomes, Crit. Care Med. source
Not checked in the free run.
May combine two claims.
InterpretationcitationThese pulse oximeters have probably been calibrated on a population sample with light-skin pigmentation, which miss relevant interaction effects of confounding variables and skin pigmentation.
“Besides, these pulse oximeters have probably been calibrated on a population sample with light-skin pigmentation, which miss relevant interaction effects of confounding variables and skin pigmentation [23,33,109].”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [23] Bickler, 2005, Effects of skin pigmentation on pulse oximeter accuracy at low saturation, Anesthesiology source
Not checked in the free run.
- cites [33] Colvonen, 2021, Response to: Investigating sources of inaccuracy in wearable optical heart rate sensors, NPJ Digit. Med. source
Not checked in the free run.
- cites [109] Xiao, 2017, Characterising the variations in ethnic skin colours: A new calibrated data base for human skin, Skin Res. Technol. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationA study in 298 patients classified by the Munsell color system concluded that skin pigmentation does not affect the bias or precision of pulse oximetry.
“In contrast, a study conducted in 298 patients classified according to the Munsell color system [108] as light group (51%), intermediate group (37%), and dark group (12%) conclude that skin pigmentation does not affect the bias or precision of pulse oximetry.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [108] Munsell, A.H. (1915). Atlas of the Munsell Color System, Wadsworth, Howland & Co., Inc., Printers. source
Not checked in the free run.
May combine two claims.
PremisecitationTwo old studies (1997 with 50 children, 15 dark-skinned; 1996 with 100 dark-skinned patients) did not find any difference in accuracy between dark and light-skinned subjects.
“Two more old studies, one in 1997 with a sample of 50 children, 15 of whom were dark-skinned [56] and another in 1996 with 100 dark-skinned patients [46], did not find any difference in accuracy between dark and light-skinned subjects.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [56] Avant, 1997, G; Lowe N.; Torres Jr., A. Comparison of accuracy and signal consistency of two reusable pulse oximeter probes in critically ill children, Respir. Care
Not checked in the free run.
- cites [46] Bothma, 1996, Accuracy of pulse oximetry in pigmented patients, S. Afr. Med. J.
Not checked in the free run.
May combine two claims.
PremisecitationBaek et al. found large errors in the dark-skinned subjects with the pulse oximeter that did not prevent crosstalk.
“They found large errors in the dark-skinned subjects with the pulse oximeter that did not prevent crosstalk.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [75] Baek, 2018, The Effect of Optical Crosstalk on Accuracy of Reflectance-Type Pulse Oximeter for Mobile Healthcare, J. Healthc. Eng. source
Not checked in the free run.
May combine two claims.
PremisecitationOne study asserted that there is no significant difference in systematic bias based on skin pigment for pulse oximetry, but this study was not statistically significant since it had a small sample size and a stringent inclusion criterion.
“Although one study asserted that there is no significant difference in systematic bias based on skin pigment for pulse oximetry [48], this study was not statistically significant since it had a small sample size and a stringent inclusion criterion that was not typical for premature infants.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [48] Foglia, 2017, The Effect of Skin Pigmentation on the Accuracy of Pulse Oximetry in Infants with Hypoxemia, J. Pediatr. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationFeiner et al. advised that a significant bias in dark-skinned patients with saturation below 80% should be considered.
“They advised that a significant bias in dark-skinned patients with saturation below 80% should be considered.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [24] Feiner, 2007, Dark skin decreases the accuracy of pulse oximeters at low oxygen saturation: The effects of oximeter probe type and gender, Anesth. Analg. source
Not checked in the free run.
May combine two claims.
InterpretationreasoningAlthough some old studies did not find any bias related to dark skin pigmentation, the latter studies have shown that pulse oximeters devices have some limitations with dark-skinned subjects, especially at low saturation or hypoxemia conditions.
“Although some old studies did not find any bias related to dark skin pigmentation, the latter studies have shown that pulse oximeters devices have some limitations with dark-skinned subjects, especially at low saturation or hypoxemia conditions.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationTwo studies found a small but consistent skin pigmentation disparity in SpO2 measurements with values that may be falsely high in dark-skinned preterm infants.
“The other two studies found a small but consistent skin pigmentation disparity in SpO2 measurements with values that may be falsely high in dark-skinned preterm infants [51,74].”
Quote in the paperSubstantial claim0 of 2 sources checked
The paper relies on this without support.
Dark skin contains more melanin.
“Because dark skin contains more melanin, several studies assert that pulse oximetry may not be reliable in dark-skinned patients [21,23,24,50,51,53,69,73,75,76,78,80,87,88,89,93].”
The paper relies on this without support.
Occult hypoxemia is associated with increased mortality risk.
“Since occult hypoxemia is associated with increased mortality risk, there is a need to determine such errors exactly.”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
There have been systematic reviews on PPG skin pigmentation bias, but these have been limited to either consumer device PR or pulse oximeter SpO2.
“There have been systematic reviews on PPG skin pigmentation bias, but these have been limited to either consumer device PR [35] or pulse oximeter SpO2 [36-38].”
From Impact of skin pigmentation on pulse oximetry blood oxygenation and wearable pulse rate accuracy: systematic review and meta-analysis (2024) · cites this paper as [37]
The use of secondary measurement methods, including HRV derived from heart rate values, produces assessments that can be inaccurate for stress and recovery evaluation.
“The use of secondary measurement methods, including heart rate variability (HRV) derived from heart rate values, produces assessments that can be inaccurate for stress and recovery evaluation [32].”
From Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring (2025) · cites this paper as [32]
The Introduction discusses pulse oximetry during exercise stress tests and respiratory rehabilitation, while the Final Remarks describe inaccuracies in oxygen-saturation readings, particularly for dark-skinned individuals. Neither section addresses HRV derived from heart-rate values or the accuracy of stress and recovery assessments. Searches for those terms and for secondary measurement methods found no passage addressing the claim. (LLM, conf=0.75)
Experimental — accuracy not yet measuredSkin pigmentation affects the accuracy of PO.
“A recent flurry in publications identifies that skin pigmentation affects the accuracy of PO.11,134–140”
From Effect of skin color on optical properties and the implications for medical optical technologies: a review (2024) · cites this paper as [134]
Black and ethnic minority groups had poorest outcomes from COVID-19 infection and this was compounded by the overestimation of oxygen saturation.
“Black and ethnic minority groups had poorest outcomes from COVID-19 infection and this was compounded by the overestimation of oxygen saturation.11,134”
From Effect of skin color on optical properties and the implications for medical optical technologies: a review (2024) · cites this paper as [134]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Incorrect percentage of dark-skinned participants in Foglia study
MinorFourteen out of 36 participants is 38.89%, not 40.00%.
Reported 40 · recomputed 38.888889
“| Foglia E.E. et al. [48], 2017 | 36 | 14 (40.00%)”
Wiles study gender counts exceed its sample size
MinorThe listed male and female counts total 294, whereas the table gives a sample size of 194.
Reported 194 · recomputed 294
“| Wiles M.D. et al. [17], 2021 | 194 | 53 (27.32%) | – | 140/154 |”
Ebmeier study gender counts fall short of its sample size
MinorThe listed male and female counts total 294, whereas the table gives a sample size of 394. The table does not indicate whether gender was unavailable for the remaining participants.
Reported 394 · recomputed 294
“| Ebmeier S.J et al. [76], 2018 | 394 | 65 (16.09%) | 47–77 | 150/144 |”
Incorrect percentage of dark-skinned participants in Ebmeier study
MinorThe table lists 65 dark-skinned participants out of 394, which is 16.50%, not 16.09%.
Reported 16.09 · recomputed 16.497462
“| Ebmeier S.J et al. [76], 2018 | 394 | 65 (16.09%)”
Adler study has two different sample sizes
MinorTable 3 gives 284 participants for Adler et al. [47], while the discussion describes the same study as having 298 patients.
“| Adler J.N. et al. [47], 1998 | 284 |”
“a study conducted in 298 patients classified according to the Munsell color system [108] as light group (51%), intermediate group (37%), and dark group (12%) conclude that skin pigmentation does not affect the bias or precision of pulse oximetry.”
Ebmeier study has two different sample sizes
MinorTable 3 lists 394 participants for Ebmeier et al. [76], but the discussion says its skin-phototype classification covered 404 ICU patients.
“| Ebmeier S.J et al. [76], 2018 | 394 |”
“The FSP scale was also employed to classify 404 intensive care unit patients as light (phototype I or II), medium (phototype III or IV), or dark (phototype V or VI) [76].”
Fallow study age ranges disagree
MinorTable 3 reports ages 19-43 for Fallow et al. [80], while the discussion describes the same 23-person study as ages 20-59.
“| Fallow B.A. et al. [80], 2013 | 23 | 10 (43.48%) | 19–43 | 11/12 |”
“Fallow et al. classified 23 healthy subjects (11 males and 12 females, 20–59 years old)”
Fallow skin-classification method disagrees between table and discussion
MinorTable 3 calls Fallow et al.'s method a subjective dark-or-light scale, whereas the discussion says the study used the Fitzpatrick Skin Pigmentation scale.
“| Fallow B.A. et al. [80], 2013 | 23 | 10 (43.48%) | 19–43 | 11/12 | Healthy adults | Subjective scale: dark or light |”
“The FSP scale is commonly used to describe skin pigmentation according to the skin tanning response to UVR [105,106,107]. This scale is divided into six phototypes (SPT) from the lightest (tanning-resistant) phototype-I (SPT-I) to the darkest phototype-VI (SPT-VI). Fallow et al. classified 23 healthy subjects”
Low-risk study outcome categories total twelve, not eleven
MinorThe paragraph says there were 11 low-risk studies, but assigns 8 to inaccuracies, 3 to no demonstrable loss of accuracy, and 1 to a further-studies conclusion-a total of 12.
Reported 11 · recomputed 12
“from the 11/35 studies classified as being at low risk of bias 8/11 (72.73%) reported inaccuracies related to skin pigmentation, 3/11 (27.27%) reported no demonstrable loss of accuracy caused by skin pigmentation [46,48,70], and 1/12 (8.33%) assessed that further studies”
Wiles ethnicity counts do not total the stated patient count
MinorThe three reported ethnicity counts total 188, not the stated 194 patients; the text does not account for the other six.
Reported 194 · recomputed 188
“for 194 patients of “White” (135), “Asian” (34) and “Black” (19) ethnic origin”
Cites
Software survey: VOSviewer, a computer program for bibliometric mapping
van Eck, N. J., & Waltman, L. (2010) · Scientometrics · doi:10.1007/s11192-009-0146-3 · Cited by 21,484 · see citing works
bibliometrix: An R-tool for comprehensive science mapping analysis
Aria, M., & Cuccurullo, C. (2017) · Journal of Informetrics · doi:10.1016/j.joi.2017.08.007 · Cited by 16,064 · see citing works
Tranfield, D., Denyer, D., & Smart, P. (2003) · British Journal of Management · doi:10.1111/1467-8551.00375 · Cited by 15,386 · see citing works
QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies
Whiting, P. F., Rutjes, A. W. S., Westwood, M. E., Mallett, S., Deeks, J. J., Reitsma, J. B., Leeflang, M. M. G., Sterne, J. A. C., Bossuyt, P. M. M., & the QUADAS-2 Group* (2011) · Annals of Internal Medicine · doi:10.7326/0003-4819-155-8-201110180-00009 · Cited by 14,199 · see citing works
Chen, C. (2005) · Journal of the American Society for Information Science and Technology · doi:10.1002/asi.20317 · Cited by 6,689 · see citing works
McGuinness, L. A., & Higgins, J. P. T. (2021) · Research Synthesis Methods · doi:10.1002/jrsm.1411 · Cited by 5,579 · see citing works
The validity and practicality of sun-reactive skin types I through VI
Fitzpatrick, T. B. (1988) · Archives of Dermatology · doi:10.1001/archderm.1988.01670060015008 · Cited by 4,339 · see citing works
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967 · see citing works
STARD 2015: an updated list of essential items for reporting diagnostic accuracy studies
Bossuyt, P. M., Reitsma, J. B., Bruns, D. E., Gatsonis, C. A., Glasziou, P. P., Irwig, L., Lijmer, J. G., Moher, D., Rennie, D., de Vet, H. C. W., et al. (2015) · BMJ · doi:10.1136/bmj.h5527 · Cited by 3,660 · see citing works
Telehealth transformation: COVID-19 and the rise of virtual care
Wosik, J., Fudim, M., Cameron, B., Gellad, Z. F., Cho, A., Phinney, D., Curtis, S., Roman, M., Poon, E. G., Ferranti, J., et al. (2020) · Journal of the American Medical Informatics Association · doi:10.1093/jamia/ocaa067 · Cited by 1,581 · see citing works
Racial bias in pulse oximetry measurement
Sjoding, M. W., Dickson, R. P., Iwashyna, T. J., Gay, S. E., & Valley, T. S. (2020) · New England Journal of Medicine · doi:10.1056/nejmc2029240 · Cited by 1,114 · see citing works
Ghamari, M. (2018) · International Journal of Biosensors & Bioelectronics · doi:10.15406/ijbsbe.2018.04.00125 · Cited by 945 · see citing works
Wearable photoplethysmographic sensors—past and present
Tamura, T., Maeda, Y., Sekine, M., & Yoshida, M. (2014) · Electronics · doi:10.3390/electronics3020282 · Cited by 912 · cited 1× · see citing works
“, a greater absorbance for both deoxyhemoglobin and oxyhemoglobin than infrared light [40].”
BackgroundInvestigating sources of inaccuracy in wearable optical heart rate sensors
Bent, B., Goldstein, B. A., Kibbe, W. A., & Dunn, J. P. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0226-6 · Cited by 673 · see citing works
Jubran, A. (2015) · Critical Care · doi:10.1186/s13054-015-0984-8 · Cited by 511 · see citing works
Improved motion robustness of remote-PPG by using the blood volume pulse signature
de Haan, G., & van Leest, A. (2014) · Physiological Measurement · doi:10.1088/0967-3334/35/9/1913 · Cited by 491 · see citing works
Pulse oximetry: Principles and limitations
Sinex, J. E. (1999) · The American Journal of Emergency Medicine · doi:10.1016/s0735-6757(99)90019-0 · Cited by 482 · see citing works
Zonios, G., Bykowski, J., & Kollias, N. (2001) · Journal of Investigative Dermatology · doi:10.1046/j.0022-202x.2001.01577.x · Cited by 477 · see citing works
Pulse oximetry: fundamentals and technology update
Nitzan, M., Romem, A., & Koppel, R. (2014) · Medical Devices: Evidence and Research · doi:10.2147/mder.s47319 · Cited by 438 · see citing works
Effects of skin pigmentation on pulse oximeter accuracy at low saturation
Bickler, P. E., Feiner, J. R., & Severinghaus, J. W. (2005) · Anesthesiology · doi:10.1097/00000542-200504000-00004 · Cited by 427 · see citing works
DistancePPG: Robust non-contact vital signs monitoring using a camera
Kumar, M., Veeraraghavan, A., & Sabharwal, A. (2015) · Biomedical Optics Express · doi:10.1364/boe.6.001565 · Cited by 417 · see citing works
Feiner, J. R., Severinghaus, J. W., & Bickler, P. E. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000285988.35174.d9 · Cited by 400 · see citing works
Relating oxygen partial pressure, saturation and content: the haemoglobin–oxygen dissociation curve
Collins, J.-A., Rudenski, A. S., Gibson, J., et al. (2015) · Breathe · doi:10.1183/20734735.001415 · Cited by 387 · see citing works
Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348 · see citing works
Jubran, A., & Tobin, M. J. (1990) · Chest · doi:10.1378/chest.97.6.1420 · Cited by 296 · see citing works
Alian, A. A., & Shelley, K. H. (2014) · Best Practice & Research Clinical Anaesthesiology · doi:10.1016/j.bpa.2014.08.006 · Cited by 289 · see citing works
Wong, A.-K. I., Charpignon, M., Kim, H., Josef, C., de Hond, A. A. H., Fojas, J. J., Tabaie, A., Liu, X., Mireles-Cabodevila, E., Carvalho, L., et al. (2021) · JAMA Network Open · doi:10.1001/jamanetworkopen.2021.31674 · Cited by 268 · see citing works
Luks, A. M., & Swenson, E. R. (2020) · Annals of the American Thoracic Society · doi:10.1513/annalsats.202005-418fr · Cited by 263 · see citing works
New principle for measuring arterial blood oxygenation, enabling motion-robust remote monitoring
van Gastel, M., Stuijk, S., & de Haan, G. (2016) · Scientific Reports · doi:10.1038/srep38609 · Cited by 253 · see citing works
Noninvasive pulse oximetry utilizing skin reflectance photoplethysmography
Mendelson, Y., & Ochs, B. D. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.7286 · Cited by 251 · see citing works
Scherrenberg, M., Wilhelm, M., Hansen, D., Völler, H., Cornelissen, V., Frederix, I., Kemps, H., & Dendale, P. (2021) · European Journal of Preventive Cardiology · doi:10.1177/2047487320939671 · Cited by 251 · see citing works
Exploring Two Decades of Research on Classroom Dialogue by Using Bibliometric Analysis
Song, Y., Chen, X., Hao, T., Liu, Z., & Lan, Z. (2019) · Computers & Education · doi:10.1016/j.compedu.2019.04.002 · Cited by 245 · see citing works
The advantages of wearable green reflected photoplethysmography
Maeda, Y., Sekine, M., & Tamura, T. (2011) · Journal of Medical Systems · doi:10.1007/s10916-010-9506-z · Cited by 209 · see citing works
The Rise of Wearable Devices during the COVID-19 Pandemic: A Systematic Review
Channa, A., Popescu, N., Skibinska, J., & Burget, R. (2021) · Sensors · doi:10.3390/s21175787 · Cited by 209 · see citing works
Atlas of the Munsell color system
Munsell, A. H. (1915) · doi:10.5479/sil.129262.39088002718880 · Cited by 201 · see citing works
(1963) · PubMed · Cited by 194 · see citing works
Potential errors in pulse oximetry: I. Pulse oximeter evaluation*
RALSTON, A. C., WEBB, R. K., & RUNCIMAN, W. B. (1991) · Anaesthesia · doi:10.1111/j.1365-2044.1991.tb09410.x · Cited by 166 · see citing works
Thirty Years of theInternational Journal of Intelligent Systems: A Bibliometric Review
Merigó, J. M., Blanco‐Mesa, F., Gil‐Lafuente, A. M., et al. (2016) · International Journal of Intelligent Systems · doi:10.1002/int.21859 · Cited by 164 · see citing works
Influence of skin type and wavelength on light wave reflectance
Fallow, B. A., Tarumi, T., & Tanaka, H. (2013) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-013-9436-7 · Cited by 162 · see citing works
Shi, C., Goodall, M., Dumville, J., Hill, J., Norman, G., Hamer, O., Clegg, A., Watkins, C. L., Georgiou, G., Hodkinson, A., et al. (2022) · doi:10.1101/2022.02.16.22271062 · Cited by 158 · see citing works
Shah, S., Majmudar, K., Stein, A., Gupta, N., Suppes, S., Karamanis, M., Capannari, J., Sethi, S., & Patte, C. (2020) · Academic Emergency Medicine · doi:10.1111/acem.14053 · Cited by 152 · see citing works
Non-contact measurement of oxygen saturation with an RGB camera
Guazzi, A., Villarroel, M., Jorge, J., Daly, J., Frise, M. C., Robbins, P. A., & Tarassenko, L. (2015) · Biomedical Optics Express · doi:10.1364/boe.6.003320 · Cited by 151 · see citing works
Recent Progress in Wearable Biosensors: From Healthcare Monitoring to Sports Analytics
Ye, S., Feng, S., Huang, L. J., & Bian, S. (2020) · Biosensors · doi:10.3390/bios10120205 · Cited by 146 · see citing works
Ries, A. L., Prewitt, L. M., & Johnson, J. J. (1989) · Chest · doi:10.1378/chest.96.2.287 · Cited by 145 · see citing works
Neonatal Pulse Oximetry: Accuracy and Reliability
Hay, W. W., Brockway, J., & Eyzaguirre, M. (1989) · PEDIATRICS · doi:10.1542/peds.83.5.717 · Cited by 141 · see citing works
Henry, N. R., Hanson, A. C., Schulte, P. J., Warner, N. S., Manento, M. N., Weister, T. J., & Warner, M. A. (2022) · Critical Care Medicine · doi:10.1097/ccm.0000000000005394 · Cited by 133 · see citing works
Motion Robust Remote-PPG in Infrared
van Gastel, M., Stuijk, S., & de Haan, G. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2390261 · Cited by 126 · see citing works
Reliability of Near-Infrared Spectroscopy in People With Dark Skin Pigmentation
Wassenaar, E. B., & van den Brand, J. G. H. (2005) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-005-1655-0 · Cited by 126 · see citing works
Algorithms for Monitoring Heart Rate and Respiratory Rate From the Video of a User’s Face
Sanyal, S., & Nundy, K. K. (2018) · IEEE Journal of Translational Engineering in Health and Medicine · doi:10.1109/jtehm.2018.2818687 · Cited by 125 · see citing works
Racial discrepancy in pulse oximeter accuracy in preterm infants
Vesoulis, Z., Tims, A., Lodhi, H., Lalos, N., & Whitehead, H. (2022) · Journal of Perinatology · doi:10.1038/s41372-021-01230-3 · Cited by 120 · see citing works
Valbuena, V. S. M., Barbaro, R. P., Claar, D. D., Valley, T. S., Dickson, R. P., Gay, S. E., Sjoding, M. W., & Iwashyna, T. J. (2021) · CHEST Journal · doi:10.1016/j.chest.2021.09.025 · Cited by 114 · see citing works
Reliability of noninvasive oximetry in black subjects during exercise and hypoxia
Zeballos, R. J., & Weisman, I. M. (1991) · American Review of Respiratory Disease · doi:10.1164/ajrccm/144.6.1240 · Cited by 108 · see citing works
Ear oximetry: accuracy and practicability in the assessment of arterial oxygenation.
Saunders, N. A., Powles, A. C. P., & Rebuck, A. S. (1976) · PubMed · doi:10.1164/arrd.1976.113.6.745 · Cited by 105 · see citing works
Effect of skin pigmentation on pulse oximetry accuracy in the emergency department
Adler, J. N., Hughes, L. A., Vtvilecchia, R., & Jr., C. A. C. (1998) · Academic Emergency Medicine · doi:10.1111/j.1553-2712.1998.tb02772.x · Cited by 102 · see citing works
Reflectance Photoplethysmography as Noninvasive Monitoring of Tissue Blood Perfusion
Abay, T. Y., & Kyriacou, P. A. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2417863 · Cited by 96 · see citing works
The Effect of Skin Pigmentation on the Accuracy of Pulse Oximetry in Infants with Hypoxemia
Foglia, E. E., Whyte, R. K., Chaudhary, A., Mott, A., Chen, J., Propert, K. J., & Schmidt, B. (2017) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2016.11.043 · Cited by 96 · see citing works
Pulse Oximeter Performance, Racial Inequity, and the Work Ahead
Okunlola, O. E., Lipnick, M. S., Batchelder, P. B., Bernstein, M., Feiner, J. R., & Bickler, P. E. (2022) · Respiratory Care · doi:10.4187/respcare.09795 · Cited by 96 · see citing works
Characterising the variations in ethnic skin colours: a new calibrated data base for human skin
Xiao, K., Yates, J. M., Zardawi, F., Sueeprasan, S., Liao, N., Gill, L., Li, C., & Wuerger, S. (2017) · Skin Research and Technology · doi:10.1111/srt.12295 · Cited by 95 · see citing works
Ebmeier, S. J., Barker, M., Bacon, M., Beasley, R. C., Bellomo, R., Chong, C. K., Eastwood, G. M., Gilchrist, J., Kagaya, H., Pilcher, J., et al. (2018) · Anaesthesia and Intensive Care · doi:10.1177/0310057x1804600307 · Cited by 92 · see citing works
The Use of Pulse Oximetry in the Assessment of Acclimatization to High Altitude
Dünnwald, T., Kienast, R., Niederseer, D., & Burtscher, M. (2021) · Sensors · doi:10.3390/s21041263 · Cited by 79 · see citing works
Response To: Investigating sources of inaccuracy in wearable optical heart rate sensors
Colvonen, P. J. (2021) · npj Digital Medicine · doi:10.1038/s41746-021-00408-5 · Cited by 65 · see citing works
Khanam, F.-T.-Z., Perera, A. G., Al‐Naji, A., Gibson, K., & Chahl, J. (2021) · Journal of Imaging · doi:10.3390/jimaging7080122 · Cited by 63 · see citing works
Non-Contact SpO2 Prediction System Based on a Digital Camera
Al-Naji, A., Khalid, G. A., Mahdi, J. F., & Chahl, J. (2021) · Applied Sciences · doi:10.3390/app11094255 · Cited by 61 · see citing works
GABRIELCZYK, M. R., & BUIST, R. J. (1988) · Anaesthesia · doi:10.1111/j.1365-2044.1988.tb09025.x · Cited by 59 · see citing works
Moço, A., & Verkruysse, W. (2021) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-019-00449-y · Cited by 55 · see citing works
Harskamp, R. E., Bekker, L., Himmelreich, J. C. L., Clercq, L. D., Karregat, E. P. M., Sleeswijk, M. E., & Lucassen, W. A. M. (2021) · BMJ Open Respiratory Research · doi:10.1136/bmjresp-2021-000939 · Cited by 55 · see citing works
Systematic Review on Human Skin-Compatible Wearable Photoplethysmography Sensors
Lee, I., Park, N., Lee, H., Hwang, C., Kim, J. H., & Park, S. (2021) · Applied Sciences · doi:10.3390/app11052313 · Cited by 52 · see citing works
A clinical evaluation of the accuracy of the Nellcor N-100 and Ohmeda 3700 pulse oximeters
Cecil, W. T., Thorpe, K. J., Fibuch, E. E., & Tuohy, G. F. (1988) · Journal of Clinical Monitoring · doi:10.1007/bf01618105 · Cited by 51 · see citing works
Accuracy of pulse oximetry in pigmented patients.
Bothma, P. A., Joynt, G. M., Lipman, J., Hon, H., Mathala, B., Scribante, J., & Kromberg, J. G. R. (1996) · PubMed · Cited by 50 · see citing works
COVID-19: Pulse oximeters in the spotlight
Michard, F., Shelley, K., & L’Her, E. (2021) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-020-00550-7 · Cited by 49 · see citing works
Non-Contact Oxygen Saturation Measurement Using YCgCr Color Space with an RGB Camera
Kim, N. H., Yu, S. G., Kim, S. E., & Lee, E. C. (2021) · Sensors · doi:10.3390/s21186120 · Cited by 48 · see citing works
Accuracy of pulse oximeters: the European multi-center trial.
Wouters, P. F., Gehring, H., Meyfroidt, G., et al. (2002) · PubMed · Cited by 47 · see citing works
Wiles, M. D., El‐Nayal, A., Elton, G., Malaj, M., Winterbottom, J., Gillies, C., Moppett, I. K., & Bauchmuller, K. (2022) · Anaesthesia · doi:10.1111/anae.15581 · Cited by 45 · see citing works
Yan, L., Hu, S., Alzahrani, A., Alharbi, S., & Blanos, P. (2017) · Biosensors · doi:10.3390/bios7020022 · Cited by 40 · see citing works
A characterization of motion affecting pulse oximetry in 350 patients.
Tobin, R. M., Pologe, J. A., & Batchelder, P. B. (2002) · PubMed · Cited by 38 · see citing works
Philip, K. E. J., Bennett, B., Fuller, S., Lonergan, B., McFadyen, C., Burns, J., Tidswell, R., & Vlachou, A. (2020) · BMJ Open Respiratory Research · doi:10.1136/bmjresp-2020-000778 · Cited by 35 · see citing works
Visible light and human skin pigmentation: The importance of skin phototype
Moreiras, H., O'Connor, C., Bell, M., & Tobin, D. J. (2021) · Experimental Dermatology · doi:10.1111/exd.14400 · Cited by 32 · see citing works
Lee, K. H., Hui, K. P., Tan, W. C., & Lim, T. K. (1993) · PubMed · Cited by 30 · see citing works
Çıçek, H., Gümüş, S., Deniz, O. F., et al. (2010) · Emergency Medicine Journal · doi:10.1136/emj.2010.096073 · Cited by 30 · see citing works
The correct measurement of oxygen saturation at high altitude
Tannheimer, M., & Lechner, R. (2019) · Sleep And Breathing · doi:10.1007/s11325-019-01784-9 · Cited by 30 · see citing works
Skin pigmentation as an influence on the accuracy of pulse oximetry.
Emery, J. R. (1987) · PubMed · Cited by 29 · see citing works
Oxygen Saturation Behavior by Pulse Oximetry in Female Athletes: Breaking Myths
Martín-Escudero, P., Cabanas, A. M., Fuentes, M. E., & Galindo-Canales, M. (2021) · Biosensors · doi:10.3390/bios11100391 · Cited by 25 · see citing works
Self-citationNoninvasive Monitoring and Assessment of Oxygenation in Infants
Poets, C. F. (2019) · Clinics in Perinatology · doi:10.1016/j.clp.2019.05.010 · Cited by 24 · see citing works
The Effect of Optical Crosstalk on Accuracy of Reflectance-Type Pulse Oximeter for Mobile Healthcare
Baek, H. J., Shin, J., & Cho, J. (2018) · Journal of Healthcare Engineering · doi:10.1155/2018/3521738 · Cited by 22 · see citing works
Accuracy of Pulse Oximetry in the Presence of Fetal Hemoglobin—A Systematic Review
Pritišanac, E., Urlesberger, B., Schwaberger, B., & Pichler, G. (2021) · Children · doi:10.3390/children8050361 · Cited by 22 · see citing works
Allado, E., Poussel, M., Moussu, A., Saunier, V., Bernard, Y., Albuisson, E., & Chenuel, B. (2021) · BMJ Open · doi:10.1136/bmjopen-2020-047896 · Cited by 22 · see citing works
Association between Pulpal-Periapical Pathology and Autoimmune Diseases: A Systematic Review
Gironés, J. G., Ros-Valverde, A., Pecci‐Lloret, M. P., et al. (2021) · Journal of Clinical Medicine · doi:10.3390/jcm10214886 · Cited by 20 · see citing works
A novel method based on two cameras for accurate estimation of arterial oxygen saturation
Liu, H., Ivanov, K., Wang, Y., & Wang, L. (2015) · BioMedical Engineering OnLine · doi:10.1186/s12938-015-0045-1 · Cited by 17 · see citing works
Kugelman, A., Wasserman, Y., Mor, F., et al. (2004) · Journal of Perinatology · doi:10.1038/sj.jp.7211102 · Cited by 16 · see citing works
The effect of varying LED intensity on pulse oximeter accuracy
De Kock, J. P., Reynolds, K. J., Tarassenko, L., & Moyles, J. T. B. (1991) · Journal of Medical Engineering & Technology · doi:10.3109/03091909109016208 · Cited by 16 · see citing works
Racial bias in pulse oximetry: More statistical detail may help tackle the problem
Philip, K. E. J., Tidswell, R., & McFadyen, C. (2021) · BMJ · doi:10.1136/bmj.n298 · Cited by 16 · see citing works
Nematswerani, N., Collie, S., Chen, T., et al. (2021) · South African Medical Journal · doi:10.7196/samj.2021.v111i10.15880 · Cited by 16 · see citing works
Lee, T. Q., Barnett, S. L., Shanfield, S. L., & Anzel, S. H. (1990) · The Journal of Rehabilitation Research and Development · doi:10.1682/jrrd.1990.10.0363 · Cited by 15 · see citing works
Gaskin, L., & Thomas, J. (1995) · Physiotherapy · doi:10.1016/s0031-9406(05)66819-8 · Cited by 14 · see citing works
The effect of henna paste on oxygen saturation reading obtained by pulse oximetry.
Al-Majed, S. A., & Harakati, M. S. (1994) · PubMed · Cited by 13 · see citing works
Best Practice & Research Clinical Anaesthesiology
Palmer, P. P., Royal, M. A., & Miller, R. D. (2009) · Cited by 13 · see citing works
Blanos, P., Hu, S., Mulvaney, D. J., & Alharbi, S. (2018) · Design and Quality for Biomedical Technologies · doi:10.1117/12.2287854 · Cited by 13 · see citing works
The Big Consequences of Small Discrepancies: Why Racial Differences in Pulse Oximetry Errors Matter
Holder, A. L., & Wong, A.-K. I. (2022) · Critical Care Medicine · doi:10.1097/ccm.0000000000005447 · Cited by 13 · see citing works
Stell, D., Noble, J. J., Kay, R. H., Kwong, M. T., Jeffryes, M. J. R., Johnston, L., Glover, G. W., & Akinluyi, E. A. (2022) · BMJ Open Respiratory Research · doi:10.1136/bmjresp-2021-001159 · Cited by 13 · see citing works
Noninvasive oximetry in pigmented patients.
Wang, Y. T., & Poh, S. C. (1985) · PubMed · Cited by 10 · see citing works
Henna dye: A cause of erroneous pulse oximetry readings
Bensghir, M., Houba, A., Hila, J. E., et al. (2013) · Saudi Journal of Anaesthesia · doi:10.4103/1658-354x.121052 · Cited by 9 · see citing works
Breuer, H.-W. M., Groeben, H., Schöndeling, H., et al. (1990) · International Journal of Sports Medicine · doi:10.1055/s-2007-1024756 · Cited by 8 · see citing works
Nasseri, N., Kleiser, S., Wolf, U., et al. (2018) · Journal of Biophotonics · doi:10.1002/jbio.201700367 · Cited by 6 · see citing works
Nematswerani, N., Collie, S., Chen, T., et al. (2021) · SSRN Electronic Journal · doi:10.2139/ssrn.3804758 · Cited by 6 · see citing works
Hunasikatti, M. (2021) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.9178 · Cited by 5 · see citing works
La oximetría en registro continuo en el esfuerzo máximo en distintas especialidades deportivas
Escudero, M. D. P. M. (2012) · Addiction · Cited by 3 · see citing works
Double junction photodiodes for multiwavelength photoplethysmography
Costa, J., Vieira, H., Louro, P., et al. (2018) · doi:10.1117/12.2307286 · Cited by 3 · see citing works
Flagged first, then by citations
Research works printed but not linked: 2
- Avant, 1997, G; Lowe N.; Torres Jr., A. Comparison of accuracy and signal consistency of two reusable pulse oximeter probes in critically ill children, Respir. Care, 42, 698
- Hameedullah, 2002, Henna paste and pulse oximetry: Effect of different methods of application, J. Anaesthesiol. Clin. Pharmacol., 18, 193
Cited by
Sorted by citations
Setchfield, K., Gorman, A., Simpson, A. H. R. W., Somekh, M. G., & Wright, A. J. (2024) · Journal of Biomedical Optics · doi:10.1117/1.jbo.29.1.010901 · Cited by 72 · see citing works
Singh, S., Bennett, M. R., Chen, C., Shin, S., Ghanbari, H., & Nelson, B. W. (2024) · Journal of Medical Internet Research · doi:10.2196/62769 · Cited by 36 · see citing works
Haxha, S., Nwibor, C., Ali, M., Sakel, M., Saunders, K., Dyo, V., & Nabakooza, S. (2024) · Sensors · doi:10.3390/s24113301 · Cited by 12 · see citing works
Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring
Asif, S., AlSaafeen, A., Nadar, S., Nambiar, S., Dannawi, J., Korrapati, H., & Wilkhoo, H. S. (2025) · Cureus · doi:10.7759/cureus.94074 · Cited by 9 · see citing works
- Authors
- Cabanas, A. M., Fuentes-Guajardo, M., Latorre, K., León, D., & Martín-Escudero, P.
- Affiliations
- University of Tarapacá (CL)
- Departamento de Tecnología Médica, Universidad de Tarapacá, Arica 1010069, Chile
- Universidad Complutense de Madrid (ES)
- Corresponding author
- Cabanas, Ana M.
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2022-04-29
- Volume
- 22
- Issue
- 9
- Pages
- 3402
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC9102088
- OpenAlex
- W4225276305
- Semantic Scholar
- cb2ae9e7d44263c7ee8c31d947e317d0aae398f7
- Cited by
- 124
- References
- 107
- Citations by year
- 2026: 15 · 2025: 29 · 2024: 43 · 2023: 34 · 2022: 3
- FWCI
- 2.43
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- UTA-Mayor Project (7721-20)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
- Keywords
- pulse oximetry, skin pigmentation, oxygen saturation measurement, measurement error, dark skin, low oxygen saturation, COVID-19 triage, bibliometric analysis, scientometrics
- MeSH terms
- COVID-19, Humans, Oximetry, Oxygen, Retrospective Studies, Skin Pigmentation, Bibliometrics, Pandemics
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review