The relationship between vacuum and hemolysis during catheter blood collection: a retrospective analysis of six large cohorts
The DOI matches its Crossref record; 26 of the 26 references it cites were found in a registry or an index.
Abstract
Background: Blood collection through intravenous (IV) catheters is a common practice at emergency departments (EDs). This technique is associated with higher in vitro hemolysis rates and may even be amplified by the use of vacuum collection tubes. Our aim was to investigate the association of five different vacuum tubes with hemolysis rates in comparison to an aspiration system under real-life conditions and to propose an equation to estimate the amount of hemolysis, depending on the vacuum collection tube type. Methods: We retrospectively evaluated hemolysis data of plasma samples from our ED, where blood is drawn through IV catheters. Over the past 5 years, we compared 19,001 hemolysis index values amongst each other and against the respective vacuum pressure (P v) of the collection tubes, which were used within the six observational periods. Results: The highest hemolysis rates were associated with full-draw evacuated tubes. Significantly reduced hemolysis was observed for two kinds of partial-draw tubes. The hemolysis rate of one partial-draw blood collection tube was comparable to those of the aspiration system. Regression analysis of P v and mean free hemoglobin (fHb) values yielded the formula fHb (g/L)=0.0082*P v2–0.1143*P v+ 0.5314 with an R 2 of 0.99. Conclusions: If IV catheters are used for blood collection, hemolysis rates directly correlate with the vacuum within the tubes and can be estimated by the proposed formula. By the use of partial-draw vacuum blood collection tubes, hemolysis rates in IV catheter collections can be reduced to levels comparable with collections performed by aspiration systems.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“We discourage blood sampling from an intravenous catheter, as it has been shown by many studies that catheter blood collection increases the risk of hemolysis [2], [3], [4].”
From Joint EFLM-COLABIOCLI Recommendation for venous blood sampling (2018) · cites this paper as [3]
Citations
Cites
Haemolysis: an overview of the leading cause of unsuitable specimens in clinical laboratories
Lippi, G., Blanckaert, N., Bonini, P., Green, S., Kitchen, S., Palicka, V., Vassault, A. J., & Plebani, M. (2008) · Clinical Chemistry and Laboratory Medicine · doi:10.1515/cclm.2008.170 · Cited by 466
Influence of hemolysis on routine clinical chemistry testing
Lippi, G., Luca Salvagno, G., Montagnana, M., Brocco, G., & Cesare Guidi, G. (2006) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2006.054 · Cited by 344
Hemolyzed specimens: a major challenge for emergency departments and clinical laboratories
Lippi, G., Plebani, M., Di Somma, S., & Cervellin, G. (2011) · Critical Reviews in Clinical Laboratory Sciences · doi:10.3109/10408363.2011.600228 · Cited by 226
Hemolyzed Specimens: A Reason for Rejection or a Clinical Challenge?
Carraro, P., Servidio, G., & Plebani, M. (2000) · Clinical Chemistry · doi:10.1093/clinchem/46.2.306 · Cited by 218
The IFCC Working Group on laboratory errors and patient safety
Sciacovelli, L., & Plebani, M. (2009) · Clinica Chimica Acta · doi:10.1016/j.cca.2009.03.025 · Cited by 161
Heyer, N. J., Derzon, J. H., Winges, L. D., Shaw, C. S., Mass, D., Snyder, S. R., Epner, P. L., Nichols, J. H., Gayken, J., Ernst, D. J., et al. (2012) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2012.08.002 · Cited by 141
Harmonization of automated hemolysis index assessment and use: Is it possible?
Dolci, A., & Panteghini, M. (2013) · Clinica Chimica Acta · doi:10.1016/j.cca.2013.10.012 · Cited by 114
Harmonization of pre-analytical quality indicators
Plebani, M., Sciacovelli, L., Aita, A., & Chiozza, M. L. (2014) · Biochemia Medica · doi:10.11613/bm.2014.012 · Cited by 113
Pre-analytical errors management in the clinical laboratory: a five-year study
Marín, Á. G., Rivas‐Ruiz, F., Hidalgo, M. D. M. P., & Molina-Mendoza, P. (2014) · Biochemia Medica · doi:10.11613/bm.2014.027 · Cited by 82
Lippi, G., Cervellin, G., & Mattiuzzi, C. (2013) · Biochemia Medica · doi:10.11613/bm.2013.022 · Cited by 66
Research works printed but not linked: 33
- 2008, [Factors related to haemolysis in the extraction of blood samples], An Sist Sanit Navar, 31, 153
- 2010, Hemolysis by peripheral intravenous catheters: materials comparison. RIMeL/IJLaM, 6, 216
- 2013, Low vacuum and discard tubes reduce hemolysis in samples drawn from intravenous catheters, Clin Biochem, 46, 1142, 10.1016/j.clinbiochem.2013.04.005 · DOI · Paper page
- 2016, Hemolysis rates in blood samples: differences between blood collected by clinicians and nurses and the effect of phlebotomy training, Clin Chem Lab Med, 54, 1987, 10.1515/cclm-2016-0175 · DOI
- 2004, Blood samples drawn from IV catheters have less hemolysis when 5-mL (vs 10-mL) collection tubes are used, J Emerg Nurs, 30, 529, 10.1016/j.jen.2004.10.004 · DOI
- 2013, Critical review and meta-analysis of spurious hemolysis in blood samples collected from intravenous catheters, Biochem Med (Zagreb), 23, 193
- 2009, The ifcc working group on laboratory errors and patient safety, Clin Chim Acta, 404, 79, 10.1016/j.cca.2009.03.025 · DOI · Paper page
- 2014, Harmonization of automated hemolysis index assessment and use: Is it possible?, Clin Chim Acta, 432, 38, 10.1016/j.cca.2013.10.012 · DOI
- 2006, Influence of hemolysis on routine clinical chemistry testing, Clin Chem Lab Med, 44, 311
- 2011, Hemolyzed specimens: a major challenge for emergency departments and clinical laboratories, Crit Rev Cl Lab Sci, 48, 143, 10.3109/10408363.2011.600228 · DOI · Paper page
- 2014, Low volume tubes are not effective to reduce the rate of hemolyzed specimens from the emergency department, Clin Biochem, 47, 227, 10.1016/j.clinbiochem.2013.12.004 · DOI
- 2014, Pre-analytical errors management in the clinical laboratory: a five-year study, Biochem Med (Zagreb), 24, 248, 10.11613/bm.2014.027 · DOI
- 2014, Low vacuum versus low volume tubes and the reduction of hemolyzed specimens from the emergency department, Clin Biochem, 47, 1344, 10.1016/j.clinbiochem.2014.06.016 · DOI
- 2016, In-vitro hemolysis and its financial impact using different blood collection systems, J Lab Med, 40, 49
- 2014, Low vacuum versus low volume tubes and the reduction of hemolyzed specimens from the emergency department, Clin Biochem, 47, 1344, 10.1016/j.clinbiochem.2014.06.016 · DOI
- 1861, A textbook on chemistry
- 2013, Use of butterfly needles to draw blood is independently associated with marked reduction in hemolysis compared to intravenous catheter, Acad Emerg Med, 20, 1151, 10.1111/acem.12245 · DOI
- 2013, Prevention of hemolysis in blood samples collected from intravenous catheters, Clin Biochem, 46, 561, 10.1016/j.clinbiochem.2013.01.021 · DOI
- 2012, Speed of sample transportation by a pneumatic tube system can influence the degree of hemolysis, Clin Chem Lab Med, 50, 471
- 2008, [Factors related to haemolysis in the extraction of blood samples], An Sist Sanit Navar, 31, 153
- 2014, Prevalence and cost of hemolyzed samples in a large urban emergency department, Int J Lab Hematol, 36, e24, 10.1111/ijlh.12135 · DOI · Paper page
- 2008, Haemolysis: an overview of the leading cause of unsuitable specimens in clinical laboratories, Clin Chem Lab Med, 46, 764
- 2015, The economic burden of hemolysis, Clin Chem Lab Med, 53, e285
- 2007, Measurements of free hemoglobin and hemolysis index: edta- or lithium-heparinate plasma?, Clin Chem, 53, 1717, 10.1373/clinchem.2007.091421 · DOI
- 2012, Effectiveness of practices to reduce blood sample hemolysis in EDS: a laboratory medicine best practices systematic review and meta-analysis, Clin Biochem, 45, 1012, 10.1016/j.clinbiochem.2012.08.002 · DOI
- 2009, A prospective study of causes of haemolysis during venepuncture: tourniquet time should be kept to a minimum, Ann Clin Biochem, 46, 244, 10.1258/acb.2009.008228 · DOI · Paper page
- 2011, Use of separate venipunctures for IV access and laboratory studies decreases hemolysis rates, Intern Emerg Med, 6, 357, 10.1007/s11739-011-0568-9 · DOI
- 2014, Low vacuum versus low volume tubes and the reduction of hemolyzed specimens from the emergency department, Clin Biochem, 47, 1344, 10.1016/j.clinbiochem.2014.06.016 · DOI
- 2015, The use of s-monovette is effective to reduce the burden of hemolysis in a large urban emergency department, Biochem Med (Zagreb), 25, 69
- 2000, Hemolyzed specimens: A reason for rejection or a clinical challenge?, Clin Chem, 46, 306, 10.1093/clinchem/46.2.306 · DOI · Paper page
- 2010, Hemolysis by peripheral intravenous catheters: materials comparison. RIMeL/IJLaM, 6, 216
- 1861, A textbook on chemistry
- 2014, Harmonization of pre–analytical quality indicators, Biochem Med (Zagreb), 24, 105
Cited by
Joint EFLM-COLABIOCLI Recommendation for venous blood sampling
Simundic, A.-M., Bölenius, K., Cadamuro, J., Church, S., Cornes, M. P., van Dongen-Lases, E. C., Eker, P., Erdeljanovic, T., Grankvist, K., Guimaraes, J. T., et al. (2018) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2018-0602 · Cited by 240
Self-citationCauses, consequences and management of sample hemolysis in the clinical laboratory
Heireman, L., Geel, P. V., Musger, L., Heylen, E., Uyttenbroeck, W., & Mahieu, B. (2017) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2017.09.013 · Cited by 172
Managing hemolyzed samples in clinical laboratories
Šimundić, A.-M., Baird, G. S., Cadamuro, J., Costelloe, S. J., & Lippi, G. (2019) · Critical Reviews in Clinical Laboratory Sciences · doi:10.1080/10408363.2019.1664391 · Cited by 165
Self-citationCadamuro, J., Cornes, M. P., Šimundić, A.-M., la Salle, B. D., Kristensen, G. B. B., Guimarães, J. T., Grankvist, K., Vermeersch, P., Nybo, M., van Dongen-Lases, E. C., et al. (2019) · Biochemia Medica · doi:10.11613/bm.2019.020705 · Cited by 71
Self-citationJoint EFLM-COLABIOCLI recommendation for venous blood sampling
Šimundić, A.-M., Bölenius, K., Cadamuro, J., Church, S., Cornes, M. P., van Dongen-Lases, E. C., Eker, P., Erdeljanovic, T., Grankvist, K., Guimarães, J. T., et al. (2019) · Annales de biologie clinique · doi:10.1684/abc.2019.1419 · Cited by 40
Self-citation
Metadata
- Authors
- Mrazek, C., Simundic, A.-M., Wiedemann, H., Krahmer, F., Felder, T. K., Kipman, U., Hoppe, U., Haschke-Becher, E., & Cadamuro, J.
- Affiliations
- Paracelsus Medical University (AT)
- Corresponding author
- Cadamuro, Janne
- Venue
- Clinical Chemistry and Laboratory Medicine (CCLM)
- Publisher
- Walter de Gruyter GmbH
- Published
- 2017-07-26
- Volume
- 55
- Issue
- 8
- Pages
- 1129-1134
- ISSN
- 1434-6621 (ISSN-L)
- 1437-4331
- OpenAlex
- W2579451927
- Semantic Scholar
- 3fb4f5bec1650dbf41a58839983b09fa5c540a03
- Cited by
- 24
- References
- 26
- Citations by year
- 2025: 3 · 2024: 3 · 2022: 3 · 2021: 1 · 2020: 3 · 2019: 6 · 2018: 3 · 2017: 2
- FWCI
- 2.83
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Clinical Chemistry and Laboratory Medicine (CCLM) (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Hematology › Blood transfusion and management
- Health Sciences › Medicine › Critical Care and Intensive Care Medicine › Central Venous Catheters and Hemodialysis
- Keywords
- hemolysis, vacuum blood collection tubes, vacuum pressure, free hemoglobin, hemolysis index, aspiration systems
- MeSH terms
- Blood Specimen Collection, Hemoglobins, Hemolysis, Humans, Retrospective Studies, Vacuum, Cohort Studies, Catheters
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: The relationship between vacuum and hemolysis during catheter blood collection: a retrospective analysis of six large cohorts. https://app.cochat.ai/papers/10.1515/cclm-2016-0940
- The paper
- Mrazek, C., Simundic, A.-M., Wiedemann, H., Krahmer, F., Felder, T. K., Kipman, U., Hoppe, U., Haschke-Becher, E., & Cadamuro, J. (2017). The relationship between vacuum and hemolysis during catheter blood collection: a retrospective analysis of six large cohorts. Clinical Chemistry and Laboratory Medicine (CCLM). https://doi.org/10.1515/cclm-2016-0940
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.