Practical recommendations for managing hemolyzed samples in clinical chemistry testing
The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
We suggest here a pragmatic approach for managing results of clinical chemistry testing in hemolyzed samples collected from adults/older children, attempting to balance the need to produce quality laboratory data with clinical urgency of releasing test results. Automatic measurement of the hemolysis index (H-index) in serum or plasma is highly advisable, whilst low-quality assessment of this test remains less good than a visual inspection. Regarding its practical use, when the H-index value does not generate an analytically significant bias, results can be released, whilst when the value is associated with analyte variation in a range between analytically and clinically significant bias (i.e. variation does not exceed the reference change value [RCV]), results of hemolysis-sensitive tests can be released in association with a comment describing the direction in which data are potentially altered, suggesting the need to collect another sample. When the H-index is associated with analyte variation exceeding clinically significant bias (i.e. variation exceeds the RCV), results of hemolysis-sensitive tests should be suppressed and replaced with a comment that biased results cannot be released because the sample is preanalytically compromised and advising the recollection of another sample. If H-index values reach an even higher critical cut-off (i.e. H-index corresponding to a cell-free hemoglobin concentration ≥10 g/L), all laboratory data may be unreliable and should hence be suppressed and replaced with a comment that all data cannot be released because the sample is grossly hemolyzed, also suggesting the recollection of another sample. Due to inaccuracy and imprecision, the use of corrective formulas for adjusting data of hemolysis-sensitive tests is discouraged.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Sandberg, S., Fraser, C. G., Horvath, A. R., Jansen, R., Jones, G., Oosterhuis, W., Petersen, P. H., Schimmel, H., Sikaris, K., & Panteghini, M. (2015) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2015-0067 · Cited by 563
Influence of hemolysis on routine clinical chemistry testing
Lippi, G., Luca Salvagno, G., Montagnana, M., Brocco, G., & Cesare Guidi, G. (2006) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2006.054 · Cited by 344 · cited 1×
“A useful example of this strategy, regarding practical management of potassium test results in a hemolyzed sample, is shown in Figure 4 [60].”
Self-citationFraser, C. G. (2012) · Clinical Chemistry and Laboratory Medicine · doi:10.1515/cclm.2011.733 · Cited by 286 · cited 3×
“Originally proposed by Harris and Yasaka [33] in 1983, the RCV was then largely endorsed by Fraser [34] in the following years.”
Ceriotti, F., Fernandez-Calle, P., Klee, G. G., Nordin, G., Sandberg, S., Streichert, T., Vives-Corrons, J.-L., & Panteghini, M. (2017) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2016-0091 · Cited by 181
A checklist for critical appraisal of studies of biological variation
Bartlett, W. A., Braga, F., Carobene, A., Coşkun, A., Prusa, R., Fernandez-Calle, P., Røraas, T., Jonker, N., Sandberg, S., & Variati, O. B. O. T. B. (2015) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2014-1127 · Cited by 151 · cited 2×
“and endorsed by the EFLM Working Group for Biological Variation (WG-BV) [35, 36], as shown in Figure 1A.”
Carobene, A., Marino, I., Coşkun, A., Serteser, M., Unsal, I., Guerra, E., Bartlett, W. A., Sandberg, S., Aarsand, A. K., Sylte, M. S., et al. (2017) · Clinical Chemistry · doi:10.1373/clinchem.2017.275115 · Cited by 77
Cornes, M. P., Church, S., van Dongen-Lases, E., Grankvist, K., Guimarães, J. T., Ibarz, M., Kovalevskaya, S., Kristensen, G. B., Lippi, G., Nybo, M., et al. (2016) · Annals of Clinical Biochemistry: International Journal of Laboratory Medicine · doi:10.1177/0004563216643969 · Cited by 43
Self-citation
Research works printed but not linked: 50
- Lippi G, Plebani M, Graber ML. Building a bridge to safe diagnosis in health care. The role of the clinical laboratory. Clin Chem Lab Med 2016;54:1–3. · DOI · Paper page
- Lippi G, Baird GS, Banfi G, Bölenius K, Cadamuro J, Church S, et al. Improving quality in the preanalytical phase through innovation, on behalf of the European Federation for Clinical Chemistry and Laboratory Medicine (EFLM) Working Group for Preanalytical Phase (WG-PRE). Clin Chem Lab Med 2017;55:489–500. · DOI
- Dikmen ZG, Pinar A, Akbiyik F. Specimen rejection in laboratory medicine: necessary for patient safety? Biochem Med (Zagreb) 2015;25:377–85. · DOI
- Lippi, G, Cervellin G, Favaloro EJ, Plebani M. In vitro and in vivo hemolysis. An unresolved dispute in laboratory medicine. Berlin, Germany: Walter de Gruyter GmbH, 2012. · doi:10.1515/9783110246148
- Simundic AM, Nikolac N, Guder WG. Preanalytical variation and preexamination processes. In: Rifai N, Horvath R, Wittwer C, editors. Tietz textbook of clinical chemistry and molecular diagnostics, 6th ed. Amsterdam, The Netherlands: Elsevier, 2018:81–120.
- Lippi G, Plebani M, Di Somma S, Cervellin G. Hemolyzed specimens: a major challenge for emergency departments and clinical laboratories. Crit Rev Clin Lab Sci 2011;48:143–53. · DOI · Paper page
- Cadamuro J, Fiedler GM, Mrazek C, Felder TK, Oberkofler H, Kipman U, et al. In-vitro hemolysis and its financial impact using different blood collection systems. J Lab Med 2016;40:49–55. · DOI
- Badrick T, Barden H, Callen S, Dimeski G, Gay S, Graham P, et al. Consensus statement for the management and reporting of haemolysed specimens. Clin Biochem Rev 2016;37:140–2.
- Lippi G, Banfi G, Buttarello M, Ceriotti F, Daves M, Dolci A, et al. Recommendations for detection and management of unsuitable samples in clinical laboratories. Clin Chem Lab Med 2007;45:728–36. · DOI
- Simundic AM, Topic E, Nikolac N, Lippi G. Hemolysis detection and management of hemolysed specimens. Biochem Med (Zagreb) 2010;20:154–9. · DOI · Paper page
- Cadamuro J, Mrazek C, Haschke-Becher E, Sandberg S. To report or not to report: a proposal on how to deal with altered test results in hemolytic samples. Clin Chem Lab Med 2017;55:1109–11. · DOI
- Lippi G, Cervellin G, Plebani M. Reporting altered test results in hemolyzed samples: is the cure worse than the disease? Clin Chem Lab Med 2017;55:1112–4. · DOI
- Cadamuro J, Simundic AM, Ajzner E, Sandberg S. A pragmatic approach to sample acceptance and rejection. Clin Biochem 2017;50:579–81. · DOI
- Ryder KW, Glick MR. Erroneous laboratory results from hemolyzed, icteric, and lipemic specimens. Clin Chem 1993;39:175–6. · doi:10.1093/clinchem/39.1.175
- Nikolac N. Lipemia: causes, interference mechanisms, detection and management. Biochem Med (Zagreb) 2014;24:57–67. · DOI · Paper page
- Antonelli G, Padoan A, Aita A, Sciacovelli L, Plebani M. Verification of examination procedures in clinical laboratory for imprecision, trueness and diagnostic accuracy according to ISO 15189:2012: a pragmatic approach. Clin Chem Lab Med 2017;55:1501–8. · DOI
- Plebani M, Lippi G. Uncertainty, quality, safety and accreditation in laboratory medicine. J Lab Precis Med 2017;2:80. · doi:10.21037/jlpm.2017.09.06
- Lippi G, Simundic AM. The EFLM strategy for harmonization of the preanalytical phase. Clin Chem Lab Med 2017. doi: 10.1515/cclm-2017-0277. [Epub ahead of print]. · DOI
- Simundic AM, Nikolac N, Ivankovic V, Ferenec-Ruzic D, Magdic B, Kvaternik M, et al. Comparison of visual vs. automated detection of lipemic, icteric and hemolyzed specimens: can we rely on a human eye? Clin Chem Lab Med 2009;47:1361–5. · DOI · Paper page
- Luksic AH, Nikolac Gabaj N, Miler M, Dukic L, Bakliza A, Simundic AM. Visual assessment of hemolysis affects patient safety. Clin Chem Lab Med 2018;56:574–81. · DOI · Paper page
- Simundic AM, Bilic-Zulle L, Nikolac N, Supak-Smolcic V, Honovic L, Avram S, et al. The quality of the extra-analytical phase of laboratory practice in some developing European countries and Mexico – a multicentric study. Clin Chem Lab Med 2011;49:215–28. · DOI · Paper page
- Li L, Vecellio E, Gay S, Lake R, Mackay M, Burnett L, et al. Making sense of a haemolysis monitoring and reporting system: a nationwide longitudinal multimethod study of 68 Australian laboratory participant organizations. Clin Chem Lab Med 2018;56:565–73. · DOI
- Howanitz PJ, Lehman CM, Jones BA, Meier FA, Horowitz GL. Practices for Identifying and Rejecting Hemolyzed Specimens Are Highly Variable in Clinical Laboratories. Arch Pathol Lab Med 2015;139:1014–9. · doi:10.5858/arpa.2014-0161-cp
- Lippi G, Cadamuro J. Visual assessment of sample quality: quo usque tandem? Clin Chem Lab Med 2018;56:513–5. · DOI
- Lippi G. Systematic assessment of the hemolysis index: pros and cons. Adv Clin Chem 2015;71:157–70. · DOI
- Perich C, Minchinela J, Ricós C, Fernández-Calle P, Alvarez V, Doménech MV, et al. Biological variation database: structure and criteria used for generation and update. Clin Chem Lab Med 2015;53:299–305. · DOI · Paper page
- Pineda-Tenor D, Laserna-Mendieta EJ, Timón-Zapata J, Rodelgo-Jiménez L, Ramos-Corral R, Recio-Montealegre A, et al. Biological variation and reference change values of common clinical chemistry and haematologic laboratory analytes in the elderly population. Clin Chem Lab Med 2013;51:851–62. · DOI · Paper page
- Harris EK, Yasaka T. On the calculation of a “reference change” for comparing two consecutive measurements. Clin Chem 1983;29:25–30. · DOI · Paper page
- Simundic AM, Kackov S, Miler M, Fraser CG, Petersen PH. Terms and symbols used in studies on biological variation: the need for harmonization. Clin Chem 2015;61:438–9. · DOI · Paper page
- Fraser CG, Lippi G, Plebani M. Reference change values may need some improvement but are invaluable tools in laboratory medicine. Clin Chem Lab Med 2012;50:963–4. · DOI
- Vermeer HJ, Thomassen E, de Jonge N. Automated processing of serum indices used for interference detection by the laboratory information system. Clin Chem 2005;51:244–7. · doi:10.1373/clinchem.2004.036301
- Monneret D, Mestari F, Atlan G, Corlouer C, Ramani Z, Jaffre J, et al. Hemolysis indexes for biochemical tests and immunoassays on Roche analyzers: determination of allowable interference limits according to different calculation methods. Scand J Clin Lab Invest 2015;75:162–9. · DOI
- Goyal T, Schmotzer CL. Validation of hemolysis index thresholds optimizes detection of clinically significant hemolysis. Am J Clin Pathol 2015;143:579–83. · DOI
- Feitosa MS, Bücker DH, Santos SM, Vasconcellos LS. Implementation of criteria for automatic release of clinical chemistry test results in a laboratory at an academic public hospital. J Bras Patol Med Lab 2016;52:149–56. · doi:10.5935/1676-2444.20160026
- Ko DH, Park HI, Hyun J, Kim HS, Park MJ, Shin DH. Utility of reference change values for delta check limits. Am J Clin Pathol 2017;148:323–9. · DOI
- Lippi G. Interference studies: focus on blood cell lysates preparation and testing. Clin Lab 2012;58:351–5.
- Carraro P, Servidio G, Plebani M. Hemolyzed specimens: a reason for rejection or a clinical challenge? Clin Chem 2000;46:306–7. · DOI · Paper page
- Lippi G, Pavesi F, Benegiamo A, Pipitone S. What do hemolyzed whole-blood specimens look like? Analysis with a CellaVision DM96 automated image analysis system. J Lab Autom 2015;20:60–3. · doi:10.1177/2211068214559644
- Vermeer HJ, Steen G, Naus AJ, Goevaerts B, Agricola PT, Schoenmakers CH. Correction of patient results for Beckman Coulter LX-20 assays affected by interference due to hemoglobin, bilirubin or lipids: a practical approach. Clin Chem Lab Med 2007;45:114–9. · DOI
- Hawkins RC. Correction and reporting of potassium results in haemolysed samples. Ann Clin Biochem 2006;43:88–9. · doi:10.1258/000456306775141849
- Jeffery J, Sharma A, Ayling RM. Detection of haemolysis and reporting of potassium results in samples from neonates. Ann Clin Biochem 2009;46:222–5. · doi:10.1258/acb.2009.008241
- Mansour MM, Azzazy HM, Kazmierczak SC. Correction factors for estimating potassium concentrations in samples with in vitro hemolysis: a detriment to patient safety. Arch Pathol Lab Med 2009;133:960–6. · DOI
- Lippi G, Avanzini P, Pavesi F, Bardi M, Ippolito L, Aloe R, et al. Studies on in vitro hemolysis and utility of corrective formulas for reporting results on hemolyzed specimens. Biochem Med (Zagreb) 2011;21:297–305. · DOI
- Lippi G, Salvagno GL, Guidi GC. Mean corpuscular volume and red blood cell distribution width are independent predictors of serum potassium concentration in healthy individuals. Clin Chim Acta 2015;446:117–8. · doi:10.1016/j.cca.2015.04.017
- Zwart A, van Assendelft OW, Bull BS, England JM, Lewis SM, Zijlstra WG. ICSH Recommendations for reference method for hemoglobinometry in human blood (ICSH standards 1995) and specifications for international hemoglobincyanamyde standard. J Clin Pathol 1996;49:271–4. · doi:10.1136/jcp.49.4.271
- Lippi G, Luca Salvagno G, Blanckaert N, Giavarina D, Green S, Kitchen S, et al. Multicenter evaluation of the hemolysis index in automated clinical chemistry systems. Clin Chem Lab Med 2009;47:934–9. · DOI
- Cadamuro J, von Meyer A, Wiedemann H, Klaus Felder T, Moser F, Kipman U, et al. Hemolysis rates in blood samples: differences between blood collected by clinicians and nurses and the effect of phlebotomy training. Clin Chem Lab Med 2016;54:1987–92. · DOI
- Sciacovelli L, Panteghini M, Lippi G, Sumarac Z, Cadamuro J, Galoro CA, et al. Defining a roadmap for harmonizing quality indicators in Laboratory Medicine: a consensus statement on behalf of the IFCC Working Group “Laboratory Error and Patient Safety” and EFLM Task and Finish Group “Performance specifications for the extra-analytical phases”. Clin Chem Lab Med 2017;55:1478–88. · DOI
- Lippi G, Sciacovelli L, Simundic AM, Plebani M. Innovative software for recording preanalytical errors in accord with the IFCC quality indicators. Clin Chem Lab Med 2017;55:e51–3. · DOI
- Lippi G, Adcock D, Simundic AM, Tripodi A, Favaloro EJ. Critical laboratory values in hemostasis: toward consensus. Ann Med 2017;49:455–61. · DOI
Cited by
Lippi, G., von Meyer, A., Cadamuro, J., & Simundic, A.-M. (2019) · Diagnosis · doi:10.1515/dx-2018-0018 · Cited by 135
Self-citationA deep learning-based system for assessment of serum quality using sample images
Yang, C., Li, D., Sun, D., Zhang, S., Zhang, P., Xiong, Y., Zhao, M., Qi, T., Situ, B., & Zheng, L. (2022) · Clinica Chimica Acta · doi:10.1016/j.cca.2022.04.010 · Cited by 19
Measuring haemolysis in cattle serum by direct UV–VIS and RGB digital image-based methods
Larrán, B., López-Alonso, M., Miranda, M., Pereira, V., Rigueira, L., Suárez, M. L., & Herrero-Latorre, C. (2022) · Scientific Reports · doi:10.1038/s41598-022-17842-4 · Cited by 7
Metadata
- Authors
- Lippi, G., Cadamuro, J., von Meyer, A., & Simundic, A.-M.
- Venue
- Clinical Chemistry and Laboratory Medicine (CCLM)
- Publisher
- Walter de Gruyter GmbH
- Published
- 2018-01-26
- Volume
- 56
- Issue
- 5
- Pages
- 718-727
- ISSN
- 1434-6621 (ISSN-L)
- 1437-4331
- OpenAlex
- W2782172861
- Cited by
- 144
- References
- 7
- FWCI
- 10.24
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Practical recommendations for managing hemolyzed samples in clinical chemistry testing. https://app.cochat.ai/papers/10.1515/cclm-2017-1104
- The paper
- Lippi, G., Cadamuro, J., von Meyer, A., & Simundic, A.-M. (2018). Practical recommendations for managing hemolyzed samples in clinical chemistry testing. Clinical Chemistry and Laboratory Medicine (CCLM). https://doi.org/10.1515/cclm-2017-1104
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.