Provisional standardization of hepcidin assays: creating a traceability chain with a primary reference material, candidate reference method and a commutable secondary reference material
Open accessThe DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Abstract
Background: Hepcidin concentrations measured by various methods differ considerably, complicating interpretation. Here, a previously identified plasma-based candidate secondary reference material (csRM) was modified into a serum-based two-leveled sRM. We validated its functionality to increase the equivalence between methods for international standardization. Methods: We applied technical procedures developed by the International Consortium for Harmonization of Clinical Laboratory Results. The sRM, consisting of lyophilized serum with cryolyoprotectant, appeared commutable among nine different measurement procedures using 16 native human serum samples in a first round robin (RR1). Harmonization potential of the sRM was simulated in RR1 and evaluated in practice in RR2 among 11 measurement procedures using three native human plasma samples. Comprehensive purity analysis of a candidate primary RM (cpRM) was performed by state of the art procedures. The sRM was value assigned with an isotope dilution mass spectrometry-based candidate reference method calibrated using the certified pRM. Results: The inter-assay CV without harmonization was 42.1% and 52.8% in RR1 and RR2, respectively. In RR1, simulation of harmonization with sRM resulted in an inter-assay CV of 11.0%, whereas in RR2 calibration with the material resulted in an inter-assay CV of 19.1%. Both the sRM and pRM passed international homogeneity criteria and showed long-term stability. We assigned values to the low (0.95±0.11 nmol/L) and middle concentration (3.75±0.17 nmol/L) calibrators of the sRM. Conclusions: Standardization of hepcidin is possible with our sRM, which value is assigned by a pRM. We propose the implementation of this material as an international calibrator for hepcidin.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Immunoassay for human serum hepcidin
Ganz, T., Olbina, G., Girelli, D., Nemeth, E., & Westerman, M. (2008) · Blood · doi:10.1182/blood-2008-02-139915 · Cited by 688
Self-citationHepcidin in the diagnosis of iron disorders
Girelli, D., Nemeth, E., & Swinkels, D. W. (2016) · Blood · doi:10.1182/blood-2015-12-639112 · Cited by 417 · cited 6×
“for guidance in oral iron therapy and diagnosis of iron refractory iron deficiency anemia [2, 6, 7].”
BackgroundMethodInfluentialSelf-citationA dual-monoclonal sandwich ELISA specific for hepcidin-25
Butterfield, A. M., Luan, P., Witcher, D. R., Manetta, J., Murphy, A. T., Wroblewski, V. J., & Konrad, R. J. (2010) · Clinical Chemistry · doi:10.1373/clinchem.2010.151522 · Cited by 59 · cited 1×
“[23] developed by ELISA of Corgenix Inc.”
Cited by
Pasricha, S.-R., Tye-Din, J., Muckenthaler, M. U., & Swinkels, D. W. (2021) · The Lancet · doi:10.1016/s0140-6736(20)32594-0 · Cited by 1,069
Self-citationHepcidin and Iron in Health and Disease
Nemeth, E., & Ganz, T. (2023) · Annual Review of Medicine · doi:10.1146/annurev-med-043021-032816 · Cited by 457
Unraveling Hepcidin Plasma Protein Binding: Evidence from Peritoneal Equilibration Testing
Diepeveen, L. E., Laarakkers, C. M., Peters, H. P. E., van Herwaarden, A. E., Groenewoud, H., IntHout, J., Wetzels, J. F., van Swelm, R. P. L., & Swinkels, D. W. (2019) · Pharmaceuticals · doi:10.3390/ph12030123 · Cited by 11
“of quantification of the freely circulating hepcidin relative to the protein bound compound between assays may influence interpretation and comparability of hepcidin measurement and hampers recent achievements on standardization of hepcidin assays [13].”
BackgroundSelf-citationAlfaro-Magallanes, V., Benito, P., Rael, B., Barba-Moreno, L., Romero-Parra, N., Cupeiro, R., Swinkels, D., Laarakkers, C., Peinado, A., & on behalf of the IronFEMME Study Group (2020) · Nutrients · doi:10.3390/nu12123866 · Cited by 5
“Serum hepcidin measurements were performed by a combination of weak cation exchange chromatography and time-of-flight mass spectrometry (WCX-TOF MS) using a stable hepcidin-25+40 isotope and a secondary reference material [48] as internal standard for quantification [49].”
MethodSelf-citation
Metadata
- Authors
- Diepeveen, L. E., Laarakkers, C. M. M., Martos, G., Pawlak, M. E., Uğuz, F. F., Verberne, K. E. S. A., van Swelm, R. P. L., Klaver, S., de Haan, A. F. J., Pitts, K. R., Bansal, S. S., Abbas, I. M., Fillet, M., Lefebvre, T., Geurts-Moespot, A. J., Girelli, D., Castagna, A., Herkert, M., Itkonen, O., Olbina, G., Tomosugi, N., Westerman, M. E., Delatour, V., Weykamp, C. W., & Swinkels, D. W.
- Venue
- Clinical Chemistry and Laboratory Medicine (CCLM)
- Publisher
- Walter de Gruyter GmbH
- Published
- 2018-11-29
- Volume
- 57
- Issue
- 6
- Pages
- 864-872
- ISSN
- 1434-6621 (ISSN-L)
- 1437-4331
- OpenAlex
- W2902581126
- Cited by
- 63
- References
- 3
- FWCI
- 3.86
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.