Wearable Lactate Threshold Predicting Device is Valid and Reliable in Runners
Open accessThe DOI matches its Crossref record; 23 of the 23 references it cites were found in a registry or an index; none of the 10 sources its claims cite is checked for support yet.
Abstract
Borges, NR and Driller, MW. Wearable lactate threshold predicting device is valid and reliable in runners. J Strength Cond Res 30(8): 2212–2218, 2016—A commercially available device claiming to be the world's first wearable lactate threshold predicting device (WLT), using near-infrared LED technology, has entered the market. The aim of this study was to determine the levels of agreement between the WLT-derived lactate threshold workload and traditional methods of lactate threshold (LT) calculation and the interdevice and intradevice reliability of the WLT. Fourteen (7 male, 7 female; mean ± SD ; age: 18–45 years, height: 169 ± 9 cm, mass: 67 ± 13 kg, V̇ o 2 max: 53 ± 9 ml·kg −1 ·min −1 ) subjects ranging from recreationally active to highly trained athletes completed an incremental exercise test to exhaustion on a treadmill. Blood lactate samples were taken at the end of each 3-minute stage during the test to determine lactate threshold using 5 traditional methods from blood lactate analysis which were then compared against the WLT predicted value. In a subset of the population ( n = 12), repeat trials were performed to determine both inter-reliability and intrareliability of the WLT device. Intraclass correlation coefficient (ICC) found high to very high agreement between the WLT and traditional methods (ICC > 0.80), with TEMs and mean differences ranging between 3.9–10.2% and 1.3–9.4%. Both interdevice and intradevice reliability resulted in highly reproducible and comparable results (CV < 1.2%, TEM <0.2 km·h −1 , ICC > 0.97). This study suggests that the WLT is a practical, reliable, and noninvasive tool for use in predicting LT in runners.
Overview
CoChat Summary
TL;DR
The wearable lactate threshold predicting device (WLT; BSXinsight multisport edition) showed no significant differences from five traditional blood lactate methods for treadmill workload at lactate threshold and had high test-retest and interdevice reliability in runners. It was most closely correlated with the 4 mmol/L method, though it tended to slightly overestimate LT workload compared with traditional methods.
Key findings
- One-way ANOVA found no statistically significant differences between LT workloads determined by the WLT and traditional methods (p = 0.60); ICCs ranged from 0.80 to 0.98.
- Highest agreement was with the 4 mmol/L method: ICC = 0.98, TEM = 0.48 km/h (3.9%), mean difference = 0.18 km/h (1.3%), 95% limits of agreement −0.82 to 1.18 km/h.
- Interdevice reliability (2 devices during one test, n = 12): CV = 1.2% (0.9–2.1), TEM = 0.16 km/h (0.12–0.28), ICC = 0.97 (0.90–0.99); intradevice reliability (same device, two tests ~24 h apart): CV = 1.2% (0.9–2.1), TEM = 0.16 km/h (0.11–0.27), ICC = 0.97 (0.90–0.99).
- The WLT tended to overestimate LT workload versus traditional methods; Bland-Altman plots showed outliers versus Dmax and modified Dmax and a downward trend versus the 1 mmol/L method with increasing workload.
Methods
- Validity/reliability study in 14 runners (7 male, 7 female; VO2max range 39.7–67.0 ml/kg/min; LT speeds 10.9–16.2 km/h); reliability subset n = 12.
- Treadmill graded exercise test to exhaustion with 3-minute stages set by WLT software, warm-up at 4.8 km/h, second stage 9.3–11.7 km/h, increments 0.3–1.1 km/h until the subject could not maintain the workload; minimum 4 stages required. Blood lactate was collected by finger-prick capillary samples in the final 10 seconds of each stage and 1 minute postexercise; the WLT NIRS device was fitted over the gastrocnemius.
- Traditional LT methods: linear spline fitting, Dmax, modified Dmax, fixed 4 mmol/L, and first rise >1 mmol/L. Analysis used 1-way ANOVA, ICC with 95% CI, TEM, Bland-Altman plots; reliability data were log-transformed and reported as CV%, TEM, and ICC.
Limitations
- Stated: the patented WLT design and algorithm make the exact LT prediction methodology difficult to evaluate; this is the first study comparing the WLT with traditional methods, so direct comparisons with existing research are difficult; heterogeneous NIRS methodology (device, measurement site, parameter) may affect compatibility of results.
- Stated: the range of LT values produced by traditional methods (12.3–13.3 km/h) makes it difficult to conclude whether the WLT provides a valid prediction of LT; the WLT showed small overestimation and method-dependent agreement (outliers vs Dmax/modified Dmax, trend vs 1 mmol/L).
- Evident: small sample (14 total; 12 reliability) and testing limited to treadmill running, gastrocnemius placement, and threshold range 10.9–16.2 km/h, which may limit generalizability.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
0 Valid · 0 Not valid · 8 Not checked · 23 Own results and interpretations
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“One-way ANOVA showed no statistically significant differences between the LT workloads determined by the different methods (p = 0.60).”
§ Wearable Lactate Threshold Predicting Device is Valid and Reliable in Runners > RESULTS
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Intraclass correlation coefficients with 95% CI ranged from 0.80 to 0.98 (20.10 to 0.99) for workload at LT calculated by the WLT and traditional methods, and are shown in Table 1.”
§ Wearable Lactate Threshold Predicting Device is Valid and Reliable in Runners > RESULTS
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“When comparing the interdevice reliability during a single test in 12 separate trials, the WLT resulted in a CV and 95% CI of 1.2% (0.9-2.1), a TEM of 0.16 km$h 21 (0.12-0.28) and an ICC of 0.97 (0.90-0.99).”
§ Wearable Lactate Threshold Predicting Device is Valid and Reliable in Runners > RESULTS
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“However, this magnitude of agreement was not apparent with all traditional methods.”
§ Wearable Lactate Threshold Predicting Device is Valid and Reliable in Runners > DISCUSSION
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Although showing high levels of agreement through ICCs, mean difference values reveal a tendency of the WLT to overestimate LT workload when compared with traditional methods (Figure 3).”
§ Wearable Lactate Threshold Predicting Device is Valid and Reliable in Runners > DISCUSSION
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
7 more claims
Create a free account to see all 31 claims and every source verdict.
- It may state two separate claims as one. 13
- Probably not a substantial claim. 10
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“BSX Insight, which determines the threshold by making a comparison with blood lactate concentration, presented good values of ICC, although this used a patented method to determine MOT based on the inflection point of SmO2 during incremental testing45.”
From Reliability of threshold determination using portable muscle oxygenation monitors during exercise testing: a systematic review and meta-analysis (2023) · cites this paper as [45]
In the citing paper
“The articles included in this systematic review use different methods for determining thresholds: BSX Insight (20%, N = 3)45,47,49, double linear regression (46%, N = 7)16,26,39,41–44, visual method17,25,40, Dmax or modified Dmax46 and applications of devices Humon Beta48.”
From Reliability of threshold determination using portable muscle oxygenation monitors during exercise testing: a systematic review and meta-analysis (2023) · cites this paper as [45]
In the citing paper
“The second threshold was determined using the blood lactate concentration and muscle oxygenation in different sports such as cycling16,17,46–48, running44,45,49 and rowing26.”
From Reliability of threshold determination using portable muscle oxygenation monitors during exercise testing: a systematic review and meta-analysis (2023) · cites this paper as [45]
In the citing paper
“Vastus lateralis was the muscle most assessed16,25,26,40–44,46, although other muscles such as gastrocnemius44,45,47,49, rectus femoris17, biceps femoris44, lateral deltoid43 or intercostal39 were also evaluated.”
From Reliability of threshold determination using portable muscle oxygenation monitors during exercise testing: a systematic review and meta-analysis (2023) · cites this paper as [45]
In the citing paper
“Some of the articles included in this review demonstrate mean bias between MOT2 and LT2 or VT2 ranging from 0.01 and 0.4 km·h−125,44,45,49, between 3.9 and 15.4 W39,41, 0.05 W·kg−117 and 10.7% of the power output26.”
From Reliability of threshold determination using portable muscle oxygenation monitors during exercise testing: a systematic review and meta-analysis (2023) · cites this paper as [45]
1 more claim about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Endurance exercise performance: the physiology of champions
Joyner, M. J., & Coyle, E. F. (2008) · The Journal of Physiology · doi:10.1113/jphysiol.2007.143834 · Cited by 1,307 · cited 1×
“Alongside a high V_ O2max and efficiency, a high LT is essential for endurance performance as it allows an athlete to perform at a higher work-rate for an extended duration (19).”
BackgroundMonitoring tissue oxygen availability with near infrared spectroscopy (NIRS) in health and disease
Boushel, R., Langberg, H., Olesen, J., Gonzales‐Alonzo, J., Bülow, J., & Kjær, M. (2001) · Scandinavian Journal of Medicine & Science in Sports · doi:10.1034/j.1600-0838.2001.110404.x · Cited by 469 · cited 1×
“Although the use of NIRS is relatively new to sports science, it has proven to be a viable noninvasive tool to determine peripheral muscle oxygenation during exercise (4,22).”
MethodGrassi, B., Pogliaghi, S., Rampichini, S., Quaresima, V., Ferrari, M., Marconi, C., & Cerretelli, P. (2003) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00695.2002 · Cited by 429 · cited 5×
“This study provides the first data on the validity and reliability of the WLT’s ability to predict lactate threshold in runners.”
ResultInfluentialA New Approach for the Determination of Ventilatory and Lactate Thresholds
Cheng, B., Kuipers, H., Snyder, A., Keizer, H., Jeukendrup, A., & Hesselink, M. (1992) · International Journal of Sports Medicine · doi:10.1055/s-2007-1021309 · Cited by 371 · cited 2×
“Cond Res 30(8): 2212–2218, 2016—A commercially avail-”
MethodEvaluation of three portable blood lactate analysers: Lactate Pro, Lactate Scout and Lactate Plus
Tanner, R. K., Fuller, K. L., & Ross, M. L. R. (2010) · European Journal of Applied Physiology · doi:10.1007/s00421-010-1379-9 · Cited by 274 · cited 1×
“2% (26), an intradevice and interdevice CVof ;1.”
BackgroundBlood lactate accumulation and muscle deoxygenation during incremental exercise
Grassi, B., Quaresima, V., Marconi, C., Ferrari, M., & Cerretelli, P. (1999) · Journal of Applied Physiology · doi:10.1152/jappl.1999.87.1.348 · Cited by 248 · cited 7×
“This study provides the first data on the validity and reliability of the WLT’s ability to predict lactate threshold in runners.”
BackgroundInfluentialAccuracy and reliability of the ParvoMedics TrueOne 2400 and MedGraphics VO2000 metabolic systems
Crouter, S. E., Antczak, A., Hudak, J. R., DellaValle, D. M., & Haas, J. D. (2006) · European Journal of Applied Physiology · doi:10.1007/s00421-006-0255-0 · Cited by 240 · cited 1×
“Given, it has been suggested that a CV of ,10% is acceptable for reliability in other sport science measurement tools (1), such as metabolic analyzers (9), and the intradevice and interdevice CV of the one of the most commonly used blood lactate analyzers (Lactate Pro; Arkray Global Business Inc.”
MethodTest Validation in Sport Physiology: Lessons Learned From Clinimetrics
Impellizzeri, F. M., & Marcora, S. M. (2009) · International Journal of Sports Physiology and Performance · doi:10.1123/ijspp.4.2.269 · Cited by 210
Blood lactate parameters related to aerobic capacity and endurance performance
Yoshida, T., Chida, M., Ichioka, M., & Suda, Y. (1987) · European Journal of Applied Physiology and Occupational Physiology · doi:10.1007/bf00696368 · Cited by 183 · cited 1×
“F or decades, lactate threshold (LT) has been utilized by endurance athletes as a key performance predictor and to dictate training practices (29).”
MethodPhysiological Responses to Shuttle Repeated-Sprint Running
Buchheit, M., Bishop, D., Haydar, B., Nakamura, F. Y., & Ahmaidi, S. (2010) · International Journal of Sports Medicine · doi:10.1055/s-0030-1249620 · Cited by 163 · cited 1×
“With further technological advancements, the use of portable NIRS systems to monitor peripheral muscle oxygenation in field testing has become more common (6).”
Research works printed but not linked: 1
- What is an acceptable amount of measurement error? The application of meaningful "analytical goals" to the reliability of sports science measurements made on a ratio scale. G Atkinson, A Nevill, B Edwards, J Sports Sci. 1718Atkinson, G, Nevill, A, and Edwards, B. What is an acceptable amount of measurement error? The application of meaningful "analytical goals" to the reliability of sports science measurements made on a ratio scale. J Sports Sci 17: 18, 1999.
Cited by
Peake, J. M., Kerr, G., & Sullivan, J. P. (2018) · Frontiers in Physiology · doi:10.3389/fphys.2018.00743 · Cited by 523
“The BSX Insight wearable sleeve has been tested independently (Borges and Driller, 2016).”
MethodWearable Training-Monitoring Technology: Applications, Challenges, and Opportunities
Cardinale, M., & Varley, M. C. (2017) · International Journal of Sports Physiology and Performance · doi:10.1123/ijspp.2016-0423 · Cited by 221
Muscle Oximetry in Sports Science: A Systematic Review
Perrey, S., & Ferrari, M. (2018) · Sports Medicine · doi:10.1007/s40279-017-0820-1 · Cited by 220
Dose-Response Matters! - A Perspective on the Exercise Prescription in Exercise-Cognition Research
Herold, F., Müller, P., Gronwald, T., & Müller, N. G. (2019) · Frontiers in Psychology · doi:10.3389/fpsyg.2019.02338 · Cited by 196
Van Hooren, B., Goudsmit, J., Restrepo, J., & Vos, S. (2020) · Journal of Sports Sciences · doi:10.1080/02640414.2019.1690960 · Cited by 141
Düking, P., Achtzehn, S., Holmberg, H., & Sperlich, B. (2018) · Sensors · doi:10.3390/s18051632 · Cited by 102
“Levels of lactate can be monitored easily by both POCT devices and Wearables [59] and different lactate thresholds have been utilized as estimates of endurance performance [60].”
MethodFarzam, P., Starkweather, Z., & Franceschini, M. A. (2018) · Physiological Reports · doi:10.14814/phy2.13664 · Cited by 107
Noninvasive In-Situ Measurement of Blood Lactate Using Microwave Sensors
Mason, A., Korostynska, O., Louis, J. B., Cordova-Lopez, L. E., Abdullah, B. M., Greene, J., Connell, R., & Hopkins, J. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2017.2715071 · Cited by 65
“This is a validated [22] wear-”
Predictors of feelings of energy differ from predictors of fatigue
Boolani, A., O’Connor, P. J., Reid, J., Ma, S., & Mondal, S. (2018) · Fatigue Biomedicine Health & Behavior · doi:10.1080/21641846.2018.1558733 · Cited by 57
“Athletics, Texas, USA) which has been supported by independent reliability and validity data [91,92].”
BackgroundPerrey, S. (2022) · Frontiers in Sports and Active Living · doi:10.3389/fspor.2022.864825 · Cited by 53
“It has been emphasized that NIRS measurements may represent a simple, safe, reliable, and fast way to determine training intensity zones based on the determination of several thresholds representing shifts in the metabolic status of the working muscle (Grassi et al., 1999; Racinais et al., 2014; Borges and Driller, 20…”
BackgroundSendra-Pérez, C., Sanchez-Jimenez, J. L., Marzano-Felisatti, J. M., Encarnación-Martínez, A., Salvador-Palmer, R., & Priego-Quesada, J. I. (2023) · Scientific Reports · doi:10.1038/s41598-023-39651-z · Cited by 52
“The second threshold was determined using the blood lactate concentration and muscle oxygenation in different sports such as cycling 16,17,46–48 , running 44,45,49 and rowing 26 .”
BackgroundMethodInfluentialWilkinson, T. J., White, A. E. M., Nixon, D. G. D., Gould, D. W., Watson, E. L., & Smith, A. C. (2018) · Clinical and Experimental Nephrology · doi:10.1007/s10157-018-1612-0 · Cited by 30
Klusiewicz, A., Rębiś, K., Ozimek, M., & Czaplicki, A. (2021) · Biology of Sport · doi:10.5114/biolsport.2021.103571 · Cited by 24
Active video gaming in primary ciliary dyskinesia: a randomized controlled trial
Sonbahar‐Ulu, H., İnce, D. İ., Sağlam, M., Cakmak-Onal, A., Vardar‐Yağlı, N., Kutukcu, E. C., Sümer, E., & Özçelik, U. (2022) · European Journal of Pediatrics · doi:10.1007/s00431-022-04490-z · Cited by 24
“The SMO2 during ISWT and Glittre ADL test was evaluated using a wearable lactate threshold device (BSXinsight, BSX Athletics, Boston, USA) [21].”
MethodEstimation of lactate threshold with machine learning techniques in recreational runners
Etxegarai, U., Portillo, E., Irazusta, J., Arriandiaga, A., & Cabanes, I. (2017) · Applied Soft Computing · doi:10.1016/j.asoc.2017.11.036 · Cited by 25
“[20] is among this type of strategies.”
Sendra-Pérez, C., Encarnacion-Martinez, A., Salvador-Palmer, R., Murias, J. M., & Priego-Quesada, J. I. (2025) · European Journal of Applied Physiology · doi:10.1007/s00421-024-05593-1 · Cited by 16
“…2021), most of the studies examining NIRS break-points have only assessed one muscle, such as the vastus lateralis (Cayot et al. 2021), rectus femoris (Salas-Montoro et al. 2022), deltoideus (Yogev et al. 2022), gastrocnemius (Borges & Driller 2016), or intercostals (Contreras-Briceño et al. 2022).”
Dik, J., Sağlam, M., Tekerlek, H., Vardar‐Yağlı, N., Kutukcu, E. C., İnce, D. İ., Arıkan, H., Polat, S. E., & Doğru, D. (2020) · Pediatric Pulmonology · doi:10.1002/ppul.24903 · Cited by 18
“Dyspnea, fatigue, and leg fatigue perceptions were evaluated using the Modified Borg Scale(16); muscle oxygenation (SmO2) was recorded using a wearable lactate threshold (WLT) device (BSXinsight; BSX Athletics, Boston).(17) The completed distance during ISWT was recorded as meters, and peak oxygen consumption (VO2) wa…”
MethodCarrier, B., Helm, M. M., Cruz, K., Barrios, B., & Navalta, J. W. (2023) · Technologies · doi:10.3390/technologies11030071 · Cited by 21
“Investigations into WT to determine LT has had positive results, with the Humon Hex (Humon, Cambridge, MA, USA) and the BSX Insight (BSX Athletics, Austin, TX, USA) both being found sufficiently accurate in their estimates [25,26].”
BackgroundMethod(2021) · International Journal of Exercise Science · doi:10.70252/hbra1900 · Cited by 19
“However, previous investigations that have demonstrated valid NIRS threshold measurements, have done so in trained populations using proprietary algorithms that identify the NIRS threshold from the exercising StO2 data (2, 9).”
BackgroundMethodInfluentialEvaluating a new wearable lactate threshold sensor in recreational to highly trained cyclists
Driller, M., Borges, N., & Plews, D. (2016) · Sports Engineering · doi:10.1007/s12283-016-0198-6 · Cited by 15
Self-citationRębiś, K., Sadowska, D., Starczewski, M., & Klusiewicz, A. (2022) · Frontiers in Physiology · doi:10.3389/fphys.2022.809864 · Cited by 12
“, 2007) and to determine so-called training intensity zones (Borges and Driller, 2016; Driller et al., 2016; Crum et al., 2017).”
BackgroundMuscle blood content and muscle oxygen saturation in response to head down and head up tilt
Çotuk, H. B., Duru, A. D., Pelvan, O., & Akbaş, S. (2018) · Acta Astronautica · doi:10.1016/j.actaastro.2018.09.011 · Cited by 13
Boolani, A., Yager, C. C., Reid, J., Lackman, J., & Smith, M. L. (2021) · Journal of American College Health · doi:10.1080/07448481.2021.1960843 · Cited by 13
A Validation Study of a Noninvasive Lactate Threshold Device
McMorries, R. M., Joubert, D. P., Jones, E. J., & Faries, M. D. (2019) · International Journal of Exercise Science · doi:10.70252/loed6863 · Cited by 14
“Our sample of 21 participants is similar to other validation studies with this specific device (6).”
ResultSleep and Subjective Recovery in Amateur Trail Runners After the Ultra-Trail du Mont Blanc® (UTMB®)
Baron, P., Hermand, É., Elsworth-Edelsten, C., Pezé, T., Bourlois, V., Mauvieux, B., & Hurdiel, R. (2022) · Journal of Science in Sport and Exercise · doi:10.1007/s42978-021-00154-w · Cited by 13
“This findings also corroborate recent works suggesting that athletes age should be taken into account for recovery kinetics [4]: aging may produce an increased depletion of energy substrates, greater muscle damage and an altered feeling of fatigue after a race [8, 9, 25].”
ResultSendra-Pérez, C., Encarnación-Martínez, A., Oficial-Casado, F., Salvador-Palmer, R., & Priego-Quesada, J. I. (2023) · Physiological Measurement · doi:10.1088/1361-6579/ad1457 · Cited by 9
Wearable near Infrared Sensor for Determining an Athlete's Lactate Threshold during Exercise
Driller, M., Plews, D., & Borges, N. (2016) · NIR news · doi:10.1255/nirn.1609 · Cited by 9
Self-citationHeiber, M., Schittenhelm, A., Schlie, J., Beckert, M., Graf, P., & Schmidt, A. (2024) · Open Access Journal of Sports Medicine · doi:10.2147/oajsm.s444568 · Cited by 7
Snarr, R. L., Tolusso, D. V., Hallmark, A. V., & Esco, M. R. (2018) · The Journal of Strength and Conditioning Research · doi:10.1519/jsc.0000000000002721 · Cited by 8
Rębiś, K., Klusiewicz, A., Długołęcka, B., Różański, P., Kowieski, K., & Kowalski, T. (2024) · Journal of Clinical Medicine · doi:10.3390/jcm13175340 · Cited by 7
“Also, Borges and Driller [18] observed in runners no significant differences in load for the AT determined using the aforementioned NIRS sensor compared to five traditional methods, based on blood lactate determinations.”
MethodWearables in Sports Cardiology
Beavers, D. L., & Chung, E. H. (2022) · Clinics in Sports Medicine · doi:10.1016/j.csm.2022.02.004 · Cited by 8
Mežan, L. Ž. (2025) · Frontiers in Sports and Active Living · doi:10.3389/fspor.2025.1707991 · Cited by 6
Bagci, R., Vardar‐Yağlı, N., Sağlam, M., Kutukcu, E. C., İnce, D. İ., Sener, F., & Damadoğlu, E. (2022) · Physiotherapy Theory and Practice · doi:10.1080/09593985.2022.2041780 · Cited by 5
“Muscle oxygenation (SmO2) was recorded using a wearable lactate threshold (WLT) mMRC: Modified Medical Research Dyspnea Scale; BMI: Body mass index; ISWT: Incremental shuttle walk test; SF-36: Short form-36 questionnaire; HADS: Hospital anxiety and depression scale. device (BSX insight; BSX Athletics, Boston, USA) (Bo…”
MethodInfluentialEtxegarai, U., Portillo, E., Irazusta, J., Koefoed, L., & Kasabov, N. K. (2019) · European Journal of Operational Research · doi:10.1016/j.ejor.2019.08.023 · Cited by 5
“So far, significant efforts have been made to overcome some of these operational constraints, and solutions such as non-invasive hardware (HW) [5], analytical models [6] and empirical models [7] have been proposed.”
BackgroundNewer Perspectives in Lactate Threshold Estimation for Endurance Sports – A Mini-Review
Krishnan, A., Guru, C. S., Sivaraman, A., Alwar, T., Sharma, D., & Angrish, P. (2021) · Central European Journal of Sport Sciences and Medicine · doi:10.18276/cej.2021.3-09 · Cited by 6
Grachan, J. J., & Essick, E. E. (2019) · BIOS · doi:10.1893/0005-3155-90.4.263 · Cited by 4
Bozdemir-Özel, C., Arıkan, H., Kutukcu, E. C., Karadüz, B. N., Inal-Ince, D., Kabakçı, G., Sümer, E., & Dağdelen, S. (2020) · Canadian Journal of Diabetes · doi:10.1016/j.jcjd.2020.02.007 · Cited by 3
Non-Invasive Real-Time Monitoring of Blood Lactate Using a Low-Power Microwave Sensor
Mason, A., Korostynska, O., Dixon, J., & Connell, R. (2024) · Proceedings of IEEE Sensors/Proceedings of IEEE Sensors ... · doi:10.1109/sensors60989.2024.10784548 · Cited by 3
“Notably, a device from BSX Insight [14] which was heavily marketed in ca. 2015 has since been discontinued.”
Non-invasive measurement of blood lactate in humans using microwave sensors
Mason, A., Louis, J. B., Greene, J., Korostynska, O., Cordova-Lopez, L. E., Abdullah, B. M., Connell, R., & Hopkins, J. (2017) · IEEE Ukraine Conference on Electrical and Computer Engineering (UKRCON) · doi:10.1109/ukrcon.2017.8100481 · Cited by 3
“This is a validated [20] wearable system to predict lactate threshold, the point at which the concentration of blood lactate begins to exponentially increase during exercise.”
BackgroundGreene, J. (2021) · Liverpool John Moores University · doi:10.24377/ljmu.t.00014802 · Cited by 5
Signal extraction and monitoring of motion loads based on wearable online device
Gong, X., & He, H. (2020) · Computer Communications · doi:10.1016/j.comcom.2020.02.072 · Cited by 2
Hasegawa, J. S., Azevedo, R. D. A., Silveira, A. C., Lima‐Silva, A. E., & Bertuzzi, R. (2024) · Brazilian Journal of Medical and Biological Research · doi:10.1590/1414-431x2024e13102 · Cited by 1
“However, while there are a large number of studies showing the reliability and sensitivity of NIRS (13,14), data showing the reliability and sensitivity of wearable NIRS devices (wNIRS) are lacking (15,16).”
BackgroundMuscle Near-Infrared Spectroscopy: mNIRS the Next Generation of Applied Muscle Physiology
Feldmann, A. (2024) · Springer eBooks · doi:10.1007/978-3-662-68241-8_17-1 · Cited by 1
Muscle Oxygenation Threshold in More and Less Active Muscles and 3,000-m Running Pace
Cirino, C., Breda, F. L., Polisel, E. E. C., Lourenço, T. F., Papoti, M., Gobatto, C. A., & Manchado-Gobatto, F. D. B. (2025) · International Journal of Sports Medicine · doi:10.1055/a-2611-3388 · Cited by 1
A Resonator-Based Active Sensing Tag for Lactate Detection in Saline Solutions
Zhu, W., Baghelani, M., Light, P. E., & Iyer, A. K. (2026) · IEEE Sensors Journal · doi:10.1109/jsen.2026.3664297 · Cited by 1
“Consequently, sensors that can detect lactate levels surpass such thresholds are highly valuable for sports physiology, performance monitoring, and athlete conditioning [8], [9], [10], [11].”
Beziehung zwischen Laktat- und anderen Schwellenkonzepten
Heck, H., & Bartmus, U. (2022) · Springer eBooks · doi:10.1007/978-3-662-59835-1_10 · Cited by 1
Zhu, Z., Cui, W., Lu, C., Shen, Y., & Pan, B. (2025) · Engineering Applications of Artificial Intelligence · doi:10.1016/j.engappai.2025.112879 · Cited by 2
STOP THE PRICKS: A VALIDATION STUDY OF A NONINVASIVE LACTATE THRESHOLD DEVICE
McMorries, R. M. (2016) · Index of Texas Archaeology Open Access Grey Literature from the Lone Star State
Information technologies in training process modelling
Plećaš, J. (2019) · TIMS Acta · doi:10.5937/timsacta1902089p
“Decenijama se laktatni prag (LP) koristi kao ključni prediktor fizičke sposobnosti sportista i na osnovu koga se modeluje trenažni proces (Borges & Driller, 2016).”
BackgroundInfluentialA Newly Developed Wearable Chest Strap for Measuring Pulmonary Ventilation During Exercise
Harel, G., Reuveni, R., Segel, M. J., Inbar, O., Dubnov‐Raz, G., & Scheinowitz, M. (2021) · Acta Scientific Pharmaceutical Sciences · doi:10.31080/asps.2021.05.0699
Study of the local oxygen saturation in cycling and differences between sexes
Sendra‐Pérez, C. (2025) · Repository of Digital Objects for Teaching Research and Culture (University of Valencia)
Muscle Near-Infrared Spectroscopy: mNIRS the Next Generation of Applied Muscle Physiology
Feldmann, A. (2025) · Springer eBooks · doi:10.1007/978-3-662-70475-2_17 · Cited by 1
Zhang, Q., Ghahjaverestan, N. M., Francisco, C. D. O., Kabir, M., Hoque, M. S., Masouleh, S. K., Oh, P., Marzolini, S., & Yadollahi, A. (2026) · BioMedical Engineering OnLine · doi:10.1186/s12938-026-01540-7
“Finally, lactate threshold measured using near-infrared LED technology embedded in a compression calf sleeve was used to estimate AT with an error of less than 10% [28].”
Jalonen, J. (2024) · Tampere University Institutional Repository (Tampere University)
Syaifudin, A., Pamungkas, Y., Ridhoi, A., Wibowo, S., Herawati, L., Handrito, R. P., Adi, S., & Hashim, U. (2026) · Human Movement · doi:10.5114/hm/217211
“However, some research, us-ing a sample of 14 recreationally active adults with no formal cycling experience who completed a maximal incremental cycling test, indicates inconsistencies in threshold detection [149].”
Muscle Near-Infrared Spectroscopy: mNIRS the Next Generation of Applied Muscle Physiology
Feldmann, A. (2026) · Springer eBooks · doi:10.1007/978-3-662-73013-3_14
Maffulli, N., Padulo, J., & Oliva, F. (2026) · British Medical Bulletin · doi:10.1093/bmb/ldag037
“In earlier work , the muscle oxygenation breakpoint approximated the MLSS in distance runners and triathletes [ 82], and a wearable near-infrared device predicted the LT in runners [ 83].”
Metadata
- Authors
- Borges, N. R., & Driller, M. W.
- Affiliations
- Central Queensland University (AU)
- University of Waikato (NZ)
- Corresponding author
- Borges, Nattai R.
- Venue
- Journal of Strength and Conditioning Research
- Published
- 2016-08
- Volume
- 30
- Issue
- 8
- Pages
- 2212-2218
- ISSN
- 1064-8011 (ISSN-L)
- OpenAlex
- W2319419678
- Semantic Scholar
- cd8dc0d61aa1e4620865d8d6f4b3ac19eb7f0cea
- Cited by
- 59
- References
- 23
- Citations by year
- 2026: 5 · 2025: 5 · 2024: 8 · 2023: 3 · 2022: 7 · 2021: 6 · 2020: 4 · 2019: 6 · 2018: 7 · 2017: 4 · 2016: 4
- FWCI
- 4.93
- Type
- journal article
- Language
- en
- Open access
- green
- figshare.com
- Licence
- cc-by
- Locations
- The Journal of Strength and Conditioning Research (publishedVersion)
- PubMed (publishedVersion)
- Acquire (CQUniversity) (submittedVersion)
- Figshare (submittedVersion, cc-by)
- Topics
- Health Sciences › Medicine › Orthopedics and Sports Medicine › Sports Performance and Training
- Health Sciences › Medicine › Physiology (medical) › Cardiovascular and exercise physiology
- Physical Sciences › Engineering › Biomedical Engineering › Lower Extremity Biomechanics and Pathologies
- Keywords
- lactate threshold, near-infrared LEDs, runners, incremental exercise test, blood lactate, device reliability, inter-device agreement, treadmill running, endurance training
- MeSH terms
- Adolescent, Adult, Exercise Test, Female, Humans, Male, Middle Aged, Running, Reproducibility of Results, Anaerobic Threshold, Biomarkers, Spectroscopy, Near-Infrared, Lactic Acid, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
A language model drew the claims from the paper on 2026-11-23, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-11-23 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.