Using failure mode and effects analysis to plan implementation of smart i.v. pump technology
The DOI matches its Crossref record; 33 of the 35 references it cites were found in a registry or an index.
Abstract
PURPOSE: Failure mode and effects analysis (FMEA) was used to evaluate a smart i.v. pump as it was implemented into a redesigned medication-use process. SUMMARY: A multidisciplinary team conducted a FMEA to guide the implementation of a smart i.v. pump that was designed to prevent pump programming errors. The smart i.v. pump was equipped with a dose-error reduction system that included a pre-defined drug library in which dosage limits were set for each medication. Monitoring for potential failures and errors occurred for three months postimplementation of FMEA. Specific measures were used to determine the success of the actions that were implemented as a result of the FMEA. The FMEA process at the hospital identified key failure modes in the medication process with the use of the old and new pumps, and actions were taken to avoid errors and adverse events. I.V. pump software and hardware design changes were also recommended. Thirteen of the 18 failure modes reported in practice after pump implementation had been identified by the team. A beneficial outcome of FMEA was the development of a multidisciplinary team that provided the infrastructure for safe technology implementation and effective event investigation after implementation. With the continual updating of i.v. pump software and hardware after implementation, FMEA can be an important starting place for safe technology choice and implementation and can produce site experts to follow technology and process changes over time. CONCLUSION: FMEA was useful in identifying potential problems in the medication-use process with the implementation of new smart i.v. pumps. Monitoring for system failures and errors after implementation remains necessary.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Citations
Cites
To Err Is Human: Building a Safer Health System
(2000) · National Academies Press eBooks · doi:10.17226/9728 · Cited by 14,044
Wall, A. (2000) · British Journal of Healthcare Management · doi:10.12968/bjhc.2000.6.9.19311 · Cited by 5,903
Incidence of Adverse Drug Events and Potential Adverse Drug Events: Implications for Prevention
Bates, D. W. (1995) · JAMA · doi:10.1001/jama.1995.03530010043033 · Cited by 2,127 · cited 1×
“medication administration is especially vulnerable due to the lack of built-in double checks and the low likelihood of event interception before the drug reaches the patient.(1,6) In a systems analysis of error, misuse of infusion pumps and other parenteral delivery systems was identified as a common proximal cause of…”
BackgroundMedication Errors and Adverse Drug Events in Pediatric Inpatients
Kaushal, R. (2001) · JAMA · doi:10.1001/jama.285.16.2114 · Cited by 1,917 · cited 4×
“Correct insertion of the tubing in the new pump required three separate sequential actions: (1) placement of the tubing top fitment into the pump, ( 2 ) insertion of a plastic side clamp into the pump to disengage the anti-free-flow mechanism, and (3) flossing of the tubing through an air-in-line sensor at the bottom…”
BackgroundMethodInfluentialMaking health care safer: a critical analysis of patient safety practices.
Shojania, K. G., Duncan, B. W., McDonald, K. M., Wachter, R. M., & Markowitz, A. J. (2001) · PubMed Central · Cited by 1,612
Systems Analysis of Adverse Drug Events
Leape, L. L. (1995) · JAMA: The Journal of the American Medical Association · doi:10.1001/jama.1995.03530010049034 · Cited by 1,404
DeRosier, J., Stalhandske, E., Bagian, J. P., & Nudell, T. (2002) · The Joint Commission Journal on Quality Improvement · doi:10.1016/s1070-3241(02)28025-6 · Cited by 611 · cited 4×
“This form of FMEA was thought to be easier to apply to health care situations due to a modified hazard matrix scoring system for failure modes based on health care outcomes.(29) HFMEA uses a 4-point scale to rate severity (minor, moderate, major, catastrophic) anchored to patient outcomes and a 4-point scale for proba…”
MethodInfluentialIncidence of adverse drug events and potential adverse drug events
Bates, D. W., Cullen, D. J., & Laird, N. (1995) · International Journal of Trauma Nursing · doi:10.1016/s1075-4210(05)80011-2 · Cited by 534 · cited 4×
“The team underwent training in the use of Healthcare FMEA (HFMEA), developed by the Veterans Administration National Center for Patient Safety.29 The steps in the HFMEA process include ( 1 ) team selection, (2) process identification, (3) process mapping, (4) failure-mode identification and scoring, and (5) determinat…”
BackgroundMethodInfluentialA balance theory of job design for stress reduction
Smith, M. J., & Sainfort, P. C. (1989) · International Journal of Industrial Ergonomics · doi:10.1016/0169-8141(89)90051-6 · Cited by 488 · cited 2×
“The work system is made up of five elements: people, tasks performed, environment, organization, and tools and technologies used to complete tasks.(16) With the addition of a new technology, the effects of the work system on all elements should be considered.”
BackgroundMethodA controlled trial of smart infusion pumps to improve medication safety in critically ill patients*
Rothschild, J. M., Keohane, C. A., Cook, E. F., Orav, E. J., Burdick, E., Thompson, S., Hayes, J., & Bates, D. W. (2005) · Critical Care Medicine · doi:10.1097/01.ccm.0000155912.73313.cd · Cited by 301
Research works printed but not linked: 13
- 2002, Health Devices, 31, 353
- 2002, Biomed Instrum Technol, 36, 84
- 2002, Nurs Econ, 20, 80
- 2001, Jt Comm J Qual Improv, 27, 430
- 1994, Hosp Pharm, 29, 319
- 2004, Hosp Peer Rev, 29, 13
- 2002, J Am Pharm Assoc, 42, S54
- 2002, Jt Comm J Qual Improv, 28, 331
- 2004, Qual Saf Health Care, 13, 265, 10.1136/qshc.2003.007443 · doi:10.1136/qshc.2003.007443 · Paper page
- 2004, Clin Leadersh Manag Rev, 18, 37
- 2002, Jt Comm J Qual Improv, 28, 248
- 2004, Clin Leadersh Manag Rev, 18, 32
- 1998, PDA J Pharm Sci Technol, 52, 52
Cited by
Items 101–143 of 143
Risk Assessment in Radiotherapy - Lessons from Systems Engineering
Fallon, E. F., Chadwick, L. M., & van der Putten, W. J. (2009) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-642-03893-8_71 · Cited by 4
“Techniques drawn from Human Factors (HF) engineering, reliability engineering and risk analysis in the medical domain have begun to be utilized [8-10].”
BackgroundTotal failure mode and effect analysis: a powerful technique for overcoming failures
Krishnaraj, C., Mohanasundram, K. M., Devadasan, S. R., & Sivaram, N. M. (2012) · International Journal of Productivity and Quality Management · doi:10.1504/ijpqm.2012.048295 · Cited by 4
“6 Wettereck et al. (2006) Used FMEA in i.”
Kharazmi, E., Salehi, A., Hashemi, N., Ghaderi, S., & Hatam, N. (2016) · Asia Pacific Journal of Health Management · doi:10.24083/apjhm.v11i2.187 · Cited by 3
“[19] These studies indicate the usefulness of FMEA in identifying malfunction of different healthcare services and providing possible intervention programs.”
BackgroundMethodSirvent, M., Calvo, M. V. M., Pérez-Pons, J. C., Penín, I. R., Martí‐Bonmatí, E., Vázquez, A., Romero, R., Crespo, C. L., & Tejada, P. E. (2014) · Farmacia Hospitalaria · doi:10.7399/fh.2014.38.5.8085 · Cited by 6
A Healthcare Failure Mode and Effect Analysis on the Safety of Secondary Infusions
Yue, R. Y. K., Trbovich, P., & Easty, T. (2012) · PsycEXTRA Dataset · doi:10.1037/e572172013-184 · Cited by 3
Thellier, S. (2017) · theses.fr (ABES) · doi:10.70675/d0db2228ze3cbz4d50zba3bz18e3cf1a55d6 · Cited by 3
RISK, CAUSES AND PREVENTIVE ACTION ASSESSMENT IN THE ICU OF A TEACHING HOSPITAL
Hosein, Y. M., Jazi, M. J., Elahe, K., & Atighechian, G. (2014) · Cited by 2
Literature review of interprofessional research on inpatient pharmacy operations
Abel, S. R., Intrevado, P., & Ozen, L. (2013) · American Journal of Health-System Pharmacy · doi:10.2146/ajhp120112 · Cited by 2
Chadwick, L. M., Fallon, E. F., & van der Putten, W. J. (2012) · Proceedings of the International Symposium on Human Factors and Ergonomics in Health Care · doi:10.1518/hcs-2012.945289401.002 · Cited by 3
“…number of negative aspects of HFMEA® have been identified that give rise to questions regarding its fundamental design and ability to facilitate and achieve its intended purpose (Chadwick & Fallon, In Press; Habraken, Van der Schaaf, Leistikow, & Reijnders-Thijssen, 2009; Wetterneck et al., 2006).”
Human-Centered Systems Engineering: Managing Stakeholder Dissonance in Healthcare Delivery
Samaras, G. (2011) · IGI Global eBooks · doi:10.4018/978-1-60960-872-9.ch007 · Cited by 6
“Wetterneck et al (2006) report data that indicates less than 75% of infusion pump failure modes identified in actual practice were captured, in advance, by their FMEA.”
BackgroundPalisoc, J. I. L., & Noroña, M. I. (2020) · doi:10.1145/3396743.3396795 · Cited by 2
Silva, A. E. B. D. C. (2008) · Digital Library of Theses and Dissertations (Universidade de São Paulo) · doi:10.11606/t.83.2008.tde-12012009-145608 · Cited by 13
Get Smart: Effective Use of Smart Pump Technology
Sanborn, M. D., & Cohen, T. (2009) · Hospital Pharmacy · doi:10.1310/hpj4404-348 · Cited by 2
“Finally, the use of failure mode and effects analysis (FMEA) has been described as a means for assisting with the implementation of smart pump technology.(18) This technique was found useful for identifying potential problems in the medication-use process related to pump implementation, but it was concluded that monit…”
MethodRequirements for prospective risk analysis in health care
Kessels-Habraken, M. M., & Schaaf, V. D. T. T. (2008) · TU/e Research Portal · Cited by 1
Grelier, S., Thetio, M., Quentin, V., Achache, V., Sánchez, N. C., Leroux, V., Durand, E., & Péquignot, R. (2011) · Annals of Physical and Rehabilitation Medicine · doi:10.1016/j.rehab.2011.01.002 · Cited by 1
“Now we see emerging the notion of risk assessment before the event occurs and several studies have been conducted on the relevance of combining these two approaches [1–3,8,11,14,17,20,23] or working on safety barriers [9,15].”
BackgroundNoyes, K., Fleming, F. J., Iannuzzi, J. C., & Monson, J. R. T. (2015) · Springer eBooks · doi:10.1007/978-1-4899-7673-4_8-1 · Cited by 1
Thellier, S., & Falzon, P. (2016) · doi:10.4267/2042/61756 · Cited by 1
Identifying appropriate interruption-handling strategies during healthcare care processes
Cure, L., Elliott, A., & Nicks, S. (2017) · IISE Transactions on Healthcare Systems Engineering · doi:10.1080/24725579.2017.1380087 · Cited by 1
“…(Bonnabry et al., 2008), its popularity (Arenas VillaFranca et al., 2014), its adaptability to different implementation strategies and sources of process and risk information (Faye et al., 2010; Wetterneck et al., 2006), and its systematic and multidisciplinary nature (Woodhouse et al., 2004).”
BackgroundMalins, R. J. (2018) · INCOSE International Symposium · doi:10.1002/j.2334-5837.2018.00523.x · Cited by 1
HFMEA™ of a Radiotherapy Information System – Challenges and Recommendations for Future Studies
Chadwick, L. M., & Fallon, E. F. (2010) · Advances in human factors and ergonomics series · doi:10.1201/ebk1439834978-c70 · Cited by 1
Rothwell, S. (2009) · TSpace (University of Toronto) · Cited by 2
Chan, K. (2013) · TSpace (University of Toronto) · Cited by 1
Noyes, K., Fleming, F. J., Iannuzzi, J. C., & Monson, J. R. T. (2019) · Health services research · doi:10.1007/978-1-4939-8715-3_8 · Cited by 1
Digital Twin Synchronization: towards a data-centric architecturePreprint
Freitas, E., Filho, A. T. D. O., do Carmo, P. R. X., Sadok, D. F. H., & Kelner, J. (2026) · arXiv (Cornell University) · doi:10.48550/arxiv.2601.23051 · Cited by 2
“A detailed logging of each update attempt should be performed, containing details of the device, the nature of the update, and the reasons for failure, for example Chu et al. (2018); Wetterneck et al. (2006).”
Exploring how nurses make sense of the safety features of smart infusion pump technology
Kirkbride, G. (2014) · doi:10.17077/etd.f5zv4p71 · Cited by 1
Understanding and Preventing Errors
Apkon, M. (2009) · Computers in health care · doi:10.1007/978-0-387-76446-7_29
Management of Complex Clinical System
Sloane, E. B. (2016) · Elsevier eBooks · doi:10.1016/b978-0-12-803767-6.00006-4
Changing Smart Pump Vendors: Lessons Learned
Arthur, K. J., Catlin, A. C., Quebe, A. K., & Washington, A. (2016) · Hospital Pharmacy · doi:10.1310/hpj5109-782
“As shown in Table 1, the limited number of profiles and significantly decreased number of total medications entries in Alaris (2,500) compared with Hospira (16,000) forced the combination of many units that had previously been divided.”
ResultChapter 16. Automation of Acute Care Pharmacy Operations
Tribble, D. A. (2010) · The American Pharmacists Association eBooks · doi:10.21019/9781582121444.ch16
Development to High-volume Manufacturing: Reducing the Risks
Drain, D. C., Grantham, K., & Cudney, E. A. (2007)
Malfará, M. R. M. (2018) · Digital Library of Theses and Dissertations (Universidade de São Paulo) · doi:10.11606/d.17.2018.tde-24042018-165341
Raphemot, C. (2017) · Carolina Digital Repository (University of North Carolina at Chapel Hill) · doi:10.17615/r1za-pc70
Tully, M. P., & Franklin, B. D. (2015) · CRC Press eBooks · doi:10.1201/b18773-28
Communicative Technologies in Nursing Homes: A Framework for Evaluation (Preprint)Preprint
Ali, H. B., & Ahmed, A. (2019) · doi:10.2196/preprints.17254
Clinical equipment standardisation and its impact on patient safety
Greenwood, K. (2011) · Carleton University's Institutional Repository (MacOdrum Library, Carleton University) · doi:10.22215/etd/2011-06847
Park, J., You, S. B., Ryu, G. W., & Kim, Y. (2024) · Figshare · doi:10.6084/m9.figshare.26653188
Park, J., You, S. B., Ryu, G. W., & Kim, Y. (2024) · Figshare · doi:10.6084/m9.figshare.26653182
Park, J., You, S. B., Ryu, G. W., & Kim, Y. (2024) · Figshare · doi:10.6084/m9.figshare.26653185
Quality and Safety of Intermittent Intravenous Infusions
Morrow Martha, S. (2018) · OhioLink ETD Center (Ohio Library and Information Network) · Cited by 5
Preventing and predicting medical accidents using systems theory
Samost, A. (2015) · DSpace@MIT (Massachusetts Institute of Technology)
deVries, M. (2022) · TSpace (University of Toronto) · Cited by 2
Li, J., Richard, S. A., Lan, Z., & Feng, L. (2026) · Scientific Reports · doi:10.1038/s41598-026-70689-x
“For example, Wetterneck et al 30 . applied HFMEA in operating room management, reducing surgical instrument preparation error rates by 60%.”
Essa, A. S., & Mohammed, F. (2013) · Figshare
Metadata
- Authors
- Wetterneck, T. B., Skibinski, K. A., Roberts, T. L., Kleppin, S. M., Schroeder, M. E., Enloe, M., Rough, S. S., Hundt, A. S., & Carayon, P.
- Affiliations
- University of Wisconsin–Madison (US)
- University of Wisconsin Health (US)
- Department of Anesthesiology, UWSMPH
- UW Health University Hospital (US)
- Corresponding author
- Wetterneck, Tosha B.
- ORCID
- Carayon, Pascale
- Venue
- American Journal of Health-System Pharmacy
- Publisher
- Oxford University Press (OUP)
- Published
- 2006-08-15
- Volume
- 63
- Issue
- 16
- Pages
- 1528-1538
- ISSN
- 1079-2082 (ISSN-L)
- 1535-2900
- OpenAlex
- W1983207700
- Cited by
- 134
- References
- 35
- Citations by year
- 2025: 1 · 2024: 3 · 2023: 3 · 2022: 2 · 2021: 3 · 2020: 8 · 2019: 12 · 2018: 6 · 2017: 5 · 2016: 8 · 2015: 2 · 2014: 8 · 2013: 11 · 2012: 10
- FWCI
- 2.18
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- American Journal of Health-System Pharmacy (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Agency for Healthcare Research and Quality (1uc1hs014253-01)
- National Center for Research Resources (1 K12-RR01764-01)
- Topics
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Patient Safety and Medication Errors
- Physical Sciences › Engineering › Biomedical Engineering › Intravenous Infusion Technology and Safety
- Health Sciences › Medicine › Surgery › Healthcare Technology and Patient Monitoring
- Keywords
- failure mode and effects analysis, smart infusion pumps, medication errors, drug library, patient safety, adverse event investigation, multidisciplinary team, technology implementation
- MeSH terms
- Equipment Safety, Humans, Infusion Pumps, Infusions, Intravenous, Reproducibility of Results, Risk Assessment
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.