Investigating the effects of temperature on photoplethysmography
Open accessAbstract
Pulse oximetry is a commonly used technique to determine arterial oxygen saturation (SpO 2 ) and heart rate. This method uses photoplethysmography (PPG) signals to estimate SpO 2 . However, pulse oximeters can be less accurate at cold temperatures. In this research, the effects of temperature on PPG signal quality was investigated for a group of 20 adult volunteers. Raw PPG data was obtained using a custom pulse oximeter (PO) system. Three tests were performed with the subject's hand maintained at baseline, cold, and warm temperatures. Analysis showed the median PPG signal RMS, as an indicator of signal-to-noise, for the warm test was 0.0289 V [IQR = 0.0262 - 0.0315 V] compared to the baseline test's 0.0174 V [IQR = 0.0104 - 0.0285 V] and the cold test's 0.0066 V [IQR = 0.0053 - 0.0106V]. This improvement of up to 4x in signal quality was associated with a closer match between the SpO 2 values estimated by the PO system and a commercial pulse oximeter. Paired RMS value comparison between the three tests showed a statistically significant difference for all P-values (P < 0.004). Results indicate that warm temperatures improve PPG quality and thus pulse oximeter accuracy.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
PPG quality is improved with warm temperatures.
“The authors conclude that PPG quality is improved with warm temperatures, yet do not mention the effect of thermoregulatory responses [155].”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [154]
There is a reduction in PPG Root Mean Square values as ambient temperatures went from warm to cold temperatures.
“Khan et al. investigated the effects of finger temperature on PPG signal on 20 healthy adult volunteers (24.5 ± 4.1 years of age), reporting a reduction in PPG Root Mean Square values as ambient temperatures went from warm to cold temperatures [154].”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [154]
Cites
(1997) · Institute of Physics eBooks · doi:10.1887/0750304677 · Cited by 455 · see citing works
Photoplethysmography: Beyond the Calculation of Arterial Oxygen Saturation and Heart Rate
Shelley, K. H. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000269512.82836.c9 · Cited by 450 · see citing works
Pulse oximetry: fundamentals and technology update
Nitzan, M., Romem, A., & Koppel, R. (2014) · Medical Devices: Evidence and Research · doi:10.2147/mder.s47319 · Cited by 438 · see citing works
Feiner, J. R., Severinghaus, J. W., & Bickler, P. E. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000285988.35174.d9 · Cited by 400 · see citing works
Jubran, A. (2004) · Intensive Care Medicine · doi:10.1007/s00134-004-2399-x · Cited by 400 · see citing works
The effect of temperature on blood flow and deep temperature in the human forearm
Barcroft, H., & Edholm, O. G. (1943) · The Journal of Physiology · doi:10.1113/jphysiol.1943.sp004009 · Cited by 344 · see citing works
Pulse Oximetry: Theory and Applications for Noninvasive Monitoring
Mendelson, Y. (1992) · Clinical Chemistry · doi:10.1093/clinchem/38.9.1601 · Cited by 320 · see citing works
Masimo Signal Extraction Pulse Oximetry
Goldman, J. M., Petterson, M. T., Kopotic, R. J., & Barker, S. J. (2000) · Journal of Clinical Monitoring and Computing · doi:10.1023/a:1011493521730 · Cited by 283 · see citing works
The peripheral pulse wave: Information overlooked
Murray, W. B., & Foster, P. A. (1996) · The Journal of Clinical Monitoring · doi:10.1007/bf02077634 · Cited by 231 · see citing works
Signal processing methods for pulse oximetry
Rusch, T. L., Sankar, R. T., & Scharf, J. E. (1996) · Computers in Biology and Medicine · doi:10.1016/0010-4825(95)00049-6 · Cited by 185 · see citing works
The Effect of Motion on Pulse Oximetry and Its Clinical Significance
Petterson, M. T., Begnoche, V. L., & Graybeal, J. M. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000278134.47777.a5 · Cited by 177 · see citing works
Influence of Cold Water Immersion on Limb and Cutaneous Blood Flow at Rest
Gregson, W. A., Black, M. A., Jones, H., Milson, J., Morton, J. P., DAWSON, B. T., Atkinson, G., & GREEN, D. J. (2011) · The American Journal of Sports Medicine · doi:10.1177/0363546510395497 · Cited by 171 · see citing works
Fortnightly Review: Pulse oximetry: a practical review
Hanning, C. D., & Alexander-Williams, J. (1995) · BMJ · doi:10.1136/bmj.311.7001.367 · Cited by 151 · see citing works
Respiration Signals from Photoplethysmography
Nilsson, L. M. (2013) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e31828098b2 · Cited by 129 · see citing works
Pulse Oximetry: Uses and Limitations
DeMeulenaere, S. (2007) · The Journal for Nurse Practitioners · doi:10.1016/j.nurpra.2007.02.021 · Cited by 109 · see citing works
The effect of local cold application on intramuscular blood flow at rest and after running
Thorsson, O., Lilja, B., Ahlgren, L., et al. (1985) · Medicine & Science in Sports & Exercise · doi:10.1249/00005768-198512000-00016 · Cited by 70 · see citing works
GALLEN, I. W., & Macdonald, I. A. (1990) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.1990.259.5.e639 · Cited by 64 · see citing works
A non-invasive measure of changes in blood flow in the human anterior tibial muscle
Zhang, Q., Lindberg, L., Kadefors, R., & STYF, J. R. (2001) · European Journal of Applied Physiology · doi:10.1007/s004210100413 · Cited by 63 · see citing works
Non-invasive continuous estimation of blood flow changes in human patellar bone
Näslund, J., Pettersson, J., Lundeberg, T., Linnarsson, D., & Lindberg, L. (2006) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-006-0070-0 · Cited by 57 · see citing works
Blood flow measurements at different depths using photoplethysmography and laser Doppler techniques
Bergstrand, S., Lindberg, L. G., Ek, A., Lindén, M., & Lindgren, M. (2009) · Skin Research and Technology · doi:10.1111/j.1600-0846.2008.00337.x · Cited by 57 · see citing works
Does Ambient Light Affect the Accuracy of Pulse Oximetry?
Fluck, R. R., Schroeder, C., Frani, G., et al. (2003) · Respiratory Care · doi:10.4187/respcare.03480677 · Cited by 56 · see citing works
Measuring venous oxygenation using the photoplethysmograph waveform
Walton, Z. D., Kyriacou, P. A., Silverman, D. G., & Shelley, K. H. (2010) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-010-9248-y · Cited by 53 · see citing works
Valdez-Lowe, C., Ghareeb, S. A., & Artinian, N. T. (2009) · AJN American Journal of Nursing · doi:10.1097/01.naj.0000352474.55746.81 · Cited by 36 · see citing works
Hypohydration effect on finger skin temperature and blood flow during cold-water finger immersion
O’Brien, C., & Montain, S. J. (2003) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00678.2002 · Cited by 23 · see citing works
Dickson, J. L., Gunn, C. A., & Chase, J. G. (2014) · International Journal of Clinical & Medical Images · Cited by 19 · see citing works
Self-citationModulation of finger photoplethysmographic traces during forced respiration: Venous blood in motion?
Phillips, J. P., Belhaj, A., Shafqat, K., et al. (2012) · doi:10.1109/embc.2012.6346756 · Cited by 17 · see citing works
Jubran, A. (2009) · Springer eBooks · doi:10.1007/978-3-642-01769-8_11 · Cited by 9 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348 · see citing works
Askarian, B., Jung, K., & Chong, J. W. (2019) · Sensors · doi:10.3390/s19132846 · Cited by 45 · see citing works
Singh, S., Bennett, M. R., Chen, C., Shin, S., Ghanbari, H., & Nelson, B. W. (2024) · Journal of Medical Internet Research · doi:10.2196/62769 · Cited by 36 · see citing works
“Researchers have identified several factors that influence the accuracy of PPG-based measurements, including skin temperature, sensor contact pressure, skin thickness, and hydration level [5,6].”
- Authors
- Khan, M., Pretty, C. G., Amies, A. C., Elliott, R., Shaw, G. M., & Chase, J. G.
- Affiliations
- University of Canterbury (NZ)
- Christchurch Hospital (NZ)
- Corresponding author
- Khan, Musabbir
- Chase, J. Geoffrey
- Venue
- IFAC-PapersOnLine
- Publisher
- Elsevier BV
- Published
- 2015
- Volume
- 48
- Issue
- 20
- Pages
- 360-365
- ISSN
- 2405-8963 (ISSN-L)
- OpenAlex
- W2212541523
- Cited by
- 41
- References
- 27
- Citations by year
- 2026: 4 · 2025: 8 · 2024: 6 · 2023: 4 · 2022: 4 · 2021: 8 · 2020: 2 · 2019: 2 · 2018: 2 · 2016: 1
- FWCI
- 0.50
- Type
- journal article
- Language
- en
- Open access
- diamond
- doi.org
- Locations
- IFAC-PapersOnLine (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Keywords
- photoplethysmography, pulse oximetry, temperature effects, arterial oxygen saturation, signal quality, signal-to-noise ratio, low temperature, pulse oximeter accuracy, hand temperature
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review