Validation and Acceptability of a Cuffless Wrist-Worn Wearable Blood Pressure Monitoring Device Among Users and Health Care Professionals: Mixed Methods Study
Open accessPublished version of a preprint
Abstract
Background: Blood pressure (BP) is an important modifiable cardiovascular risk factor, yet its long-term monitoring remains problematic. Wearable cuffless devices enable the capture of multiple BP measures during everyday activities and could improve BP monitoring, but little is known about their validity or acceptability. Objective: This study aimed to validate a wrist-worn cuffless wearable BP device (Model T2; TMART Technologies Limited) and assess its acceptability among users and health care professionals. Methods: A mixed methods study was conducted to examine the validity and comparability of a wearable cuffless BP device against ambulatory and home devices. BP was measured simultaneously over 24 hours using wearable and ambulatory devices and over 7 days using wearable and home devices. Pearson correlation coefficients compared the degree of association between the measures, and limits of agreement (LOA; Bland-Altman plots) were generated to assess measurement bias. Semistructured interviews were conducted with users and 10 health care professionals to assess acceptability, facilitators, and barriers to using the wearable device. Interviews were audio recorded, transcribed, and analyzed. Results: A total of 9090 BP measurements were collected from 20 healthy volunteers (mean 20.3 years, SD 5.4; N=10 females). Mean (SD) systolic BP (SBP)/diastolic BP (DBP) measured using the ambulatory (24 hours), home (7 days), and wearable (7 days) devices were 126 (SD 10)/75 (SD 6) mm Hg, 112 (SD 10)/71 (SD 9) mm Hg and 125 (SD 4)/77 (SD 3) mm Hg, respectively. Mean (LOA) biases and precision between the wearable and ambulatory devices over 24 hours were 0.5 (−10.1 to 11.1) mm Hg for SBP and 2.24 (−17.6 to 13.1) mm Hg for DBP. The mean biases (LOA) and precision between the wearable and home device over 7 days were −12.7 (−28.7 to 3.4) mm Hg for SBP and −5.6 (−20.5 to 9.2) mm Hg for DBP. The wearable BP device was well accepted by participants who found the device easy to wear and use. Both participants and health care providers agreed that the wearable cuffless devices were easy to use and that they could be used to improve BP monitoring. Conclusions: Wearable BP measures compared well against a gold-standard ambulatory device, indicating potential for this user-friendly method to augment BP management, particularly by enabling long-term monitoring that could improve treatment titration and increase understanding of users’ BP response during daily activity and stressors.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In a small study, mean biases between cuff-less wearable and ambulatory devices over 24 h were 0.5 mmHg for systolic BP and 2.24 mmHg for diastolic BP, with mean bias widening over 7 days to -12.7 mmHg for systolic BP and -5.6 mmHg for diastolic BP.
“Mean biases between the wearable and ambulatory devices over 24 h were 0.5 mmHg (–10.1 mmHg to 11.1 mmHg) for systolic BP and 2.24 mmHg (–17.6 mmHg to 13.1 mmHg) for diastolic BP, with the mean bias widening over 7 days to –12.7 mmHg for systolic BP and –5.6 mmHg for diastolic BP.”
From Smart wearable devices in cardiovascular care: Where we are and how to move forward (2021) · cites this paper as [28]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Part of this paper’s reference list could not be split into single entries.
Three Approaches to Qualitative Content Analysis
Hsieh, H.-F., & Shannon, S. E. (2005) · Qualitative Health Research · doi:10.1177/1049732305276687 · Cited by 44,601 · see citing works
James, P. A., Oparil, S., Carter, B. L., Cushman, W. C., Dennison-Himmelfarb, C., Handler, J., Lackland, D. T., LeFevre, M. L., MacKenzie, T. D., Ogedegbe, O., et al. (2014) · JAMA · doi:10.1001/jama.2013.284427 · Cited by 7,950 · see citing works
Atkinson, G., & Nevill, A. M. (1998) · Sports Medicine · doi:10.2165/00007256-199826040-00002 · Cited by 3,817 · see citing works
Gakidou, E., Afshin, A., Abajobir, A. A., Abate, K. H., Abbafati, C., Abbas, K. M., Abd-Allah, F., Abdulle, A. M., Abera, S. F., Aboyans, V., et al. (2017) · The Lancet · doi:10.1016/s0140-6736(17)32366-8 · Cited by 2,406 · see citing works
Global Burden of Hypertension and Systolic Blood Pressure of at Least 110 to 115 mm Hg, 1990-2015
Forouzanfar, M. H., Liu, P., Roth, G. A., Ng, M., Biryukov, S., Marczak, L., Alexander, L., Estep, K., Hassen Abate, K., Akinyemiju, T. F., et al. (2017) · JAMA · doi:10.1001/jama.2016.19043 · Cited by 2,381 · see citing works
Chow, C. K. (2013) · JAMA · doi:10.1001/jama.2013.184182 · Cited by 2,000 · see citing works
Self-citationEuropean Society of Hypertension practice guidelines for ambulatory blood pressure monitoring
Parati, G., Stergiou, G., O’Brien, E., Asmar, R., Beilin, L., Bilo, G., Clement, D., de la Sierra, A., de Leeuw, P., Dolan, E., et al. (2014) · Journal of Hypertension · doi:10.1097/hjh.0000000000000221 · Cited by 1,021 · see citing works
European Society of Hypertension Practice Guidelines for home blood pressure monitoring
Parati, G., Stergiou, G. S., Asmar, R., Bilo, G., de Leeuw, P., Imai, Y., Kario, K., Lurbe, E., Manolis, A., Mengden, T., et al. (2010) · Journal of Human Hypertension · doi:10.1038/jhh.2010.54 · Cited by 482 · see citing works
2013 ESH/ESC Guidelines for the management of arterial hypertension
Mancia, G., Fagard, R. H., Narkiewicz, K., Redón, J., Zanchetti, A., Böhm, M., Christiaens, T. C., Cífková, R., Backer, G. D., Dominiczak, A. F., et al. (2013) · Blood Pressure · doi:10.3109/08037051.2013.812549 · Cited by 425 · see citing works
Uhlig, K., Patel, K., Ip, S., Kitsios, G. D., & Balk, E. M. (2013) · Annals of Internal Medicine · doi:10.7326/0003-4819-159-3-201308060-00008 · Cited by 368 · see citing works
Measuring Blood Pressure Accurately
Jones, D. W., Appel, L. J., Sheps, S. G., Roccella, E. J., & Lenfant, C. J. M. (2003) · JAMA · doi:10.1001/jama.289.8.1027 · Cited by 263 · see citing works
Peter, L., Noury, N., & Cerny, M. (2014) · IRBM · doi:10.1016/j.irbm.2014.07.002 · Cited by 261 · see citing works
Kim, S.-H., Lilot, M., Sidhu, K. S., Rinehart, J. B., Yu, Z., Canales, C., & Cannesson, M. P. (2014) · Anesthesiology · doi:10.1097/aln.0000000000000226 · Cited by 206 · see citing works
Imai, Y., Kario, K., Shimada, K., Kawano, Y., Hasebe, N., Matsuura, H., Tsuchihashi, T., Ohkubo, T., Kuwajima, I., & Miyakawa, M. (2012) · Hypertension Research · doi:10.1038/hr.2012.56 · Cited by 204 · see citing works
An Armband Wearable Device for Overnight and Cuff-Less Blood Pressure Measurement
Zheng, Y.-L., Yan, B. P., Zhang, Y., & Poon, C. C. Y. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2318779 · Cited by 176 · see citing works
Palatini, P., Frigo, Bertolo, O., Roman, E., Da Cortà, R., & Winnicki, M. (1998) · PubMed · Cited by 134 · see citing works
Zheng, Y., Poon, C. C. Y., Yan, B. P., & Lau, J. Y. W. (2016) · Journal of Medical Systems · doi:10.1007/s10916-016-0558-6 · Cited by 82 · see citing works
A New Cuffless Device for Measuring Blood Pressure: A Real-Life Validation Study
Schoot, T. S., Weenk, M., van de Belt, T. H., Engelen, L. J., van Goor, H., & Bredie, S. J. H. (2016) · Journal of Medical Internet Research · doi:10.2196/jmir.5414 · Cited by 52 · see citing works
on behalf of the EVIDENT Group, García‐Ortiz, L., Recio-Rodriguez, J. I., Canales-Reina, J. J., Sánchez, A. C., Gómez-Arranz, A., Magdalena-Belio, J. F., Guenaga-Saenz, N., Agudo‐Conde, C., & Gómez-Marcos, M. Á. (2012) · Hypertension Research · doi:10.1038/hr.2012.3 · Cited by 47 · see citing works
[Mobile Health: IEEE Standard for Wearable Cuffless Blood Pressure Measuring Devices].
Zhou, X., Wu, W., & Bao, S. (2015) · PubMed · Cited by 29 · see citing works
A review of the latest guidelines for NIBP device validation
Alpert, B. S., Quinn, D. E., & Friedman, B. A. (2013) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000007 · Cited by 17 · see citing works
Islam, S. M. S. (2016) · Journal of Hypertension · doi:10.1097/01.hjh.0000501232.99910.05 · Cited by 3 · see citing works
Self-citationEgan, B. M. (2019) · Hypertension · doi:10.1161/hypertensionaha.118.11712 · Cited by 3 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Smart wearable devices in cardiovascular care: Where we are and how to move forward
Bayoumy, K., Gaber, M., Elshafeey, A., Mhaimeed, O., Dineen, E. H., Marvel, F. A., Martin, S. S., Muse, E. D., Turakhia, M. P., Tarakji, K. G., et al. (2021) · Nature Reviews Cardiology · doi:10.1038/s41569-021-00522-7 · Cited by 829 · see citing works
Wearable health devices in health care: narrative systematic review
Lu, L., Zhang, J., Xie, Y., Gao, F., Xu, S., Wu, X., & Ye, Z. (2020) · JMIR mHealth and uHealth · doi:10.2196/18907 · Cited by 701 · see citing works
- Authors
- Islam, S. M. S., Cartledge, S., Karmakar, C., Rawstorn, J. C., Fraser, S. F., Chow, C., & Maddison, R.
- Affiliations
- Deakin University (AU)
- The University of Sydney (AU)
- Corresponding author
- Islam, Sheikh Mohammed Shariful
- ORCID
- Venue
- JMIR mHealth and uHealth
- Publisher
- JMIR Publications Inc.
- Published
- 2019-09-14
- Volume
- 7
- Issue
- 10
- Pages
- e14706
- ISSN
- 2291-5222 (ISSN-L)
- PMC
- PMC6827985
- OpenAlex
- W2954015722
- Cited by
- 57
- References
- 23
- Citations by year
- 2026: 5 · 2025: 9 · 2024: 5 · 2023: 9 · 2022: 14 · 2021: 7 · 2020: 8
- FWCI
- 2.64
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- Locations
- JMIR mhealth and uhealth (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Figshare (submittedVersion, cc-by)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Keywords
- cuffless blood pressure measurement, ambulatory blood pressure monitoring, home blood pressure monitoring, device validation, measurement bias, limits of agreement, acceptability, user experience, healthcare professionals' perceptions, cardiovascular risk, long-term blood pressure monitoring
- MeSH terms
- Wearable Electronic Devices, Adolescent, Adult, Blood Pressure, Blood Pressure Determination, Female, Health Personnel, Humans, Male, Patients
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review