CoChat

A New Cuffless Device for Measuring Blood Pressure: A Real-Life Validation Study

Open access
Schoot, Tessa S, Weenk, Mariska, van de Belt, Tom H, Engelen, Lucien JLPG, van Goor, Harry, Bredie, Sebastian JH · 2016 · Journal of Medical Internet Research
Cited by 52

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

BACKGROUND: Cuffless blood pressure (BP) monitoring devices, based on pulse transit time, are being developed as an easy-to-use, more convenient, fast, and relatively cheap alternative to conventional BP measuring devices based on cuff occlusion. Thereby they may provide a great alternative to BP self-measurement. OBJECTIVE: The objective of our study was to evaluate the performance of the first release of the Checkme Health Monitor (Viatom Technology), a cuffless BP monitor, in a real-life setting. Furthermore, we wanted to investigate whether the posture of the volunteer and the position of the device relative to the heart level would influence its outcomes. METHODS: Study volunteers fell into 3 BP ranges: high (>160 mmHg), normal (130-160 mmHg), and low (<130 mmHg). All requirements for test environment, observer qualification, volunteer recruitment, and BP measurements were met according to the European Society of Hypertension International Protocol (ESH-IP) for the validation of BP measurement devices. After calibrating the Checkme device, we measured systolic BP with Checkme and a validated, oscillometric reference BP monitor (RM). Measurements were performed in randomized order both in supine and in sitting position, and with Checkme at and above heart level. RESULTS: We recruited 52 volunteers, of whom we excluded 15 (12 due to calibration failure with Checkme, 3 due to a variety of reasons). The remaining 37 volunteers were divided into low (n=14), medium (n=13), and high (n=10) BP ranges. There were 18 men and 19 women, with a mean age of 54.1 (SD 14.5) years, and mean recruitment systolic BP of 141.7 (SD 24.7) mmHg. BP results obtained by RM and Checkme correlated well. In the supine position, the difference between the RM and Checkme was >5 mmHg in 17 of 37 volunteers (46%), of whom 9 of 37 (24%) had a difference >10 mmHg and 5 of 37 (14%) had a difference >15 mmHg. CONCLUSIONS: BP obtained with Checkme correlated well with RM BP, particularly in the position (supine) in which the device was calibrated. These preliminary results are promising for conducting further research on cuffless BP measurement in the clinical and outpatient settings.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.2196/jmir.5414

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 2 references

Research works printed but not linked: 9

  1. Perloff, D, 1993, Circulation, 88, 2460, 10.1161/01.CIR.88.5.2460 · DOI · Paper page
  2. 10.1016/j.jash.2014.08.014 · DOI · Paper page
  3. 10.1097/MBP.0b013e3283360e98 · DOI · Paper page
  4. 10.1097/MBP.0b013e3283624b3b · DOI
  5. Nichols, WW, 2011, McDonald's Blood Flow in Arteries, 6th edition: Theoretical, Experimental and Clinical Principles
  6. 10.1109/TBME.2015.2440291 · DOI · Paper page
  7. 10.1016/j.jelectrocard.2010.11.019 · DOI
  8. 10.1007/s00421-011-1983-3 · DOI · Paper page
  9. 10.1155/2014/571623 · DOI · Paper page

Cited by

Metadata

Authors
Schoot, T. S., Weenk, M., van de Belt, T. H., Engelen, L. J., van Goor, H., & Bredie, S. J.
Venue
Journal of Medical Internet Research
Published
2016-05-05
Volume
18
Issue
5
Pages
e85
ISSN
1438-8871 (ISSN-L)
OpenAlex
W2346057711
Cited by
52
References
2
FWCI
3.50
Type
journal article
Language
en
Open access
gold

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.