Minimizing artifacts caused by metallic implants at MR imaging: experimental and clinical studies.
Open accessThe DOI matches its Crossref record; 9 of the 9 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: The purpose of this study was to investigate the effect of metallic implant positioning on MR imaging artifacts, to determine the optimal imaging conditions for minimizing artifacts, and to show the usefulness of artifact-minimizing methods in imaging of the knee. MATERIALS AND METHODS: Using MR images of experimental phantoms (titanium alloy and stainless steel screws), we compared the magnitude of metal-induced artifacts for various pulse sequences, different imaging parameters for the fast spin-echo sequence, and different imaging parameters for several incremental angles between the long axis of the screw and the direction of the main magnetic field. In clinical MR imaging of knees with metallic implants (n = 19), we assessed geometric distortion of anatomic structures to compare the influence of different pulse sequences (n = 19), frequency-encoding directions (n = 7), and knee positions (n = 15). RESULTS: Titanium alloy screws consistently produced smaller artifacts than did stainless steel screws. In experimental MR studies, artifacts were reduced with fast spin-echo sequences, with a screw orientation as closely parallel to the main magnetic field as possible, and, particularly, with smaller voxels that correlated positively with artifact size (R2 = .88, p < .01). In clinical MR studies, fast spin-echo MR imaging obscured articular structures less than did spin-echo imaging (8/19 patients). In particular, the anterior-posterior frequency-encoding direction (3/7 patients) and the flexion position of the knee (5/15 patients) were effective in reducing artifacts. CONCLUSION: MR artifacts can be minimized by optimally positioning in the magnet subjects with metallic implants and by choosing fast spin-echo sequences with an anterior-posterior frequency-encoding direction and the smallest voxel size.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Schenck, J. F. (1996) · Medical Physics · doi:10.1118/1.597854 · Cited by 1,522
Susceptibility artefacts in NMR imaging
Lüdeke, K. M., Röschmann, P., & Tischler, R. (1985) · Magnetic Resonance Imaging · doi:10.1016/0730-725x(85)90397-2 · Cited by 445
Sources of intensity nonuniformity in spin echo images at 1.5 T
Simmons, A., Tofts, P. S., Barker, G. J., & Arridge, S. R. (1994) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910320117 · Cited by 277
Reduction of susceptibility artifact in gradient‐echo imaging
Cho, Z. H., & Ro, Y. M. (1992) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910230120 · Cited by 165
Analysis of machine-dependent and object-induced geometric distortion in 2DFT MR imaging
Bakker, C. J. G., Moerland, M. A., Bhawandien, R., & Beersma, R. (1992) · Magnetic Resonance Imaging · doi:10.1016/0730-725x(92)90011-n · Cited by 96
Magnetic Resonance Imaging for the Determination of Magnetic Susceptibility of Materials
Beuf, O., Briguet, A., Lissac, M., & Davis, R. (1996) · Journal of Magnetic Resonance Series B · doi:10.1006/jmrb.1996.0120 · Cited by 68
MR Imaging of Large Nonferromagnetic Metallic Implants at 1.5 T
Augustiny, N., von Schulthess, G. K., Meier, D., & Boesiger, P. (1987) · Journal of Computer Assisted Tomography · doi:10.1097/00004728-198707000-00024 · Cited by 43
Evaluation of the susceptibility effect on the phase images of a simple gradient echo.
Yamada, N., Imakita, S., Sakuma, T., Nishimura, Y., Yamada, Y., Naito, H., Nishimura, T., & Takamiya, M. (1990) · Radiology · doi:10.1148/radiology.175.2.2326482 · Cited by 35
Fat-suppressed three-dimensional fast spoiled gradient-recalled echo imaging
Wang, S., Cheng, H., & Chang, C. (1999) · Clinical Imaging · doi:10.1016/s0899-7071(99)00127-8 · Cited by 27
Cited by
Lee, M.-J., Kim, S., Lee, S.-A., Song, H.-T., Huh, Y.-M., Kim, D.-H., Han, S. H., & Suh, J.-S. (2007) · RadioGraphics · doi:10.1148/rg.273065087 · Cited by 513
Self-citationDual-Energy CT for the Musculoskeletal System
Mallinson, P. I., Coupal, T. M., McLaughlin, P. D., Nicolaou, S., Munk, P. L., & Ouellette, H. A. (2016) · Radiology · doi:10.1148/radiol.2016151109 · Cited by 242
Stradiotti, P., Curti, A., Castellazzi, G., & Zerbi, A. (2009) · European Spine Journal · doi:10.1007/s00586-009-0998-5 · Cited by 241
“In addition, the artifact size is affected by the implant size with larger implants producing more obtrusive artifacts [18, 19].”
BackgroundJungmann, P. M., Agten, C. A., Pfirrmann, C. W., & Sutter, R. (2017) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.25708 · Cited by 227
“constant field of view) reduces both in-plane distortions and intravoxel dephasing.(66) Matrix sizes of 512 are recommended.”
BackgroundComplications of Total Hip Arthroplasty: MR Imaging—Initial Experience
White, L. M., Kim, J. K., Mehta, M., Merchant, N., Schweitzer, M. E., Morrison, W. B., Hutchison, C. R., & Gross, A. E. (2000) · Radiology · doi:10.1148/radiology.215.1.r00ap11254 · Cited by 226
“As a result, the frequency-encoding gradient was oriented parallel to the long axis of the femoral prostheses during our clinical study to decrease the degree of artifact along the bone prosthetic interface of the femoral prosthetic stem (8,12,13).”
BackgroundMethod
Metadata
- Authors
- Suh, J. S., Jeong, E. K., Shin, K. H., Cho, J. H., Na, J. B., Kim, D. H., & Han, C. D.
- Venue
- American Journal of Roentgenology
- Publisher
- American Roentgen Ray Society
- Published
- 1998-11
- Volume
- 171
- Issue
- 5
- Pages
- 1207-1213
- ISSN
- 0002-9580 (ISSN-L)
- 0361-803X
- 1546-3141
- OpenAlex
- W1981391986
- Cited by
- 183
- References
- 9
- FWCI
- 4.73
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.