Insulin Pump Therapy With Automated Insulin Suspension in Response to Hypoglycemia
Open accessThe DOI matches its Crossref record, with a small difference in title or year; 12 of the 14 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: To evaluate a sensor-augmented insulin pump with a low glucose suspend (LGS) feature that automatically suspends basal insulin delivery for up to 2 h in response to sensor-detected hypoglycemia. RESEARCH DESIGN AND METHODS: The LGS feature of the Paradigm Veo insulin pump (Medtronic, Inc., Northridge, CA) was tested for 3 weeks in 31 adults with type 1 diabetes. RESULTS: There were 166 episodes of LGS: 66% of daytime LGS episodes were terminated within 10 min, and 20 episodes lasted the maximum 2 h. LGS use was associated with reduced nocturnal duration ≤2.2 mmol/L in those in the highest quartile of nocturnal hypoglycemia at baseline (median 46.2 vs. 1.8 min/day, P = 0.02 [LGS-OFF vs. LGS-ON]). Median sensor glucose was 3.9 mmol/L after 2-h LGS and 8.2 mmol/L at 2 h after basal restart. CONCLUSIONS: Use of an insulin pump with LGS was associated with reduced nocturnal hypoglycemia in those at greatest risk and was well accepted by patients.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“This device shuts off insulin delivery for up to 2 h once the interstitial glucose concentration reaches a preset threshold and reduces the duration of nocturnal hypoglycemia (123).”
From Hypoglycemia and Diabetes: A Report of a Workgroup of the American Diabetes Association and The Endocrine Society (2013) · cites this paper as [123]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“This device shuts off insulin delivery for up to 2 h once the interstitial glucose concentration reaches a preset threshold and reduces the duration of nocturnal hypoglycemia (123).”
From Hypoglycemia and Diabetes: A Report of a Workgroup of the American Diabetes Association and The Endocrine Society (2013) · cites this paper as [123]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Continuous glucose monitoring and intensive treatment of type 1 diabetes
(2008) · New England Journal of Medicine · doi:10.1056/nejmoa0805017 · Cited by 1,559
Risk of hypoglycaemia in types 1 and 2 diabetes: Effects of treatment modalities and their duration
UK Hypoglycaemia Study Group (2007) · Diabetologia · doi:10.1007/s00125-007-0599-y · Cited by 825
Pickup, J. C., & Sutton, A. J. (2008) · Diabetic Medicine · doi:10.1111/j.1464-5491.2008.02486.x · Cited by 617
Self-citationDecreased Epinephrine Responses to Hypoglycemia during Sleep
Jones, T. W., Porter, P., Sherwin, R. S., Davis, E. A., O'Leary, P., Frazer, F., Byrne, G., Stick, S., & Tamborlane, W. V. (1998) · New England Journal of Medicine · doi:10.1056/nejm199806043382303 · Cited by 301
Raccah, D., Sulmont, V., Reznik, Y., Guerci, B., Renard, E., Hanaire, H., Jeandidier, N., & Nicolino, M. (2009) · Diabetes Care · doi:10.2337/dc09-0750 · Cited by 237
O’Connell, M. A., Donath, S., O’Neal, D. N., Colman, P. G., Ambler, G. R., Jones, T. W., Davis, E. A., & Cameron, F. J. (2009) · Diabetologia · doi:10.1007/s00125-009-1365-0 · Cited by 211
Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group (2010) · Diabetes Care · doi:10.2337/dc09-2081 · Cited by 180
Monsod, T. P., Flanagan, D. E., Rife, F., Saenz, R., Caprio, S., Sherwin, R. S., & Tamborlane, W. V. (2002) · Diabetes Care · doi:10.2337/diacare.25.5.889 · Cited by 173
Response to Nocturnal Alarms Using a Real-Time Glucose Sensor
Buckingham, B. A., Block, J. M., Burdick, J., Kalajian, A., Kollman, C., Choy, M., Wilson, D. M., & Chase, P. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.440 · Cited by 115
Accuracy of a New Real-Time Continuous Glucose Monitoring Algorithm
Keenan, D. B., Cartaya, R., & Mastrototaro, J. J. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400114 · Cited by 82
Guerci, B., Meyer, L., Sallé, A., Charrié, A., Dousset, B., Ziegler, O., & Drouin, P. (1999) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jcem.84.8.5912 · Cited by 58
Pickup, J. C., Viberti, G. C., Bilous, R. W., KEEN, H., Alberti, K. G. M. M., Home, P. D., & Binder, C. (1982) · Diabetologia · doi:10.1007/bf00283748 · Cited by 47
Self-citation
Cited by
Seaquist, E. R., Anderson, J., Childs, B., Cryer, P., Dagogo-Jack, S., Fish, L., Heller, S. R., Rodriguez, H., Rosenzweig, J., & Vigersky, R. (2013) · Diabetes Care · doi:10.2337/dc12-2480 · Cited by 1,389
Threshold-based insulin-pump interruption for reduction of hypoglycemia
Bergenstal, R. M., Klonoff, D. C., Garg, S. K., Bode, B. W., Meredith, M., Slover, R. H., Ahmann, A. J., Welsh, J. B., Lee, S. W., & Kaufman, F. R. (2013) · New England Journal of Medicine · doi:10.1056/nejmoa1303576 · Cited by 652
Seaquist, E. R., Anderson, J., Childs, B., Cryer, P., Dagogo-Jack, S., Fish, L., Heller, S. R., Rodriguez, H., Rosenzweig, J., & Vigersky, R. (2013) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2012-4127 · Cited by 412
Ly, T. T., Nicholas, J. A., Retterath, A., Lim, E. M., Davis, E. A., & Jones, T. W. (2013) · JAMA · doi:10.1001/jama.2013.277818 · Cited by 411
Hu, X., Yu, J., Qian, C., Lu, Y., Kahkoska, A. R., Xie, Z., Jing, X., Buse, J. B., & Gu, Z. (2017) · ACS Nano · doi:10.1021/acsnano.6b06892 · Cited by 313
Metadata
- Authors
- Choudhary, P., Shin, J., Wang, Y., Evans, M. L., Hammond, P. J., Kerr, D., Shaw, J. A. M., Pickup, J. C., & Amiel, S. A.
- Affiliations
- King's College London (GB)
- Medtronic (United States) (US)
- University of Cambridge (GB)
- Harrogate District Hospital (GB)
- Bournemouth Diabetes and Endocrine Centre, Bournemouth District General Hospital, Bournemouth, U.K.
- Newcastle University (GB)
- Corresponding author
- Choudhary, Pratik
- ORCID
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 2011-08-19
- Volume
- 34
- Issue
- 9
- Pages
- 2023-2025
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- PMC
- PMC3161284
- OpenAlex
- W2071497593
- Cited by
- 175
- References
- 14
- Citations by year
- 2025: 3 · 2024: 4 · 2022: 2 · 2021: 3 · 2020: 7 · 2019: 5 · 2018: 14 · 2017: 12 · 2016: 16 · 2015: 17 · 2014: 24 · 2013: 30 · 2012: 33
- FWCI
- 23.88
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by-nc-nd
- Locations
- Diabetes Care (publishedVersion, cc-by-nc-nd)
- Europe PMC (PubMed Central) (publishedVersion, cc-by-nc-nd)
- PubMed Central (publishedVersion, cc-by-nc-nd)
- PubMed (publishedVersion)
- Research Portal (King's College London) (submittedVersion)
- Newcastle University ePrints (Newcastle Univesity)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Keywords
- insulin pump therapy, low glucose suspend, sensor-augmented pump, type 1 diabetes, nocturnal hypoglycemia, continuous glucose monitoring, hypoglycemia detection
- MeSH terms
- Adult, Circadian Rhythm, Female, Humans, Hypoglycemia, Hypoglycemic Agents, Insulin, Insulin Infusion Systems, Male, Middle Aged
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.