Insulin Pump Therapy With Automated Insulin Suspension in Response to Hypoglycemia
Open accessThe DOI matches its Crossref record, with a small difference in title or year; 12 of the 14 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: To evaluate a sensor-augmented insulin pump with a low glucose suspend (LGS) feature that automatically suspends basal insulin delivery for up to 2 h in response to sensor-detected hypoglycemia. RESEARCH DESIGN AND METHODS: The LGS feature of the Paradigm Veo insulin pump (Medtronic, Inc., Northridge, CA) was tested for 3 weeks in 31 adults with type 1 diabetes. RESULTS: There were 166 episodes of LGS: 66% of daytime LGS episodes were terminated within 10 min, and 20 episodes lasted the maximum 2 h. LGS use was associated with reduced nocturnal duration ≤2.2 mmol/L in those in the highest quartile of nocturnal hypoglycemia at baseline (median 46.2 vs. 1.8 min/day, P = 0.02 [LGS-OFF vs. LGS-ON]). Median sensor glucose was 3.9 mmol/L after 2-h LGS and 8.2 mmol/L at 2 h after basal restart. CONCLUSIONS: Use of an insulin pump with LGS was associated with reduced nocturnal hypoglycemia in those at greatest risk and was well accepted by patients.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“This device shuts off insulin delivery for up to 2 h once the interstitial glucose concentration reaches a preset threshold and reduces the duration of nocturnal hypoglycemia (123).”
From Hypoglycemia and Diabetes: A Report of a Workgroup of the American Diabetes Association and The Endocrine Society (2013) · cites this paper as [123]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“This device shuts off insulin delivery for up to 2 h once the interstitial glucose concentration reaches a preset threshold and reduces the duration of nocturnal hypoglycemia (123).”
From Hypoglycemia and Diabetes: A Report of a Workgroup of the American Diabetes Association and The Endocrine Society (2013) · cites this paper as [123]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Continuous glucose monitoring and intensive treatment of type 1 diabetes
(2008) · New England Journal of Medicine · doi:10.1056/nejmoa0805017 · Cited by 1,559
Risk of hypoglycaemia in types 1 and 2 diabetes: Effects of treatment modalities and their duration
UK Hypoglycaemia Study Group (2007) · Diabetologia · doi:10.1007/s00125-007-0599-y · Cited by 825
Pickup, J. C., & Sutton, A. J. (2008) · Diabetic Medicine · doi:10.1111/j.1464-5491.2008.02486.x · Cited by 617
Self-citationDecreased Epinephrine Responses to Hypoglycemia during Sleep
Jones, T. W., Porter, P., Sherwin, R. S., Davis, E. A., O'Leary, P., Frazer, F., Byrne, G., Stick, S., & Tamborlane, W. V. (1998) · New England Journal of Medicine · doi:10.1056/nejm199806043382303 · Cited by 301
Raccah, D., Sulmont, V., Reznik, Y., Guerci, B., Renard, E., Hanaire, H., Jeandidier, N., & Nicolino, M. (2009) · Diabetes Care · doi:10.2337/dc09-0750 · Cited by 237
O’Connell, M. A., Donath, S., O’Neal, D. N., Colman, P. G., Ambler, G. R., Jones, T. W., Davis, E. A., & Cameron, F. J. (2009) · Diabetologia · doi:10.1007/s00125-009-1365-0 · Cited by 211
Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group (2010) · Diabetes Care · doi:10.2337/dc09-2081 · Cited by 180
Monsod, T. P., Flanagan, D. E., Rife, F., Saenz, R., Caprio, S., Sherwin, R. S., & Tamborlane, W. V. (2002) · Diabetes Care · doi:10.2337/diacare.25.5.889 · Cited by 173
Response to Nocturnal Alarms Using a Real-Time Glucose Sensor
Buckingham, B. A., Block, J. M., Burdick, J., Kalajian, A., Kollman, C., Choy, M., Wilson, D. M., & Chase, P. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.440 · Cited by 115
Accuracy of a New Real-Time Continuous Glucose Monitoring Algorithm
Keenan, D. B., Cartaya, R., & Mastrototaro, J. J. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400114 · Cited by 82
Cited by
Seaquist, E. R., Anderson, J., Childs, B., Cryer, P., Dagogo-Jack, S., Fish, L., Heller, S. R., Rodriguez, H., Rosenzweig, J., & Vigersky, R. (2013) · Diabetes Care · doi:10.2337/dc12-2480 · Cited by 1,389
Threshold-based insulin-pump interruption for reduction of hypoglycemia
Bergenstal, R. M., Klonoff, D. C., Garg, S. K., Bode, B. W., Meredith, M., Slover, R. H., Ahmann, A. J., Welsh, J. B., Lee, S. W., & Kaufman, F. R. (2013) · New England Journal of Medicine · doi:10.1056/nejmoa1303576 · Cited by 652
Seaquist, E. R., Anderson, J., Childs, B., Cryer, P., Dagogo-Jack, S., Fish, L., Heller, S. R., Rodriguez, H., Rosenzweig, J., & Vigersky, R. (2013) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2012-4127 · Cited by 412
Ly, T. T., Nicholas, J. A., Retterath, A., Lim, E. M., Davis, E. A., & Jones, T. W. (2013) · JAMA · doi:10.1001/jama.2013.277818 · Cited by 411
Hu, X., Yu, J., Qian, C., Lu, Y., Kahkoska, A. R., Xie, Z., Jing, X., Buse, J. B., & Gu, Z. (2017) · ACS Nano · doi:10.1021/acsnano.6b06892 · Cited by 313
Little, S. A., Leelarathna, L., Walkinshaw, E., Tan, H. K., Chapple, O., Lubina-Solomon, A., Chadwick, T. J., Barendse, S., Stocken, D. D., Brennand, C., et al. (2014) · Diabetes Care · doi:10.2337/dc14-0030 · Cited by 237
Self-citationAssessing Sensor Accuracy for Non-Adjunct Use of Continuous Glucose Monitoring
Kovatchev, B. P., Patek, S. D., Ortiz, E. A., & Breton, M. D. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0272 · Cited by 234
Abraham, M. B., Jones, T. W., Naranjo, D., Karges, B., Oduwole, A., Tauschmann, M., & Maahs, D. M. (2018) · Pediatric Diabetes · doi:10.1111/pedi.12698 · Cited by 226
Pickup, J. C., Ford Holloway, M., & Samsi, K. (2015) · Diabetes Care · doi:10.2337/dc14-1855 · Cited by 189
Self-citationOral delivery of insulin for treatment of diabetes: status quo, challenges and opportunities
Wong, C. Y., Martinez, J., & Dass, C. R. (2016) · Journal of Pharmacy and Pharmacology · doi:10.1111/jphp.12607 · Cited by 183
Challenges and recent progress in the development of a closed-loop artificial pancreas
Bequette, B. W. (2012) · Annual Reviews in Control · doi:10.1016/j.arcontrol.2012.09.007 · Cited by 183
Choudhary, P., Ramasamy, S., Green, L., Gallen, G., Pender, S., Brackenridge, A., Amiel, S. A., & Pickup, J. C. (2013) · Diabetes Care · doi:10.2337/dc13-0939 · Cited by 160
Self-citationTechnology in the management of type 1 diabetes mellitus — current status and future prospects
Tauschmann, M., & Hovorka, R. (2018) · Nature Reviews Endocrinology · doi:10.1038/s41574-018-0044-y · Cited by 183
Thabit, H., Lubina-Solomon, A., Stadler, M., Leelarathna, L., Walkinshaw, E., Pernet, A., Allen, J. M., Iqbal, A., Choudhary, P., Kumareswaran, K., et al. (2014) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(14)70114-7 · Cited by 162
Self-citationGarg, S., Brazg, R. L., Bailey, T. S., Buckingham, B. A., Slover, R. H., Klonoff, D. C., Shin, J., Welsh, J. B., & Kaufman, F. R. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0292 · Cited by 138
Self-citationBiester, T., Kordonouri, O., Holder, M., Remus, K., Kieninger-Baum, D., Wadien, T., & Danne, T. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0349 · Cited by 128
Choudhary, P., Olsen, B. S., Conget, I., Welsh, J. B., Vorrink, L., & Shin, J. J. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0324 · Cited by 118
Self-citationBuckingham, B. A., Cameron, F., Calhoun, P. M., Maahs, D. M., Wilson, D. M., Chase, H. P., Bequette, B. W., Lum, J. W., Sibayan, J., Beck, R. W., et al. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0040 · Cited by 104
Closed-loop insulin delivery for treatment of type 1 diabetes
Elleri, D., Dunger, D. B., & Hovorka, R. (2011) · BMC Medicine · doi:10.1186/1741-7015-9-120 · Cited by 114
Evolution of Diabetes Technology
Dovč, K., & Battelino, T. (2019) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2019.10.009 · Cited by 132
Danne, T., Tsioli, C., Kordonouri, O., Bläsig, S., Remus, K., Roy, A., Keenan, D. B., Lee, S. W., & KAUFMAN, F. R. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0327 · Cited by 93
Continuous glucose monitoring in type 1 diabetes
Mauras, N., Fox, L. A., ENGLERT, K. A., & Beck, R. W. (2012) · Endocrine · doi:10.1007/s12020-012-9765-1 · Cited by 98
Bertachi, A. H., Viñals, C., Biagi, L., Contreras, I., Vehí, J., Conget, I., & Giménez, M. (2020) · Sensors · doi:10.3390/s20061705 · Cited by 96
Clinical Use of Continuous Glucose Monitoring in Adults with Type 1 Diabetes
Slattery, D., & Choudhary, P. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0051 · Cited by 87
Self-citationLittle, S. A., Speight, J., Leelarathna, L., Walkinshaw, E., Tan, H. K., Bowes, A., Lubina-Solomon, A., Chadwick, T. J., Stocken, D. D., Brennand, C., et al. (2018) · Diabetes Care · doi:10.2337/dc17-2682 · Cited by 86
Self-citationZhong, A., Choudhary, P., McMahon, C., Agrawal, P. K., Welsh, J. B., Cordero, T. L., & KAUFMAN, F. R. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0216 · Cited by 80
Self-citationPsychology, technology, and diabetes management.
Gonder-Frederick, L. A., Shepard, J. A., Grabman, J. H., & Ritterband, L. M. (2016) · American Psychologist · doi:10.1037/a0040383 · Cited by 85
Management of diabetes mellitus: is the pump mightier than the pen?
Pickup, J. C. (2012) · Nature Reviews Endocrinology · doi:10.1038/nrendo.2012.28 · Cited by 90
Self-citationClosed-Loop Insulin Delivery in Type 1 Diabetes
Thabit, H., & Hovorka, R. (2012) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2011.12.003 · Cited by 83
Advances in management of type 1 diabetes mellitus
Ravindranath, A., & Jain, V. (2014) · World Journal of Diabetes · doi:10.4239/wjd.v5.i5.689 · Cited by 91
Continuous Glucose Monitoring: Evidence and Consensus Statement for Clinical Use
Liebl, A., Henrichs, H. R., Heinemann, L., Freckmann, G., Biermann, E., & Thomas, A. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700227 · Cited by 77
Closed-Loop System in the Management of Diabetes: Past, Present, and Future
Shah, V. N., Shoskes, A., Tawfik, B., & Garg, S. K. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0193 · Cited by 84
Morbidity and Mortality of Diabetic Ketoacidosis With and Without Insulin Pump Care
Realsen, J., Goettle, H., & Chase, H. P. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0161 · Cited by 81
Riemsma, R., Corro Ramos, I., Birnie, R., Büyükkaramikli, N., Armstrong, N., Ryder, S., Duffy, S., Worthy, G., Al, M., Severens, J., et al. (2016) · Health Technology Assessment · doi:10.3310/hta20170 · Cited by 73
The risks of nocturnal hypoglycaemia in insulin-treated diabetes
Graveling, A. J., & Frier, B. M. (2017) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2017.08.012 · Cited by 85
Heller, S., White, D., Lee, E., Lawton, J., Pollard, D., Waugh, N., Amiel, S., Barnard, K., Beckwith, A., Brennan, A., et al. (2017) · Health Technology Assessment · doi:10.3310/hta21200 · Cited by 73
Self-citationBoscari, F., & Avogaro, A. (2021) · Reviews in Endocrine and Metabolic Disorders · doi:10.1007/s11154-021-09635-3 · Cited by 80
A review of artificial pancreas technologies with an emphasis on bi‐hormonal therapy
Bakhtiani, P. A., Zhao, L., Youssef, J. E., Castle, J. R., & Ward, W. K. (2013) · Diabetes Obesity and Metabolism · doi:10.1111/dom.12107 · Cited by 69
Uncovering undetected hypoglycemic events
Unger, J. (2012) · Diabetes Metabolic Syndrome and Obesity · doi:10.2147/dmso.s29367 · Cited by 71
Thabit, H., & Hovorka, R. (2016) · Expert Opinion on Drug Delivery · doi:10.1517/17425247.2016.1115013 · Cited by 58
Sensor-Augmented Insulin Pumps and Hypoglycemia Prevention in Type 1 Diabetes
Steineck, I. I. K., Ranjan, A. G., Nørgaard, K., & Schmidt, S. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816672689 · Cited by 63
Inpatient Studies of a Kalman-Filter-Based Predictive Pump Shutoff Algorithm
Cameron, F., Wilson, D. M., Buckingham, B. A., Arzumanyan, H., Clinton, P. K., Chase, H. P., Lum, J. W., Maahs, D. M., Calhoun, P. M., & Bequette, B. W. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600519 · Cited by 56
Hovorka, R., Nodale, M., Haidar, A., & Wilinska, M. E. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0185 · Cited by 61
Severe Hypoglycemia: Is It Still a Threat for Children and Adolescents With Type 1 Diabetes?
Urakami, T. (2020) · Frontiers in Endocrinology · doi:10.3389/fendo.2020.00609 · Cited by 59
Schaepelynck, P., Darmon, P., Molines, L., Jannot-Lamotte, M.-F., Tréglia, C., & Raccah, D. (2011) · Diabetes & Metabolism · doi:10.1016/s1262-3636(11)70972-7 · Cited by 60
Hypoglycemia Reduction and Changes in Hemoglobin A1c in the ASPIRE In-Home Study
Weiss, R., Garg, S. K., Bode, B. W., Bailey, T. S., Ahmann, A. J., Schultz, K. A., Welsh, J. B., Shin, J. J., & for the ASPIRE In-Home Study Group (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0306 · Cited by 56
Self-citationEvaluation of a Predictive Low-Glucose Management System In-Clinic
Buckingham, B. A., Bailey, T. S., Christiansen, M., Garg, S., Weinzimer, S., Bode, B., Anderson, S. M., Brazg, R., Ly, T. T., & Kaufman, F. R. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0319 · Cited by 52
Choudhary, P., & Amiel, S. A. (2018) · Diabetologia · doi:10.1007/s00125-018-4566-6 · Cited by 52
Self-citationVettoretti, M., & Facchinetti, A. (2019) · BioMedical Engineering OnLine · doi:10.1186/s12938-019-0658-x · Cited by 56
Analysis of Glucose Responses to Automated Insulin Suspension With Sensor-Augmented Pump Therapy
Ly, T. T., Nicholas, J. A., Retterath, A., Davis, E. A., & Jones, T. W. (2012) · Diabetes Care · doi:10.2337/dc12-0052 · Cited by 49
Little, S. A., Leelarathna, L., Barendse, S. M., Walkinshaw, E., Tan, H. K., Lubina Solomon, A., de Zoysa, N., Rogers, H., Choudhary, P., Amiel, S. A., et al. (2014) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.2492 · Cited by 48
Self-citationZecchin, C., Facchinetti, A., Sparacino, G., & Cobelli, C. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0208 · Cited by 51
Benhamou, P., Catargi, B., Delenne, B., Guerci, B., Hanaire, H. P., Jeandidier, N., Leroy, R., Meyer, L., Penfornis, A., Radermecker, R. P., et al. (2012) · Diabetes & Metabolism · doi:10.1016/s1262-3636(12)71538-0 · Cited by 46
Thieu, V. T., Mitchell, B. D., Varnado, O. J., & Frier, B. M. (2020) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.13941 · Cited by 44
Beck, R. W., Raghinaru, D., Wadwa, R. P., Chase, H. P., Maahs, D. M., Buckingham, B. A., & for the In Home Closed Loop Study Group (2014) · Diabetes Care · doi:10.2337/dc13-2775 · Cited by 44
Roze, S., Smith‐Palmer, J., Valentine, W. J., Cook, M., Jethwa, M., de Portu, S., & Pickup, J. C. (2015) · Journal of Medical Economics · doi:10.3111/13696998.2015.1113979 · Cited by 42
Self-citationAgrawal, P. K., Zhong, A., Welsh, J. B., Shah, R. V., & KAUFMAN, F. R. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0257 · Cited by 40
DEHENNIS, A. D., Mortellaro, M. A., & Ioacără, S. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815596760 · Cited by 44
Continuous Glucose Monitoring: Impact on Hypoglycemia
van Beers, C. A. J., & DeVries, J. H. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816653411 · Cited by 46
Abraham, M. B., Davey, R., O'Grady, M. J., Ly, T. T., Paramalingam, N., Fournier, P. A., Roy, A., Grosman, B., Kurtz, N., Fairchild, J. M., et al. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0141 · Cited by 40
Lucidi, P., Porcellati, F., Bolli, G. B., & Fanelli, C. G. (2018) · Current Diabetes Reports · doi:10.1007/s11892-018-1065-6 · Cited by 38
Technology to Optimize Pediatric Diabetes Management and Outcomes
Markowitz, J. T., Harrington, K. R., & Laffel, L. M. (2013) · Current Diabetes Reports · doi:10.1007/s11892-013-0419-3 · Cited by 41
Elbarbary, N. S. (2016) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.2781 · Cited by 37
Abraham, M. B., de Bock, M. I., Paramalingam, N., O'Grady, M. J., Ly, T. T., George, C., Roy, A., Spital, G., Karula, S., Heels, K., et al. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0364 · Cited by 36
Hering, B. J., & O’Connell, P. J. (2016) · Xenotransplantation · doi:10.1111/xen.12228 · Cited by 37
Sherr, J. L., Palau Collazo, M., Cengiz, E., Michaud, C., Carria, L., Steffen, A. T., Weyman, K., Zgorski, M., Tichy, E., Tamborlane, W. V., et al. (2014) · Diabetes Care · doi:10.2337/dc13-1608 · Cited by 35
Feupe, S. F., Frias, P. F., Mednick, S. C., McDevitt, E. A., & Heintzman, N. D. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700525 · Cited by 36
Asgharzadeh, A., Patel, M., Connock, M. J., Damery, S. L., Ghosh, I., Jordan, M., Freeman, K., Brown, A., Court, R. A., Baldwin, S., et al. (2024) · Health Technology Assessment · doi:10.3310/jypl3536 · Cited by 37
Hypo- and Hyperglycemic Alarms: Devices and Algorithms
Howsmon, D. P., & Bequette, B. W. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815583507 · Cited by 35
Wang, W., Wang, S., Wang, X., Liu, D., Geng, Y., & Wu, T. (2020) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2020.3015199 · Cited by 33
Hypoglycemia Begets Hypoglycemia: The Order Effect in the ASPIRE In-Clinic Study
Garg, S. K., Brazg, R., Bailey, T. S., Buckingham, B. A., Slover, R. H., Klonoff, D. C., Shin, J., Welsh, J. B., & KAUFMAN, F. R. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0219 · Cited by 31
Self-citationÖzcan, Ş., Amiel, S. A., Rogers, H., Choudhary, P., Cox, A., de Zoysa, N., Hopkins, D., & Forbes, A. (2014) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2014.07.023 · Cited by 28
Self-citationShah, V. N., Moser, E. G., Blau, A., Dhingra, M., & Garg, S. K. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0228 · Cited by 27
Outpatient Management of Pediatric Type 1 Diabetes
Beck, J. K., & Cogen, F. R. (2015) · The Journal of Pediatric Pharmacology and Therapeutics · doi:10.5863/1551-6776-20.5.344 · Cited by 25
Immune Intervention in Type I Diabetes Mellitus
Ludvigsson, J. (2013) · InTech eBooks · doi:10.5772/52801 · Cited by 24
Banting Memorial Lecture 2014 Technology and diabetes care: appropriate and personalized
Pickup, J. C. (2014) · Diabetic Medicine · doi:10.1111/dme.12613 · Cited by 22
Self-citationCrenier, L., Lytrivi, M., Dalem, A. V., Keymeulen, B., & Corvilain, B. (2016) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2015-4035 · Cited by 22
Insulin pump therapy in youth with type 1 diabetes: toward closed-loop systems
Tauschmann, M., & Hovorka, R. (2014) · Expert Opinion on Drug Delivery · doi:10.1517/17425247.2014.910192 · Cited by 24
Detecting sensor and insulin infusion set anomalies in an artificial pancreas
Baysal, N., Cameron, F., Buckingham, B. A., Wilson, D. M., & Bequette, B. W. (2013) · Proceedings of the ... American Control Conference/Proceedings of the American Control Conference · doi:10.1109/acc.2013.6580279 · Cited by 21
Who Should Be Considered for Islet Transplantation Alone?
Othonos, N., & Choudhary, P. (2017) · Current Diabetes Reports · doi:10.1007/s11892-017-0847-6 · Cited by 24
Self-citationJENDLE, J. H., Smith‐Palmer, J., Delbaere, A., de Portu, S., Papo, N. L., Valentine, W. J., & Roze, S. (2017) · Diabetes Therapy · doi:10.1007/s13300-017-0294-z · Cited by 22
Dassau, E., Atlas, E., & Phillip, M. (2010) · International Journal of Clinical Practice · doi:10.1111/j.1742-1241.2009.02274.x · Cited by 22
Abraham, M. B., Nicholas, J. A., Ly, T. T., Roby, H. C., Paramalingam, N., Fairchild, J. M., King, B. R., Ambler, G. R., Cameron, F. J., Davis, E. A., et al. (2016) · BMJ Open · doi:10.1136/bmjopen-2016-011589 · Cited by 19
Nimri, R., Yakoob, H., Schoenberg, B. E., & Dassau, E. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.2504 · Cited by 20
Inpatient studies of a Kalman-filter-based predictive pump shutoff algorithm
Cameron, F., Wilson, D. M., Buckingham, B. A., Arzumanyan, H., Clinton, P. K., Chase, H. P., Lum, J. W., Maahs, D. M., Calhoun, P. M., & Bequette, B. W. (2012) · Diabetes Technology & Therapeutics · Cited by 17
Davis, T., Salahi, A., Welsh, J. B., & Bailey, T. S. (2015) · Diabetes Obesity and Metabolism · doi:10.1111/dom.12542 · Cited by 19
Porcellati, F., Mauro, S. D., Mazzieri, A., Scamporrino, A., Filippello, A., Fano, M. D., Fanelli, C. G., Purrello, F., Malaguarnera, R., & Piro, S. (2021) · Biomolecules · doi:10.3390/biom11091281 · Cited by 19
Sperling, M. A., Tamborlane, W. V., Battelino, T., WEINZIMER, S. A., & Phillip, M. (2014) · Elsevier eBooks · doi:10.1016/b978-1-4557-4858-7.00028-7 · Cited by 16
Closed-loop artificial pancreas: current studies and promise for the future
WEINZIMER, S. A. (2012) · Current Opinion in Endocrinology Diabetes and Obesity · doi:10.1097/med.0b013e3283514e6b · Cited by 16
Challenges and progress in the development of a closed-loop artificial pancreas
Bequette, B. W. (2012) · Proceedings of the ... American Control Conference/Proceedings of the American Control Conference · doi:10.1109/acc.2012.6315593 · Cited by 15
Dassau, E., Atlas, E., & Phillip, M. (2011) · International Journal of Clinical Practice · doi:10.1111/j.1742-1241.2010.02575.x · Cited by 15
Insulin Pump Therapy With Automated Insulin Suspension: Toward Freedom From Nocturnal Hypoglycemia
Choudhary, P. (2013) · JAMA · doi:10.1001/jama.2013.278576 · Cited by 15
Self-citationThe latest pharmacotherapy options for type 1 diabetes
Ludvigsson, J. (2013) · Expert Opinion on Pharmacotherapy · doi:10.1517/14656566.2014.855197 · Cited by 13
Schöttler, H., Auzanneau, M., Best, F., Braune, K., Freff, M., Heidtmann, B., Jung, R., Karges, B., Klee, D., Müller, A., et al. (2020) · Diabetologie und Stoffwechsel · doi:10.1055/a-1259-1190 · Cited by 15
Treatment of patients with type 1 diabetes – Insulin pumps or multiple injections?
Krzymień, J., Rachuta, M., Kozlowska, I., Ładyżyński, P., & Foltyński, P. (2015) · Journal of Applied Biomedicine · doi:10.1016/j.bbe.2015.10.002 · Cited by 12
Weiss, R., Garg, S. K., Bergenstal, R. M., Klonoff, D. C., Bode, B. W., Bailey, T. S., Thrasher, J. R., Schwartz, F., Welsh, J. B., KAUFMAN, F. R., et al. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815586014 · Cited by 12
Klonoff, D. C., Bergenstal, R. M., Garg, S. K., Bode, B. W., Meredith, M., Slover, R. H., Ahmann, A., Welsh, J. B., & Lee, S. W. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700424 · Cited by 12
An Expert Opinion on Advanced Insulin Pump Use in Youth with Type 1 Diabetes
Bode, B. W., KAUFMAN, F. R., & Vint, N. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0354 · Cited by 12
Medical care of pregnant women with type 1 diabetes: current guidelines and clinical practice
Cyganek, K., Klupa, T., Szopa, M., Katra, B., & Malecki, M. T. (2013) · Polskie Archiwum Medycyny Wewnętrznej · doi:10.20452/pamw.1595 · Cited by 14
Technology to Reduce Hypoglycemia
Yeoh, E. C. K., & Choudhary, P. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815574547 · Cited by 12
Self-citation
Metadata
- Authors
- Choudhary, P., Shin, J., Wang, Y., Evans, M. L., Hammond, P. J., Kerr, D., Shaw, J. A. M., Pickup, J. C., & Amiel, S. A.
- Affiliations
- King's College London (GB)
- Medtronic (United States) (US)
- University of Cambridge (GB)
- Harrogate District Hospital (GB)
- Bournemouth Diabetes and Endocrine Centre, Bournemouth District General Hospital, Bournemouth, U.K.
- Newcastle University (GB)
- Corresponding author
- Choudhary, Pratik
- ORCID
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 2011-08-19
- Volume
- 34
- Issue
- 9
- Pages
- 2023-2025
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- PMC
- PMC3161284
- OpenAlex
- W2071497593
- Cited by
- 175
- References
- 14
- Citations by year
- 2025: 3 · 2024: 4 · 2022: 2 · 2021: 3 · 2020: 7 · 2019: 5 · 2018: 14 · 2017: 12 · 2016: 16 · 2015: 17 · 2014: 24 · 2013: 30 · 2012: 33
- FWCI
- 23.88
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by-nc-nd
- Locations
- Diabetes Care (publishedVersion, cc-by-nc-nd)
- Europe PMC (PubMed Central) (publishedVersion, cc-by-nc-nd)
- PubMed Central (publishedVersion, cc-by-nc-nd)
- PubMed (publishedVersion)
- Research Portal (King's College London) (submittedVersion)
- Newcastle University ePrints (Newcastle Univesity)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Keywords
- insulin pump therapy, low glucose suspend, sensor-augmented pump, type 1 diabetes, nocturnal hypoglycemia, continuous glucose monitoring, hypoglycemia detection
- MeSH terms
- Adult, Circadian Rhythm, Female, Humans, Hypoglycemia, Hypoglycemic Agents, Insulin, Insulin Infusion Systems, Male, Middle Aged
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.