Sensor-Augmented Pump Therapy for A1C Reduction (STAR 3) Study
Open accessAbstract
OBJECTIVE To examine the effects of crossing over from optimized multiple daily injection (MDI) therapy to sensor-augmented pump (SAP) therapy for 6 months, and the effects of 18 months' sustained use of SAP. RESEARCH DESIGN AND METHODS The 6-month, single-crossover continuation phase of Sensor-Augmented Pump Therapy for A1C Reduction (STAR 3) provided SAP therapy to 420 subjects who completed the 1-year randomized study. The primary outcome was change in A1C in the crossover group. RESULTS A1C values were initially lower in the continuing-SAP group than in the crossover group (7.4 vs. 8.0%, P < 0.001). A1C values remained reduced in the SAP group. After 3 months on the SAP system, A1C decreased to 7.6% in the crossover group (P < 0.001); this was a significant and sustained decrease among both adults and children (P < 0.05). CONCLUSIONS Switching from optimized MDI to SAP therapy allowed for rapid and safe A1C reductions. Glycemic benefits of SAP therapy persist for at least 18 months.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
A trial comparing a sensor-augmented pump with multiple daily injection therapy showed significant improvement in HbA1C reduction with less hypoglycaemia in the sensor-augmented pump cohort.
“A trial comparing a sensor-augmented pump with multiple daily injection therapy showed significant improvement in HbA1C reduction with less hypoglycaemia in the sensor-augmented pump cohort.88,94”
From Type 1 diabetes (2014) · cites this paper as [94]
The Abstract Conclusions report greater A1C reductions with sensor-augmented pump therapy than with multiple daily injections. The Results do not support the claim of less hypoglycaemia: in the continuation phase, severe hypoglycaemia rates were not significantly different between the continuing-pump and crossover groups. Our copy does not establish a lower hypoglycaemia rate for the pump cohort versus the injection cohort. (LLM, conf=0.88)
Experimental — accuracy not yet measuredA1C fails to identify the magnitude and frequency of intra- and inter-day glucose variation.
“A1C also fails to identify the magnitude and frequency of intra- and inter-day glucose variation (34; 35).”
From Clinical Targets for Continuous Glucose Monitoring Data Interpretation: Recommendations From the International Consensus on Time in Range (2019) · cites this paper as [34]
The Results and Conclusions (page unavailable) report changes in A1C and their association with CGM sensor use. Our copy does not address whether A1C fails to identify the magnitude or frequency of within-day or between-day glucose variation; searches for that phrase and for glucose variability found no such passage. (LLM, conf=0.76)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Defining and Reporting Hypoglycemia in Diabetes
American Diabetes Association Workgroup on Hypoglycemia (2005) · Diabetes Care · doi:10.2337/diacare.28.5.1245 · Cited by 1,146 · see citing works
Effectiveness of Sensor-Augmented Insulin-Pump Therapy in Type 1 Diabetes
Bergenstal, R. M., Tamborlane, W. V., Ahmann, A., Buse, J. B., Dailey, G., Davis, S. N., Joyce, C., Peoples, T., Perkins, B. A., Welsh, J. B., et al. (2010) · New England Journal of Medicine · doi:10.1056/nejmoa1002853 · Cited by 894 · see citing works
Self-citationRaccah, D., Sulmont, V., Reznik, Y., Guerci, B., Renard, É., Hanaire, H. P., Jeandidier, N., & Nicolino, M. P. (2009) · Diabetes Care · doi:10.2337/dc09-0750 · Cited by 236 · see citing works
Slover, R. H., Welsh, J. B., Criego, A. B., WEINZIMER, S. A., Willi, S. M., Wood, M. A., & Tamborlane, W. V. (2011) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2011.00793.x · Cited by 135 · see citing works
Self-citationKowalski, A. J. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0031 · Cited by 127 · see citing works
Agrawal, P. K., Welsh, J. B., Kannard, B., Askari, S., Yang, Q., & KAUFMAN, F. R. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500514 · Cited by 86 · see citing works
Self-citationDavis, S. N., Horton, E. S., Battelino, T., Rubin, R. R., Schulman, K. A., & Tamborlane, W. V. (2010) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0145 · Cited by 40 · see citing works
Self-citationThe Use of an Energy Monitor in the Management of Diabetes: A Pilot Study
Voon, R., Celler, B. G., & Lovell, N. H. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0021 · Cited by 19 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Battelino, T., Danne, T., Bergenstal, R. M., Amiel, S. A., Beck, R., Biester, T., Bosi, E., Buckingham, B. A., Cefalu, W. T., Close, K. L., et al. (2019) · Diabetes Care · doi:10.2337/dci19-0028 · Cited by 4,032 · see citing works
Self-citationAtkinson, M. A., Eisenbarth, G. S., & Michels, A. W. (2014) · The Lancet · doi:10.1016/s0140-6736(13)60591-7 · Cited by 2,713 · see citing works
Bolinder, J., Antuna, R., Geelhoed-Duijvestijn, P., Kröger, J., & Weitgasser, R. (2016) · The Lancet · doi:10.1016/s0140-6736(16)31535-5 · Cited by 893 · see citing works
Lind, M., Polonsky, W., Hirsch, I. B., Heise, T., Bolinder, J., Dahlqvist, S., Schwarz, E., Ólafsdóttir, A. F., Frid, A., Wedel, H., et al. (2017) · JAMA · doi:10.1001/jama.2016.19976 · Cited by 705 · see citing works
Continuous glucose monitoring: A review of the technology and clinical use
Klonoff, D. C., Ahn, D., & Drincic, A. (2017) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2017.08.005 · Cited by 320 · see citing works
ISPAD Clinical Practice Consensus Guidelines 2018: diabetes technologies
Sherr, J. L., Tauschmann, M., Battelino, T., de Bock, M., Forlenza, G., Roman, R., Hood, K. K., & Maahs, D. M. (2018) · Pediatric Diabetes · doi:10.1111/pedi.12731 · Cited by 197 · see citing works
Fonseca, V. A., Grunberger, G., Anhalt, H., Bailey, T. S., Blevins, T., Garg, S. K., Handelsman, Y., Hirsch, I. B., Orzeck, E. A., Roberts, V. L., et al. (2016) · Endocrine Practice · doi:10.4158/ep161392.cs · Cited by 195 · see citing works
Self-citationPredicting Blood Glucose with an LSTM and Bi-LSTM Based Deep Neural Network
Sun, Q., Jankovic, M. V., Bally, L., & Mougiakakou, S. G. (2018) · Symposium on Neural Networks and Applications (NEUREL) · doi:10.1109/neurel.2018.8586990 · Cited by 195 · see citing works
Messer, L. H., Berget, C., Vigers, T., Pyle, L. L., Geno, C. R., Wadwa, R. P., Driscoll, K. A., & Forlenza, G. P. (2019) · Pediatric Diabetes · doi:10.1111/pedi.12971 · Cited by 161 · see citing works
Maahs, D. M., Hermann, J. M., DuBose, S. N., Miller, K. M., Heidtmann, B., DiMeglio, L. A., Rami-Merhar, B. M., Beck, R. W., Schober, E., Tamborlane, W. V., et al. (2014) · Diabetologia · doi:10.1007/s00125-014-3272-2 · Cited by 160 · see citing works
Self-citationTechnology in the management of type 1 diabetes mellitus — current status and future prospects
Tauschmann, M., & Hovorka, R. (2018) · Nature Reviews Endocrinology · doi:10.1038/s41574-018-0044-y · Cited by 161 · see citing works
Christiansen, M. P., Klaff, L. J., Brazg, R., Chang, A. R., Levy, C. J., Lam, D. W.-H., Denham, D. S., Atiee, G. J., Bode, B. W., Walters, S. J., et al. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0142 · Cited by 146 · see citing works
Abraham, M. B., Nicholas, J. A., Smith, G. J., Fairchild, J. M., King, B. R., Ambler, G. R., Cameron, F. J., Davis, E. A., Jones, T. W., & on behalf of the PLGM Study Group (2017) · Diabetes Care · doi:10.2337/dc17-1604 · Cited by 141 · see citing works
Continuous glucose monitoring: The achievement of 100 years of innovation in diabetes technology
Galindo, R. J., & Aleppo, G. (2020) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2020.108502 · Cited by 138 · see citing works
Naranjo, D. M., Tanenbaum, M. L., Iturralde, E., & Hood, K. K. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816650900 · Cited by 109 · see citing works
Toward a Run-to-Run Adaptive Artificial Pancreas: In Silico Results
Toffanin, C., VISENTIN, R., Messori, M., Palma, F. D., Magni, L., & Cobelli, C. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2017.2652062 · Cited by 107 · see citing works
Akinetic mutism: disconnection of frontal-subcortical circuits.
Mega, M. S., & COHENOUR, R. C. (1997) · PubMed · Cited by 95 · see citing works
Single- and Dual-Hormone Artificial Pancreas for Overnight Glucose Control in Type 1 Diabetes
Haidar, A., RABASA-LHORET, R. P. R., Legault, L., Lovblom, L. E., Rakheja, R., Messier, V., D'Aoust, É., Falappa, C. M., Justice, T. D., Orszag, A., et al. (2015) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2015-3003 · Cited by 89 · see citing works
Self-citationUltrafast-Acting Insulins: State of the Art
Heinemann, L., & Muchmore, D. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600402 · Cited by 90 · see citing works
Laffel, L. M. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0380 · Cited by 86 · see citing works
Laffel, L. M., Aleppo, G., Buckingham, B. A., Forlenza, G. P., Rasbach, L. E., Tsalikian, E., Weinzimer, S. A., & Harris, D. R. (2017) · Journal of the Endocrine Society · doi:10.1210/js.2017-00389 · Cited by 79 · see citing works
Uncovering undetected hypoglycemic events
Unger, J. (2012) · Diabetes Metabolic Syndrome and Obesity · doi:10.2147/dmso.s29367 · Cited by 62 · see citing works
Thabit, H., & Hovorka, R. (2015) · Expert Opinion on Drug Delivery · doi:10.1517/17425247.2016.1115013 · Cited by 58 · see citing works
Routine use of continuous glucose monitoring in 10 501 people with diabetes mellitus
Battelino, T., Liabat, S., Veeze, H. J., Castañeda, J., Arrieta, A., & Cohen, O. (2015) · Diabetic Medicine · doi:10.1111/dme.12825 · Cited by 57 · see citing works
Erie, C., Name, M. A. V., Weyman, K., WEINZIMER, S. A., Finnegan, J., Sikes, K. A., Tamborlane, W. V., & Sherr, J. L. (2017) · Pediatric Diabetes · doi:10.1111/pedi.12518 · Cited by 59 · see citing works
Self-citationSensor-Augmented Insulin Pumps and Hypoglycemia Prevention in Type 1 Diabetes
Steineck, I. I. K., Ranjan, A. G., Nørgaard, K., & Schmidt, S. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816672689 · Cited by 57 · see citing works
Levetan, C. S., & Pierce, S. M. (2012) · Endocrine Practice · doi:10.4158/ep12138.ra · Cited by 50 · see citing works
Wearable systems and mobile applications for diabetes disease management
Georga, E. I., Protopappas, V. C., Bellos, C. V., & Fotiadis, D. I. (2014) · Health and Technology · doi:10.1007/s12553-014-0082-y · Cited by 46 · see citing works
Nationwide reduction in the frequency of severe hypoglycemia by half
The Danish Society for Diabetes in Childhood and Adolescence, Fredheim, S., Johansen, A., Thorsen, S. U., Kremke, B., Nielsen, L. B., Olsen, B. S., Lyngsøe, L., Sildorf, S. M., Pipper, C. B., et al. (2014) · Acta Diabetologica · doi:10.1007/s00592-014-0697-5 · Cited by 46 · see citing works
Benhamou, P., Catargi, B., Delenne, B., Guerci, B., Hanaire, H. P., Jeandidier, N., Leroy, R., Meyer, L., Penfornis, A., Radermecker, R. P., et al. (2012) · Diabetes & Metabolism · doi:10.1016/s1262-3636(12)71538-0 · Cited by 44 · see citing works
Diabetes Technology Use in Adults with Type 1 and Type 2 Diabetes
Kravarusic, J., & Aleppo, G. (2019) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2019.10.006 · Cited by 43 · see citing works
Braune, K., Boß, K., Schmidt-Herzel, J., Gajewska, K. A., Thieffry, A., Schulze, L., Posern, B., & Raile, K. (2020) · JMIR mhealth and uhealth · doi:10.2196/24374 · Cited by 45 · see citing works
Buse, J. B., Kudva, Y. C., Battelino, T., Davis, S. N., Shin, J., & Welsh, J. B. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0294 · Cited by 41 · see citing works
Self-citationSmith, M. B., Albanese-O’Neill, A., Macieira, T. G. R., Yao, Y., Abbatematteo, J. M., Lyon, D., Wilkie, D. J., Haller, M. J., & Keenan, G. M. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0136 · Cited by 39 · see citing works
Automatic adaptation of basal therapy for Type 1 diabetic patients: A Run-to-Run approach
Toffanin, C., Messori, M., Cobelli, C., & Magni, L. (2016) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2016.09.002 · Cited by 36 · see citing works
New technologies for diabetes: a review of the present and the future
Ramchandani, N., & Heptulla, R. A. (2012) · International Journal of Pediatric Endocrinology · doi:10.1186/1687-9856-2012-28 · Cited by 34 · see citing works
Closed-loop for type 1 diabetes – an introduction and appraisal for the generalist
Bally, L., Thabit, H., & Hovorka, R. (2017) · BMC Medicine · doi:10.1186/s12916-017-0794-8 · Cited by 35 · see citing works
Pleus, S., Schmid, C., Link, M., Zschornack, E., Klötzer, H.-M., Haug, C., & Freckmann, G. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700405 · Cited by 34 · see citing works
Gómez, A. M., Carrillo, L. F. M., Múñoz, Ó., Rondón, M., Correa, C. M. A., Garzón, E. M., Diaz, M. C. C., & Henao, D. C. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0332 · Cited by 32 · see citing works
Lombardo, F., Passanisi, S., Caminiti, L., Barbalace, A., Marino, A., Iannelli, M., Messina, M. F., Pajno, G. B., & Salzano, G. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0236 · Cited by 32 · see citing works
Wood, M. A., Shulman, D. I., Forlenza, G. P., Bode, B. W., Pinhas‐Hamiel, O., Buckingham, B. A., Kaiserman, K. B., Liljenquist, D. R., Bailey, T. S., Shin, J., et al. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0209 · Cited by 30 · see citing works
Self-citationFrontino, G., Bonfanti, R., Scaramuzza, A. E., Rabbone, I., Meschi, F., Rigamonti, A., Battaglino, R., Favalli, V., Bonura, C., Sicignano, S., et al. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0044 · Cited by 29 · see citing works
Picard, S., Hanaire, H. P., Baillot-Rudoni, S., Bonnemaison, É., Not, D., Reznik, Y., & Guerci, B. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0240 · Cited by 29 · see citing works
A systematic review on clinical implication of continuous glucose monitoring in diabetes management
Gillani, S. W., Azhar, A., Mohiuddin, G., & Majeed, R. A. (2020) · Journal of Pharmacy And Bioallied Sciences · doi:10.4103/jpbs.jpbs_7_20 · Cited by 29 · see citing works
Update on Clinical Utility of Continuous Glucose Monitoring in Type 1 Diabetes
Haviland, N., Walsh, J. S. A., Roberts, R., & Bailey, T. S. (2016) · Current Diabetes Reports · doi:10.1007/s11892-016-0808-5 · Cited by 27 · see citing works
Manning, M. L., Singh, H., STONER, K. C., & HABIF, S. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819897976 · Cited by 24 · see citing works
Diabetes mellitus type 1 in childhood
Peterkova, V. A., Shestakova, M. V., Bezlepkina, O. B., Лаптев, Д. Н., Кураева, Т. Л., Mayorov, A. Y., Титович, Е. В., Emel'yanov, A. O., Svetlova, G. N., Eremina, I. A., et al. (2020) · Diabetes Mellitus · doi:10.14341/dm12504 · Cited by 22 · see citing works
Choudhary, P., de Portu, S., Arrieta, A., Castañeda, J., & Campbell, F. (2019) · Diabetic Medicine · doi:10.1111/dme.14043 · Cited by 21 · see citing works
Sensor-Augmented Pump Therapy at 36 Months
Schmidt, S., & Nørgaard, K. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0148 · Cited by 19 · see citing works
Patient Perception and Satisfaction With Insulin Pump System: Pilot User Experience Survey
Grando, A., Bayuk, M., Karway, G. K., Corrette, K., Groat, D., Cook, C. B., & Thompson, B. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819843146 · Cited by 20 · see citing works
Automated Insulin Delivery: Easy Enough to Use in Primary Care?
HEILE, M. K., Hollstegge, B., Broxterman, L., Cai, A., & Close, K. L. (2020) · Clinical Diabetes · doi:10.2337/cd20-0050 · Cited by 18 · see citing works
William, J., McCluskey, J. T., & Gleeson, N. (2022) · Endocrinology Diabetes & Metabolism · doi:10.1002/edm2.324 · Cited by 18 · see citing works
Gómez, A. M., Grizales, A. M., Veloza, A., Sánchez, A. M., Múñoz, Ó., & Rondón, M. (2013) · Avances en Diabetología · doi:10.1016/j.avdiab.2013.04.001 · Cited by 17 · see citing works
Deeb, A., Yousef, H., Qahtani, N. A., Artan, I., Suliman, S., Tomy, M., Abdulrahman, L., Suwaidi, H. A., Attia, S., & Nagelkerke, N. J. D. (2019) · Journal of Diabetes & Metabolic Disorders · doi:10.1007/s40200-018-0351-9 · Cited by 17 · see citing works
Impact of new technologies on diabetes care
Giani, E. (2015) · World Journal of Diabetes · doi:10.4239/wjd.v6.i8.999 · Cited by 17 · see citing works
Continuous Glucose Monitoring Frequency and Glycemic Control in People With Type 2 Diabetes
Hirsch, I. B., Garg, S. K., Repetto, E., Snell‐Bergeon, J. K., Ulmer, B. J., Perkins, C., & Bergenstal, R. M. (2025) · JAMA Network Open · doi:10.1001/jamanetworkopen.2025.39278 · Cited by 16 · see citing works
Self-citationNauck, M. A., Lindmeyer, A. M., Mathieu, C., & Meier, J. J. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819882752 · Cited by 16 · see citing works
Diabetes Technology Experiences Among Latinx and Non-Latinx Youth with Type 1 Diabetes
Tsai, D., Garcia, J. F., Fogel, J. L., Wee, C. P., Reid, M. W., & Raymond, J. K. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211029260 · Cited by 16 · see citing works
Barnard, K. D., Bromba, M., de Lange-Ranzijn, M., Halbron, M., Levy, B. L., Lévy, M., Lippmann-Grob, B., Walshe, K., & Ziegler, R. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814567893 · Cited by 16 · see citing works
Tanenberg, R. J., & Welsh, J. B. (2014) · Endocrine Practice · doi:10.4158/ep14187.or · Cited by 14 · see citing works
Self-citation“Brittleness” in Diabetes: Easier Spoken than Broken
Voulgari, C., Pagoni, S., Paximadas, S., & Vinik, A. I. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0058 · Cited by 13 · see citing works
Smith, M. B., Albanese-O’Neill, A., Yao, Y., Wilkie, D. J., Haller, M. J., & Keenan, G. M. (2020) · JMIR Diabetes · doi:10.2196/15410 · Cited by 13 · see citing works
Treatment of patients with type 1 diabetes – Insulin pumps or multiple injections?
Krzymień, J., Rachuta, M., Kozlowska, I., Ładyżyński, P., & Foltyński, P. (2015) · Journal of Applied Biomedicine · doi:10.1016/j.bbe.2015.10.002 · Cited by 12 · see citing works
van Beers, C. A. J., Kleijer, S. J., Serné, E. H., Geelhoed-Duijvestijn, P. H. L. M., Snoek, F. J., Kramer, M. H. H., & Diamant, M. (2015) · BMC Endocrine Disorders · doi:10.1186/s12902-015-0040-3 · Cited by 12 · see citing works
Physiological principles of Starling-like control of rotary ventricular assist devices
Stephens, A. F., Gregory, S. D., Burrell, A. J. C., Marasco, S., Stub, D. A., & Salamonsen, R. F. (2020) · Expert Review of Medical Devices · doi:10.1080/17434440.2020.1841631 · Cited by 12 · see citing works
Majithia, A. R., Wiltschko, A. B., Zheng, H., Walford, G. A., & Nathan, D. M. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817725756 · Cited by 11 · see citing works
Martín-Vaquero, P., Martínez-Brocca, M. A., & García-López, J. M. (2014) · Endocrinologia y Nutricion · doi:10.1016/j.endonu.2014.08.003 · Cited by 8 · see citing works
Bénéfices de l’insulinothérapie par pompe chez les enfants diabétiques de type 1
Hasselmann, C., Bonnemaison, É., Faure, N., Mercat, I., Pépin-Donat, M., Magontier, N., Chantepie, A., & Labarthe, F. (2012) · Archives de Pédiatrie · doi:10.1016/j.arcped.2012.03.051 · Cited by 11 · see citing works
Tweden, K. S., Deiss, D., Rastogi, R., Addaguduru, S., & KAUFMAN, F. R. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0342 · Cited by 10 · see citing works
Can Improved Glycemic Control Slow Renal Function Decline at All Stages of Diabetic Nephropathy?
Goel, G., & Perkins, B. A. (2012) · Seminars in Nephrology · doi:10.1016/j.semnephrol.2012.07.004 · Cited by 10 · see citing works
Self-citationPrice, D. A., Graham, C., Parkin, C. G., & Peyser, T. A. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815607863 · Cited by 9 · see citing works
A Data-Driven Behavior Modeling and Analysis Framework for Diabetic Patients on Insulin Pumps
Chen, S., Feng, L., Rickels, M. R., Peleckis, A. J., Sokolsky, O. V., & Lee, I. (2015) · IEEE International Conference on Healthcare Informatics (ICHI) · doi:10.1109/ichi.2015.33 · Cited by 8 · see citing works
Martín-Vaquero, P., Martínez-Brocca, M. A., & García-López, J. M. (2014) · Avances en Diabetología · doi:10.1016/j.avdiab.2014.07.002 · Cited by 4 · see citing works
Sperling, M. A., Wolfsdorf, J. I., Menon, R. K., Tamborlane, W. V., Maahs, D. M., Battelino, T., & Phillip, M. (2021) · Elsevier eBooks · doi:10.1016/b978-0-323-62520-3.00021-x · Cited by 9 · see citing works
Self-citationMolfetta, S. D., Rossi, A., Gesuita, R., Faragalli, A., Cutruzzolà, A., Irace, C., Minuto, N., Pitocco, D., Cardella, F., Arnaldi, C., et al. (2024) · Acta Diabetologica · doi:10.1007/s00592-024-02381-3 · Cited by 6 · see citing works
Glycaemic control and novel technology management strategies in pregestational diabetes mellitus
Newman, C., Ero, A., & Dunne, F. (2023) · Frontiers in Endocrinology · doi:10.3389/fendo.2022.1109825 · Cited by 6 · see citing works
Perkins, B. A., Halpern, E. M., Orszag, A., Weisman, A., Houlden, R. L., Bergenstal, R. M., & Joyce, C. (2014) · Canadian Journal of Diabetes · doi:10.1016/j.jcjd.2014.03.003 · Cited by 5 · see citing works
Self-citationHanaire, H. P., Atlan, C., Benhamou, P., Bismuth, É., Bonnemaison, É., Borot, S., Catargi, B., Charpentier, G., Farret, A., Filhol, N., et al. (2017) · Médecine des Maladies Métaboliques · doi:10.1016/s1957-2557(18)30030-0 · Cited by 4 · see citing works
Omeara, M., Acuña, J. C. M., & Uribe, A. (2023) · JMIR Diabetes · doi:10.2196/46880 · Cited by 4 · see citing works
Guideline Development for Medical Device Technology: Issues for Consideration
Battelino, T., Brosius, F. C., Ceriello, A., Cosentino, F., Green, J., Kellerer, M., Koob, S., Kosiborod, M., Lalic, N. M., Marx, N., et al. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221093355 · Cited by 4 · see citing works
Лаптев, Д. Н., Князева, Т. Т., & Bezlepkina, O. B. (2024) · Diabetes Mellitus · doi:10.14341/dm13129 · Cited by 3 · see citing works
Miller, V. A., Xiao, R., & Willi, S. M. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2020.0668 · Cited by 3 · see citing works
Self-citationContinuous glucose monitoring systems in well-controlled children with type 1 diabetes mellitus
Kowalczyk, E., Adamczyk, M., Pietrzyk, J., Jastrzębska, B., & Szypowska, A. (2021) · Pediatric Endocrinology Diabetes and Metabolism · doi:10.5114/pedm.2021.107717 · Cited by 3 · see citing works
Stadler, M., Zlamal-Fortunat, S., Schütz-Fuhrmann, I., Rami-Merhar, B. M., Schober, E., Kautzky‐Willer, A., Weitgasser, R., Prager, R., & Bischof, M. G. (2012) · Wiener klinische Wochenschrift · doi:10.1007/s00508-012-0287-x · Cited by 3 · see citing works
HTA report: new devices for the management of glycaemia in young diabetics
Scalzo, A. L., Lenzi, L., Chiarolla, E., Maltoni, S., Negro, A., Ballini, L., Casino, D., Ghedi, A., Pace, N. L., Scondotto, S., et al. (2012) · Cited by 2 · see citing works
Tecce, N., Menafra, D., Proganò, M., Tecce, M. F., Pivonello, R., & Colao, A. (2024) · Healthcare · doi:10.3390/healthcare12181823 · Cited by 2 · see citing works
Insulinpumpentherapie bei Kindern, Jugendlichen und Erwachsenen (Update 2019)
Schütz-Fuhrmann, I., Stadler, M., Zlamal-Fortunat, S., Rami-Merhar, B. M., Fröhlich-Reiterer, E. E., Hofer, S. E., Mader, J. K., Resl, M., Bischof, M. G., Kautzky‐Willer, A., et al. (2019) · Wiener klinische Wochenschrift · doi:10.1007/s00508-019-1485-6 · Cited by 2 · see citing works
Alissa, N., Alhumaidi, S., Alzaid, S., Aldibasi, O. S., Alfaraidi, H., & Almutair, A. (2024) · Pediatric Diabetes · doi:10.1155/2024/5529674 · Cited by 1 · see citing works
La mesure continue du glucose : présent et avenir Mesure continue du glucose au long cours en 2012
Sola‐Gazagnes, A., & Riveline, J. (2012) · Médecine des Maladies Métaboliques · doi:10.1016/s1957-2557(12)70467-4 · Cited by 1 · see citing works
The Application of Digital Health in Diabetes Management
Ye, J. (2023) · Highlights in Science Engineering and Technology · doi:10.54097/hset.v66i.11711 · Cited by 1 · see citing works
Implantable pumps and artificial and bio‐artificial pancreas system
Tamborlane, W. V., & Palau‐Collazo, M. M. (2015) · doi:10.1002/9781118387658.ch51
Self-citationLithovius, R. (2015) · Työväentutkimus Vuosikirja
A perspective of sensor-augmented insulin pump therapy in the treatment of type 1 diabetes.
Rayfield, E. (2015) · Endocrine Practice · doi:10.4158/ep14443.co
Kontinuierliche Glukosemessung im Alltag
Jecht, M. (2016) · Der Diabetologe · doi:10.1007/s11428-015-0056-3
Insulinpumpentherapie bei Kindern, Jugendlichen und Erwachsenen
Stadler, M., Zlamal-Fortunat, S., Schütz-Fuhrmann, I., Rami-Merhar, B. M., Fröhlich-Reiterer, E. E., Hofer, S. E., Mader, J. K., Resl, M., Kautzky‐Willer, A., Weitgasser, R., et al. (2016) · Wiener klinische Wochenschrift · doi:10.1007/s00508-016-0966-0
Nationwide reduction in the frequency of severe hypoglycemia by half
Italia, S.-V. (2015)
Siegmund, T. (2012) · Der Diabetologe · doi:10.1007/s11428-011-0854-1
Kontinuierliche Glukosemessung (CGM) bei Erwachsenen
Gehr, B. (2017) · Der Diabetologe · doi:10.1007/s11428-017-0203-0
Diabetes mellitus Typ 1 – neue Entwicklungen in der intensivierten Insulintherapie
Gerber, P. A., Zimmermann, D., & Spinas, G. A. (2013) · Swiss Medical Forum ‒ Schweizerisches Medizin-Forum · doi:10.4414/smf.2013.01525
Avancées technologiques et traitement du diabète
Roy‐Fleming, A., & Taleb, N. (2018) · Nutrition, science en évolution · doi:10.7202/1048943ar
- Authors
- Bergenstal, R. M., Tamborlane, W. V., Ahmann, A., Buse, J. B., Dailey, G., Davis, S. N., Joyce, C., Perkins, B. A., Welsh, J. B., Willi, S. M., Wood, M. A., & for the STAR 3 Study Group
- Affiliations
- Park Nicollet Clinic (US)
- Yale University (US)
- Oregon Health & Science University (US)
- University of North Carolina at Chapel Hill (US)
- Scripps Clinic (US)
- University of Maryland, Baltimore (US)
- Memorial University of Newfoundland (CA)
- University of Toronto (CA)
- Medtronic (United States) (US)
- Children's Hospital of Philadelphia (US)
- Spectrum Health (US)
- Corresponding author
- Bergenstal, Richard M.
- ORCID
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 2011-10-15
- Volume
- 34
- Issue
- 11
- Pages
- 2403-2405
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- PMC
- PMC3198292
- OpenAlex
- W2105443410
- Cited by
- 118
- References
- 8
- Citations by year
- 2026: 1 · 2025: 3 · 2024: 7 · 2023: 4 · 2022: 5 · 2021: 4 · 2020: 12 · 2019: 12 · 2018: 6 · 2017: 12 · 2016: 12 · 2015: 12 · 2014: 8 · 2013: 4 · 2012: 15
- FWCI
- 7.08
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- Locations
- Diabetes Care (publishedVersion, cc-by-nc-nd)
- PubMed Central (publishedVersion, cc-by-nc-nd)
- PubMed (publishedVersion)
- Funders
- American Diabetes Association
- Bristol-Myers Squibb
- Eli Lilly and Company
- Pfizer
- AstraZeneca
- GlaxoSmithKline
- Sanofi
- Children's Hospital of Philadelphia
- Intarcia Therapeutics
- Insulet Corporation
- MannKind Corporation
- Halozyme
- Dexcom
- ResMed
- UnitedHealth Group
- Bayer HealthCare
- Novo Nordisk
- Memorial University of Newfoundland
- Abbott Diabetes Care
- School of Medicine, University of North Carolina at Chapel Hill
- Daiichi Sankyo Europe
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Keywords
- sensor-augmented pump therapy, continuous glucose monitoring, insulin pump, multiple daily insulin injections, glycemic control, HbA1c reduction, type 1 diabetes, treatment switching
- MeSH terms
- Adolescent, Adult, Aged, Blood Glucose, Child, Diabetes Mellitus, Type 1, Female, Glycated Hemoglobin, Humans, Hyperglycemia, Hypoglycemic Agents, Injections, Subcutaneous, Insulin, Insulin Infusion Systems, Male, Middle Aged, Treatment Outcome, Cross-Over Studies, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review