Physiological Differences Between Interstitial Glucose and Blood Glucose Measured in Human Subjects
Open accessAbstract
OBJECTIVE—This study investigated whether glucose readings from a sensor sampling in interstitial fluid differ substantially from blood glucose (BG) values measured at the same time. RESEARCH DESIGN AND METHODS—We have evaluated the relationship between BG and glucose extracted from interstitial fluid using the GlucoWatch (Cygnus, Redwood City, CA) biographer, a device that collects glucose from subcutaneous interstitial space through intact skin by application of a low electric current. We evaluated the relative change in the interstitial glucose (IG) signal (IGS) as measured by the biographer versus BG using a normalized two-point sensitivity index (NSI). RESULTS—The results show that biographer measures of IG differ in time and magnitude from the corresponding BG values. In particular, the biographer values were shifted in time due to instrumental and physiological lag. Results show an average total lag of 17.2 ± 7.2 min for all subjects evaluated. The instrumental lag was 13.5 min, suggesting that physiological lag is ∼5 min. In addition, when glucose was increasing, the change in IGS was less than that in BG, while when BG was decreasing, the change in IGS was greater than that in BG. CONCLUSIONS—Similar results have been reported for other measures of IG, suggesting that differences reflect physiological variation in glucose uptake, utilization, and elimination in blood and interstitial space. This further evidence of the difference between IG and BG should be considered when interpreting glucose measurements from devices that sample interstitial fluid.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Electrochemical Glucose Sensors and Their Applications in Diabetes Management
Heller, A., & Feldman, B. (2008) · Chemical Reviews · doi:10.1021/cr068069y · Cited by 1,707
Risk of hypoglycaemia in types 1 and 2 diabetes: Effects of treatment modalities and their duration
UK Hypoglycaemia Study Group (2007) · Diabetologia · doi:10.1007/s00125-007-0599-y · Cited by 825
Biosensors for real-time in vivo measurements
Wilson, G. S., & Gifford, R. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.12.003 · Cited by 660
Goldstein, D. E., Little, R. R., Lorenz, R. A., Malone, J. I., Nathan, D., Peterson, C. M., & Sacks, D. B. (2004) · Diabetes Care · doi:10.2337/diacare.27.7.1761 · Cited by 625
Artificial Pancreas: Past, Present, Future
Cobelli, C., Renard, E., & Kovatchev, B. (2011) · Diabetes · doi:10.2337/db11-0654 · Cited by 600
Metadata
- Authors
- Kulcu, E., Tamada, J. A., Reach, G., Potts, R. O., & Lesho, M. J.
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 2003-08-01
- Volume
- 26
- Issue
- 8
- Pages
- 2405-2409
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- OpenAlex
- W2139536521
- Cited by
- 251
- References
- 0
- FWCI
- 6.35
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Physiological Differences Between Interstitial Glucose and Blood Glucose Measured in Human Subjects. https://app.cochat.ai/papers/10.2337/diacare.26.8.2405
- The paper
- Kulcu, E., Tamada, J. A., Reach, G., Potts, R. O., & Lesho, M. J. (2003). Physiological Differences Between Interstitial Glucose and Blood Glucose Measured in Human Subjects. Diabetes Care. https://doi.org/10.2337/diacare.26.8.2405
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.