Physiological Differences Between Interstitial Glucose and Blood Glucose Measured in Human Subjects
Open accessThe DOI matches its Crossref record; 6 of the 6 references it cites were found in a registry or an index.
Abstract
OBJECTIVE—This study investigated whether glucose readings from a sensor sampling in interstitial fluid differ substantially from blood glucose (BG) values measured at the same time. RESEARCH DESIGN AND METHODS—We have evaluated the relationship between BG and glucose extracted from interstitial fluid using the GlucoWatch (Cygnus, Redwood City, CA) biographer, a device that collects glucose from subcutaneous interstitial space through intact skin by application of a low electric current. We evaluated the relative change in the interstitial glucose (IG) signal (IGS) as measured by the biographer versus BG using a normalized two-point sensitivity index (NSI). RESULTS—The results show that biographer measures of IG differ in time and magnitude from the corresponding BG values. In particular, the biographer values were shifted in time due to instrumental and physiological lag. Results show an average total lag of 17.2 ± 7.2 min for all subjects evaluated. The instrumental lag was 13.5 min, suggesting that physiological lag is ∼5 min. In addition, when glucose was increasing, the change in IGS was less than that in BG, while when BG was decreasing, the change in IGS was greater than that in BG. CONCLUSIONS—Similar results have been reported for other measures of IG, suggesting that differences reflect physiological variation in glucose uptake, utilization, and elimination in blood and interstitial space. This further evidence of the difference between IG and BG should be considered when interpreting glucose measurements from devices that sample interstitial fluid.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Rebrin, K., Steil, G. M., van Antwerp, W. P., & Mastrototaro, J. J. (1999) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.1999.277.3.e561 · Cited by 366 · cited 1×
“7 min (2), which is in agreement with previously published estimates of time lag between BG and IG (8).”
Noninvasive Glucose Monitoring: Comprehensive Clinical Results
Tamada, J. A., Garg, S., Jovanovic, L., Pitzer, K. R., Fermi, S., Potts, R. O., & and the Cygnus Research Team (1999) · JAMA · doi:10.1001/jama.282.19.1839 · Cited by 282 · cited 1×
“IGS(T Ti Tp 60) IGS(T Ti Tp) IGS(T Ti Tp 60) IGS(T Ti Tp) (4)”
Self-citationGarg, S. K., Potts, R. O., Ackerman, N. R., Fermi, S. J., Tamada, J. A., & Chase, H. P. (1999) · Diabetes Care · doi:10.2337/diacare.22.10.1708 · Cited by 201
Self-citationAussedat, B., Dupire-Angel, M., Gifford, R., Klein, J. C., Wilson, G. S., & Reach, G. (2000) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.2000.278.4.e716 · Cited by 186 · cited 3×
“CONCLUSIONS — The results from a variety of studies show that continuous measurements of IG differ in time and magnitude from corresponding BG values (1,2).”
Self-citationThe GlucoWatch®biographer: a frequent, automatic and noninvasive glucose monitor
Tierney, M. J., Tamada, J. A., Potts, R. O., Eastman, R. C., Pitzer, K., Ackerman, N. R., & Fermi, S. J. (2000) · Annals of Medicine · doi:10.3109/07853890009002034 · Cited by 154 · cited 1×
“RESEARCH DESIGN AND METHODS — Data from existing clinical studies were used for this analysis (5).”
MethodSelf-citationMoberg, E., Toft, E., Amer, P., & Bolinder, J. (1997) · Diabetologia · doi:10.1007/s001250050827 · Cited by 80
Cited by
Electrochemical Glucose Sensors and Their Applications in Diabetes Management
Heller, A., & Feldman, B. (2008) · Chemical Reviews · doi:10.1021/cr068069y · Cited by 1,707
Risk of hypoglycaemia in types 1 and 2 diabetes: Effects of treatment modalities and their duration
UK Hypoglycaemia Study Group (2007) · Diabetologia · doi:10.1007/s00125-007-0599-y · Cited by 825
“The relationship between blood and interstitial glucose concentrations is variable, according to whether they are falling or rising and circulating insulin concentrations may also be important [23].”
BackgroundKlonoff, D. C. (2005) · Diabetes Care · doi:10.2337/diacare.28.5.1231 · Cited by 724
Biosensors for real-time in vivo measurements
Wilson, G. S., & Gifford, R. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.12.003 · Cited by 660
Goldstein, D. E., Little, R. R., Lorenz, R. A., Malone, J. I., Nathan, D., Peterson, C. M., & Sacks, D. B. (2004) · Diabetes Care · doi:10.2337/diacare.27.7.1761 · Cited by 625
Artificial Pancreas: Past, Present, Future
Cobelli, C., Renard, E., & Kovatchev, B. (2011) · Diabetes · doi:10.2337/db11-0654 · Cited by 600
Diabetes: Models, Signals, and Control
Cobelli, C., Dalla Man, C., Sparacino, G., Magni, L., De Nicolao, G., & Kovatchev, B. P. (2009) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2009.2036073 · Cited by 511
Chang, H., Zheng, M., Yu, X., Than, A., Seeni, R. Z., Kang, R., Tian, J., Khanh, D. P., Liu, L., Chen, P., et al. (2017) · Advanced Materials · doi:10.1002/adma.201702243 · Cited by 462
Sampling interstitial fluid from human skin using a microneedle patch
Samant, P. P., Niedzwiecki, M. M., Raviele, N., Tran, V., Mena-Lapaix, J., Walker, D. I., Felner, E. I., Jones, D. P., Miller, G. W., & Prausnitz, M. R. (2020) · Science Translational Medicine · doi:10.1126/scitranslmed.aaw0285 · Cited by 379
Lee, I., Probst, D., Klonoff, D., & Sode, K. (2021) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2021.113054 · Cited by 288
Technologies for Continuous Glucose Monitoring: Current Problems and Future Promises
Vaddiraju, S., Burgess, D. J., Tomazos, I., Jain, F. C., & Papadimitrakopoulos, F. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400632 · Cited by 285
A Wireless-Implantable Microsystem for Continuous Blood Glucose Monitoring
Ahmadi, M. M., & Jullien, G. A. (2009) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2009.2016844 · Cited by 266
A Hydrogel Microneedle Patch for Point‐of‐Care Testing Based on Skin Interstitial Fluid
He, R., Niu, Y., Li, Z., Li, A., Yang, H., Xu, F., & Li, F. (2020) · Advanced Healthcare Materials · doi:10.1002/adhm.201901201 · Cited by 257
Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors
Kovatchev, B., Anderson, S., Heinemann, L., & Clarke, W. (2008) · Diabetes Care · doi:10.2337/dc07-2401 · Cited by 236
Assessing Sensor Accuracy for Non-Adjunct Use of Continuous Glucose Monitoring
Kovatchev, B. P., Patek, S. D., Ortiz, E. A., & Breton, M. D. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0272 · Cited by 234
Continuous Glucose Monitoring: Real-Time Algorithms for Calibration, Filtering, and Alarms
Bequette, B. W. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400222 · Cited by 192
Mazze, R. S., Strock, E., Wesley, D., Borgman, S., Morgan, B., Bergenstal, R., & Cuddihy, R. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2007.0293 · Cited by 182
Design, Development, and Evaluation of a Novel Microneedle Array-based Continuous Glucose Monitor
Jina, A., Tierney, M. J., Tamada, J. A., McGill, S., Desai, S., Chua, B., Chang, A., & Christiansen, M. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814526191 · Cited by 176
“(which have high blood flow due to their role in thermal regulation) are in close communication with the sampling site.(15) Devices utilizing minimally invasive transdermal interstitial fluid sampling to date have demonstrated limited operating lifetime.”
BackgroundSelf-citationA practical guide to the development of microneedle systems – In clinical trials or on the market
Lee, K. J., Jeong, S. S., Roh, D. H., Kim, D. Y., Choi, H.-K., & Lee, E. H. (2020) · International Journal of Pharmaceutics · doi:10.1016/j.ijpharm.2019.118778 · Cited by 173
Analysis, Modeling, and Simulation of the Accuracy of Continuous Glucose Sensors
Breton, M., & Kovatchev, B. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200517 · Cited by 173
Tear Glucose Dynamics in Diabetes Mellitus
Lane, J. D., Krumholz, D. M., Sack, R. A., & Morris, C. (2006) · Current Eye Research · doi:10.1080/02713680600976552 · Cited by 161
Time Delay of CGM Sensors: Relevance, Causes, and Countermeasures
Schmelzeisen-Redeker, G., Schoemaker, M., Kirchsteiger, H., Freckmann, G., Heinemann, L., & del Re, L. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815590154 · Cited by 161
Choudhary, P., Ramasamy, S., Green, L., Gallen, G., Pender, S., Brackenridge, A., Amiel, S. A., & Pickup, J. C. (2013) · Diabetes Care · doi:10.2337/dc13-0939 · Cited by 160
Improving the Clinical Value and Utility of CGM Systems: Issues and Recommendations
Petrie, J. R., Peters, A. L., Bergenstal, R. M., Holl, R. W., Fleming, G. A., & Heinemann, L. (2017) · Diabetes Care · doi:10.2337/dci17-0043 · Cited by 135
“lower when glucose was increasing and ≅20% higher when glucose was decreasing [3a] (40).”
BackgroundWearable Skin Sensors and Their Challenges: A Review of Transdermal, Optical, and Mechanical Sensors
Ahmad Tarar, A., Mohammad, U., & K. Srivastava, S. (2020) · Biosensors · doi:10.3390/bios10060056 · Cited by 119
A Century of Diabetes Technology: Signals, Models, and Artificial Pancreas Control
Kovatchev, B. (2019) · Trends in Endocrinology & Metabolism · doi:10.1016/j.tem.2019.04.008 · Cited by 80
Modeling Glucose Transport From Systemic Circulation to Sweat
La Count, T. D., Jajack, A., Heikenfeld, J., & Kasting, G. B. (2019) · Journal of Pharmaceutical Sciences · doi:10.1016/j.xphs.2018.09.026 · Cited by 60
Robust Fault Detection System for Insulin Pump Therapy Using Continuous Glucose Monitoring
Herrero, P., Calm, R., Vehí, J., Armengol, J., Georgiou, P., Oliver, N., & Tomazou, C. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600518 · Cited by 52
A Century-long Journey From the Discovery of Insulin to the Implantation of Stem Cell–derived Islets
Ramzy, A., Belmonte, P. J., Braam, M. J. S., Ida, S., Wilts, E. M., Levings, M. K., Rezania, A., & Kieffer, T. J. (2023) · Endocrine Reviews · doi:10.1210/endrev/bnac021 · Cited by 34
Beyond A1C: exploring continuous glucose monitoring metrics in managing diabetes
Friedman, J. G., Coyne, K., Aleppo, G., & Szmuilowicz, E. D. (2023) · Endocrine Connections · doi:10.1530/ec-23-0085 · Cited by 22
Metadata
- Authors
- Kulcu, E., Tamada, J. A., Reach, G., Potts, R. O., & Lesho, M. J.
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 2003-08-01
- Volume
- 26
- Issue
- 8
- Pages
- 2405-2409
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- OpenAlex
- W2139536521
- Cited by
- 251
- References
- 6
- FWCI
- 6.35
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.