Validity and Between-Day Reliability of the Cervical Range of Motion (CROM) Device
Abstract
Study Design: Clinical measurement, validity and intrarater reliability study. Objectives: (1) To confirm the validity and assess between-day test-retest reliability of cervical spine motion measurements made with the cervical range of motion (CROM) device in flexion, extension, bilateral rotation, and bilateral side flexion; (2) to provide meaningful information to clinicians about the standard error of measurement and the minimal detectable change for the CROM device. Background: Range of motion is a common outcome measure used in the assessment of the cervical spine. The CROM device is one of the tools used to measure cervical range of motion in the clinical setting. However, its psychometric properties are not well established, especially for measurements taken on separate days. Methods: Quasi-experimental design with 1 group comparison. Twenty healthy adults (9 men and 11 women) participated in this study. Cervical range of motion was simultaneously recorded with the CROM device and the Fastrak motion analysis system for all 6 cervical movements mentioned above. The CROM device was placed on the participant's head consistent with standard clinical procedures. Two Fastrak sensors were positioned with 1 on the forehead and 1 over the spinous process of the T6 vertebra. Test-retest reliability of measurements made with the CROM device was assessed, as well as its standard error of measurement and minimal detectable change, with measures taken on 2 separate days spaced 48 hours apart. Results: Values obtained by the 2 measuring devices yielded Pearson correlation coefficients ranging between 0.93 and 0.98. Test-retest reliability of measurements of cervical range of motion using the CROM was found to be good, with ICCs ranging between 0.89 and 0.98. The standard errors of measurement across the 6 movements ranged from 1.6° to 2.8° and the minimal detectable changes across the 6 movements ranged from 3.6° to 6.5°. Conclusions: The measurements made with the CROM were shown to be reliable in all movement directions. J Orthop Sports Phys Ther 2010;40(5):318–323.doi:10.2519/jospt.2010.3180
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Foundations of Clinical Research: Applications to Practice,
(2002) · Survey of Ophthalmology · doi:10.1016/s0039-6257(02)00362-4 · Cited by 9,489
Spitzer, W. O., Skovron, M. L., Salmi, L., Cassidy, J. D., DURANCEAU, J., Suissa, S., & Zeiss, E. E. (1995) · PubMed · Cited by 1,853
Foundations of clinical research : applications to practice, 2nd ed.
Portney, L. G. (2000) · Cited by 932
de Vet, H. C., Terwee, C. B., Ostelo, R. W., Beckerman, H., Knol, D. L., & Bouter, L. M. (2006) · Health and Quality of Life Outcomes · doi:10.1186/1477-7525-4-54 · Cited by 829
Childs, J. D., Cleland, J. A., Elliott, J. M., Teyhen, D. S., Wainner, R. S., Whitman, J. M., Sopky, B. J., Godges, J. J., Flynn, T. W., Delitto, A., et al. (2008) · Journal of Orthopaedic & Sports Physical Therapy · doi:10.2519/jospt.2008.0303 · Cited by 684
Côté, P., Cassidy, D. J., Carroll, L. J., & Kristman, V. (2004) · Pain · doi:10.1016/j.pain.2004.09.004 · Cited by 624
Wainner, R. S., Fritz, J. M., Irrgang, J. J., Boninger, M. L., Delitto, A., & Allison, S. (2003) · Spine · doi:10.1097/00007632-200301010-00014 · Cited by 598
Reliability of Measurements of Cervical Spine Range of Motion—Comparison of Three Methods
Youdas, J. W., Carey, J. R., & Garrett, T. R. (1991) · Physical Therapy · doi:10.1093/ptj/71.2.98 · Cited by 514
Physician Views About Treating Low Back Pain: The Results of a National Survey
Cherkin, D. C., Deyo, R. A., Wheeler, K., & Ciol, M. A. (1995) · Spine · doi:10.1097/00007632-199501000-00001 · Cited by 513
Normal Range of Motion of the Cervical Spine: An Initial Goniometric Study
Youdas, J. W., Garrett, T. R., Suman, V. J., Bogard, C. L., Hallman, H. O., & Carey, J. R. (1992) · Physical Therapy · doi:10.1093/ptj/72.11.770 · Cited by 404
Physical and psychological factors predict outcome following whiplash injury
Sterling, M., Jull, G., Vicenzino, B., Kenardy, J., & Darnell, R. (2005) · Pain · doi:10.1016/j.pain.2004.12.005 · Cited by 389
Miller, D., & Hardwick, J. E. (1973) · Cited by 384
Fletcher, J. P., & Bandy, W. D. (2008) · Journal of Orthopaedic & Sports Physical Therapy · doi:10.2519/jospt.2008.2680 · Cited by 304
Stratford, P. W. (2004) · Physiotherapy Canada · doi:10.2310/6640.2004.15377 · Cited by 302
de Koning, C. H. P., van den Heuvel, S. P., Staal, J. B., Smits-Engelsman, B. C. M., & Hendriks, E. J. M. (2008) · European Spine Journal · doi:10.1007/s00586-008-0656-3 · Cited by 179
Hole, D. E., Cook, J. M., & Bolton, J. E. (1995) · Manual Therapy · doi:10.1054/math.1995.0248 · Cited by 171
Meta-Analysis of Normative Cervical Motion
Chen, J., Solinger, A. B., Poncet, J. F., & Lantz, C. A. (1999) · Spine · doi:10.1097/00007632-199908010-00011 · Cited by 168
González-Iglesias, J., Fernández‐de‐las‐Peñas, C., Cleland, J. A., & Gutiérrez-Vega, M. D. R. (2008) · Journal of Orthopaedic and Sports Physical Therapy · doi:10.2519/jospt.2009.2914 · Cited by 149
Lau, H. M. C., Chiu, T. T. W., & Lam, T.-H. (2010) · Manual Therapy · doi:10.1016/j.math.2010.08.003 · Cited by 143
Tousignant, M., Smeesters, C., Breton, A.-M., Breton, É., & Corriveau, H. (2006) · Journal of Orthopaedic and Sports Physical Therapy · doi:10.2519/jospt.2006.36.4.242 · Cited by 143
Neck Mobility in Different Headache Disorders
Zwart, J. (1997) · Headache The Journal of Head and Face Pain · doi:10.1046/j.1526-4610.1997.3701006.x · Cited by 131
Jordan, K. P. (2000) · British journal of rheumatology · doi:10.1093/rheumatology/39.4.382 · Cited by 122
Lumbar Range of Motion: Influence of Time of Day and Individual Factors on Measurements
Ensink, F. B. M., Saur, P. M. M., Frese, K., Seeger, D., & Hildebrandt, J. (1996) · Spine · doi:10.1097/00007632-199606010-00012 · Cited by 83
Amiri, M., Jull, G. A., & Bullock-Saxton, J. E. (2003) · Manual Therapy · doi:10.1016/s1356-689x(03)00009-2 · Cited by 81
Mannion, A. F., Klein, G. N., Dvořák, J., & Lanz, C. (2000) · European Spine Journal · doi:10.1007/s005860000187 · Cited by 78
Do inertial sensors represent a viable method to reliably measure cervical spine range of motion?
Theobald, P. S., Jones, M. D., & Williams, J. M. (2011) · Manual Therapy · doi:10.1016/j.math.2011.06.007 · Cited by 75
Characterization of Acute and Chronic Whiplash-Associated Disorders
Elliott, J. M., Noteboom, J. T., Flynn, T. W., & Sterling, M. M. (2009) · Journal of Orthopaedic and Sports Physical Therapy · doi:10.2519/jospt.2009.2826 · Cited by 75
Petersen, C., Johnson, R. D., & Schuit, D. (2000) · Manual Therapy · doi:10.1054/math.2000.0232 · Cited by 51
Jordan, K. P., Haywood, K. L., Dziedzic, K., Garratt, A., Jones, P. W., Ong, B. N., & Dawes, P. T. (2004) · PubMed · Cited by 40
Knowledge to Action: A Challenge for Neck Pain Treatment
Gross, A., Haines, T., Goldsmith, C. H., Santaguida, L., Mclaughlin, L. M., Peloso, P. M. J., Burnie, S. J., Hoving, J. L., & Cervical Overview Group (2008) · Journal of Orthopaedic and Sports Physical Therapy · doi:10.2519/jospt.2009.2831 · Cited by 40
Howell, E. R., Hudes, K., Vernon, H. T., & Soave, D. M. (2011) · Journal of Musculoskeletal Pain · doi:10.3109/10582452.2011.635849 · Cited by 27
Cited by
Campa-Moran, I., Rey-Gudin, E., Fernández-Carnero, J., Paris-Alemany, A., Gil-Martinez, A., Lerma Lara, S., Prieto-Baquero, A., Alonso-Perez, J. L., & La Touche, R. (2015) · Pain Research and Treatment · doi:10.1155/2015/327307 · Cited by 81
Dunleavy, K., Neil, J., Tallon, A., & Adamo, D. E. (2015) · Journal of Manual & Manipulative Therapy · doi:10.1179/2042618614y.0000000070 · Cited by 23
Metadata
- Authors
- Audette, I., Dumas, J.-P., Côté, J. N., & De Serres, S. J.
- Venue
- Journal of Orthopaedic & Sports Physical Therapy
- Published
- 2010-05
- Volume
- 40
- Issue
- 5
- Pages
- 318-323
- ISSN
- 0190-6011 (ISSN-L)
- 1938-1344
- OpenAlex
- W2076850657
- Cited by
- 396
- References
- 32
- FWCI
- 5.94
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.