Reproducibility and Accuracy of Quantitative Myocardial Blood Flow Assessment with 82Rb PET: Comparison with 13N-Ammonia PET
Open accessAbstract
UNLABELLED: (82)Rb cardiac PET allows the assessment of myocardial perfusion with a column generator in clinics that lack a cyclotron. There is evidence that the quantitation of myocardial blood flow (MBF) and coronary flow reserve (CFR) with dynamic (82)Rb PET is feasible. The objectives of this study were to determine the accuracy and reproducibility of MBF estimates from dynamic (82)Rb PET by using our methodology for generalized factor analysis (generalized factor analysis of dynamic sequences [GFADS]) and compartment analysis. METHODS: Reproducibility was evaluated in 22 subjects undergoing dynamic rest and dipyridamole stress (82)Rb PET studies at a 2-wk interval. The inter- and intraobserver variability of MBF quantitation with dynamic (82)Rb PET was assessed with 4 repeated estimations by each of 4 observers. Accuracy was evaluated in 20 subjects undergoing dynamic rest and dipyridamole stress PET studies with (82)Rb and (13)N-ammonia, respectively. The left ventricular and right ventricular blood pool and left ventricular tissue time-activity curves were estimated by GFADS. MBF was estimated by fitting the blood pool and tissue time-activity curves to a 2-compartment kinetic model for (82)Rb and to a 3-compartment model for (13)N-ammonia. CFR was estimated as the ratio of peak MBF to baseline MBF. RESULTS: The reproducibility of the MBF estimates in repeated (82)Rb studies was very good at rest and during peak stress (R(2)= 0.935), as was the reproducibility of the CFR estimates (R(2) = 0.841). The slope of the correlation line was very close to one for the estimation of MBF (0.986) and CFR (0.960) in repeated (82)Rb studies. The intraobserver reliability was less than 3% for the estimation of MBF at rest and during peak stress as well as for the estimation of CFR. The interobserver reliabilities were 0.950 at rest and 0.975 at peak stress. The correlation between myocardial flow estimates obtained at rest and those obtained during peak stress in (82)Rb and (13)N-ammonia studies was very good (R(2) = 0.857). Bland-Altman plots comparing CFR estimated with (82)Rb and CFR estimated with (13)N-ammonia revealed an underestimation of CFR with (82)Rb compared with (13)N-ammonia; the underestimation was within +/-1.96 SD. CONCLUSION: MBF quantitation with GFADS and dynamic (82)Rb PET demonstrated excellent reproducibility as well as intra- and interobserver reliability. The accuracy of the absolute quantitation of MBF with factor and compartment analyses and dynamic (82)Rb PET was very good, compared with that achieved with (13)N-ammonia, for MBF of up to 2.5 mL/g/min.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Numerical recipes in C: the art of scientific computing
(1993) · Choice Reviews Online · doi:10.5860/choice.30-5638 · Cited by 17,933
Numerical recipes in Pascal: the art of scientific computing
(1990) · Choice Reviews Online · doi:10.5860/choice.27-5158 · Cited by 11,811
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1191/096228099673819272 · Cited by 9,619
American Heart Association Writing Group on Myocardial Segmentation and Registration for Cardiac Imaging:, Cerqueira, M. D., Weissman, N. J., Dilsizian, V., Jacobs, A. K., Kaul, S., Laskey, W. K., Pennell, D. J., Rumberger, J. A., Ryan, T., et al. (2002) · Circulation · doi:10.1161/hc0402.102975 · Cited by 6,990
The Advanced Theory of Statistics.
Clarke, G. M., Kendall, M., & Stuart, A. (1978) · Journal of the Royal Statistical Society. Series A (General) · doi:10.2307/2344782 · Cited by 6,658
The Advanced Theory of Statistics.
Watson, G. S., Kendall, M., Stuart, A., & Ord, K. (1984) · Journal of the American Statistical Association · doi:10.2307/2288735 · Cited by 921
Cerqueira, M. D., Weissman, N. J., Dilsizian, V., Jacobs, A. K., Kaul, S., Laskey, W. K., Pennell, D. J., Rumberger, J. A., Ryan, T., & Verani, M. S. (2002) · Journal of Nuclear Cardiology · doi:10.1067/mnc.2002.123122 · Cited by 770
Albro, P. C., Gould, K. L., Westcott, R. J., Hamilton, G. W., Ritchie, J. L., & Williams, D. L. (1978) · The American Journal of Cardiology · doi:10.1016/0002-9149(78)90094-2 · Cited by 536
Cited by
Fihn, S. D., Gardin, J. M., Abrams, J., Berra, K., Blankenship, J. C., Dallas, A. P., Douglas, P. S., Foody, J. M., Gerber, T. C., Hinderliter, A. L., et al. (2012) · Circulation · doi:10.1161/cir.0b013e3182776f83 · Cited by 2,410
Fihn, S. D., Gardin, J. M., Abrams, J., Berra, K., Blankenship, J. C., Dallas, A. P., Douglas, P. S., Foody, J. M., Gerber, T. C., Hinderliter, A. L., et al. (2012) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2012.07.013 · Cited by 1,555
Improved Cardiac Risk Assessment With Noninvasive Measures of Coronary Flow Reserve
Murthy, V. L., Naya, M., Foster, C. R., Hainer, J., Gaber, M., Di Carli, G., Blankstein, R., Dorbala, S., Sitek, A., Pencina, M. J., et al. (2011) · Circulation · doi:10.1161/circulationaha.111.050427 · Cited by 850
Self-citationFihn, S. D., Gardin, J. M., Abrams, J., Berra, K., Blankenship, J. C., Dallas, A. P., Douglas, P. S., Foody, J. M., Gerber, T. C., Hinderliter, A. L., et al. (2012) · Circulation · doi:10.1161/cir.0b013e318277d6a0 · Cited by 845
Effects of Sex on Coronary Microvascular Dysfunction and Cardiac Outcomes
Murthy, V. L., Naya, M., Taqueti, V. R., Foster, C. R., Gaber, M., Hainer, J., Dorbala, S., Blankstein, R., Rimoldi, O., Camici, P. G., et al. (2014) · Circulation · doi:10.1161/circulationaha.113.008507 · Cited by 618
Self-citation
Metadata
- Authors
- El Fakhri, G., Kardan, A., Sitek, A., Dorbala, S., Abi-Hatem, N., Lahoud, Y., Fischman, A., Coughlan, M., Yasuda, T., & Di Carli, M. F.
- Venue
- Journal of Nuclear Medicine
- Publisher
- Society of Nuclear Medicine
- Published
- 2009-06-12
- Volume
- 50
- Issue
- 7
- Pages
- 1062-1071
- ISSN
- 0161-5505 (ISSN-L)
- 2159-662X
- OpenAlex
- W2124144555
- Cited by
- 266
- References
- 21
- FWCI
- 14.56
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Reproducibility and Accuracy of Quantitative Myocardial Blood Flow Assessment with 82Rb PET: Comparison with 13N-Ammonia PET. https://app.cochat.ai/papers/10.2967/jnumed.104.007831
- The paper
- El Fakhri, G., Kardan, A., Sitek, A., Dorbala, S., Abi-Hatem, N., Lahoud, Y., Fischman, A., Coughlan, M., Yasuda, T., & Di Carli, M. F. (2009). Reproducibility and Accuracy of Quantitative Myocardial Blood Flow Assessment with 82Rb PET: Comparison with 13N-Ammonia PET. Journal of Nuclear Medicine. https://doi.org/10.2967/jnumed.104.007831
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.