Performance Measurement of the microPET Focus 120 Scanner
Open accessThe DOI matches its Crossref record; 25 of the 25 references it cites were found in a registry or an index.
Abstract
UNLABELLED: The microPET Focus 120 scanner is a third-generation animal PET scanner dedicated to rodent imaging. Here, we report the results of scanner performance testing. METHODS: A (68)Ge point source was used to measure energy resolution, which was determined for each crystal and averaged. Spatial resolution was measured using a (22)Na point source with a nominal size of 0.25 mm at the system center and various off-center positions. Absolute sensitivity without attenuation was determined by extrapolating the data measured using an (18)F line source and multiple layers of absorbers. Scatter fraction and counting rate performance were measured using 2 different cylindric phantoms simulating rat and mouse bodies. Sensitivity, scatter fraction, and noise equivalent counting rate (NECR) experiments were repeated under 4 different conditions (energy window, 250 approximately 750 keV or 350 approximately 650 keV; coincidence window, 6 or 10 ns). A performance phantom with hot-rod inserts of various sizes was scanned, and several animal studies were also performed. RESULTS: Energy resolution at a 511-keV photopeak was 18.3% on average. Radial, tangential, and axial resolution of images reconstructed with the Fourier rebinning (FORE) and filtered backprojection (FBP) algorithms were 1.18 (radial), 1.13 (tangential), and 1.45 mm full width at half maximum (FWHM) (axial) at center and 2.35 (radial), 1.66 (tangential), and 2.00 mm FWHM (axial) at a radial offset of 2 cm. Absolute sensitivities at transaxial and axial centers were 7.0% (250 approximately 750 keV, 10 ns), 6.7% (250 approximately 750 keV, 6 ns), 4.0% (350 approximately 650 keV, 10 ns), and 3.8% (350 approximately 650 keV, 6 ns). Scatter fractions were 15.9% (mouse phantom) and 35.0% (rat phantom) for 250 approximately 750 keV and 6 ns. Peak NECR was 869 kcps at 3,242 kBq/mL (mouse phantom) and 228 kcps at 290 kBq/mL (rat phantom) at 250 approximately 750 keV and 6 ns. Hot-rod inserts of 1.6-mm diameter were clearly identified, and animal studies illustrated the feasibility of this system for studies of whole rodents and mid-sized animal brains. CONCLUSION: The results of this independent field test showed the improved physical characteristics of the F120 scanner over the previous microPET series systems. This system will be useful for imaging studies on small rodents and brains of larger animals.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Molecular imaging of cancer with positron emission tomography
Gambhir, S. S. (2002) · Nature Reviews Cancer · doi:10.1038/nrc882 · Cited by 1,656
THE INTERNATIONAL COMMISSION ON RADIATION UNITS AND MEASUREMENTS
Wambersie, A. (2004) · Journal of the ICRU · doi:10.1093/jicru/ndh019 · Cited by 869
National electrical manufacturers association
(1926) · Journal of the A.I.E.E. · doi:10.1109/jaiee.1926.6537344 · Cited by 860
Clinical Applications of PET in Oncology
Rohren, E. M., Turkington, T. G., & Coleman, R. E. (2004) · Radiology · doi:10.1148/radiol.2312021185 · Cited by 859
Analytic 3D image reconstruction using all detected events
Kinahan, P. E., & Rogers, J. G. (1989) · IEEE Transactions on Nuclear Science · doi:10.1109/23.34585 · Cited by 767 · cited 1×
“The PET image was reconstructed using 3DRP (16).”
MethodExact and approximate rebinning algorithms for 3-D PET data
Defrise, M., Kinahan, P. E., Townsend, D. W., Michel, C., Sibomana, M., & Newport, D. F. (1997) · IEEE Transactions on Medical Imaging · doi:10.1109/42.563660 · Cited by 758 · cited 6×
“3D sinograms were rebinned into 2D sinograms using FORE and reconstructed using FBP with a ramp filter cutoff at the Nyquist frequency after detector normalization using the component-based normalization method (21).”
MethodInfluentialHigh-resolution 3D Bayesian image reconstruction using the microPET small-animal scanner
Qi, J., Leahy, R. M., Cherry, S. R., Chatziioannou, A., & Farquhar, T. H. (1998) · Physics in Medicine and Biology · doi:10.1088/0031-9155/43/4/027 · Cited by 635
PET imaging of amyloid in Alzheimer's disease
Nordberg, A. (2004) · The Lancet Neurology · doi:10.1016/s1474-4422(04)00853-1 · Cited by 464
(2005) · PubMed · Cited by 395
Treatment of axial data in three-dimensional PET.
Daube-Witherspoon, M. E., & Muehllehner, G. (1987) · PubMed · Cited by 357
Performance evaluation of the GE healthcare eXplore VISTA dual-ring small-animal PET scanner.
Wang, Y., Seidel, J., Tsui, B. M. W., Vaquero, J. J., & Pomper, M. G. (2006) · PubMed · Cited by 329
[18F]FLT-PET in oncology: current status and opportunities
Been, L. B., Suurmeijer, A. J. H., Cobben, D. C. P., Jager, P. L., Hoekstra, H. J., & Elsinga, P. H. (2004) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-004-1687-6 · Cited by 284
Performance evaluation of the microPET R4 PET scanner for rodents
Knoess, C., Siegel, S., Smith, A., Newport, D., Richerzhagen, N., Winkeler, A., Jacobs, A., Goble, R. N., Graf, R., Wienhard, K., et al. (2003) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-002-1052-6 · Cited by 256 · cited 8×
“84 mm, respectively) at the FOV centers (19).”
BackgroundMethodResultInfluentialPET performance measurements using the NEMA NU 2-2001 standard.
Daube-Witherspoon, M. E., Karp, J. S., Casey, M., DiFilippo, F. P., Hines, H., Muehllehner, G., Simcic, V. J., Stearns, C. W., Adam, L.-E., Kohlmyer, S. G., et al. (2002) · PubMed · Cited by 222
A method for measuring the absolute sensitivity of positron emission tomographic scanners
Bailey, D. L., Jones, T., & Spinks, T. J. (1991) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/bf02258426 · Cited by 136 · cited 2×
“Although we measured the absolute sensitivity free of photon attenuation and scatter using an extrapolation method (27), the sensitivity of the F220 in (12) was measured using a point source encapsulated in a stainless steel sleeve.”
MethodPET Scanners Dedicated to Molecular Imaging of Small Animal Models
Chatziioannou, A. F. (2002) · Molecular Imaging and Biology · doi:10.1016/s1095-0397(01)00067-x · Cited by 110 · cited 1×
“Because the linear dimensions of organs in these animals are 10 or more times smaller in each dimension than in human subjects, the spatial resolution of a suitable PET scanner must be similarly higher and the voxel size should also be reduced according to sampling theory (9,10).”
BackgroundPerformance Evaluation of the microPET®—FOCUS-F120
Laforest, R., Longford, D., Siegel, S., Newport, D. F., & Yap, J. (2007) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2006.889312 · Cited by 93
Erdi, Y. E., Nehmeh, S. A., Mulnix, T., Humm, J. L., & Watson, C. C. (2004) · PubMed · Cited by 93
NEMA NU 2 performance tests for scanners with intrinsic radioactivity.
Watson, C. C., Casey, M., Eriksson, L., Mulnix, T., Adams, D., & Bendriem, B. (2004) · PubMed · Cited by 90
Small Animal Imaging in Drug Development
Pomper, M. G., & Lee, J. S. (2005) · Current Pharmaceutical Design · doi:10.2174/138161205774424681 · Cited by 77
Self-citationSmall animal PET: aspects of performance assessment
Beer, S., & Bauer, A. (2004) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-004-1683-x · Cited by 73 · cited 1×
“Spatial resolution measurements were not corrected for source dimension, positron range, or acolinearity of positron annihilation (12,20).”
BackgroundYao, R., Seidel, J., Johnson, C. A., Daube-Witherspoon, M. E., Green, M. V., & Carson, R. E. (2000) · IEEE Transactions on Medical Imaging · doi:10.1109/42.876305 · Cited by 64
Evaluation of maximum-likelihood based attenuation correction in positron emission tomography
Nuyts, J. L., Dupont, P. J., Stroobants, S. G., Maes, A., Mortelmans, L., & Suetens, P. (1999) · IEEE Transactions on Nuclear Science · doi:10.1109/23.790847 · Cited by 40
Can small animal imaging accelerate drug development?
Pomper, M. G. (2002) · Journal of Cellular Biochemistry · doi:10.1002/jcb.10443 · Cited by 28
Performance evaluation of the microPET-Focus - F120
Laforest, R. A., Longford, D., Siegel, S., Newport, D. F., & Yap, J. T. (2005) · IEEE Symposium Conference Record Nuclear Science 2004. · doi:10.1109/nssmic.2004.1466308 · Cited by 22
Cited by
Davies, B. R., Greenwood, H., Dudley, P., Crafter, C., Yu, D.-H., Zhang, J., Li, J., Gao, B., Ji, Q., Maynard, J., et al. (2012) · Molecular Cancer Therapeutics · doi:10.1158/1535-7163.mct-11-0824-t · Cited by 471
89Zr-DFO-J591 for ImmunoPET of Prostate-Specific Membrane Antigen Expression In Vivo
Holland, J. P., Divilov, V., Bander, N. H., Smith-Jones, P. M., Larson, S. M., & Lewis, J. S. (2010) · Journal of Nuclear Medicine · doi:10.2967/jnumed.110.076174 · Cited by 418
Standardized methods for the production of high specific-activity zirconium-89
Holland, J. P., Sheh, Y., & Lewis, J. S. (2009) · Nuclear Medicine and Biology · doi:10.1016/j.nucmedbio.2009.05.007 · Cited by 408
Resolution modeling in PET imaging: Theory, practice, benefits, and pitfalls
Rahmim, A., Qi, J., & Sossi, V. (2013) · Medical Physics · doi:10.1118/1.4800806 · Cited by 357
Bao, Q., Newport, D., Chen, M., Stout, D. B., & Chatziioannou, A. F. (2009) · Journal of Nuclear Medicine · doi:10.2967/jnumed.108.056374 · Cited by 306
Metadata
- Authors
- Kim, J. S., Lee, J. S., Im, K. C., Kim, S. J., Kim, S.-Y., Lee, D. S., & Moon, D. H.
- Venue
- Journal of Nuclear Medicine
- Publisher
- Society of Nuclear Medicine
- Published
- 2007-08-17
- Volume
- 48
- Issue
- 9
- Pages
- 1527-1535
- ISSN
- 0161-5505 (ISSN-L)
- OpenAlex
- W2137277859
- Cited by
- 180
- References
- 25
- FWCI
- 14.82
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.