Performance Evaluation of the uEXPLORER Total-Body PET/CT Scanner Based on NEMA NU 2-2018 with Additional Tests to Characterize PET Scanners with a Long Axial Field of View
Open accessAbstract
The world’s first total-body PET scanner with an axial field of view (AFOV) of 194 cm is now in clinical and research use at our institution. The uEXPLORER PET/CT system is the first commercially available total-body PET scanner. Here we present a detailed physical characterization of this scanner based on National Electrical Manufacturers Association (NEMA) NU 2-2018 along with a new set of measurements devised to appropriately characterize the total-body AFOV. Methods: Sensitivity, count-rate performance, time-of-flight resolution, spatial resolution, and image quality were evaluated following the NEMA NU 2-2018 protocol. Additional measurements of sensitivity and count-rate capabilities more representative of total-body imaging were performed using extended-geometry phantoms based on the world-average human height (∼165 cm). Lastly, image quality throughout the long AFOV was assessed with the NEMA image quality (IQ) phantom imaged at 5 axial positions and over a range of expected total-body PET imaging conditions (low dose, delayed imaging, short scan duration). Results: Our performance evaluation demonstrated that the scanner provides a very high sensitivity of 174 kcps/MBq, a count-rate performance with a peak noise-equivalent count rate of approximately 2 Mcps for total-body imaging, and good spatial resolution capabilities for human imaging (≤3.0 mm in full width at half maximum near the center of the AFOV). Excellent IQ, excellent contrast recovery, and low noise properties were illustrated across the AFOV in both NEMA IQ phantom evaluations and human imaging examples. Conclusion: In addition to standard NEMA NU 2-2018 characterization, a new set of measurements based on extending NEMA NU 2-2018 phantoms and experiments was devised to characterize the physical performance of the first total-body PET system. The rationale for these extended measurements was evident from differences in sensitivity, count-rate–activity relationships, and noise-equivalent count-rate limits imposed by differences in dead time and randoms fraction between the NEMA NU 2 70-cm phantoms and the more representative total-body imaging phantoms. Overall, the uEXPLORER PET system provides ultra-high sensitivity that supports excellent spatial resolution and IQ throughout the field of view in both phantom and human imaging.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“This becomes especially evident with recent PET systems with extra-large axial FOV of >25 cm and their potential to substantially reduce required acquisition times [47,149,150].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [150]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
A century of trends in adult human height
NCD Risk Factor Collaboration (NCD-RisC) (2016) · eLife · doi:10.7554/elife.13410 · Cited by 1,558
Performance Measurements of the Siemens mMR Integrated Whole-Body PET/MR Scanner
Delso, G., Fürst, S., Jakoby, B., Ladebeck, R., Ganter, C., Nekolla, S. G., Schwaiger, M., & Ziegler, S. I. (2011) · Journal of Nuclear Medicine · doi:10.2967/jnumed.111.092726 · Cited by 870
Cherry, S. R., Jones, T., Karp, J. S., Qi, J., Moses, W. W., & Badawi, R. D. (2018) · Journal of Nuclear Medicine · doi:10.2967/jnumed.116.184028 · Cited by 763
Self-citationExact and approximate rebinning algorithms for 3-D PET data
Defrise, M., Kinahan, P. E., Townsend, D. W., Michel, C. J., Sibomana, M., & Newport, D. F. (1997) · IEEE Transactions on Medical Imaging · doi:10.1109/42.563660 · Cited by 757
First Human Imaging Studies with the EXPLORER Total-Body PET Scanner*
Badawi, R. D., Shi, H., Hu, P., Chen, S., Xu, T., Price, P., Ding, Y., Spencer, B. A., Nardo, L., Liu, W., et al. (2019) · Journal of Nuclear Medicine · doi:10.2967/jnumed.119.226498 · Cited by 725
Self-citationPhysical and clinical performance of the mCT time-of-flight PET/CT scanner
Jakoby, B., Bercier, Y., Conti, M., Casey, M. E., Bendriem, B., & Townsend, D. W. (2011) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/8/004 · Cited by 487
Performance Characteristics of the Digital Biograph Vision PET/CT System
van Sluis, J., de Jong, J., Schaar, J., Noordzij, W., van Snick, P., Dierckx, R., Borra, R., Willemsen, A., & Boellaard, R. (2019) · Journal of Nuclear Medicine · doi:10.2967/jnumed.118.215418 · Cited by 484
Physical Performance of the new hybrid PET/CT Discovery-690
Bettinardi, V., Presotto, L., Rapisarda, E., Picchio, M., Gianolli, L., & Gilardi, M. C. (2011) · Medical Physics · doi:10.1118/1.3635220 · Cited by 377
Total-body imaging: Transforming the role of positron emission tomography
Cherry, S. R., Badawi, R. D., Karp, J. S., Moses, W. W., Price, P., & Jones, T. (2017) · Science Translational Medicine · doi:10.1126/scitranslmed.aaf6169 · Cited by 251
Self-citationGrant, A. M., Deller, T. W., Khalighi, M. M., Maramraju, S. H., Delso, G., & Levin, C. S. (2016) · Medical Physics · doi:10.1118/1.4945416 · Cited by 246
Research works printed but not linked: 2
- Poon JK . The Performance Limits of Long Axial Field of View PET Scanners. Thesis. University of California, Davis; 2013.
- Tang S Zhao Y Wang J Liu Y Dong Y . Geometric correction in normalization for long axial field-of-view, whole-body PET scanner. Poster presented at: the 2019 Nuclear Science Symposium (NSS) and Medical Imaging Conference (MIC), and the International Symposium on Room Temperature Semiconductor Detectors (RTSD); October 26 to November 2, 2019; Manchester, U.K.
Cited by
Sorted by citations
Items 201–300 of 404
First positronium lifetime imaging with scandium-44 on a long axial field-of-view PET/CT
Mercolli, L., Steinberger, W. M., Grundler, P. V., Moiseeva, A. N., Braccini, S., Conti, M., Moskal, P., Rathod, N., Rominger, A., Sari, H., et al. (2025) · Frontiers in Nuclear Medicine · doi:10.3389/fnume.2025.1648621 · Cited by 6
Leung, E. K., Abdelhafez, Y. G., Berg, E., Xie, Z., Zhang, X., Bayerlein, R., Spencer, B. A., Li, E. J., Omidvari, N., Selfridge, A. R., et al. (2022) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ac72f1 · Cited by 6
Self-citationFerrara, D., Sundar, L. K. S., Chalampalakis, Z., Geist, B. K., Gompelmann, D., Gutschmayer, S., Hacker, M., Kertész, H., Kluge, K., Idzko, M., et al. (2024) · Frontiers in Physics · doi:10.3389/fphy.2024.1378521 · Cited by 6
Tang, H., Wu, Y., Cheng, Z., Song, S., Dong, Q., Zhou, Y., Shu, Z., Hu, Z., & Zhu, X. (2024) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-024-06926-0 · Cited by 6
Feasibility of ultra-low-activity 18F-FDG PET/CT imaging in children
Wu, H., Liu, G., Yu, H., Zheng, Z., He, Y., & Shi, H. (2024) · British Journal of Radiology · doi:10.1093/bjr/tqae208 · Cited by 6
Total Body Positron Emission Tomography/Computed Tomography: Current Status in Oncology
Nanni, C., Farolfi, A., Castellucci, P., & Fanti, S. (2024) · Seminars in Nuclear Medicine · doi:10.1053/j.semnuclmed.2024.10.006 · Cited by 6
Montgomery, M. E., Andersen, F. L., Mathiasen, R., Borgwardt, L., Andersen, K. F., & Ladefoged, C. N. (2024) · Diagnostics · doi:10.3390/diagnostics14242788 · Cited by 6
Abi-Akl, M., Dadgar, M., Toufique, Y., Bouhali, O., & Vandenberghe, S. (2023) · EJNMMI Physics · doi:10.1186/s40658-023-00559-2 · Cited by 6
Muntean, A., Shoshan, Y., Yuzhaninov, S., Ripiccini, E., Bruschini, C. E., Fish, A., & Edoardo Charbon, A. (2023) · IEEE Journal of Selected Topics in Quantum Electronics · doi:10.1109/jstqe.2023.3346957 · Cited by 6
Phantom study for 90Y liver radioembolization dosimetry with a long axial field-of-view PET/CT
Mercolli, L., Zeimpekis, K., Prenosil, G. A., Sari, H., Rathke, H. G., Rominger, A., & Shi, K. (2024) · Physica Medica · doi:10.1016/j.ejmp.2024.103296 · Cited by 6
Non-invasive quantification of 18F-florbetaben with total-body EXPLORER PET
Holy, E. N., Li, E. J., Bhattarai, A., Fletcher, E. M., Alfaro, E. R., Harvey, D. J., Spencer, B. A., Cherry, S. R., DeCarli, C. S., & Fan, A. P. (2024) · EJNMMI Research · doi:10.1186/s13550-024-01104-7 · Cited by 5
Self-citationMehadji, B., Spencer, B. A., Alberts, I., Karakatsanis, N. A., Roncali, E., & Nardo, L. (2025) · PET Clinics · doi:10.1016/j.cpet.2025.07.003 · Cited by 5
Self-citationLymphoma Therapy Response Assessment with Low-Dose [ 18 F]FDG Total-Body PET/CT
Mingels, C., Chung, K. J., Nalbant, H., Abdelhafez, Y. G., Esteghamat, N., Rokni, M., Zoghi, S., Tuscano, J. M., Rominger, A., Badawi, R. D., et al. (2025) · Journal of Nuclear Medicine · doi:10.2967/jnumed.124.268841 · Cited by 5
Self-citationThe role of 18F-FDG PET/CT in primary cutaneous lymphoma: an educational review
McDonald, E. O., Amanullah, A. A., Park, P. S. U., Song, W. B., Werner, T. J., Alavi, A. A., & Revheim, M.-E. R. (2023) · Annals of Nuclear Medicine · doi:10.1007/s12149-023-01830-3 · Cited by 5
Long-Axial Field-of-View PET Imaging in Patients with Lymphoma
Mingels, C., Nalbant, H., Sari, H., Godinez, F., Şen, F., Spencer, B. A., Esteghamat, N., Tuscano, J. M., & Nardo, L. (2024) · PET Clinics · doi:10.1016/j.cpet.2024.05.005 · Cited by 5
Self-citationHe, Y., Rusho, M. A., Menon, S. V., Kaur, M., Jabir, M. S., Jawad, S. F., Hasan, T. F., Najm, M. A. A., Jawad, M. A., & Khelief, A. J. (2024) · Journal of Radioanalytical and Nuclear Chemistry · doi:10.1007/s10967-024-09804-z · Cited by 5
Zhang, X., Xiang, Z., Wang, F., Pan, X., Zhang, Q., Wang, P., Jiang, L., & Yuan, H. (2025) · EJNMMI Physics · doi:10.1186/s40658-025-00729-4 · Cited by 5
Development of Whole-Body TOF-DOI PET Detector Using Dual-Ended Readout Method
Wang, H. B., Xie, J., & Du, J. (2025) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2025.3570343 · Cited by 5
Gamma Ray Emission Imaging in the Medical and Nuclear Safeguards Fields
Dendooven, P. G., & Bubba, T. A. (2022) · Lecture notes in physics · doi:10.1007/978-3-031-10751-1_7 · Cited by 5
Zhu, Y., Tran, Q. T. T., Wang, Y., Badawi, R. D., Cherry, S. R., Qi, J., Abbaszadeh, S., & Wang, G. (2024) · NeuroImage · doi:10.1016/j.neuroimage.2024.120611 · Cited by 5
Self-citationPerformance Benefits of Extending the AFOV of PET Scanners
Viswanath, V., Daube-Witherspoon, M. E., Pantel, A. R., Parma, M. J., Werner, M. E., & Karp, J. S. (2020) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nss/mic42677.2020.9507858 · Cited by 5
Fanta, O., Patil, S., Werner, T. J., Torigian, D. A., & Alavi, A. A. (2025) · Current Cardiology Reports · doi:10.1007/s11886-025-02240-9 · Cited by 4
Bhattarai, A., Holy, E. N., Wang, Y., Spencer, B. A., Wang, G., DeCarli, C. S., & Fan, A. P. (2025) · EJNMMI Research · doi:10.1186/s13550-025-01260-4 · Cited by 4
Self-citationFranco, M. D., Giorgio, A. D., Cuzzani, G., Farolfi, A., Castellucci, P., Nanni, C., & Fanti, S. (2025) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-025-07611-6 · Cited by 4
Long axial field-of-view (LAFOV) PET in the era of multi-parametric imaging and theranostics
Karakatsanis, N. A. (2026) · British Journal of Radiology · doi:10.1093/bjr/tqag060 · Cited by 4
Abi-Akl, M., Maebe, J., Vervenne, B., Bouhali, O., Vanhove, C., & Vandenberghe, S. (2025) · EJNMMI Physics · doi:10.1186/s40658-025-00766-z · Cited by 4
The Impact of PET Imaging on Translational Medicine: Insights from Large-Animal Disease Models
Deng, Z., Xi, P., Zheng, D., Xie, Z., Meng, X., & Ren, Q. (2025) · Biomolecules · doi:10.3390/biom15070919 · Cited by 4
Self-citationFranco, M. D., Giorgio, A. D., Farolfi, A., Michelle, A. M. O. N., Mingels, C., Nardo, L., & Triumbari, E. K. A. (2025) · The Quarterly Journal of Nuclear Medicine and Molecular Imaging · doi:10.23736/s1824-4785.25.03656-8 · Cited by 4
Evaluation of the timing performance of the DIET ASIC
Lang, Z., Deng, Z., & Gong, H. (2023) · Journal of Instrumentation · doi:10.1088/1748-0221/18/03/p03010 · Cited by 4
Self-citationValidation of a Monte Carlo model of the uEXPLORER total-body PET scanner using GATE code
Katib, M. E., Chakir, E. M., Sebihi, R., Saikouk, H., & Nhila, O. (2023) · Radiation Physics and Chemistry · doi:10.1016/j.radphyschem.2023.111052 · Cited by 4
Jiang, J., Hua, J., Wang, H., Yuan, Z., Meng, Y., Lu, H., Liu, S. H., Chen, Y., & Tai, Y. (2023) · Physics in Medicine and Biology · doi:10.1088/1361-6560/acdfaf · Cited by 4
Numerical investigation reveals challenges in measuring the contrast recovery coefficients in PET
Bayerlein, R., Spencer, B. A., Abdelhafez, Y. G., Cherry, S. R., Badawi, R. D., & Omidvari, N. (2023) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ad00fa · Cited by 4
Self-citationCañizares, G., Jiménez-Serrano, S., Lucero, A., Morera-Ballester, C., Muñoz, E., Benlloch, J., & González, A. J. (2024) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2024.3365911 · Cited by 4
Morris, E. D., Emvalomenos, G. M., Hoye, J., & Meikle, S. R. (2024) · Journal of Nuclear Medicine · doi:10.2967/jnumed.124.267494 · Cited by 4
Yang, H., Wang, L., Tang, G., Zhou, W., Tian, Y., Zhang, Y., Huang, Y., Chen, L., Wang, M., Han, Y., et al. (2025) · Journal of Nuclear Medicine Technology · doi:10.2967/jnmt.125.269735 · Cited by 4
Realistic total‐body J‐PET geometry optimization: Monte Carlo study
Baran, J., Krzemień, W., Parzych, S., Raczyński, L., Bała, M., Coussat, A., Chug, N., Czerwiński, E., Curceanu, C. O., Dadgar, M., et al. (2025) · Medical Physics · doi:10.1002/mp.17627 · Cited by 4
Artificial Intelligence-Guided PET Image Reconstruction and Multi-Tracer Imaging
Dassanayake, M., Lopez, A., Reader, A. J., Cook, G., Mingels, C., Rahmim, A., Seifert, R., Alberts, I., & Yousefirizi, F. (2025) · PET Clinics · doi:10.1016/j.cpet.2025.07.005 · Cited by 3
Xu, H., Sun, Y., He, L., Liu, Y., Li, N., Tang, S., Ding, Y., Wang, X., Bi, D., Liu, H., et al. (2023) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmicrtsd49126.2023.10338795 · Cited by 3
Self-citationGao, M., Daube-Witherspoon, M. E., Karp, J. S., & Surti, S. (2025) · Journal of Nuclear Medicine · doi:10.2967/jnumed.124.267769 · Cited by 3
Giacomo, S. D., Ronchi, M., Amadori, M., Borghi, G., Carminati, M., & Fiorini, C. E. (2025) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2025.3530774 · Cited by 3
Zabihi, A., Li, X., Ramirez, A., Ahmad, I., Rolo, M. D. D. R., Franco, D., Gabriele, F., Galbiati, C., Lai, M., Marlow, D. R., et al. (2025) · Physics in Medicine and Biology · doi:10.1088/1361-6560/adbaac · Cited by 3
Bharathi, P. G., Abbaszadeh, S., Moskal, P., & Alavi, A. A. (2025) · Bio-Algorithms and Med-Systems · doi:10.5604/01.3001.0055.2175 · Cited by 3
Huang, Y., Tang, S., Wang, T., Li, S., Cao, L., Chen, W., Liu, L., Wu, T., Zhao, Y., Fan, W., et al. (2025) · European Radiology · doi:10.1007/s00330-025-11847-z · Cited by 3
Self-citationInternational Conference Cybernetic Modelling of Biological Systems MCSB 2021
Wiśniowski, Z. (2021) · Bio-Algorithms and Med-Systems · doi:10.1515/bams-2021-0050 · Cited by 3
Liu, H., & Morris, E. D. (2021) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ac195d · Cited by 3
Fiz, F., Cirone, A., Righi, S., Massollo, M., Amoroso, L., Bottoni, G., Conte, M., Gambaro, M., Massone, F., Orengo, S., et al. (2023) · Pediatric Blood & Cancer · doi:10.1002/pbc.30806 · Cited by 3
Recent advances in combined Positron Emission Tomography and Magnetic Resonance Imaging
Galve, P., Rodríguez-Vila, B., Herraiz, J. L., García‐Vázquez, V., Malpica, N., Udías, J. M., & Torrado-Carvajal, Á. (2024) · Journal of Instrumentation · doi:10.1088/1748-0221/19/01/c01001 · Cited by 3
Dadgar, M., Maebe, J., & Vandenberghe, S. (2024) · EJNMMI Physics · doi:10.1186/s40658-024-00645-z · Cited by 3
Total-body dynamic PET/CT imaging reveals kinetic distribution of [13N]NH3 in normal organs
Liu, G., Gu, T., Chen, S., Gu, Y., Yu, H., & Shi, H. (2024) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-024-06826-3 · Cited by 3
Zhang, X., Xiang, Z., Wang, F., Han, C., Zhang, Q., Liu, E., Yuan, H., & Jiang, L. (2024) · EJNMMI Physics · doi:10.1186/s40658-024-00689-1 · Cited by 3
An improved Patlak-based Ki parametric imaging approach for clinical 18F-FDG total-body PET
Gu, W., Zhu, Z., Liu, Z., Wang, Y., Li, Y., Xu, T., Liu, W., Wang, K., Luo, G., & Zhou, Y. (2024) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ad9ce4 · Cited by 3
Self-citationTOF MLEM Adaptation for the Total-Body J-PET With a Realistic Analytical System Response Matrix
Shopa, R. Y., Baran, J., Klimaszewski, K., Krzemień, W., Raczyński, L., Wiślicki, W., Brzeziński, K., Chug, N., Coussat, A., Curceanu, C., et al. (2023) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2023.3243735 · Cited by 3
Choudhary, D., & Nag, S. (2025) · Journal of Instrumentation · doi:10.1088/1748-0221/20/05/p05015 · Cited by 2
Roya, M., Tsoumpas, C., van Snick, J. H., Dao, V., Willemsen, A. T. M., Slart, R. H. J. A., Boellaard, R., Glaudemans, A. W. J. M., & van Sluis, J. (2025) · Journal of Nuclear Medicine · doi:10.2967/jnumed.125.270078 · Cited by 2
Advancing Positron Emission Tomography Image Quantification
Yousefirizi, F., Dassanayake, M., Lopez, A., Reader, A. J., Cook, G., Mingels, C., Rahmim, A., Seifert, R., & Alberts, I. (2025) · PET Clinics · doi:10.1016/j.cpet.2025.07.010 · Cited by 2
Bai, S., Shao, H., Yang, H., Wang, K., Zhang, C., Li, Q., Ding, J., Xie, C., Xue, T., Hao, Y., et al. (2025) · Neurotherapeutics · doi:10.1016/j.neurot.2025.e00720 · Cited by 2
Wang, Y., Zhu, L., Xu, L., Dong, L., Zhao, H., Li, L., Huang, G., Chen, R., & Liu, J. (2025) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-025-07546-y · Cited by 2
Future perspectives of functional connectivity: Advances in data acquisition and analysis
Di, X., Xu, T., & Miao, X. (2025) · Elsevier eBooks · doi:10.1016/b978-0-443-19099-5.00020-9 · Cited by 2
Rauscher, L., Allen, M. S., Judenhofer, M. S., Byars, L. G., Scipioni, M., Udías, J. M., Catana, C., Pichler, B. J., & Schmidt, F. P. (2025) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2025.3624609 · Cited by 2
Mercolli, L., Steinberger, W. M., Läppchen, T., Amon, M., Bregenzer, C., Conti, M., Cardoso, Â. R. F., Mingels, C., Moskal, P., Rathod, N., et al. (2025) · EANM Innovation · doi:10.1016/j.eanmi.2025.100017 · Cited by 2
Saraste, A., & Knuuti, J. M. (2026) · Journal of Nuclear Cardiology · doi:10.1016/j.nuclcard.2026.102652 · Cited by 2
First positronium lifetime imaging using 52Mn and 55Co with a plastic-scintillator-based PET scanner
Das, M., Sharma, S., Beyene, E., Bilewicz, A., Choiński, J., Chug, N., Curceanu, C., Czerwiński, E., Hajduga, J., Jalali, S., et al. (2026) · Scientific Reports · doi:10.1038/s41598-026-43965-z · Cited by 2
Tarantal, A. F. (2020) · Human Gene Therapy · doi:10.1089/hum.2020.29145.ata · Cited by 2
Super-resolution reconstruction of γ-ray CT images for PET-enabled dual-energy CT imaging
Zhu, Y., Spencer, B. A., Xie, Z., Leung, E. K., Bayerlein, R., Omidvari, N., Cherry, S. R., Qi, J., Badawi, R. D., & Wang, G. (2023) · SPIE Medical Imaging: Physics of Medical Imaging · doi:10.1117/12.2654431 · Cited by 2
Self-citationRevealing Anatomical Structures in PET to Generate CT for Attenuation Correction
Pan, Y., Liu, F., Jiang, C., Huang, J., Xia, Y., & Shen, D. G. (2023) · Lecture notes in computer science · doi:10.1007/978-3-031-43999-5_3 · Cited by 2
Su, X., Geng, J., Liu, J., Liu, F., Wu, Y., Zheng, R., & Wang, X. (2023) · Radiation Detection Technology and Methods · doi:10.1007/s41605-023-00441-w · Cited by 2
A cross-staged gantry for total-body PET and CT imaging
Kaplanoglu, T., & Moskal, P. (2023) · Bio-Algorithms and Med-Systems · doi:10.5604/01.3001.0054.1941 · Cited by 2
Optimization of the WLS design for positron emission mammography and Total-Body J-PET systems
Georgadze, A. S., Shivani, S., Ardebili, K. T., & Moskal, P. (2023) · Bio-Algorithms and Med-Systems · doi:10.5604/01.3001.0054.1942 · Cited by 2
List-Mode PET Image Reconstruction Using Dykstra-Like Splitting
Ote, K., Hashimoto, F., Onishi, Y., & Ouchi, Y. (2024) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2024.3441526 · Cited by 2
Modelization of extended axial field-of-view PET scanners to analyze the performance improvement
Peña-Acosta, M. M., Gallardo, S., Lorduy-Alós, M., & Verdú, G. (2024) · Radiation Physics and Chemistry · doi:10.1016/j.radphyschem.2024.111793 · Cited by 2
Gu, T., Liu, S., Hou, X., Zhao, L., Ng, Y. L., Wang, J., & Shi, H. (2024) · European Radiology · doi:10.1007/s00330-024-11039-1 · Cited by 2
Performance evaluation of a small‐animal PET scanner with 213 mm axis using NEMA NU 4‐2008
Wei, Q., Wang, Y., Huang, X., Li, D., Zheng, Y., Wang, P., Sun, X.-L., Chai, P., Han, X., Liu, S., et al. (2024) · Medical Physics · doi:10.1002/mp.17469 · Cited by 2
Luo, F., Tang, H., Li, W., Wang, H., Chen, R., Liu, J., Zhou, C., Zhang, X., Fan, W., Zhao, Y., et al. (2025) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2025.3540112 · Cited by 2
Dai, B., Krishnamoorthy, S., Morales, E., Surti, S., & Karp, J. S. (2025) · Physics in Medicine and Biology · doi:10.1088/1361-6560/adc96d · Cited by 2
Advances in nuclear medicine imaging equipment
Hong, H., & Geng, J. (2025) · Chinese Science Bulletin (Chinese Version) · doi:10.1360/tb-2025-0048 · Cited by 2
Rashid, Z. M., Al-Sawaff, Z. H., Yousif, A. S., & Al-Nima, R. R. O. (2025) · Journal of Electrical and Computer Engineering · doi:10.1155/jece/8108611 · Cited by 2
Angelis, G. I., Ockenden, K., Meikle, S. R., & Calamante, F. (2025) · EJNMMI Reports · doi:10.1186/s41824-025-00251-x · Cited by 2
Cao, Y., Cai, D., Sui, X., Wang, X., Song, J. W., Tan, H., Hu, P., Zhang, Y., Yu, H., & Shi, H. (2023) · Quantitative Imaging in Medicine and Surgery · doi:10.21037/qims-22-440 · Cited by 2
Zhao, M., Zan, K., Cui, X., Chai, L., Ge, M., Cheng, Z., Sun, H., & Duan, Y. (2023) · Nuclear Medicine Communications · doi:10.1097/mnm.0000000000001777 · Cited by 2
He, Y., Gu, Y., Yu, H., Wu, B., Wang, S., Tan, H., Cao, Y., Chen, S., Sui, X., Zhang, Y., et al. (2021) · Research Square · doi:10.21203/rs.3.rs-1144588/v1 · Cited by 2
A Deep Neural Network for Parametric Image Reconstruction on A Large-Axial Field of View PETPreprint
Li, Y., Hu, J., Sari, H., Xue, S., Ma, R., Kandarpa, S., Visvikis, D., Rominger, A., Liu, H., & Shi, K. (2022) · Research Square · doi:10.21203/rs.3.rs-1783613/v1 · Cited by 2
Collection on clinical total-body-PET studies
Slart, R. H. J. A., & Rominger, A. (2023) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-023-06161-z · Cited by 2
First positronium image of the human brain in vivoPreprint
Moskal, P., Baran, J., Bass, S. D., Choiński, J., Chug, N., Curceanu, C., Czerwiński, E., Dadgar, M., Das, M., Dulski, K., et al. (2024) · medRxiv · doi:10.1101/2024.02.01.23299028 · Cited by 2
Mercolli, L., Steinberger, W. M., Läppchen, T., Amon, M., Bregenzer, C., Conti, M., Cardoso, Â. R. F., Mingels, C., Moskal, P., Rathod, N., et al. (2025) · medRxiv · doi:10.1101/2025.05.28.25328504 · Cited by 2
DIANA - Detectability Investgations using Artificial Nodal Additions
Bayerlein, R., Xia, M., Xie, H., Spencer, B. A., Ouyang, J., Fakhri, G. E., Nardo, L., Liu, C., & Badawi, R. D. (2024) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nss/mic/rtsd57108.2024.10657528 · Cited by 1
Self-citationTayal, S., Jain, Y., Thakur, S., Shukla, V., Venkatachalam, M. M., Kumar, A., & Sinha, R. (2024) · Indian Journal of Nuclear Medicine · doi:10.4103/ijnm.ijnm_90_24 · Cited by 1
Guglielmo, P., Laudicella, R., Rovera, G., Filice, A., Panareo, S., Chierichetti, F., Zorz, A., Ferretti, S., Iudicello, A., Frantellizzi, V., et al. (2024) · Clinical and Translational Imaging · doi:10.1007/s40336-024-00675-8 · Cited by 1
Molecular imaging of psoriatic arthritis
Groothuizen, S., & van der Laken, C. J. (2025) · Current Opinion in Rheumatology · doi:10.1097/bor.0000000000001098 · Cited by 1
Feasibility of in vivo small animal imaging using a clinical total-body PET/CT system
Mannheim, J. G., Lan, W., Conti, M., Siedler, F., Krueger, M. A., Herfert, K., la Fougère, C. J., & Schmidt, F. P. (2025) · EJNMMI Physics · doi:10.1186/s40658-025-00782-z · Cited by 1
Holman, B. F., Willson, T., Ferreira, B., Davis, N., Natarajan, H., Khan, J., Wagner, T., & McCool, D. (2025) · Frontiers in Nuclear Medicine · doi:10.3389/fnume.2025.1646628 · Cited by 1
Hu, Y., Yang, Z., Liang, L., Shen, J.-Y., Zhao, Y., Lin, J., Zhang, W., Feng, J.-T., Lv, X.-F., Jia, G.-D., et al. (2025) · European Journal of Radiology · doi:10.1016/j.ejrad.2025.112422 · Cited by 1
Ghane, M. B., Sadremomtaz, A., & Saed, M. (2025) · Biomedical Physics & Engineering Express · doi:10.1088/2057-1976/ae0d93 · Cited by 1
Clinical Evaluation of BSREM Reconstruction in Pediatric Oncology Using [ 18 F]FDG PET/CT
Shkumat, N. A., Vali, R., & Shammas, A. (2025) · Journal of Nuclear Medicine Technology · doi:10.2967/jnmt.125.269555 · Cited by 1
Chen, Q., Wang, W. T., Wang, Y., Dong, L., Zhao, H., Li, L., Huang, G., Liu, L., Chen, R., & Liu, J. (2025) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-025-07581-9 · Cited by 1
Staging lung cancer: total-body PET/CT or whole-body PET/CT?
Zhang, X., Xie, J., Zheng, Y., Tian, Y., Chen, X., Chen, L., Zhou, K., Sun, P., Zhou, W., & Wu, H. (2025) · EJNMMI Research · doi:10.1186/s13550-025-01317-4 · Cited by 1
Fiz, F., Kurch, L., Garaventa, A., Battaglia, T., Livellara, V., Bottoni, G., Borgwardt, L., Treglia, G., & Piccardo, A. (2025) · The Quarterly Journal of Nuclear Medicine and Molecular Imaging · doi:10.23736/s1824-4785.25.03648-9 · Cited by 1
Application of Total-Body PET in Pediatric Patients
Wu, H., Lin, Y., Yu, H., & Shi, H. (2025) · PET Clinics · doi:10.1016/j.cpet.2025.10.001 · Cited by 1
Chen, Y., Mintzer, R. A., Cho, S., Pocci, D., Hamdi, M., Komarov, S., Cai, L., Aggarwal, S., Thomas, A., Corbeil, J. L., et al. (2025) · Journal of Nuclear Medicine · doi:10.2967/jnumed.125.270117 · Cited by 1
Dhaenens, E., Vervenne, B., Akl, M. A., Maebe, J., Vanhove, C., & Vandenberghe, S. (2026) · Frontiers in Physics · doi:10.3389/fphy.2025.1712273 · Cited by 1
Harmonisation of standard uptake values in long-axial field-of-view PET/CT: an exploratory study
Li, Y., Zhang, Z.-Y., Bai, H., Lu, X., & Pang, H. (2026) · EJNMMI Physics · doi:10.1186/s40658-025-00819-3 · Cited by 1
Metadata
- Authors
- Spencer, B. A., Berg, E., Schmall, J. P., Omidvari, N., Leung, E. K., Abdelhafez, Y. G., Tang, S., Deng, Z., Dong, Y., Lv, Y., Bao, J., Liu, W., Li, H., Jones, T., Badawi, R. D., & Cherry, S. R.
- Affiliations
- University of California, Davis (US)
- United Imaging Healthcare (China) (CN)
- ORCID
- Venue
- Journal of Nuclear Medicine
- Publisher
- Society of Nuclear Medicine
- Published
- 2020-10-02
- Volume
- 62
- Issue
- 6
- Pages
- 861-870
- ISSN
- 0161-5505 (ISSN-L)
- 2159-662X
- PMC
- PMC8729871
- OpenAlex
- W3090317854
- Cited by
- 403
- References
- 32
- Citations by year
- 2026: 50 · 2025: 78 · 2024: 96 · 2023: 78 · 2022: 66 · 2021: 32 · 2020: 3
- FWCI
- 21.80
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of Nuclear Medicine (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- eScholarship (California Digital Library) (submittedVersion)
- Funders
- National Cancer Institute (R01CA249422, R01CA206187, P30CA093373)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Cardiac Imaging and Diagnostics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
- Keywords
- total-body PET/CT, axial field of view, scanner performance evaluation, count rate performance, noise equivalent count rate, dead time, spatial resolution, image quality, time-of-flight resolution
- MeSH terms
- Positron Emission Tomography Computed Tomography, Humans, Quality Control, Time Factors, Phantoms, Imaging, Whole Body Imaging, Limit of Detection
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review