Dobutamine stress magnetocardiography for the detection of significant coronary artery stenoses—a prospective study in comparison with simultaneous 12-lead electrocardiography
The DOI matches its Crossref record; 39 of the 40 references it cites were found in a registry or an index.
Abstract
Background: Exercise electrocardiography is an imperfect test for the detection of coronary artery disease (CAD). Magnetocardiography detects cardiac electrical disturbances associated with myocardial ischemia. We prospectively investigated the accuracy of high-dose dobutamine stress magnetocardiography (DS-MCG) and simultaneous electrocardiography (DS-ECG) for the detection of significant CAD. Methods: 100 patients with an intermediate pre-test probability for CAD underwent DS-MCG using a multichannel magnetometer prior to invasive coronary angiography. Patients were examined at rest and during a standard dobutamine–atropine scheme. Significant reduction of epicardial current strength/density during stress, reconstructed from the magnetic field map and superposed on a virtual heart model indicates myocardial ischemia. A 12-lead DS-ECG was recorded simultaneously. Significant coronary artery stenosis was defined as ≥70% of lumen reduction. Results: Without β-blocker all 100 patients reached the targeted heart rate. The image quality of DS-MCG and DS-ECG was sufficient for analysis in all patients. In 19 patients CAD was ruled out angiographically. Thirty two or seven patients revealed coronary artery stenoses of 30–49% or of 50–69%, respectively. In 42 patients we found significant stenoses of ≥70%. In 41 of these patients DS-MCG revealed myocardial ischemia. The sensitivity of DS-MCG and DS-ECG for the detection of significant coronary artery stenosis was 97.6% and 26.2%, the specificity of DS-MCG and DS-ECG 82.8% and 82.8%, respectively. Conclusions: DS-MCG can be performed with a standard dobutamine/atropine stress protocol. DS-MCG yields a significantly higher accuracy for the detection of significant coronary artery stenosis than DS-ECG.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“However, despite the great clinical significance of movable MCG systems, few studies have been reported (14, 15, 23–27).”
From A movable unshielded magnetocardiography system (2023) · cites this paper as [25]
In the citing paper
“The usefulness of MCG for a better understanding of the arrhythmogenic mechanisms underlying the high-risk J syndromes (225, 327–336); the identification of atrial arrhythmogenic vulnerability (248), of left atrial dysfunction (330), and of the atrial propagation pathways (228); and the dominant frequencies (181, 183, 189, 232) in patients with atrial fibrillation, as well as providing accurate”
From Clinical magnetocardiography: the unshielded bet—past, present, and future (2023) · cites this paper as [332]
Citations
Cites
Analysis of probability as an aid in the clinical diagnosis of coronary artery disease.
Diamond, G. A., & Forrester, J. S. (1979) · New England Journal of Medicine · doi:10.1056/nejm197906143002402 · Cited by 2,664 · cited 1×
“…artery disease (CAD) the need for further evaluation by non-invasive and invasive diagnosis can be simply estimated by calculating the pre-test probability of the disease on the basis of patient’s history, physical examination, and initial tests including laboratory tests and ECG at rest [1,2].”
Committee Members, Gibbons, R. J., Balady, G. J., Timothy Bricker, J., Chaitman, B. R., Fletcher, G. F., Froelicher, V. F., Mark, D. B., McCallister, B. D., Mooss, A. N., et al. (2002) · Circulation · doi:10.1161/01.cir.0000034670.06526.15 · Cited by 2,607
Picano, E., Mathias, W., A, P., R, B., & M, P. (1994) · The Lancet · doi:10.1016/s0140-6736(94)90508-8 · Cited by 394
Paetsch, I., Jahnke, C., Wahl, A., Gebker, R., Neuss, M., Fleck, E., & Nagel, E. (2004) · Circulation · doi:10.1161/01.cir.0000138927.00357.fb · Cited by 333
Morise, A. P., & Diamond, G. A. (1995) · American Heart Journal · doi:10.1016/0002-8703(95)90072-1 · Cited by 191 · cited 2×
“Although the ECG tolerance test represents a cornerstone in non-invasive diagnostics of CAD, its accuracy is comparatively poor [3,4].”
BackgroundElhendy, A., O'Leary, E. L., Xie, F., McGrain, A. C., Anderson, J. R., & Porter, T. R. (2004) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2004.08.059 · Cited by 179 · cited 1×
“Cardiac imaging techniques like stress echocardiography [5–8], contrast echocardiography [9–13], nuclear imaging (SPECT, PET) [8,14– 18], or cardiac computed tomography [19–22] have demonstrated significantly better diagnostic values.”
BackgroundHeinle, S. K., Noblin, J., Goree-Best, P., Mello, A., Ravad, G., Mull, S., Mammen, P., & Grayburn, P. A. (2000) · Circulation · doi:10.1161/01.cir.102.1.55 · Cited by 173
Berman, D. S., Hachamovitch, R., Shaw, L. J., Friedman, J. D., Hayes, S. W., Thomson, L., Fieno, D. S., Germano, G., Slomka, P. J., Wong, N. D., et al. (2006) · PubMed · Cited by 159
Shimoni, S., Zoghbi, W. A., Xie, F., Kricsfeld, D., Iskander, S., Gobar, L. S., Mikati, I. A., Abukhalil, J. M., Verani, M. S., O’Leary, E. L., et al. (2001) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(00)01179-7 · Cited by 151
Holland, R. P., & Brooks, H. (1975) · Circulation Research · doi:10.1161/01.res.37.4.471 · Cited by 148 · cited 3×
“Nevertheless, changes of the ionic currents of the action potential are the first consequences of cellular hypoxia and occur long before discernible functional disturbances of the myocardium [37,38,52].”
BackgroundInfluential
Cited by
A movable unshielded magnetocardiography system
Xiao, W., Sun, C., Shen, L., Feng, Y., Liu, M., Wu, Y., Liu, X., Wu, T., Peng, X., & Guo, H. (2023) · Science Advances · doi:10.1126/sciadv.adg1746 · Cited by 116
Kwong, J. S. W., Leithäuser, B., Park, J.-W., & Yu, C.-M. (2013) · International Journal of Cardiology · doi:10.1016/j.ijcard.2012.12.056 · Cited by 103
“2 and 3) [38].”
MethodSelf-citationClinical magnetocardiography: the unshielded bet—past, present, and future
Brisinda, D., Fenici, P., & Fenici, R. (2023) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2023.1232882 · Cited by 80
Pena, M. E., Pearson, C. L., Goulet, M. P., Kazan, V. M., DeRita, A. L., Szpunar, S. M., & Dunne, R. B. (2020) · IJC Heart & Vasculature · doi:10.1016/j.ijcha.2019.100466 · Cited by 70
Magnetocardiography for the detection of myocardial ischemia
Her, A., Dischl, D., Kim, Y. H., Kim, S., & Shin, E. (2023) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2023.1242215 · Cited by 68
“(43) Abnormalities of the mean time isointegral MFM Acute MI/ Patients with MI with a history of prolonged cardiac pain and diagnostic enzyme level elevations who were either previously diagnosed using standard 12-lead EKG, as having anterior (4) or inferior (7) Q wave MI or non-Q wave MI (11)/Normal controls (9) Shie…”
MethodInfluential
Metadata
- Authors
- Park, J.-W., Leithäuser, B., Vršansky, M., & Jung, F.
- Affiliations
- Lausitzer Seenland Klinikum (DE)
- Institute for Heart and Circulation Research, Hoyerswerda, Germany
- Cardiology/Angiology Division, Hoyerswerda Hospital, Hoyerswerda, Germany
- Berlin-Brandenburger Centrum für Regenerative Therapien (DE)
- ORCID
- Jung, F.
- Venue
- Clinical Hemorheology and Microcirculation
- Publisher
- SAGE Publications
- Published
- 2008-06
- Volume
- 39
- Issue
- 1-4
- Pages
- 21-32
- ISSN
- 1386-0291 (ISSN-L)
- 1875-8622
- OpenAlex
- W314671638
- Cited by
- 27
- References
- 40
- Citations by year
- 2026: 2 · 2025: 6 · 2024: 1 · 2023: 4 · 2022: 1 · 2020: 2 · 2019: 1 · 2018: 1 · 2017: 1 · 2016: 1 · 2015: 1 · 2014: 1 · 2013: 1 · 2012: 2
- FWCI
- 0.51
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Clinical Hemorheology and Microcirculation (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Cardiac Imaging and Diagnostics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Acute Myocardial Infarction Research
- Keywords
- coronary artery stenosis, coronary artery disease, myocardial ischemia, 12-lead ECG, magnetic field mapping, virtual heart model, diagnostic accuracy, exercise electrocardiography
- MeSH terms
- Aged, Coronary Artery Disease, Dobutamine, Electrocardiography, Electrophysiology, Female, Humans, Male, Middle Aged, Models, Biological, Probability, Prospective Studies, Signal Processing, Computer-Assisted, Coronary Angiography, Coronary Stenosis, Magnetocardiography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Dobutamine stress magnetocardiography for the detection of significant coronary artery stenoses—a prospective study in comparison with simultaneous 12-lead electrocardiography. https://app.cochat.ai/papers/10.3233/ch-2008-1064
- The paper
- Park, J.-W., Leithäuser, B., Vršansky, M., & Jung, F. (2008). Dobutamine stress magnetocardiography for the detection of significant coronary artery stenoses—a prospective study in comparison with simultaneous 12-lead electrocardiography. Clinical Hemorheology and Microcirculation. https://doi.org/10.3233/ch-2008-1064
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.