Innovative wearable technology for continuous echocardiographic monitoring: first-in-neonate case report

Open access
Wang, Yan, Wu, Gangjun, Song, Jianqiang, Song, Haibo, Chu, Qinjun · 2025 · Frontiers in Pediatrics

The DOI matches its Crossref record; 8 of the 8 references it cites were found in a registry or an index; 2 of the 16 sources its claims cite were checked for support.

Abstract

Neonates in the early stage are in the transition from fetal to adult circulation, accompanied by complex hemodynamic changes, which are prone to aggravate the condition of the circulatory system during the perioperative period. Here, we present a case of a 3-day-old neonate with upper gastrointestinal bleeding caused by multiple esophageal and gastric ulcers. To achieve accurate circulation management, we utilized a novel wearable ultrasound device for non—invasive, continuous echocardiographic monitoring throughout general anesthesia gastroscopy and postoperative NICU care. The case demonstrates the feasibility of continuous echocardiographic monitoring in neonates and offers new ideas for circulatory monitoring and management in critical neonates.

Overview

CoChat Summary

TL;DR

This case report describes the first use of a novel wearable ultrasound device to provide continuous, non-invasive echocardiographic monitoring in a 3-day-old neonate during general anesthesia for gastroscopy and subsequent NICU care. The authors conclude that continuous echocardiographic monitoring is feasible and practical for neonatal perioperative circulatory management.

Key findings

  • A 3-day-old neonate (51 cm, 3.75 kg) with upper gastrointestinal bleeding from multiple esophageal and gastric ulcers underwent gastroscopy under general anesthesia with continuous wearable echocardiographic monitoring.
  • Preoperative ultrasound showed a ~2.4 mm mid-atrial septal left-to-right shunt, a ~1.6 mm shunt between the descending aortic root and left pulmonary artery, and mild tricuspid regurgitation, suggesting PFO and PDA; no PDA shunt was detected during the wearable examination.
  • After induction, vital signs were HR 130 beats/min, BP 58/40 mmHg, SPO2 92%, and monitoring showed slightly deficient left ventricular volume with normal contractility; based on this, 10 ml crystalloid was given and BP stabilized at 70/40 mmHg.
  • Postoperatively in NICU, monitoring continued to guide fluid and circulation management; HR varied 130–150 beats/min with SPO2 above 94% (FiO2 40%), the endotracheal tube was removed on day 2, only mild tricuspid regurgitation remained on postoperative day 5, and the neonate was discharged on postoperative day 10.

Methods

  • Design: single first-in-neonate case report.
  • Case/data: one 3-day-old neonate admitted to NICU with hypoglycemia (blood glucose as low as 1.3 mmol/L) and gastrointestinal hemorrhage, treated with fasting, gastrointestinal decompression, intragastric hemostatic medication, and 40 ml cryoprecipitate transfusion before gastroscopy.
  • Intervention/monitoring: a wearable ultrasound device (continuous phased array; model P2-5-64; 3.6-MHz center frequency, 64 array base, 0.32-mm array spacing) developed by Sichuan Wismed Medical Technology Co., Ltd., secured to the left anterior chest wall with disposable gel skin patches to obtain a parasternal long-axis view, with images wirelessly transmitted to terminal devices; anesthesia used 5 mg esketamine, 2 mg rocuronium bromide, and 3 mg propofol with endotracheal intubation.

Limitations

  • The paper states no explicit limitations.
  • Evident from the text: this is a single case (n=1) with no control or comparison group, so feasibility is demonstrated but efficacy, safety, and generalizability cannot be established.
  • Evident from the text: the authors note the device was "not specifically designed for neonates," and the report relies on a single device model and operator-dependent image acquisition.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.3389/fped.2025.1567386

Cites

8 of 8 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Figures

Figure 1 Cardiac images from four standard views taken by a wearable cardiac ultrasound monitoring imager. (A) Parasternal long-axis view; (B) parasternal LV short-axis view; (C) parasternal aortic short-axis view; (D) apical 4-chamber view. RA, right atrium; RV, right ventricle; LA, left atrium; LV, left ventricle; PA, pulmonary artery.
Figure 1 Cardiac images from four standard views taken by a wearable cardiac ultrasound monitoring imager. (A) Parasternal long-axis view; (B) parasternal LV short-axis view; (C) parasternal aortic short-axis view; (D) apical 4-chamber view. RA, right atrium; RV, right ventricle; LA, left atrium; LV, left ventricle; PA, pulmonary artery.CC BY · PubMed Central
Figure 2 Wearable ultrasonic colour flow doppler examination between the aorta and pulmonary artery did not find a PDA.
Figure 2 Wearable ultrasonic colour flow doppler examination between the aorta and pulmonary artery did not find a PDA.CC BY · PubMed Central
Figure 3 Continuous ultrasound imaging of the heart during operation. (A) Wearable ultrasound technology is utilized in anesthesia within clinical settings; (B) continuous ultrasound imaging of parasternal long-axis view.
Figure 3 Continuous ultrasound imaging of the heart during operation. (A) Wearable ultrasound technology is utilized in anesthesia within clinical settings; (B) continuous ultrasound imaging of parasternal long-axis view.CC BY · PubMed Central
Figure 4 Wearable ultrasound technology is utilized in the NICU. (A) A disposable gel skin patches was placed on the neonate's chest for the ultrasound; (B) the wearable ultrasound patch was placed on the left anterior chest of the neonate; (C) parasternal long-axis view; (D) subcostal view of the inferior vena cava.
Figure 4 Wearable ultrasound technology is utilized in the NICU. (A) A disposable gel skin patches was placed on the neonate's chest for the ultrasound; (B) the wearable ultrasound patch was placed on the left anterior chest of the neonate; (C) parasternal long-axis view; (D) subcostal view of the inferior vena cava.CC BY · PubMed Central

Verification

Claims the paper makes

2 Valid · 0 Not valid · 11 Not checked · 18 Own results and interpretations

6 more claims

Create a free account to see all 31 claims and every source verdict.

See all claims — free account
  • Probably not a substantial claim. 13
  • The quote only partly states this claim. 1
  • It may state two separate claims as one. 7

Create a free account to see these claims.

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

Each contradiction between two passages listed here was confirmed by an independent check.

1 of 1

Citations

Cites

Show all 8 references

Cited by

Metadata

Authors
Wang, Y., Wu, G., Song, J., Song, H., & Chu, Q.
Affiliations
  • Zhengzhou University (CN)
  • West China Hospital of Sichuan University (CN)
Corresponding author
Chu, Qinjun
Venue
Frontiers in Pediatrics
Published
2025-04-15
Volume
13
Pages
1567386-1567386
ISSN
2296-2360 (ISSN-L)
OpenAlex
W4409469021
Cited by
8
References
8
Citations by year
2026: 6 · 2025: 2
FWCI
7.46
Type
journal article
Language
en
Open access
Locations
  • Frontiers in Pediatrics (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
Topics
  • Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
  • Health Sciences › Medicine › Critical Care and Intensive Care Medicine › Ultrasound in Clinical Applications
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Congenital Heart Disease Studies
Keywords
wearable ultrasound device, neonates, upper gastrointestinal bleeding, esophageal ulcer, gastric ulcer, perioperative hemodynamic management, neonatal intensive care

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

The claim verdicts this report counts are experimental: their accuracy is not yet measured.

A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).

A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).

Cited sources are checked with a language model’s help (gpt-6-sol).

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.