Comparative Analysis of the Equivital EQ02 Lifemonitor with Holter Ambulatory ECG Device for Continuous Measurement of ECG, Heart Rate, and Heart Rate Variability: A Validation Study for Precision and Accuracy

Open access
Akintola, Abimbola A., van de Pol, Vera, Bimmel, Daniel, Maan, Arie C., van Heemst, Diana · 2016 · Frontiers in Physiology
Cited by 87

The DOI matches its Crossref record; 12 of the 12 references it cites were found in a registry or an index.

Abstract

Background: The Equivital (EQ02) is a multi-parameter telemetric device offering both real-time and/or retrospective, synchronized monitoring of ECG, HR and HRV, respiration, activity and temperature. Unlike the Holter, which is the gold standard for continuous ECG measurement, EQO2 continuously monitors ECG via electrodes interwoven in the textile of a wearable belt. Objective: To compare EQ02 with the Holter for continuous home measurement of ECG, heart rate (HR) and heart rate variability (HRV). Methods: Eighteen healthy participants wore, simultaneously for 24 hours, the Holter and EQ02 monitors. Per participant, averaged HR and HRV per 5 minutes from the two devices were compared using Pearson correlation, paired T-test and Blant-Altman analyses. Accuracy and precision metrics included mean absolute relative difference (MARD). Results: Artefact content of EQ02 data varied widely between (range 1.93% to 56.45%) and within (range 0.75% to 99.61%) participants. Comparing the EQ02 to the Holter, the Pearson correlations were respectively 0.724, 0.955 and 0.997 for datasets containing all data and data with <50% or <20% artefacts respectively. For datasets containing respectively all data, data with <50% or <20% artefacts, bias estimated by Bland-Altman analysis was -2.8, -1.0 and -0.8 beats per minute and 24h MARD was 7.08, 3.01 and 1.5. After selecting a three- hour stretch of data containing 1.15% artefacts, Pearson correlation was 0.786 for HRV measured as standard deviation of NN intervals (SDNN). Conclusions: Although the EQ02 can accurately measure ECG and HRV, its accuracy and precision is highly dependent on artefact content. This is a serious limitation for clinical use in individual patients. However, the advantages of the EQ02 (ability to simultaneously monitor several physiologic parameters) may outweigh its disadvantages (higher artefact load) for research purposes and/ or for home monitoring in larger groups of study participants. Further studies can be aimed at minimizing the artefacts.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.3389/fphys.2016.00391

Cites

12 of 12 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 12 references

Cited by

See all 87

Metadata

Authors
Akintola, A. A., van de Pol, V., Bimmel, D., Maan, A. C., & van Heemst, D.
Venue
Frontiers in Physiology
Published
2016-09-21
Volume
7
Pages
391-391
ISSN
1664-042X (ISSN-L)
OpenAlex
W2522398876
Cited by
87
References
12
FWCI
2.28
Type
journal article
Language
en
Open access
gold

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Comparative Analysis of the Equivital EQ02 Lifemonitor with Holter Ambulatory ECG Device for Continuous Measurement of ECG, Heart Rate, and Heart Rate Variability: A Validation Study for Precision and Accuracy. https://app.cochat.ai/papers/10.3389/fphys.2016.00391
The paper
Akintola, A. A., van de Pol, V., Bimmel, D., Maan, A. C., & van Heemst, D. (2016). Comparative Analysis of the Equivital EQ02 Lifemonitor with Holter Ambulatory ECG Device for Continuous Measurement of ECG, Heart Rate, and Heart Rate Variability: A Validation Study for Precision and Accuracy. Frontiers in Physiology. https://doi.org/10.3389/fphys.2016.00391

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.