Insulin Micropump with Embedded Pressure Sensors for Failure Detection and Delivery of Accurate Monitoring

Open access
Dumont-Fillon, Dimitry, Tahriou, Hassen, Conan, Christophe, Chappel, Eric·2014·Micromachines

Abstract

Improved glycemic control with insulin pump therapy in patients with type 1 diabetes mellitus has shown gradual reductions in nephropathy and retinopathy. More recently, the emerging concept of the artificial pancreas, comprising an insulin pump coupled to a continuous glucose meter and a control algorithm, would become the next major breakthrough in diabetes care. The patient safety and the efficiency of the therapy are directly derived from the delivery accuracy of rapid-acting insulin. For this purpose, a specific precision-oriented design of micropump has been built. The device, made of a stack of three silicon wafers, comprises two check valves and a pumping membrane that is actuated against stop limiters by a piezo actuator. Two membranes comprising piezoresistive strain gauges have been implemented to measure the pressure in the pumping chamber and at the outlet of the pump. Their high sensitivity makes possible the monitoring of the pumping accuracy with a tolerance of ±5% for each individual stroke of 200 nL. The capability of these sensors to monitor priming, reservoir overpressure, reservoir emptying, outlet occlusion and valve leakage has also been studied.

AI snapshot

Written from the paper’s full text.

TL;DR

A MEMS reciprocating displacement insulin micropump with two integrated piezoresistive pressure sensors was characterized for delivery accuracy and failure detection. The sensors detected priming, occlusion, reservoir over/underpressure, leakage, and air bubbles, with delivery accuracy better than ±5% for observation windows larger than four minutes and a mean flow error of +0.28%.

Key findings

  • The pump chip is a stack of three silicon wafers with two check valves and a pumping membrane actuated by a piezo actuator against stop limiters; nominal stroke volume is 200 nL, compression ratio is about 1.9:1, and two pressure sensors are placed in the pumping chamber and downstream of the outlet valve. The sensors have a sensitivity of 18 mV·V⁻¹·bar⁻¹, maximum offset of ±2 mV·V⁻¹·bar⁻¹, and minimum resolution of 1 mbar.
  • Total occlusion was detected during the first actuation after the event, preferably using the outer sensor; stroke volume remained constant from −0.3 bar to +0.2 bar inlet pressure and did not vary with basal rate from 0.1 U/h to 70 U/h.
  • Leakage detection is based on pressure decay after outlet valve closure and can detect an accuracy loss larger than 5%; a back-flow equivalent to a 5% decrease of nominal stroke volume changed the pressure decay. A 200 nL air bubble introduced into the pumping chamber enlarged and reduced inner-sensor pressure peaks and was progressively expelled over following strokes.
  • At 1 U/h and 38 °C, delivery accuracy was better than ±5% for observation windows larger than four minutes, with mean flow error +0.28%. Maximum flow rate for water was 4.3 mL/h at 18 Hz; for water/glycerol mixes, maximum flow rate fell to 0.85 mL/h at 10 cP and 0.46 mL/h at 25 cP.

Methods

  • The micropump was characterized in a clean room using a Sartorius MC5 precision scale with microgram resolution; a glass capillary was immersed in insulin Novorapid, with paraffin oil to prevent evaporation and temperature regulated at 20 ± 1 °C. Pressure sensor signals were recorded with NI 6143 data acquisition cards, and a Druck DPI520 pressure controller simulated occlusion, reservoir overpressure, or an empty reservoir; an oven at 38 °C simulated patch-pump in-use temperature.
  • Tests covered priming at 3.125 Hz, occlusion in basal mode at 7.2 U/h, stroke volume versus inlet pressure, flow rate, actuation voltage, and viscosity using water and water/glycerol mixes, leakage using 6 × 10⁵ particles/mL of 0.5 micron silica particles, air bubble introduction of 200 nL, and delivery accuracy via trumpet curves according to IEC60601-2-24 processed with Infuscale 7.0.

Limitations

  • Stated: air bubble results are only qualitative because fractioning of the air bubble makes estimation of the quantity of air in the pumping chamber during the test difficult.
  • Stated: delivery accuracy test setup had limitations, in particular instabilities due to the temperature gradient in the flexible tube between the scale and the pumping unit, and the norm’s stabilization period was intentionally reduced from 24 h to 1 h.
  • Evident from the text: characterization is bench/in vitro only, with no human or clinical outcomes reported; leakage testing removed the particle filter and used silica particles, and occlusion was simulated by clamping the cannula with tweezers.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.3390/mi5041161

Cites

14 of 14 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Claims the paper makes

0 of 2 cited sources checked · 0 supported · 0 flagged

  • Findingreasoning

    The device exhibits self-priming capabilities due to its high compression ratio of about 1.9:1 for a nominal stroke volume of 200 nL.

    “Due to its high compression ratio, about 1.9:1 for a nominal stroke volume of 200 nL, the device exhibits self-priming capabilities.”
    Quote in the paperNot a substantial claim
  • Findingreasoning

    The pressure sensors' high sensitivity enables monitoring of pumping accuracy with a tolerance of ±5% for each individual 200 nL stroke.

    “Their high sensitivity makes possible the monitoring of the pumping accuracy with a tolerance of ±5% for each individual stroke of 200 nL.”
    Quote in the paperSubstantial claim
  • Findingown data

    Infrared inspection revealed silica particles forming clusters on both valve seats.

    “Infrared (IR) inspection has revealed an important amount of silica particles forming clusters on both valve seats.”
    Quote in the paperNot a substantial claim
  • Findingown data

    The pressure decay after outlet valve closure during the push phase is significantly changed for a back-flow equivalent to a 5% decrease of nominal stroke volume.

    “It is worth noting that the pressure decay following the outlet valve closure, during the push phase, is significantly changed in the case of a back-flow equivalent to a decrease of 5% of the nominal stroke volume.”
    Quote in the paperSubstantial claim
  • Findingown data

    Delivery accuracy better than ±5% is obtained for observation windows larger than four minutes, despite test setup limitations.

    “Despite the limitations induced by the test setup itself (in particular, the instabilities due to the temperature gradient in the flexible tube between the scale and the pumping unit), delivery accuracy better than ±5% is obtained for observation windows larger than four minutes.”
    Quote in the paperSubstantial claim
  • May combine two claims. 9
  • Classified as probably not a substantial claim. 5
  • May combine two claims.

    Findingown data

    For water, flow rate varies linearly with actuation frequency up to 2.5 mL/h at 3.5 Hz, and maximum flow rate is 4.3 mL/h at 18 Hz.

    “For water, the flow rate varies linearly with the actuation frequency up to 2.5 mL/h (actuation frequency of 3.5 Hz), and a maximum flow rate of 4.3 mL/h is measured at 18 Hz.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingreasoning

    Total occlusion can be detected during the first actuation after the event, allowing an alarm and pump stop to prevent large pressure.

    “Total occlusion could be detected during the first actuation that follows the event, thus triggering an alarm and stopping the pump to prevent the generation of large pressure, which may damage the fluidic line.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingreasoning

    Any occlusion induces a large increase of outlet pressure that can be detected by both sensors.

    “Because the fluidic pathway downstream the micropump exhibits a very low compressibility, any occlusion will induce a large increase of the outlet pressure, which could be detected by both sensors.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Study design and conductnone given

    Stabilization of the scale is the key parameter to obtain reproducible results.

    “The stabilization of the scale is the key parameter to obtain reproducible results.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation

    Switching from multiple daily infusion (MDI) to continuous subcutaneous insulin infusion (CSII) reduces swings in blood glucose levels and limits severe hypoglycemia.

    “Switching from multiple daily infusion (MDI) to continuous subcutaneous insulin infusion (CSII) induces well-known advantages, including the reduction of swings in blood glucose levels and the limitation of severe hypoglycemia [5][6][7].”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites [5][6][7]

      Couldn’t be matched to a paper in Crossref, so it can’t be checked.

  • Classified as probably not a substantial claim.

    Interpretationreasoning

    The outer sensor is preferably used to monitor occlusion.

    “The outer sensor, which is in direct communication with the infusion site (i.e., without any valve in-between), is preferably used to monitor occlusion.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationnone given

    The height increase of the pressure peak in the pumping chamber during priming is mainly related to a change of compressibility, while the height of the pressure peak at the outer sensor location is more related to the viscosity change between gas and liquid.

    “The height increase of the pressure peak in the pumping chamber during priming is mainly related to a change of compressibility, while the height of the pressure peak at the outer sensor location is more related to the viscosity change (between gas and liquid).”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Premisecitation

    The MEMS-based micropump provides outstanding delivery accuracy and failure detection capabilities.

    “In addition, this MEMS-based micropump provides outstanding delivery accuracy and failure detection capabilities, as reported in recent in vitro and in vivo studies comparing the pumping capability of several pump systems [8].”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [8] Borot, 2014, Accuracy of a new patch pump based on a microelectromechanical system (MEMS) compared to other commercially available insulin pumps: Results of the first in vitro and in vivo studies, J. Diabetes Sci. Technol. source
  • May combine two claims.

    Interpretationreasoning

    Stroke volume is independent of viscosity for incompressible media, but maximum flow rate is reduced for viscous liquid.

    “The stroke volume is independent of viscosity for incompressible media, since the device is a volumetric pump, but the maximum flow rate is reduced for viscous liquid.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationnone given

    The device can pump with very high accuracy up to 2.5 mL/h and its integrated piezoresistive pressure sensors can detect malfunctions (occlusion, reservoir over/under pressure, air bubbles, leakage) that could lead to accuracy loss larger than 5%.

    “In addition to its capability to pump with very high accuracy up to a rate of 2.5 mL/h, the device comprises powerful integrated piezoresistive pressures sensors able to detect malfunctions, like occlusion, reservoir over/under pressure, air bubbles or leakage, which could lead to an accuracy loss larger than 5%.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S2 (checkable). May combine two claims.

    Interpretationreasoning

    A sensor in the pumping chamber able to monitor delivery accuracy is a key element to prevent severe hypoglycemic episodes and improve safety of future closed-loop systems.

    “The presence of a sensor in the pumping chamber able to monitor delivery accuracy is a key element to prevent severe hypoglycemic episodes and to improve the safety of future closed-loop systems.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Study design and conductown data

    The air bubble results are only qualitative because fractioning makes estimation of air quantity difficult.

    “These results are only qualitative, because of the fractioning of the air bubble, making the estimation of the quantity of air in the pumping chamber during the test difficult.”
    Quote in the paperNot a substantial claim
12 of 12
  • The paper relies on this without support.

    Occlusion is a common failure of insulin pumps.

    “Occlusion is a common failure of insulin pumps.”
  • The paper relies on this without support.

    Patient safety and efficiency of insulin therapy are directly derived from delivery accuracy of rapid-acting insulin.

    “The patient safety and the efficiency of the therapy are directly derived from the delivery accuracy of rapid-acting insulin.”
  • The paper relies on this without support.

    Microfluidic devices are usually not able to guarantee correct infusion, in particular in the case of failure.

    “In contrast to syringe pumps, which control the residual volume of the insulin reservoir via the monitoring of the plunger position, microfluidic devices are usually not able to guarantee correct infusion, in particular in the case of failure.”
3 of 3

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Hypoglycemia can occur if an occlusion is due to kinking or compression and the insulin infusion rate is increased to overcome the occlusion and maintain normoglycemia, because after the occlusion is relieved a relatively large volume of noninfused insulin is suddenly delivered.

    “hypoglycemia if an occlusion is due to kinking or compression and if the insulin infusion rate is increased to overcome the occlusion and maintain normoglycemia, then after the occlusion is relieved, a relatively large volume of noninfused insulin that was not infused during the occlusion will suddenly be delivered.4”

    From Insulin Pump Occlusions: For Patients Who Have Been Around the (Infusion) Block (2017) · cites this paper as [4]

    Split into 2 atoms — 0 confirmed, 2 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.

    Experimental — accuracy not yet measured

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Dumont-Fillon, D., Tahriou, H., Conan, C., & Chappel, E.
Affiliations
Debiotech (Switzerland) (CH)
Corresponding author
Chappel, Eric
Venue
Micromachines
Publisher
MDPI AG
Published
2014-11-18
Volume
5
Issue
4
Pages
1161-1172
ISSN
2072-666X (ISSN-L)
OpenAlex
W2089992465
Cited by
38
References
14
Citations by year
2026: 1 · 2025: 1 · 2024: 1 · 2023: 5 · 2022: 7 · 2021: 5 · 2020: 6 · 2019: 2 · 2018: 2 · 2017: 4 · 2016: 3 · 2015: 1
FWCI
1.41
Type
journal article
Language
en
Open access
Locations
  • Micromachines (publishedVersion, cc-by)
  • HAL (Le Centre pour la Communication Scientifique Directe) (submittedVersion, other-oa)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
Topics
  • Physical Sciences › Engineering › Electrical and Electronic Engineering › Advanced MEMS and NEMS Technologies
  • Physical Sciences › Engineering › Biomedical Engineering › Microfluidic and Capillary Electrophoresis Applications
  • Physical Sciences › Engineering › Biomedical Engineering › Intravenous Infusion Technology and Safety
Keywords
pressure sensors, piezoresistive strain gauges, piezoelectric actuators, fault detection, type 1 diabetes, artificial pancreas, continuous glucose monitoring, rapid-acting insulin, check valves, valve leakage

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-13.

Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-13.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review