Chronically implanted pressure sensors: challenges and state of the field
Open accessAbstract
Several conditions and diseases are linked to the elevation or depression of internal pressures from a healthy, normal range, motivating the need for chronic implantable pressure sensors. A simple implantable pressure transduction system consists of a pressure-sensing element with a method to transmit the data to an external unit. The biological environment presents a host of engineering issues that must be considered for long term monitoring. Therefore, the design of such systems must carefully consider interactions between the implanted system and the body, including biocompatibility, surgical placement, and patient comfort. Here we review research developments on implantable sensors for chronic pressure monitoring within the body, focusing on general design requirements for implantable pressure sensors as well as specifications for different medical applications. We also discuss recent efforts to address biocompatibility, efficient telemetry, and drift management, and explore emerging trends.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Misalignment between the coils can disrupt the continuity of the RF signal, potentially leading to data loss and reduced operational capacity of the device.
“Misalignment between the coils can further disrupt the continuity of the RF signal, potentially leading to data loss and reduced operational capacity of the device (Liu et al., 2014).”
From Comparative analysis of energy transfer mechanisms for neural implants (2024) · cites this paper as Liu et al., 2014
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Optimal Frequency for Wireless Power Transmission Into Dispersive Tissue
Poon, A. S. Y., O'Driscoll, S., & Meng, T. H. (2010) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2010.2044310 · Cited by 469 · see citing works
First steps toward noninvasive intraocular pressure monitoring with a sensing contact lens
Leonardi, M., Leuenberger, P., Bertrand, D., Bertsch, A., & Renaud, P. (2004) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.04-0015 · Cited by 244 · see citing works
Flagged first, then by citations
Research works printed but not linked: 109
- [2]Sit A.J. Continuous monitoring of intraocular pressure: Rationale and progress toward a clinical device J. Glaucoma 2009 18 272 279 19365190
- [3]Verdecchia P. Schillaci G. Borgioni C. Ciucci A. Zampi I. Gattobigio R. Sacchi N. Porcellati C. White coat hypertension and white coat effect similarities and differences Am. J. Hypertens. 1995 8 790 798 7576395
- [4]Raboel P.H. Bartek J. Jr. Andresen M. Bellander B.M. Romner B. Intracranial Pressure Monitoring: Invasive versus Non-Invasive Methods—A Review Crit. Care Res. Pract. 2012 10.1155/2012/950393 PMC3376474 22720148 · DOI
- [5]Gelabert-Gonzalez M. Ginesta-Galan V. Sernamito-Garcia R. Allut A.G. Bandin-Diéguez J. Rumbo R.M. The Camino intracranial pressure device in clinical practice. Assessment in a 1000 cases Acta Neurochir. 2006 148 435 441 16374566
- [6]Hong W.C. Tu Y.K. Chen Y.S. Lien L.M. Huang S.J. Subdural intracranial pressure monitoring in severe head injury: Clinical experience with the Codman MicroSensor Surg. Neurol. 2006 66 S8 S13 17071260
- [7]Martínez-Mañas R.M. Santamarta D. de Campos J.M. Ferrer E. Camino® intracranial pressure monitor: Prospective study of accuracy and complications J. Neurol. Neurosurg. Psychiatr. 2000 69 82 86 10864608 PMC1736998
- [8]Jacobson B. Nordberg L. Endoradiosondes for pressure telemetering IRE Trans. BioMed. Electron. 1961 8 192 196 13718594
- [10]Atkinson J.R. Shurtleff D.B. Foltz E.L. Radio Telemetry for the Measurement of Intracranial Pressure* J. Neurosurg. 1967 27 428 432 6059415
- [11]Eaton W.P. Smith J.H. Micromachined Pressure Sensors: Review and Recent Developments Proceedings of the SPIE 3046, Smart Structures and Materials San Diego, CA, USA 19 June 1997 30 41
- [13]Katuri K.C. Asrani S. Ramasubramanian M.K. Intraocular pressure monitoring sensors IEEE Sens. J. 2008 8 12 19
- [14]Tan R. McClure T. Lin C.K. Jea D. Dabiri F. Massey T. Sarrafzadeh M. Srivastava M. Montemagno C.D. Schulam P. Development of a fully implantable wireless pressure monitoring system Biomed. Microdevices 2009 11 259 264 18836836
- [15]Sepúlveda A. Pontes A. Viana J. de Villoria R.G. Fachin F. Wardle B. Rocha L.A. Flexible Sensor for Blood Pressure Measurement Proceedings of the 2011 Annual International Conference of the IEEE on Engineering in Medicine and Biology Society, EMBC Boston, MA, USA 30 August–3 September 2011 512 515 22254360
- [16]Kotzar G. Freas M. Abel P. Fleischman A. Roy S. Zorman C. Moran J.M. Melzak J. Evaluation of MEMS materials of construction for implantable medical devices Biomaterials 2002 23 2737 2750 12059024
- [18]Jiang G. Design challenges of implantable pressure monitoring system Front. Neurosci. 2010 4 10.3389/neuro.20.002.2010 PMC2891810 20582269 · DOI
- [19]Axisa F. Jourand P. Lippens E. Rymarczyk-Machal M. de Smet N. Schacht E. Vanfleteren J. Puers R. Cornelissen R. Design and Fabrication of a Low Cost Implantable Bladder Pressure Monitor Proceedings of the Annual International Conference of the IEEE on Engineering in Medicine and Biology Society, EMBC 2009 Minneapolis, MN, USA 2–6 September 2009 4864 4867 19963629
- [20]Hannan M.A. Abbas S.M. Samad S.A. Hussain A. Modulation techniques for biomedical implanted devices and their challenges Sensors 2011 12 297 319 22368470 PMC3279214
- [22]Abouei J. Brown J.D. Plataniotis K.N. Pasupathy S. Energy efficiency and reliability in wireless biomedical implant systems IEEE Trans. Inf. Technol. Biomed. 2011 15 456 466 21233054
- [24]Lo B. Yang G.Z. Key technical challenges and current implementations of body sensor networks Proceedings of the 2nd International Workshop on Body Sensor Networks (BSN 2005) London, UK 1–5 April 2005
- [25]Cong P. Ko W.H. Young D.J. Wireless batteryless implantable blood pressure monitoring microsystem for small laboratory animals IEEE Sens. J. 2010 10 243 254
- [26]Cong P. Suster M.A. Chaimanonart N. Young D.J. Wireless Power Recharging for Implantable Bladder Pressure Sensor Proceedings of the 2009 IEEE on Sensors Christchurch, New Zealand 25–28 October 2009 1670 1673
- [27]Suster M. Young D. Wireless recharging of battery over large distance for implantable bladder pressure chronic monitoring Proceedings of the 16th International on Solid-State Sensors, Actuators and Microsystems Conference (TRANSDUCERS) Beijing, China 5–9 June 2011 1208 1211
- [29]Murphy O.H. Bahmanyar M.R. Borghi A. McLeod C.N. Navaratnarajah M. Yacoub M.H. Toumazou C. Continuous in vivo blood pressure measurements using a fully implantable wireless SAW sensor Biomed. Microdevices 2013 15 737 749 23559403 PMC3764326
- [30]Kilinc E.G. Moya A.C. van Lintel H. Renaud P. Maloberti F. Wang Q. Dehollain C. Remotely powered implantable heart monitoring system for freely moving animals Proceedings of the 2013 5th IEEE International Workshop on Advances in Sensors and Interfaces (IWASI) Bari, Italy 13–14 June 2013 10 13
- [31]Chow E.Y. Chlebowski A.L. Chakraborty S. Chappell W.J. Irazoqui P.P. Fully wireless implantable cardiovascular pressure monitor integrated with a medical stent IEEE Trans. Biomed. Eng. 2010 57 1487 1496 20172781
- [32]Potkay J.A. Long term, implantable blood pressure monitoring systems Biomed. Microdevices 2008 10 379 392 18095169
- [33]Jacobson S. Rothballer A. Prolonged Measurement of Experimental Intracranial Pressure Using a Subminiature Absolute Pressure Transducer* J. Neurosurg. 1967 26 603 608 6027445
- [34]Olsen E.R. Collins C.C. Loughborough W.F. Richards V. Adams J.E. Pinto D.W. Intracranial pressure measurement with a miniature passive implanted pressure transensor Am. J. Surg. 1967 113 727 729 4960729
- [35]Shulman K. Marmamou A. Analysis of intracranial pressure in hydrocephalus Dev. Med. Child Neurol. 1968 10 11 16 5733634
- [36]Nornes H. Serck-Hanssen F. Miniature transducer for intracranial pressure monitoring in man Acta Neurol. Scand. 1970 46 203 214 5431455
- [38]Lin K.M. Sant H.J. Ambati B.K. Gale B.K. Intraocular Pressure Sensors: New Approaches for Real-Time Intraocular Pressure Measurement Using A Purely Microfluidic Chip Proceedings of the 16th International Conference on Miniaturized Systems for Chemistry and Life Sciences, MicroTAS Okinawa, Japan 28 October–1 November 2012
- [42]Koley G. Liu J. Nomani M.W. Yim M. Wen X. Hsia T.Y. Miniaturized implantable pressure and oxygen sensors based on polydimethylsiloxane thin films Mater. Sci. Eng. C 2009 29 685 690 PMC2716561 20161216
- [43]Kim J. Lee H. Cha S. Choi B. Fabrication and evaluation of implantable pressure sensor using strain gauge Proceedings of the Biomedical Engineering International Conference (BMEiCON) Ubon Ratchathani, Thailand 5–7 December 2012 1 4
- [44]Poeggel S. Tosi D. Leen G. Lewis E. Fiber optic extrinsic FPI/FBG sensor for temperature-compensated pressure measurement in medical applications Proceedings of the Fifth European Workshop on Optical Fibre Sensors Krakow, Poland 19–22 May 2013 87940J-87940J-87946
- [45]Kim A. Powell C. Ziaie B. An Implantable Pressure Sensing System with Electromechanical Interrogation Scheme IEEE Trans. BioMed. Eng. 2014 61 2209 2217 24800754
- [46]Neumann J.J. Jr. Gabriel K.J. A fully-integrated CMOS-MEMS audio microphone Proceedings of the 12th International Conference on TRANSDUCERS, Solid-State Sensors, Actuators and Microsystems Boston, MA, USA 8–12 June 2003 230 233
- [47]Sheplak M. Seiner J.M. Breuer K.S. Schmidt M.A. A MEMS microphone for aeroacoustics measurements Proceedings of the 37th Aerospace Science Meeting and Exhibit AIAA 99 Reno, NV, USA 11 January 1999 11 14
- [48]Eccardt P.C. Niederer K. Scheiter T. Hierold C. Surface micromachined ultrasound transducers in CMOS technology Proceedings of the 1996 IEEE on Ultrasonics Symposium San Antonio, TX, USA 3–6 November 1996 959 962
- [49]Kottapalli A.G.P. Asadnia M. Miao J.M. Barbastathis G. Triantafyllou M.S. A flexible liquid crystal polymer MEMS pressure sensor array for fish-like underwater sensing Smart Mater. Struct. 2012 21 10.1088/0964-1726/21/11/115030 · DOI
- [51]Kalvesten E. Smith L. Tenerz L. Stemme G. The first surface micromachined pressure sensor for cardiovascular pressure measurements Proceedings of The Eleventh Annual International Workshop on Micro Electro Mechanical Systems Heidelberg, Germany 25–29 January 1998 574 579
- [54]Shin K.H. Moon C.Y. Lee T.H. Lim C.H. Kim Y.J. Implantable flexible wireless pressure sensor module Proceedings of the IEEE on Sensors Vienna, Austria 24–27 October 2004 844 847
- [55]Fonseca M.A. Allen M.G. Kroh J. White J. Flexible wireless passive pressure sensors for biomedical applications Proceedings of the Solid-State Sensor, Actuator, and Microsystems Workshop Hilton Head Island, SC, USA 4–8 June 2006
- [56]Yu L. Meng E. A microbubble pressure transducer with bubble nucleation core Proceedings of the 27th International Conference on Micro Electro Mechanical Systems (MEMS) San Francisco, CA, USA 26–30 January 2014 104 107
- [58]Ning M. Chronis N. An X-ray Detectable Pressure Microsensor for Monitoring Coronary In-Stent Restenosis Proceedings of the MEMS San Francisco, CA, USA 26–30 January 2014
- [59]Gutierrez C.A. Meng E. Low-cost carbon thick-film strain sensors for implantable applications J. Micromech. Microeng. 2010 20 10.1088/0960-1317/20/9/095028 · DOI
- [60]Lee C. Jug L. Meng E. High strain biocompatible polydimethylsiloxane-based conductive graphene and multiwalled carbon nanotube nanocomposite strain sensors Appl. Phys. Lett. 2013 102 10.1063/1.4804580 · DOI
- [61]Gutierrez C.A. Meng E. A Subnanowatt Microbubble Pressure Sensor Based on Electrochemical Impedance Transduction in a Flexible All-Parylene package Proceedings of the 24th International Conference on Micro Electro Mechanical Systems Cancun, Mexico 23–27 January 2011 549 552
- [62]Iberti T.J. Lieber C.E. Benjamin E. Determination of intra-abdominal pressure using a transurethral bladder catheter: Clinical validation of the technique Anesthesiology 1989 70 47 50 2912315
- [63]Weaver J.N. Alspaugh J.C. Behkam B. Toward a minimally invasive bladder pressure monitoring system: Model bladder for in vitro testing Proceedings of the 2010 3rd IEEE RAS and EMBS International Conference on Biomedical Robotics and Biomechatronics (BioRob) Tokyo, Japan 26–29 September 2010 638 643
- [64]Schein M. Wittmann D. Aprahamian C. Condon R. The abdominal compartment syndrome J. Am. Coll. Surg. 1995 180 745 7773495
- [65]Heijl A. Leske M.C. Bengtsson B. Hyman L. Bengtsson B. Hussein M. Reduction of intraocular pressure and glaucoma progression: Results from the Early Manifest Glaucoma Trial Arch. Ophthalmol. 2002 120 1268 1279 12365904
- [66]Narayan R.K. Kishore P.R. Becker D.P. Ward J.D. Enas G.G. Greenberg R.P. da Silva A.D. Lipper M.H. Choi S.C. Mayhall C.G. Intracranial pressure: To monitor or not to monitor? A review of our experience with severe head injury J. Neurosurg. 1982 56 650 659 7069477
- [67]Izzedine H. Launay-Vacher V. Deray G. Abnormal blood pressure circadian rhythm: A target organ damage? Int. J. Cardiol. 2006 107 343 349 16503256
- [68]Eide P.K. Bakken A. The baseline pressure of intracranial pressure (ICP) sensors can be altered by electrostatic discharges Biomed. Eng. Online 2011 10 10.1186/1475-925X-10-75 PMC3180435 21859487 · DOI
- [69]Czosnyka M. Harland S. Piechnik S. Kirkpatrick P. Czosnyka Z. Menon D. Systematic overestimation of intracranial pressure measured using a Camino pressure monitor J. Neurol. Neurosurg. Psychiatr. 1996 61 427 428 8890796 PMC486599
- [70]Leonardi A.D.C. Bir C.A. Ritzel D.V. VandeVord P.J. Intracranial pressure increases during exposure to a shock wave J. Neurotrauma 2011 28 85 94 21091267
- [71]Piper I. Barnes A. Smith D. Dunn L. The Camino intracranial pressure sensor: Is it optimal technology? An internal audit with a review of current intracranial pressure monitoring technologies Neurosurgery 2001 49 1158 1165 11846910
- [72]Al-Tamimi Y.Z. Helmy A. Bavetta S. Price S.J. Assessment of zero drift in the Codman intracranial pressure monitor: A study from 2 neurointensive care units Neurosurgery 2009 64 94 99 19145157
- [73]Kroin J.S. McCarthy R.J. Stylos L. Miesel K. Ivankovich A.D. Penn R.D. Long-term testing of an intracranial pressure monitoring device J. Neurosurg. 2000 93 852 858 11059668
- [74]Morgalla M.H. Mettenleiter H. Bitzer M. Fretschner R. Grote E.H. ICP measurement control: Laboratory test of 7 types of intracranial pressure transducers J. Med. Eng. Technol. 1999 23 144 151 10561825
- [75]Romner B. Grande P.O. Traumatic brain injury: Intracranial pressure monitoring in traumatic brain injury Nat. Rev. Neurol. 2013 9 185 186 23478466
- [76]Hamilton R. Baldwin K. Fuller J. Vespa P. Hu X. Bergsneider M. Intracranial pressure pulse waveform correlates with aqueductal cerebrospinal fluid stroke volume J. Appl. Physiol. (1985) 2012 113 1560 1566 22995390 PMC3524656
- [77]Ednick M.D. Pagala M. Barakat J.P. Nino G. Shah P. Cunningham J.N. Jr. Vaynblat M. Kazachkov M. Telemetric recording of intrapleural pressure J. Surg. Res. 2007 138 10 14 17084413
- [78]Czosnyka M. Pickard J.D. Monitoring and interpretation of intracranial pressure J. Neurol. Neurosurg. Psychiatr. 2004 75 813 821 15145991 PMC1739058
- [79]McLaren J.W. Brubaker R.F. Fitzsimon J.S. Continuous measurement of intraocular pressure in rabbits by telemetry Investig. Ophthalmol. Vis. Sci. 1996 37 966 975 8631640
- [80]Symon L. Dorsch N. Stephens R. Pressure waves in so-called low-pressure hydrocephalus Lancet 1972 300 1291 1292 4117815
- [81]Realini T. Barber L. Burton D. Frequency of asymmetric intraocular pressure fluctuations among patients with and without glaucoma Ophthalmology 2002 109 1367 1371 12093664
- [82]Bray R. Sherwood A. Halter J. Robertson C. Grossman R. Development of a clinical monitoring system by means of ICP waveform analysis Intracranial Pressure VI Springer Berlin, Germany 1986 260 264
- [83]Fletter P.C. Majerus S. Cong P. Damaser M.S. Ko W.H. Young D.J. Garverick S.L. Wireless micromanometer system for chronic bladder pressure monitoring Proceedings of the 2009 Sixth International Conference on Networked Sensing Systems (INSS) Pittsburgh, PA, USA 17–19 June 2009 1 4
- [84]Coleman T.J. Thomsen J.C. Maass S.D. Hsu Y. Nygaard I.E. Hitchcock R.W. Development of a wireless intra-vaginal transducer for monitoring intra-abdominal pressure in women Biomed. Microdevices 2012 14 347 355 22147020 PMC3303945
- [85]Cooper R. Beale D. Constable I. Grose G. Continual monitoring of intraocular pressure: Effect of central venous pressure, respiration, and eye movements on continual recordings of intraocular pressure in the rabbit, dog, and man Bri. J. Ophthalmol. 1979 63 799 804 PMC1043629 526459
- [86]Yokota T. Kuribara K. Tokuhara T. Zschieschang U. Klauk H. Takimiya K. Sadamitsu Y. Hamada M. Sekitani T. Someya T. Flexible Low-Voltage Organic Transistors with High Thermal Stability at 250 °C Adv. Mater. 2013 25 3639 3644 23616376
- [87]Kuribara K. Wang H. Uchiyama N. Fukuda K. Yokota T. Zschieschang U. Jaye C. Fischer D. Klauk H. Yamamoto T. Organic transistors with high thermal stability for medical applications Nat. Commun. 2012 3 10.1038/ncomms1721 22395614 · DOI
- [88]Matlaga B.F. Yasenchak L.P. Salthouse T.N. Tissue response to implanted polymers: The significance of sample shape J. Biomed. Mater. Res. 1976 10 391 397 1270456
- [89]Cleven N.J. Muntjes J. Fassbender H. Urban U. Gortz M. Vogt H. Grafe M. Göttsche T. Penzkofer T. Schmitz-Rode T. A novel fully implantable wireless sensor system for monitoring hypertension patients IEEE Trans. Biomed. Eng. 2012 59 3124 3130 22955864
- [90]Voskerician G. Shive M.S. Shawgo R.S. Recum H.V. Anderson J.M. Cima M.J. Langer R. Biocompatibility and biofouling of MEMS drug delivery devices Biomaterials 2003 24 1959 1967 12615486
- [91]IEEE Standards Coordinating Committee on Non-Ionizing Radiation Hazards IEEE Standard for Safety Levels with Respect to Human Exposure to Radio Frequency Electromagnetic Fields, 3kHz to 300 GHz Institute of Electrical and Electonics Engineers, Incorporated New York, NY, USA 1992
- [92]Guideline I. Guidelines for limiting exposure to time-varying electric, magnetic, and electromagnetic fields (up to 300 GHz) Health Phys. 1998 74 494 522 9525427
- [93]Aquilina K. Thoresen M. Chakkarapani E. Pople I.K. Coakham H.B. Edwards R.J. Preliminary evaluation of a novel intraparenchymal capacitive intracranial pressure monitor: Laboratory investigation J. Neurosurg. 2011 115 561 569 21619410
- [94]Kotecha A. White E. Schlottmann P.G. Garway-Heath D.F. Intraocular Pressure Measurement Precision with the Goldmann Applanation, Dynamic Contour, and Ocular Response Analyzer Tonometers Ophthalmology 2010 117 730 737 20122737
- [95]Chihara E. Assessment of true intraocular pressure: The gap between theory and practical data Surv. Ophthalmol. 2008 53 203 218 18501267
- [96]Schäfer W. Abrams P. Liao L. Mattiasson A. Pesce F. Spangberg A. Sterling A.M. Zinner N.R. Kerrebroeck P.v. Good urodynamic practices: Uroflowmetry, filling cystometry, and pressure-flow studies** Neurourol. Urodyn. 2002 21 261 274 11948720
- [97]Arriada N. Sotelo J. Review: Treatment of hydrocephalus in adults Surg. Neurol. 2002 58 377 384 12517612
- [98]Auer L.M. Sayama I. Intracranial pressure oscillations (B-waves) caused by oscillations in cerebrovascular volume Acta Neurochir. 1983 68 93 100 6858734
- [99]Zhong J. Dujovny M. Park H.K. Perez E. Perlin A.R. Diaz F.G. Advances in ICP monitoring techniques Neurol. Res. 2003 25 339 350 12870259
- [100]Czosnyka M. Czosnyka Z. Pickard J.D. Laboratory testing of three intracranial pressure microtransducers: Technical report Neurosurgery 1996 38 219 224 8747977
- [101]Bavetta S. Norris J.S. Wyatt M. Sutcliffe J.C. Hamlyn P.J. Prospective study of zero drift in fiberoptic pressure monitors used in clinical practice J. Neurosurg. 1997 86 927 930 9171170
- [102]Springer F. Gunther R.W. Schmitz-Rode T. Aneurysm Sac pressure measurement with minimally invasive implantable pressure sensors: An alternative to current surveillance regimes after EVAR? Cardiovasc. Interv. Radiol. 2008 31 460 467 18087767
- [103]Van der Wel M.C. Buunk I.E. van Weel C. Thien T.A. Bakx J.C. A novel approach to office blood pressure measurement: 30-minute office blood pressure vs. daytime ambulatory blood pressure Ann. Fam. Med. 2011 9 128 135 21403139 PMC3056860
- [104]Ellozy S.H. Carroccio A. Lookstein R.A. Minor M.E. Sheahan C.M. Juta J. Cha A. Valenzuela R. Addis M.D. Jacobs T.S. First experience in human beings with a permanently implantable intrasac pressure transducer for monitoring endovascular repair of abdominal aortic aneurysms J. Vasc. Surg. 2004 40 405 412 15337865
- [106]Wille S. Schumacher P. Paas J. Tenholte D. Eminaga O. Muller U. Muthen N. Mehner J. Cornely O. Engelmann U. Catheterless long-term ambulatory urodynamic measurement using a novel three-device system PLoS One 2014 9 e96280 24840482 PMC4026131
- [107]Gupta A. Defreitas G. Lemack G.E. The reproducibility of urodynamic findings in healthy female volunteers: Results of repeated studies in the same setting and after short-term follow-up Neurourol. Urodyn. 2004 23 311 316 15227647
- [108]Majerus S.J.A. Garverick S.L. Suster M.A. Fletter P.C. Damaser M.S. Wireless, Ultra-Low-Power Implantable Sensor for Chronic Bladder Pressure Monitoring ACM J. Emerg. Technol. Comput. Syst. 2012 8 1 13 PMC4712728 26778926
- [109]Meslier N. Simon I. Kouatchet A. Ouksel H. Person C. Racineux J.L. Validation of a suprasternal pressure transducer for apnea classification during sleep SLEEP 2002 25 753 757 12405611
- [110]Ginggen A. Tardy Y. Crivelli R. Bork T. Renaud P. A telemetric pressure sensor system for biomedical applications IEEE Trans. Biomed. Eng. 2008 55 1374 1381 18390328
- [111]Majerus S.J. Fletter P.C. Damaser M.S. Garverick S.L. Low-power wireless micromanometer system for acute and chronic bladder-pressure monitoring IEEE Trans. Biomed. Eng. 2011 58 763 767 20934942 PMC8514143
- [113]Siwapornsathain E. Lal A. Binard J. A telemetry and sensor platform for ambulatory urodynamics Proceedings of the Biology 2nd Annual International IEEE-EMB Special Topic Conference on Microtechnologies in Medicine Madison, WI, USA 2–4 May 2002 283 287
- [114]Takahata K. DeHennis A. Wise K.D. Gianchandani Y.B. A wireless microsensor for monitoring flow and pressure in a blood vessel utilizing a dual-inductor antenna stent and two pressure sensors Proceedings of the 17th IEEE International Conference on Micro Electro Mechanical Systems Maastricht, The Netherlands 25–29 January 2004 216 219
- [115]Hierold C. Clasbrumme B. Behrend D. Scheiter T. Steger M. Oppermann K. Kapels H. Landgraf E. Wenzel D. Etuodt D. Implantable low power integrated pressure sensor system for minimal invasive telemetric patient monitoring Proceedings of the The 11th Annual International Workshop on Micro Electro Mechanical Systems Heidelberg, Germany 25–29 January 1998 568 573
- [118]Wang C.C. Huang C.C. Liou J.S. Ciou Y.J. Huang I.Y. Li C.P. Lee Y.C. Wu W.J. A mini-invasive long-term bladder urine pressure measurement ASIC and system IEEE Trans. Biomed. Circ. Syst. 2008 2 44 49 23852632
- [119]Eom J. Lee J. Lee H. Choi B. Energy harvesting for bladder pressure sensor using parametric amplification phenomenon of PVDF bimorph cantilever Proceedings of the 2013 6th Biomedical Engineering International Conference (BMEiCON) Amphur Muang, Thailand 23–25 Ocotber 2013 1 3
- [120]Chen G. Ghaed H. Haque R.U. Wieckowski M. Kim Y. Kim G. Fick D. Kim D. Seok M. Wise K. A cubic-millimeter energy-autonomous wireless intraocular pressure monitor Proceedings of the 2011 IEEE International on Solid-State Circuits Conference Digest of Technical Papers (ISSCC) San Francisco, CA, USA 20–24 February 2011 310 312
- [121]Liang B. Fang L. Tu C. Zhou C. Wang X. Wang Q. Wang P. Ye X. A novel implantable saw sensor for blood pressure monitoring Proceedings of the 16th International on Solid-State Sensors, Actuators and Microsystems Conference (TRANSDUCERS) Beijing, China 5–9 June 2011 2184 2187
- [122]Li C. Wu P.M. Shutter L.A. Narayan R.K. Dual-mode operation of flexible piezoelectric polymer diaphragm for intracranial pressure measurement Appl. Phys. Lett. 2010 96 10.1063/1.3299003 · DOI
- [124]Lee S.A. Vasquez D.J. Bergsneider M. Judy J.W. Magnetic microactuators for MEMS-enabled ventricular catheters for hydrocephalus Proceedings of the 28th Annual International Conference of the IEEE on Engineering in Medicine and Biology Society New York, NY, USA 31 August–1 September 2006 2494 2497 17946960
- [126]Someya T. Sekitani T. Iba S. Kato Y. Kawaguchi H. Sakurai T. A large-area, flexible pressure sensor matrix with organic field-effect transistors for artificial skin applications Proc. Natl. Acad. Sci. USA 2004 101 9966 9970 15226508 PMC454198
- [127]Gutierrez C.A. McCarty C. Kim B. Pahwa M. Meng E. An Implantable All-Parylene Liquid-Impedance Based MEMS Force Sensor Proceedings of the 23rd IEEE International Conference on Micro Electro Mechanical Systems Wanchai, Hong Kong 24–28 January 2010 600 603
- [128]Allen M.G. Microfabricated implantable wireless microsystems: Permanent and biodegradable implementations Proceedings of the 2014 IEEE 27th International Conference on Micro Electro Mechanical Systems (MEMS) San Francisco, CA, USA 26–30 January 2014 1 4
- [129]Tsang M. Armutlulu A. Martinez A. Herrault F. Allen S.A.B. Allen M.G. A MEMS-enabled biodegradable battery for powering transient implantable devices Proceedings of the 2014 IEEE 27th International Conference on Micro Electro Mechanical Systems (MEMS) San Francisco, CA, USA 26–30 January 2014 358 361
- [130]Yang G.Z. Yacoub M. Body Sensor Networks Springer Berlin, Germany 2006
Cited by
Sorted by citations
Emerging flexible and wearable physical sensing platforms for healthcare and biomedical applications
Kenry, Yeo, J. C., & Lim, C. T. (2016) · Microsystems & Nanoengineering · doi:10.1038/micronano.2016.43 · Cited by 550 · see citing works
Shin, J., Yan, Y., Bai, W., Xue, Y., Gamble, P., Tian, L., Kandela, I., Haney, C. R., Spees, W., Lee, Y., et al. (2018) · Nature Biomedical Engineering · doi:10.1038/s41551-018-0300-4 · Cited by 263 · see citing works
Comparative analysis of energy transfer mechanisms for neural implants
Miziev, S., Pawlak, W. A., & Howard, N. (2024) · Frontiers in Neuroscience · doi:10.3389/fnins.2023.1320441 · Cited by 21 · see citing works
- Authors
- Yu, L., Kim, B., & Meng, E.
- Affiliations
- University of Southern California (US)
- Corresponding author
- Meng, Ellis
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2014-10-31
- Volume
- 14
- Issue
- 11
- Pages
- 20620-20644
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC4279503
- OpenAlex
- W2037884469
- Cited by
- 200
- References
- 2
- Citations by year
- 2026: 13 · 2025: 16 · 2024: 16 · 2023: 15 · 2022: 9 · 2021: 19 · 2020: 24 · 2019: 20 · 2018: 26 · 2017: 19 · 2016: 17 · 2015: 6
- FWCI
- 1.20
- Type
- journal article
- Language
- en
- Open access
- gold
- dx.doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- National Science Foundation (1332394, 1231994)
- University of Southern California
- Division of Electrical, Communications and Cyber Systems (ECCS-1231994)
- Division of Emerging Frontiers in Research and Innovation (EFRI-1332394)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- implantable pressure sensor, biocompatibility, telemetry, surgical implantation, patient comfort, internal pressure, medical device design
- MeSH terms
- Equipment Design, Manometry, Telemetry, Transducers, Pressure, Equipment Failure Analysis, Prostheses and Implants
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review