Very High Bit Rate Near-Field Communication with Low-Interference Coils and Digital Single-Bit Sampling Transceivers for Biomedical Sensor Systems

Open access

The DOI matches its Crossref record; 32 of the 34 references it cites were found in a registry or an index.

Abstract

The evolution of microelectronics increased the information acquired by today’s biomedical sensor systems to an extent where the capacity of low-power communication interfaces becomes one of the central bottlenecks. Hence, this paper mathematically analyzes and experimentally verifies novel coil and transceiver topologies for near-field communication interfaces, which simultaneously allow for high data transfer rates, low power consumption, and reduced interference to nearby wireless power transfer interfaces. Data coil design is focused on presenting two particular topologies which provide sufficient coupling between a reader and a wireless sensor system, but do not couple to an energy coil situated on the same substrate, severely reducing interference between wireless data and energy transfer interfaces. A novel transceiver design combines the approaches of a minimalistic analog front-end with a fully digital single-bit sampling demodulator, in which rectangular binary signals are processed by simple digital circuits instead of sinusoidal signals being conditioned by complex analog mixers and subsequent multi-bit analog-to-digital converters. The concepts are implemented using an analog interface in discrete circuit technology and a commercial low-power field-programmable gate array, yielding a transceiver which supports data rates of up to 6.78 MBit/s with an energy consumption of just 646 pJ/bit in transmitting mode and of 364 pJ/bit in receiving mode at a bit error rate of 2×10−7, being 10 times more energy efficient than any commercial NFC interface and fully implementable without any custom CMOS technology.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.3390/s20216025

Cites

32 of 32 references verified0 citation errors · 0 not checkable

Statistics

Figures

Figure 1 Physical and electrical representations of a near-field system.
Figure 1 Physical and electrical representations of a near-field system.CC BY · PubMed Central
Figure 2 System concept of this work: A separate inductive data communication interface is implemented by two additional coils, which can be subject to an individual optimization. The near-field communication interface is based on an active communication with digital modulators and demodulators, so that data are modulated on a rectangular wave, transmitted over an analog channel and converted back to a rectangular wave by a comparator stage in the receiver, where it gets demodulated by a fully digital circuit.
Figure 2 System concept of this work: A separate inductive data communication interface is implemented by two additional coils, which can be subject to an individual optimization. The near-field communication interface is based on an active communication with digital modulators and demodulators, so that data are modulated on a rectangular wave, transmitted over an analog channel and converted back to a rectangular wave by a comparator stage in the receiver, where it gets demodulated by a fully digital circuit.CC BY · PubMed Central
Figure 3 Data coil topologies to minimize mutual inductance to an energy coil on the same substrate and magnetic flux density distribution Bz of a planar spiral coil from [28]. By incorporating flux components of opposite sign, the effective flux in the data coil structure can be minimized. Zero coupling can be achieved by appropriate selection of rc2,i and sb,i.
Figure 3 Data coil topologies to minimize mutual inductance to an energy coil on the same substrate and magnetic flux density distribution Bz of a planar spiral coil from [28]. By incorporating flux components of opposite sign, the effective flux in the data coil structure can be minimized. Zero coupling can be achieved by appropriate selection of rc2,i and sb,i.CC BY · PubMed Central
Figure 4 Positional analysis of the data coils’ mutual inductance and the cross-coupling to the energy coils.
Figure 4 Positional analysis of the data coils’ mutual inductance and the cross-coupling to the energy coils.CC BY · PubMed Central
Figure 5 Equivalent circuit of the receiver’s resonator consisting of the data coil Lc, a series resistance Rc, a parallel tuning capacitor Cp, and the load resistance RL representing the receiving amplifier with additional impedance matching circuits.
Figure 5 Equivalent circuit of the receiver’s resonator consisting of the data coil Lc, a series resistance Rc, a parallel tuning capacitor Cp, and the load resistance RL representing the receiving amplifier with additional impedance matching circuits.CC BY · PubMed Central
Figure 6 Setup of the overall coil system: A wireless power transfer link is realized by two planar spiral coils, while the data link consists of two additional butterfly-shaped coils.
Figure 6 Setup of the overall coil system: A wireless power transfer link is realized by two planar spiral coils, while the data link consists of two additional butterfly-shaped coils.CC BY · PubMed Central
Figure 7 Analog front-end of the custom NFC transceiver.
Figure 7 Analog front-end of the custom NFC transceiver.CC BY · PubMed Central
Figure 8 Frequency behavior of voltage transfer function and input impedance of the matching and filtering network.
Figure 8 Frequency behavior of voltage transfer function and input impedance of the matching and filtering network.CC BY · PubMed Central
Figure 9 System architecture of the NFC transceiver’s digital circuitry implemented in an FPGA.
Figure 9 System architecture of the NFC transceiver’s digital circuitry implemented in an FPGA.CC BY · PubMed Central
Figure 10 Signals and circuits of the custom NFC transmitter.
Figure 10 Signals and circuits of the custom NFC transmitter.CC BY · PubMed Central
Figure 11 Block diagram as well as data and control signals of the fully digital BPSK demodulator.
Figure 11 Block diagram as well as data and control signals of the fully digital BPSK demodulator.CC BY · PubMed Central
Figure 12 Schematic diagram of the BSPK demodulator core with sub-elements to perform carrier synchronization, down-conversion (mixing and filtering) as well as symbol mapping (comparator).
Figure 12 Schematic diagram of the BSPK demodulator core with sub-elements to perform carrier synchronization, down-conversion (mixing and filtering) as well as symbol mapping (comparator).CC BY · PubMed Central
Figure 13 A sinusoidal RF input signal uc,in(t) is sampled at eight points per period. As the voltage is subject to phase and amplitude noise, the acquired voltage shows a certain probability to deviate from the nominal value, expressed by a probability density function p(uc,in) being indicated by the Gaussian curves at any sampling point. Due to the hysteresis thresholds ±UH/2 of the comparator, illustrated by the two dashed lines, the probability of a value xi being sampled as ’0’ or as ’1’ is expressed by a conditional probability P(xi|xi−1) is taking the previous value xi−1 into account. The graph is plotted for u^c,in=20mV, σP=2π×0.03, σA=2mV and UH=10mV.
Figure 13 A sinusoidal RF input signal uc,in(t) is sampled at eight points per period. As the voltage is subject to phase and amplitude noise, the acquired voltage shows a certain probability to deviate from the nominal value, expressed by a probability density function p(uc,in) being indicated by the Gaussian curves at any sampling point. Due to the hysteresis thresholds ±UH/2 of the comparator, illustrated by the two dashed lines, the probability of a value xi being sampled as ’0’ or as ’1’ is expressed by a conditional probability P(xi|xi−1) is taking the previous value xi−1 into account. The graph is plotted for u^c,in=20mV, σP=2π×0.03, σA=2mV and UH=10mV.CC BY · PubMed Central
Figure 14 The possible sequences of samples Xn acquired for one period can be represented as a binary tree. The transition probabilities are given by the conditional probabilities P(xi|xi−1).
Figure 14 The possible sequences of samples Xn acquired for one period can be represented as a binary tree. The transition probabilities are given by the conditional probabilities P(xi|xi−1).CC BY · PubMed Central
Figure 15 Prototype coil systems manufactured on an FR4 substrate. The data coil structures include resistive and capacitive tuning elements, the energy coils are realized as capacitively segmented coils.
Figure 15 Prototype coil systems manufactured on an FR4 substrate. The data coil structures include resistive and capacitive tuning elements, the energy coils are realized as capacitively segmented coils.CC BY · PubMed Central
Figure 16 Prototype circuits implementing the custom NFC transceiver: A shield including the analog front-end and the coil connectors is mounted on a general purpose control unit including DC/DC converters, an RF microcontroller, and a low-power FPGA. The net area required for the NFC transceiver is enclosed by the colored contours.
Figure 16 Prototype circuits implementing the custom NFC transceiver: A shield including the analog front-end and the coil connectors is mounted on a general purpose control unit including DC/DC converters, an RF microcontroller, and a low-power FPGA. The net area required for the NFC transceiver is enclosed by the colored contours.CC BY · PubMed Central
Figure 17 Measurement setup to determine the S-parameters of the coil systems in air and pork tissue environment.
Figure 17 Measurement setup to determine the S-parameters of the coil systems in air and pork tissue environment.CC BY · PubMed Central
Figure 18 Frequency behavior of the effective coil resistances Ri and the mutual reactance X12 for the given data coil system at a coil separation distance of d34=20mm in air and pork muscle tissue.
Figure 18 Frequency behavior of the effective coil resistances Ri and the mutual reactance X12 for the given data coil system at a coil separation distance of d34=20mm in air and pork muscle tissue.CC BY · PubMed Central
Figure 19 Voltage at the comparator of the receiving stage uc,in versus mutual inductance M34 of the data coils (in air). The simulated data was provided by a Harmonic Balance simulation using Keysight Advanced Design System 2016, the measurement was performed with the prototype circuits presented above.
Figure 19 Voltage at the comparator of the receiving stage uc,in versus mutual inductance M34 of the data coils (in air). The simulated data was provided by a Harmonic Balance simulation using Keysight Advanced Design System 2016, the measurement was performed with the prototype circuits presented above.CC BY · PubMed Central
Figure 20 Transient behavior of important analog and digital signals within the near-field interface.
Figure 20 Transient behavior of important analog and digital signals within the near-field interface.CC BY · PubMed Central
Figure 21 Required resources of the NFC transceiver implemented in a Lattice iCE40UP5K FPGA. ^(*) The BPSK Demodulator also includes components of the RX Packet Handler.
Figure 21 Required resources of the NFC transceiver implemented in a Lattice iCE40UP5K FPGA. ^(*) The BPSK Demodulator also includes components of the RX Packet Handler.CC BY · PubMed Central
Figure 22 Bit error rate versus coil separation distance d34 in simulation and measurement to characterize the complete link. The simulated data are obtained from the mathematical description of this work and the parameters σP=2π×0.04, σA=8mV and UH=20mV; the experimental values of the BER include the measurement with five different receiver PCBs and 1.2×10⁸ bits of a PRBS-15 sequence each.
Figure 22 Bit error rate versus coil separation distance d34 in simulation and measurement to characterize the complete link. The simulated data are obtained from the mathematical description of this work and the parameters σP=2π×0.04, σA=8mV and UH=20mV; the experimental values of the BER include the measurement with five different receiver PCBs and 1.2×10⁸ bits of a PRBS-15 sequence each.CC BY · PubMed Central
Figure 23 BER vs. comparator input voltage for varying hysteresis, phase noise or amplitude noise parameters according to the model of Section 6. The nominal values of the blue curves are σP=2π×0.04, σA=8mV and UH=20mV. The measurement results of Figure 22 were related to the corresponding input voltage u^in and are shown in black color as a reference for the nominal noise parameters.
Figure 23 BER vs. comparator input voltage for varying hysteresis, phase noise or amplitude noise parameters according to the model of Section 6. The nominal values of the blue curves are σP=2π×0.04, σA=8mV and UH=20mV. The measurement results of Figure 22 were related to the corresponding input voltage u^in and are shown in black color as a reference for the nominal noise parameters.CC BY · PubMed Central
Figure 24 Cross-coupling analysis of data and energy coils in pork tissue environment: The ratio of M14(40.68MHz) and M34(13.56MHz) is shown versus coil distance d and lateral misalignment l, the measurement values on a grid with 5 mm resolution are shown as a numerical reference.
Figure 24 Cross-coupling analysis of data and energy coils in pork tissue environment: The ratio of M14(40.68MHz) and M34(13.56MHz) is shown versus coil distance d and lateral misalignment l, the measurement values on a grid with 5 mm resolution are shown as a numerical reference.CC BY · PubMed Central
Figure 25 Measured voltage transfer function of the primary impedance matching network.
Figure 25 Measured voltage transfer function of the primary impedance matching network.CC BY · PubMed Central

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 32 references

Research works printed but not linked: 2

  1. Rozgic, 2019, A 0.338 cm3, Artifact-Free, 64-Contact Neuromodulation Platform for Simultaneous Stimulation and Sensing, IEEE Trans. Biomed. Circuits Syst., 13, 38
  2. Pozar, D. (2012). Microwave Engineering, John Wiley & Sons. [4th ed.].

Cited by

See all 12

Metadata

Authors
Stoecklin, S., Rosch, E., Yousaf, A., & Reindl, L.
Affiliations
University of Freiburg (DE)
Corresponding author
Stoecklin, Sebastian
Venue
Sensors
Publisher
MDPI AG
Published
2020-10-23
Volume
20
Issue
21
Pages
6025
ISSN
1424-8220 (ISSN-L)
OpenAlex
W3094138147
Cited by
12
References
34
Citations by year
2025: 2 · 2024: 4 · 2023: 2 · 2022: 2 · 2021: 1 · 2020: 1
FWCI
0.51
Type
journal article
Language
en
Open access
Locations
  • Sensors (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
Funders
  • Fritz Huettinger Stiftung (Fritz Huettinger-Stipendium)
  • Deutsche Forschungsgemeinschaft (EXC1086)
  • Albert-Ludwigs-Universität Freiburg (Open Access Publishing)
Topics
  • Physical Sciences › Engineering › Electrical and Electronic Engineering › Wireless Power Transfer Systems
  • Physical Sciences › Computer Science › Computer Networks and Communications › RFID technology advancements
  • Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
Keywords
near field communication, wireless power transfer, interference suppression, low-power transceiver, energy efficiency, magnetic coupling, FPGA
MeSH terms
Communication, Energy Transfer, Equipment Design, Telemetry, Biosensing Techniques, Wireless Technology

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Very High Bit Rate Near-Field Communication with Low-Interference Coils and Digital Single-Bit Sampling Transceivers for Biomedical Sensor Systems. https://app.cochat.ai/papers/10.3390/s20216025
The paper
Stoecklin, S., Rosch, E., Yousaf, A., & Reindl, L. (2020). Very High Bit Rate Near-Field Communication with Low-Interference Coils and Digital Single-Bit Sampling Transceivers for Biomedical Sensor Systems. Sensors. https://doi.org/10.3390/s20216025

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.