Bias in clinical chemistry
Open accessAbstract
Clinical chemistry uses automated measurement techniques and medical knowledge in the interest of patients and healthy subjects. Automation has reduced repeatability and day-to-day variation considerably. Bias has been reduced to a lesser extent by reference measurement systems. It is vital to minimize clinically important bias, in particular bias within conglomerates of laboratories that measure samples from the same patients. Small and variable bias components will over time show random error properties and conventional random-error based methods for calculating measurement uncertainty can then be applied. The present overview of bias presents the general principles of error and uncertainty concepts, terminology and analysis, and suggests methods to minimize bias and measurement uncertainty in the interest of healthcare.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Even after the advent of accreditation standards ISO 17025 and 15189 demanding uncertainty calculations, top-down variance component analysis used in error/frequentist statistics are commonly used for uncertainty estimation using internal quality control samples.
“Even after the advent of accreditation standards ISO 17025 and 15189 demanding uncertainty calculations, top-down variance component analysis used in error/frequentist statistics are commonly used for uncertainty estimation using internal quality control samples.75,76”
From Uncertainty in Measurement and Total Error (2017) · cites this paper as [76]
Split into 2 atoms — 0 confirmed, 2 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Introduction to Statistical Quality Control
Sauter, R., & Montgomery, D. C. (1992) · Technometrics · doi:10.2307/1269251 · Cited by 7,881 · see citing works
GUIDE TO THE EXPRESSION OF UNCERTAINTY IN MEASUREMENT
Standard, S. (2006) · Cited by 6,460 · see citing works
Searle, S. R., Casella, G., & McCulloch, C. E. (1992) · Wiley Series in Probability and Statistics · doi:10.1002/9780470316856 · Cited by 1,926 · see citing works
Beyond Normality: The Predictive Value and Efficiency of Medical Diagnoses.
(1977) · Annals of Internal Medicine · doi:10.7326/0003-4819-86-4-513_1 · Cited by 1,377 · see citing works
Quantifying Uncertainty in Analytical Measurement
Silva, R. A., Bremser, W., Brzyski, A., Kaarls, R., Kaus, R., Magnusson, B., Amico, D. E., Robouch, P. B., Rösslein, M., van der, V. A., et al. (2012) · Cited by 1,141 · see citing works
Current databases on biological variation: pros, cons and progress
C. Ricós, V. A. (1999) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365519950185229 · Cited by 987 · see citing works
Bonini, P., Plebani, M., Ceriotti, F., & Rubboli, F. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.5.691 · Cited by 850 · see citing works
ROC Curves for Continuous Data
Krzanowski, W. J., & Hand, D. J. (2009) · Chapman & Hall/CRC eBooks · doi:10.1201/9781439800225 · Cited by 669 · see citing works
Introduction to Statistical Quality Control
Freund, R. A. (1987) · Journal of Quality Technology · doi:10.1080/00224065.1987.11979072 · Cited by 472 · see citing works
The use of Control Charts in the Clinical Laboratory*
Levey, S., & Jennings, E. R. (1950) · American Journal of Clinical Pathology · doi:10.1093/ajcp/20.11_ts.1059 · Cited by 361 · see citing works
Understanding the meaning of accuracy, trueness and precision
Menditto, A., Patriarca, M., & Magnusson, B. (2006) · Accreditation and Quality Assurance · doi:10.1007/s00769-006-0191-z · Cited by 342 · see citing works
Estimation of Variance Components and Applications.
Venables, W. N., Rao, C. R., & Kleffe, J. (1990) · Biometrics · doi:10.2307/2531661 · Cited by 313 · see citing works
The Evidence Base of Clinical Diagnosis
(2008) · Wiley eBooks · doi:10.1002/9781444300574 · Cited by 276 · see citing works
Criteria for judging precision and accuracy in method development and evaluation
Westgard, J. O., Carey, R. N., & Wold, S. (1974) · Clinical Chemistry · doi:10.1093/clinchem/20.7.825 · Cited by 236 · see citing works
Roadmap for Harmonization of Clinical Laboratory Measurement Procedures
Miller, W. G., Myers, G. L., Gantzer, M. L., Kahn, S. E., Schönbrunner, E. R., Thienpont, L. M., Bunk, D. M., Christenson, R. H., Eckfeldt, J. H., Lo, S. F., et al. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2011.164012 · Cited by 213 · see citing works
Miller, W. G., Myers, G. L., & Rej, R. (2006) · Clinical Chemistry · doi:10.1373/clinchem.2005.063511 · Cited by 178 · see citing works
Uncertainty of measurement: implications of its use in analytical science
Analytical Methods Committee (1995) · The Analyst · doi:10.1039/an9952002303 · Cited by 176 · see citing works
Harmonization in laboratory medicine: the complete picture
Plebani, M. (2013) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2013-0075 · Cited by 174 · see citing works
Traceability in Laboratory Medicine
Vesper, H. W., & Thienpont, L. M. (2009) · Clinical Chemistry · doi:10.1373/clinchem.2008.107052 · Cited by 169 · see citing works
Reference materials and commutability.
Vesper, H. W., Miller, W. G., & Myers, G. L. (2007) · PubMed Central · Cited by 159 · see citing works
Estimation of Variance Components and Applications
Bremer, R. H., Rao, C. R., Kleffe, J., & Kleffe, J. (1989) · Technometrics · doi:10.1080/00401706.1989.10488613 · Cited by 150 · see citing works
Koehler, T. L., & Natrella, M. G. (1965) · Technometrics · doi:10.1080/00401706.1965.10490307 · Cited by 139 · see citing works
Quality Control in Analytical Chemistry.
Fearn, T., Kateman, G., & Buydens, L. M. C. (1994) · Journal of the Royal Statistical Society Series A (Statistics in Society) · doi:10.2307/2983539 · Cited by 136 · see citing works
Traceability as a unique tool to improve standardization in laboratory medicine
Panteghini, M. (2009) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2008.09.098 · Cited by 104 · see citing works
Estimating uncertainties of analytical results using information from the validation process
Maroto, A., Riu, J., Boqué, R., & Rius, F. X. (1999) · Analytica Chimica Acta · doi:10.1016/s0003-2670(99)00111-7 · Cited by 100 · see citing works
Treatment of uncorrected measurement bias in uncertainty estimation for chemical measurements
Magnusson, B., & Ellison, S. L. R. (2007) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-007-1693-1 · Cited by 91 · see citing works
Guidelines for expressing the uncertainty of measurement results containing uncorrected bias
Phillips, S. D., Eberhardt, K. R., & Parry, B. (1997) · Journal of Research of the National Institute of Standards and Technology · doi:10.6028/jres.102.039 · Cited by 84 · see citing works
Simundic, A.-M., Cornes, M., Grankvist, K., Lippi, G., Nybo, M., Kovalevskaya, S., Sprongl, L., Sumarac, Z., & Church, S. (2013) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2013-0283 · Cited by 84 · see citing works
Armbruster, D., & Miller, R. R. (2007) · PubMed · Cited by 82 · see citing works
Miller, W. G., & Myers, G. L. (2013) · Clinical Chemistry · doi:10.1373/clinchem.2013.208785 · Cited by 80 · see citing works
Harmonization: the Sample, the Measurement, and the Report
Miller, W. G., Tate, J. R., Barth, J. H., & Jones, G. R. D. (2014) · Annals of Laboratory Medicine · doi:10.3343/alm.2014.34.3.187 · Cited by 80 · see citing works
Cox, D. R., & Solomon, P. J. (2002) · Chapman & Hall/CRC eBooks · doi:10.1201/9781482285949 · Cited by 80 · see citing works
Manual of Standardized Procedures for Spectrophotometric Chemistry
Levey, S. (1950) · American Journal of Clinical Pathology · doi:10.1093/ajcp/20.9.854a · Cited by 78 · see citing works
Richtlinie der Bundesärztekammer zur Qualitätssicherung laboratoriumsmedizinischer Untersuchungen
Schumann, G. (2019) · Springer Reference Medizin · doi:10.1007/978-3-662-48986-4_2697 · Cited by 78 · see citing works
Panteghini, M. (2009) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2010.020 · Cited by 76 · see citing works
Harmonization of laboratory testing — Current achievements and future strategies
Tate, J. R., Johnson, R. N., Barth, J. H., & Panteghini, M. (2013) · Clinica Chimica Acta · doi:10.1016/j.cca.2013.08.021 · Cited by 76 · see citing works
Setting Performance Goals and Evaluating Total Analytical Error for Diagnostic Assays
Krouwer, J. S. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.6.919 · Cited by 71 · see citing works
Validation and verification of measurement methods in clinical chemistry
Theodorsson, E. (2012) · Bioanalysis · doi:10.4155/bio.11.311 · Cited by 68 · see citing works
Running of Standards in Clinical Chemistry and the Use of the Control Chart
Henry, R. J., & SEGALOVE, M. (1952) · Journal of Clinical Pathology · doi:10.1136/jcp.5.4.305 · Cited by 65 · see citing works
Metrological traceability in clinical biochemistry
White, G. (2011) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1258/acb.2011.011079 · Cited by 62 · see citing works
Aakre, K. M., Langlois, M. R., Watine, J., Barth, J. H., Baum, H. H. E., Collinson, P. O., Laitinen, P. H., & Oosterhuis, W. P. (2012) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2012-0574 · Cited by 54 · see citing works
Impact of reference materials on accuracy in clinical chemistry
Franzini, C., & Ceriotti, F. (1998) · Clinical Biochemistry · doi:10.1016/s0009-9120(98)00054-x · Cited by 52 · see citing works
Petersen, P. H., Jørgensen, L. G. M., Brandslund, I., Olivarius, N. D. F., & Stahl, M. (2005) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365510500236135 · Cited by 52 · see citing works
Treatment of bias in estimating measurement uncertainty
O’Donnell, G. E., & Hibbert, D. B. (2005) · The Analyst · doi:10.1039/b414843f · Cited by 49 · see citing works
Estimating measurement uncertainty: reconciliation using a cause and effect approach
Ellison, S. L. R., & Barwick, V. J. (1998) · Accreditation and Quality Assurance · doi:10.1007/s007690050198 · Cited by 49 · see citing works
Systematic errors in analytical measurement results
Hibbert, D. B. (2007) · Journal of Chromatography A · doi:10.1016/j.chroma.2007.03.021 · Cited by 46 · see citing works
Zegers, I., Beetham, R., Keller, T., Sheldon, J., Bullock, D. G., MacKenzie, F., Trapmann, S., Emons, H., & Schimmel, H. G. (2013) · Clinical Chemistry · doi:10.1373/clinchem.2012.201954 · Cited by 44 · see citing works
Estimation of variance components and applications
Rao, C. R. 1.-2., & Kleffe, J. (1988) · Cited by 44 · see citing works
Towards a unified model of errors in analytical measurement
Thompson, M. (2000) · The Analyst · doi:10.1039/b006376m · Cited by 43 · see citing works
Total error and uncertainty: Friends or foes?
Rozet, E., Marini, R. D., Ziémons, E., Hubert, P., Dewé, W., Rudaz, S. L., & Boulanger, B. (2011) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2010.12.009 · Cited by 42 · see citing works
Collinson, P. O., van Dieijen‐Visser, M. P., Pulkki, K. J., Hammerer‐Lercher, A., Suvisaari, J. V. J., Ravkilde, J., Stavljenić-Rukavina, A., Baum, H. H. E., & Laitinen, P. H. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2011.171439 · Cited by 41 · see citing works
The influence of analytical bias on diagnostic misclassifications
Petersen, P. H., de Verdier, C., Groth, T., Fraser, C. G., Blaabjerg, O., & Hørder, M. (1997) · Clinica Chimica Acta · doi:10.1016/s0009-8981(96)06496-0 · Cited by 37 · see citing works
Estimating measurement uncertainty: reconciliation using a cause and effect approach
Ellison, S. L. R., & Barwick, V. J. (1998) · Springer eBooks · doi:10.1007/978-3-662-05173-3_21 · Cited by 36 · see citing works
Greenberg, N. L. (2014) · Clinica Chimica Acta · doi:10.1016/j.cca.2013.12.045 · Cited by 36 · see citing works
Attempts to include uncorrected bias in the measurement uncertainty
Synek, V. (2004) · Talanta · doi:10.1016/j.talanta.2004.07.038 · Cited by 35 · see citing works
The evaluation of measurement uncertainty from method validation studies
Barwick, V. J., Ellison, S. L. R., Rafferty, M. J. Q., & Gill, R. S. (2000) · Accreditation and Quality Assurance · doi:10.1007/s007690050022 · Cited by 35 · see citing works
Interpretation of clinical chemistry laboratory data
Fraser, C. (1986) · Medical Entomology and Zoology · Cited by 34 · see citing works
Interpretation of clinical chemistry laboratory data
Gw, P. (1987) · Journal of Clinical Pathology · doi:10.1136/jcp.40.3.358-b · Cited by 34 · see citing works
Commutability of reference materials in clinical chemistry.
Franzini, C. (1993) · PubMed · Cited by 34 · see citing works
Becker, D. S., & Becker, D. S. (1992) · doi:10.6028/nist.sp.829 · Cited by 33 · see citing works
Estimating total analytical error and its sources. Techniques to improve method evaluation.
Krouwer, J. S. (1992) · PubMed · Cited by 31 · see citing works
Barwick, V. J., & Ellison, S. L. R. (1998) · Analytical Communications · doi:10.1039/a806576d · Cited by 30 · see citing works
Accounting for bias in measurement uncertainty estimation
Haesselbarth, W. (2004) · Accreditation and Quality Assurance · doi:10.1007/s00769-004-0782-5 · Cited by 25 · see citing works
Stavelin, A. V., Petersen, P. H., Sølvik, U. Ø., & Sandberg, S. (2012) · Clinical Chemistry · doi:10.1373/clinchem.2012.191957 · Cited by 25 · see citing works
Metrological traceability – a concept for standardization in laboratory medicine
Siekmann, L. (2012) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2012-0710 · Cited by 24 · see citing works
Accuracy in clinical chemistry – who will kiss Sleeping Beauty awake?
Thienpont, L. M. (2008) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2008.245 · Cited by 22 · see citing works
Laboratory Statistics, Handbook of Formulas and Terms
Kallner, A. (2014) · Chemistry International · doi:10.1515/ci.2014.36.4.23 · Cited by 22 · see citing works
How to Improve Total Error Modeling by Accounting for Error Sources Beyond Imprecision and Bias
Krouwer, J. S. (2001) · Clinical Chemistry · doi:10.1093/clinchem/47.7.1329 · Cited by 21 · see citing works
Sølvik, U. Ø., Stavelin, A. V., Christensen, N. G., & Sandberg, S. (2006) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365510600684580 · Cited by 18 · see citing works
Components of Total Measurement Error for Hemoglobin A1c Determination
Phillipov, G., & Phillips, P. J. J. (2001) · Clinical Chemistry · doi:10.1093/clinchem/47.10.1851 · Cited by 15 · see citing works
Measurement uncertainty procedures revisited: Direct determination of uncertainty and bias handling
Hässelbarth, W. (1998) · Accreditation and Quality Assurance · doi:10.1007/s007690050277 · Cited by 15 · see citing works
ISO uncertainty and collaborative trial data
Ellison, S. L. R. (1998) · Accreditation and Quality Assurance · doi:10.1007/s007690050197 · Cited by 15 · see citing works
Dybkær, R. (1999) · Accreditation and Quality Assurance · doi:10.1007/s007690050395 · Cited by 15 · see citing works
Thienpont, L. M., Nuwenborg, J. E. V., & Stöckl, D. (1998) · Clinical Chemistry · doi:10.1093/clinchem/44.4.849 · Cited by 15 · see citing works
Dybkær, R. (1999) · Springer eBooks · doi:10.1007/3-540-27093-0_7 · Cited by 14 · see citing works
Kuttatharmmakul, S., Massart, D. L., & Smeyers-Verbeke, J. (2000) · Chemometrics and Intelligent Laboratory Systems · doi:10.1016/s0169-7439(00)00079-4 · Cited by 14 · see citing works
A method to estimate the uncertainty of measurements in a conglomerate of instruments/laboratories
Kallner, A., Khorovskaya, L. A., & Pettersson, T. (2005) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365510500206567 · Cited by 12 · see citing works
Krouwer, J. S. (1991) · Clinical Chemistry · doi:10.1093/clinchem/37.1.26 · Cited by 11 · see citing works
How good is the evidence base for test selection in clinical guidelines?
Misra, S., & Barth, J. H. (2014) · Clinica Chimica Acta · doi:10.1016/j.cca.2014.01.040 · Cited by 11 · see citing works
Establishing measurement traceability in clinical chemistry
Siekmann, L. (2004) · Accreditation and Quality Assurance · doi:10.1007/s00769-003-0718-5 · Cited by 10 · see citing works
Quality Control In Analytical Chemistry
Kateman, G., & Pijpers, F. W. (1981) · Andalas University Repository (Andalas University) · Cited by 10 · see citing works
Traceability in laboratory medicine
M�ller, M. M. (2003) · Accreditation and Quality Assurance · doi:10.1007/s00769-003-0666-0 · Cited by 9 · see citing works
Do we need to consider metrological meanings of different measurement uncertainty estimations?
Álvarez-Prieto, M., Jiménez-Chacón, J., & Montero-Curbelo, Á. (2009) · Accreditation and Quality Assurance · doi:10.1007/s00769-009-0570-3 · Cited by 9 · see citing works
Sobas, F., Benattar, N., Bellisario, A., Marin, S., Nougier, C., Lienhart, A., & Négrier, C. (2010) · Blood Coagulation & Fibrinolysis · doi:10.1097/mbc.0b013e328338dbd3 · Cited by 8 · see citing works
Ünsal, İ., Coşkun, A., Serteser, M., İnal, T. C., & Özpınar, A. (2010) · Accreditation and Quality Assurance · doi:10.1007/s00769-010-0694-5 · Cited by 7 · see citing works
A new approach for improved identification of measurement bias
Devanathan, S., Rollins, D. K., & Vardeman, S. B. (2000) · Computers & Chemical Engineering · doi:10.1016/s0098-1354(00)00626-8 · Cited by 6 · see citing works
Commutability of two JCTLM-listed secondary reference materials for two commercial lithium assays
Infusino, I., Frusciante, E., Ferrero, C. A., & Panteghini, M. (2012) · Clinica Chimica Acta · doi:10.1016/j.cca.2012.09.001 · Cited by 5 · see citing works
Implementation of the concept of traceability in laboratory medicine in external quality assessment
Siekmann, L. (2009) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2008.09.014 · Cited by 4 · see citing works
Matrix-dependent bias in total thyroxine measurement on the Beckman Access
Dietzen, D. J., Queen, S. F., & Solomon, S. S. (2002) · Clinica Chimica Acta · doi:10.1016/s0009-8981(01)00723-9 · Cited by 3 · see citing works
Simulating Total Error While Excluding Results Underestimates Total Error
Krouwer, J. S. (2010) · Clinical Chemistry · doi:10.1373/clinchem.2010.151936 · Cited by 3 · see citing works
[RiliBAK--a new computation concept].
Bauersfeld, W. (2006) · PubMed · Cited by 2 · see citing works
Flagged first, then by citations
1 of the 93 references couldn’t be matched to a known paper.
Research works printed but not linked: 3
- [2]2. Compendium of chemical terminology : IUPAC recommendations. (2nd). Blackwell Science, Oxford England ; Malden, MA, USA. (1997). 2. Clsi: EP09-A3 Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guideline - Third Edition . (2013).
- [49]49. Ellison SLR, Farrant TJ, Barwick V, Royal Society of Chemistry (Great Britain): Practical statistics for the analytical scientist : a bench guide. (2nd). RSC Publishing, Cambridge, UK. (2009).
- [80]80. Ellison SLR, Barwick VJ: Estimating measurement uncertainty: reconciliation using a cause and effect approach . Accredit Qual Assur 3(3), 101-105 (1998).
Cited by
Sorted by citations
The use of error and uncertainty methods in the medical laboratory
Oosterhuis, W. P., Bayat, H., Armbruster, D., Coşkun, A., Freeman, K. P., Kallner, A., Koch, D. D., MacKenzie, F., Migliarino, G. A., Orth, M., et al. (2017) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2017-0341 · Cited by 100 · see citing works
Uncertainty of Measurement in Laboratory Medicine
Milinković, N., Ignjatović, S., Šumarac, Z., & Majkić‐Singh, N. (2018) · Journal of Medical Biochemistry · doi:10.2478/jomb-2018-0002 · Cited by 60 · see citing works
Soleimani, R., Khourssaji, M., Gruson, D., Rodriguez‐Villalobos, H., Berghmans, M., Belkhir, L., Yombi, J. C., & Kabamba‐Mukadi, B. (2020) · Journal of Medical Virology · doi:10.1002/jmv.26430 · Cited by 52 · see citing works
Uncertainty in Measurement and Total Error
Theodorsson, E. (2017) · Clinics in Laboratory Medicine · doi:10.1016/j.cll.2016.09.002 · Cited by 54 · see citing works
Bias in Laboratory Medicine: The Dark Side of the Moon
Coşkun, A. (2023) · Annals of Laboratory Medicine · doi:10.3343/alm.2024.44.1.6 · Cited by 45 · see citing works
Diagnostic utility of droplet digital PCR for HIV reservoir quantification
Trypsteen, W., Kiselinova, M., Vandekerckhove, L., & Spiegelaere, W. D. (2016) · Journal of Virus Eradication · doi:10.1016/s2055-6640(20)30460-x · Cited by 42 · see citing works
Sutariya, P. G., Soni, H. N., Gandhi, S. A., Soni, S. S., & Prasad, J. (2021) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2020.129417 · Cited by 40 · see citing works
Statistical modelling of measurement error in wet chemistry soil data
van Leeuwen, C. C. E., Mulder, V. L., Batjes, N. H., & Heuvelink, G. B. M. (2021) · European Journal of Soil Science · doi:10.1111/ejss.13137 · Cited by 39 · see citing works
Quality Assurance in Clinical Chemistry: A Touch of Statistics and A Lot of Common Sense
Theodorsson, E. (2016) · Journal of Medical Biochemistry · doi:10.1515/jomb-2016-0012 · Cited by 38 · see citing works
Westgard, J. O., & Westgard, S. A. (2016) · Clinics in Laboratory Medicine · doi:10.1016/j.cll.2016.09.001 · Cited by 37 · see citing works
Braga, F., & Panteghini, M. (2018) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2018.03.007 · Cited by 37 · see citing works
Pre-analytical and analytical aspects affecting clinical reliability of plasma glucose results
Pasqualetti, S., Braga, F., & Panteghini, M. (2017) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2017.03.009 · Cited by 41 · see citing works
Measurement uncertainty: Friend or foe?
Infusino, I., & Panteghini, M. (2018) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2018.01.025 · Cited by 35 · see citing works
Measurement uncertainty for practical use
Coşkun, A., Theodorsson, E., Oosterhuis, W. P., & Sandberg, S. (2022) · Clinica Chimica Acta · doi:10.1016/j.cca.2022.04.1003 · Cited by 32 · see citing works
Ruggieri, F., Alimonti, A., & Bocca, B. (2016) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2016.03.023 · Cited by 31 · see citing works
Analysis of Nitrite and Nitrate in Foods
Merino, L., Örnemark, U., & Toldrá, F. (2016) · Advances in food and nutrition research · doi:10.1016/bs.afnr.2016.11.004 · Cited by 30 · see citing works
Irregular analytical errors in diagnostic testing – a novel concept
Vogeser, M., & Seger, C. (2017) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2017-0454 · Cited by 29 · see citing works
Establishing, evaluating and monitoring analytical quality in the traceability era
Panteghini, M., & Krintus, M. (2025) · Critical Reviews in Clinical Laboratory Sciences · doi:10.1080/10408363.2024.2434562 · Cited by 28 · see citing works
Farrance, I., Badrick, T. C., & Frenkel, R. B. (2018) · Practical Laboratory Medicine · doi:10.1016/j.plabm.2018.e00097 · Cited by 25 · see citing works
Pawlik-Sobecka, L., Sołkiewicz, K., Kokot, I., Kiraga, A., Płaczkowska, S., Schlichtinger, A. M., & Kratz, E. M. (2020) · Diagnostics · doi:10.3390/diagnostics10010051 · Cited by 24 · see citing works
Korchia, J., & Freeman, K. P. (2021) · Journal of Veterinary Diagnostic Investigation · doi:10.1177/10406387211031194 · Cited by 20 · see citing works
Pereira, P., Magnusson, B., Theodorsson, E., Westgard, J. O., & Encarnação, P. M. M. (2015) · Accreditation and Quality Assurance · doi:10.1007/s00769-015-1180-x · Cited by 21 · see citing works
Frenkel, R. B., Farrance, I., & Badrick, T. C. (2019) · Clinica Chimica Acta · doi:10.1016/j.cca.2019.03.1633 · Cited by 20 · see citing works
Korchia, J., & Freeman, K. P. (2021) · Journal of Veterinary Diagnostic Investigation · doi:10.1177/10406387211029247 · Cited by 20 · see citing works
Bietenbeck, A., Thaler, M., Luppa, P. B., & Klawonn, F. (2017) · Clinical Chemistry · doi:10.1373/clinchem.2016.269845 · Cited by 19 · see citing works
Pum, J. K. W. (2019) · Clinica Chimica Acta · doi:10.1016/j.cca.2019.07.018 · Cited by 17 · see citing works
Parvizi, O., El‐Adawy, H., Roesler, U. H., Neubauer, H., & Mertens‐Scholz, K. (2020) · Pathogens · doi:10.3390/pathogens9030165 · Cited by 17 · see citing works
External quality assurance in the era of standardization
Theodorsson, E., Meijer, P., & Badrick, T. C. (2024) · Clinica Chimica Acta · doi:10.1016/j.cca.2024.117876 · Cited by 16 · see citing works
Full method validation in clinical chemistry
Theodorsson, E., & Magnusson, B. (2017) · Accreditation and Quality Assurance · doi:10.1007/s00769-017-1275-7 · Cited by 16 · see citing works
Measurement uncertainty and the importance of correlation
Farrance, I., & Frenkel, R. B. (2020) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2020-1205 · Cited by 14 · see citing works
Coşkun, A. (2025) · Annals of Laboratory Medicine · doi:10.3343/alm.2024.0569 · Cited by 13 · see citing works
Balik, A. R., & Güçel, F. (2021) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365513.2021.1955294 · Cited by 12 · see citing works
Error Methods Are More Practical, But Uncertainty Methods May Still Be Preferred
Westgard, J. O. (2018) · Clinical Chemistry · doi:10.1373/clinchem.2017.284406 · Cited by 12 · see citing works
Repeat patient testing shows promise as a quality control method for veterinary hematology testing
Flatland, B., & Freeman, K. P. (2018) · Veterinary Clinical Pathology · doi:10.1111/vcp.12593 · Cited by 12 · see citing works
Rascón-Careaga, A., Corella-Madueño, M. A. G., Martínez, C. J. P., García-Rojas, A. M., Souflé-Vásquez, S. Z., García-Moroyoqui, M. T., Córdoba-Beltrán, L. J., Carrasco, M. G. C., & García-Alegría, A. M. (2021) · MAPAN · doi:10.1007/s12647-021-00441-5 · Cited by 12 · see citing works
Frenkel, R. B., & Farrance, I. (2018) · Advances in clinical chemistry · doi:10.1016/bs.acc.2018.02.003 · Cited by 11 · see citing works
Evaluation of the i-STAT Alinity v in a veterinary clinical setting
Burke, J. E., Nguyen, T. H. T., Davis, T., Koenig, A., Lane, S. L., Good, J., & Brainard, B. M. (2021) · Journal of Veterinary Diagnostic Investigation · doi:10.1177/10406387211019710 · Cited by 9 · see citing works
Not all biases are created equal: how to deal with bias on laboratory measurements
Panteghini, M. (2024) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2024-1208 · Cited by 9 · see citing works
Comments on Sample Preparation in Chromatography for Different Types of Materials
Moldoveanu, S. C., & David, V. (2014) · Elsevier eBooks · doi:10.1016/b978-0-444-54319-6.00012-8 · Cited by 9 · see citing works
Theodorsson, E. (2024) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2023-1208 · Cited by 8 · see citing works
Calibration – an under-appreciated component in the analytical process of the medical laboratories
Sonntag, O., & Loh, T. P. (2023) · Advances in Laboratory Medicine / Avances en Medicina de Laboratorio · doi:10.1515/almed-2023-0127 · Cited by 7 · see citing works
Studies of instrument/method comparison, as part of validation for use in veterinary species
Freeman, K. P., & Flatland, B. (2017) · Veterinary Clinical Pathology · doi:10.1111/vcp.12455 · Cited by 7 · see citing works
Reference-mean-centered statistical quality control
Yago, M., & Pla, C. (2020) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2019-1034 · Cited by 6 · see citing works
Milinković, N., Jovičić, S. Ž., & Ignjatović, S. (2020) · Critical Reviews in Clinical Laboratory Sciences · doi:10.1080/10408363.2020.1784838 · Cited by 6 · see citing works
Zhou, R., Qin, Y., Yin, H., Yang, Y., & Wang, Q. (2017) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2017-0511 · Cited by 5 · see citing works
Bias, its minimization or circumvention to simplify internal quality assurance
Haeckel, R., Gurr, E., Hoff, T., & on behalf of the working group Guide Limits of the German So (2016) · Journal of Laboratory Medicine · doi:10.1515/labmed-2016-0036 · Cited by 4 · see citing works
Jeffery, N. D., Bate, S. T., Safayi, S., Howard, M. A., Moon, L. D. F., & Jeffery, U. (2018) · European Journal of Neuroscience · doi:10.1111/ejn.13847 · Cited by 4 · see citing works
Vandra, A. B. (2026) · JMIRx Med · doi:10.2196/49657 · Cited by 3 · see citing works
Evaluation of Ethanol Interference on Routine Biochemical Tests
Gümüş, A. (2018) · American Journal of Clinical Pathology · doi:10.1093/ajcp/aqy081 · Cited by 3 · see citing works
Keleş, M. (2019) · Turkish Journal of Biochemistry · doi:10.1515/tjb-2018-0250 · Cited by 3 · see citing works
Roy, M., Kumari, S., Sachdev, S., & Sharma, R. R. (2024) · Asian Journal of Transfusion Science · doi:10.4103/ajts.ajts_37_24 · Cited by 2 · see citing works
Population reference intervals. Discussion4
Av, M., & Савельев, Л. И. (2019) · Laboratory Service · doi:10.17116/labs201980214 · Cited by 2 · see citing works
Vandra, A. B. (2023) · medRxiv · doi:10.1101/2023.05.24.23290382 · Cited by 2 · see citing works
Vandra, A. B., & Drégelyi-Kiss, Á. (2025) · Metrology · doi:10.3390/metrology5040067 · Cited by 1 · see citing works
Mutie, C. K., Gatwiri, S., & Warutere, P. (2023) · International Journal of Current Aspects · doi:10.35942/05q2nv90 · Cited by 1 · see citing works
Liang, S., Wu, H., Zhao, J., Guo, X., Qiang, Y., Zhao, X. S., Lan, M., Zhao, C., & Zhang, D. (2024) · Biochemia Medica · doi:10.11613/bm.2024.010705 · Cited by 1 · see citing works
Olkhovik, A. Y., Oksha, T. A., Vasiliev, A., Сергеевич, С. П., & Emanuel, V. L. (2019) · Laboratory Service · doi:10.17116/labs2019803120 · Cited by 1 · see citing works
Allowable bias derived from the NOBIDA reference values
Åsberg, A., & Odsæter, I. H. (2019) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365513.2019.1669219 · Cited by 1 · see citing works
Lawal, B. J. (2020) · American Journal of Biomedical Science & Research · doi:10.34297/ajbsr.2020.11.001636 · Cited by 1 · see citing works
Uncertainty of Measurement in Laboratory Medicine
Milinković, N., Ignjatović, S., Šumarac, Z., & Majkić‐Singh, N. (2018) · Journal of Medical Biochemistry · doi:10.1515/jomb-2018-0002 · Cited by 1 · see citing works
Allowable bias when monitoring reference change values
Theodorsson, E., & Magnusson, B. (2015) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.3109/00365513.2015.1057899 · Cited by 1 · see citing works
Bias analysis of proficiency testing programme results
Wong, S. K., & Yao, W. Y. (2016) · Accreditation and Quality Assurance · doi:10.1007/s00769-016-1197-9
van Toorenenbergen, A. W., & Kurstjens, J. (2017) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2016-1053
Борисовна, Т. Н., & Av, M. (2018) · Laboratory Service · doi:10.17116/labs20187143-46
Comments on sample preparation in chromatography for different types of materials
Moldoveanu, S. C., & David, V. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-821405-3.00012-5
Sánchez, J. M. (2023) · Journal of Biological Education · doi:10.1080/00219266.2023.2271508
Calibración, un componente subestimado del proceso analítico en el laboratorio clínico
Sonntag, O., & Loh, T. P. (2024) · Advances in Laboratory Medicine / Avances en Medicina de Laboratorio · doi:10.1515/almed-2023-0147
Qureshi, M. A., Niaz, A., Ali, M. A., Rahman, S. U., Ehsan, S., Nazir, S., Rafique, M., Javed, H., Ijaz, F., Nawaz, A., et al. (2024) · Pakistan Journal of Biotechnology · doi:10.34016/pjbt.2024.21.02.933
Coşkun, A., Kaya, Z. Z., & Lippi, G. (2025) · Turkish Journal of Biochemistry · doi:10.1515/tjb-2025-0270
d’Errico, A., Moirano, G., Pizzi, C., Popovic, M., Chatzi, L., Andrušaitytė, S., Gražulevičienė, R., Slama, R., McEachan, R., Yang, T. C., et al. (2026) · Journal of Exposure Science & Environmental Epidemiology · doi:10.1038/s41370-026-00859-6
Mishra, R. K., Roy, S., Mishra, A., Prakash, V., Shekhar, R., Kumar, N., & Murti, K. (2026) · Environmental Geochemistry and Health · doi:10.1007/s10653-026-03518-9
- Authors
- Theodorsson, E., Magnusson, B., & Leito, I.
- Affiliations
- Linköping University (SE)
- RISE Research Institutes of Sweden (SE)
- University of Tartu (EE)
- Venue
- Bioanalysis
- Publisher
- Informa UK Limited
- Published
- 2014-11
- Volume
- 6
- Issue
- 21
- Pages
- 2855-2875
- ISSN
- 1757-6180 (ISSN-L)
- 1757-6199
- OpenAlex
- W2037120565
- Cited by
- 73
- References
- 93
- Citations by year
- 2026: 3 · 2025: 4 · 2024: 7 · 2023: 6 · 2022: 1 · 2021: 9 · 2020: 7 · 2019: 6 · 2018: 11 · 2017: 8 · 2016: 8 · 2015: 2 · 2014: 1
- FWCI
- 0.74
- Type
- journal article
- Language
- en
- Open access
- green
- urn.kb.se
- Licence
- other-oa
- Locations
- Bioanalysis (publishedVersion)
- PubMed (publishedVersion)
- KTH Publication Database DiVA (KTH Royal Institute of Technology) (submittedVersion, other-oa)
- Topics
- Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Scientific Measurement and Uncertainty Evaluation
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Keywords
- measurement bias, clinical chemistry, measurement uncertainty, random error, repeatability, day-to-day variability, reference measurement system, laboratory network, bias mitigation
- MeSH terms
- Automation, Humans, Laboratories, Quality Control, Reproducibility of Results, Clinical Chemistry Tests, Uncertainty
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review