Survey of national guidelines, education and training on phlebotomy in 28 European countries: an original report by the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) working group for the preanalytical phase (WG-PA)
Open accessAbstract
Background: European questionnaire survey was conducted by the European Federation of Clinical Chemistry and Laboratory Medicine Working Group for the Preanalytical Phase (EFLM WG-PA) to assess how phlebotomy is performed in EFLM countries, including differences in personnel, level of education and skills, and to investigate the presence and compliance of national phlebotomy guidelines on this matter. Methods: A questionnaire was constructed containing questions elucidating different aspects of the organization behind the phlebotomy praxis on a national basis, including questions on the staff performing phlebotomy, the education of these staff members, and the existence of and adherence to national guidelines. All 39 EFLM member countries were invited to participate. Results: In total 28/39 (72%) EFLM member countries responded. Seven out of the 28 (25%) have national phlebotomy guidelines and five have implemented other guidelines. The estimated compliance with phlebotomy guidance for the laboratories in the countries that have national guidelines available is poor, regardless to whether the phlebotomy was under the laboratory control or not. Most countries were interested in EFLM guidelines and to participate in a pilot EFLM preanalytical phase external quality assessment (EQA) scheme. In the responding EFLM member countries, the majority of phlebotomy is performed by nurses and laboratory technicians. Their basic education is generally 4–5 years of high school, followed by 2–5 years of colleague or university studies. Only a third (10/28; 36%) of the participating member countries has any specific training available as a continuous educational resource. A specific training for phlebotomy is not part of the education required to become qualified in 6/28 (21%) and 9/28 (32%) of countries for nurses and laboratory technicians, respectively. In countries and professions where training is required, most require more than 5 h of training. Conclusions: Based on the results of this survey we conclude the following: 1) There is a need to assess the quality of current practices, compliance to the CLSI H3-A6 guidelines and to identify some most critical steps which occur during phlebotomy, in different healthcare settings, across Europe; 2) Existing CLSI H3-A6 phlebotomy guidelines should be adapted and used locally in all European countries which do not have their own guidelines; 3) National EFLM societies need to be engaged in basic training program development and continuous education of healthcare phlebotomy staff (implementing the certification of competence).
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Francke, A. L., Smit, M. C., de Veer, A. J. E., & Mistiaen, P. (2008) · BMC Medical Informatics and Decision Making · doi:10.1186/1472-6947-8-38 · Cited by 1,131 · see citing works
Errors in clinical laboratories or errors in laboratory medicine?
Plebani, M. (2006) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2006.123 · Cited by 737 · see citing works
How can we improve guideline use? A conceptual framework of implementability
Gagliardi, A. R., Brouwers, M. C., Palda, V. A., Lemieux‐Charles, L., & Grimshaw, J. (2011) · Implementation Science · doi:10.1186/1748-5908-6-26 · Cited by 469 · see citing works
Preanalytical variability: the dark side of the moon in laboratory testing
Lippi, G., Guidi, G. C., Mattiuzzi, C., & Plebani, M. (2006) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2006.073 · Cited by 416 · see citing works
Self-citationPreanalytical quality improvement: from dream to reality
Lippi, G., Chance, J. J., Church, S., Dazzi, P., Fontana, R., Giavarina, D., Grankvist, K., Huisman, W., Kouri, T., Palička, V., et al. (2011) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2011.600 · Cited by 390 · see citing works
Self-citationPloeg, J., Davies, B., Edwards, N. C., Gifford, W., & Elliott-Miller, P. (2007) · Worldviews on Evidence-Based Nursing · doi:10.1111/j.1741-6787.2007.00106.x · Cited by 310 · see citing works
Preanalytical quality improvement: in quality we trust
Lippi, G., Becan-McBride, K., Behúlová, D., Bowen, R. A. R., Church, S., Delanghe, J., Grankvist, K., Kitchen, S., Nybo, M., Nauck, M. A., et al. (2012) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2012-0597 · Cited by 233 · see citing works
Self-citationQuality Standards for Sample Collection in Coagulation Testing
Lippi, G., Salvagno, G. L., Montagnana, M., Lima-Oliveira, G., Guidi, G. C., & Favaloro, E. J. (2012) · Seminars in Thrombosis and Hemostasis · doi:10.1055/s-0032-1315961 · Cited by 196 · see citing works
Self-citationRisk management in the preanalytical phase of laboratory testing
Lippi, G., & Guidi, G. C. (2007) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2007.167 · Cited by 174 · see citing works
Self-citationThe IFCC Working Group on laboratory errors and patient safety
Sciacovelli, L., & Plebani, M. (2009) · Clinica Chimica Acta · doi:10.1016/j.cca.2009.03.025 · Cited by 163 · see citing works
Preanalytical phase – a continuous challenge for laboratory professionals
Šimundić, A.-M., & Lippi, G. (2012) · Biochemia Medica · doi:10.11613/bm.2012.017 · Cited by 156 · see citing works
Self-citationQuality Indicators in Laboratory Medicine: from theory to practice
Sciacovelli, L., O’Kane, M., Skaik, Y. A., Caciagli, P., Pellegrini, C., Rin, G. D., Ivanov, A., Ghys, T., & Plebani, M. (2011) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2011.128 · Cited by 141 · see citing works
Phlebotomy issues and quality improvement in results of laboratory testing.
Lippi, G., Salvagno, G. L., Montagnana, M., Franchini, M., & Guidi, G. C. (2006) · PubMed · Cited by 132 · see citing works
Self-citationEvidence-Based Guidelines in Laboratory Medicine: Principles and Methods
Oosterhuis, W. P., Bruns, D. E., Watine, J., Sandberg, S., & Horvath, A. R. (2004) · Clinical Chemistry · doi:10.1373/clinchem.2003.025528 · Cited by 83 · see citing works
Medical errors: an introduction to concepts
Kalra, J. (2004) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2004.08.007 · Cited by 81 · see citing works
Wallin, O., Söderberg, J., Guelpen, B. R. V., Stenlund, H. C., Grankvist, K., & Brulin, C. (2008) · Clinica Chimica Acta · doi:10.1016/j.cca.2008.02.016 · Cited by 72 · see citing works
Self-citationError rates in Australian chemical pathology laboratories
Khoury, M., Burnett, L., & Mackay, M. A. (1996) · The Medical Journal of Australia · doi:10.5694/j.1326-5377.1996.tb124883.x · Cited by 72 · see citing works
Lima-Oliveira, G., Lippi, G., Salvagno, G. L., Montagnana, M., Picheth, G., & Guidi, G. C. (2012) · Biochemia Medica · doi:10.11613/bm.2012.036 · Cited by 66 · see citing works
Self-citationSöderberg, J., Brulin, C., Grankvist, K., et al. (2009) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2009.048 · Cited by 64 · see citing works
Self-citationThe prevalence of preanalytical errors in a Croatian ISO 15189 accredited laboratory
Šimundić, A.-M., Nikolac, N., Vukasović, I., & Vrkić, N. (2010) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2010.221 · Cited by 64 · see citing works
Preanalytical quality control program – an overview of results (2001–2005 summary)
Alsina, M. J., Álvarez, V., Barba, N., Bullich, S., Cortés, M., Escoda, I., & Martínez-Brú, C. (2008) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2008.168 · Cited by 63 · see citing works
Laboratory errors and patient safety
Miligy, D. A. (2015) · International Journal of Health Care Quality Assurance · doi:10.1108/ijhcqa-10-2008-0098 · Cited by 59 · see citing works
Lima-Oliveira, G., Lippi, G., Salvagno, G. L., Montagnana, M., Pitanguiera Manguera, C. L., Sumita, N. M., Picheth, G., Guidi, G. C., & Scartezini, M. (2011) · Biochemia Medica · doi:10.11613/bm.2011.024 · Cited by 56 · see citing works
Self-citationQuality Improvement in Laboratory Medicine: Extra-Analytical Issues
Lippi, G., Fostini, R., & Guidi, G. C. (2008) · Clinics in Laboratory Medicine · doi:10.1016/j.cll.2007.12.007 · Cited by 55 · see citing works
Self-citationA content validated questionnaire for assessment of self reported venous blood sampling practices
Bölenius, K., Brulin, C., Grankvist, K., Lindkvist, M., & Söderberg, J. (2012) · BMC Research Notes · doi:10.1186/1756-0500-5-39 · Cited by 52 · see citing works
Self-citationEvidence-based laboratory medicine
Badrick, T. C. (2017) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2017-7051 · Cited by 48 · see citing works
Wallin, O., Söderberg, J., Guelpen, B. R. V., Stenlund, H. C., Grankvist, K., & Brulin, C. (2010) · Scandinavian Journal of Caring Sciences · doi:10.1111/j.1471-6712.2009.00753.x · Cited by 44 · see citing works
Self-citationPrevention of hemolysis in blood samples collected from intravenous catheters
Lippi, G., Avanzini, P., & Cervellin, G. (2013) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2013.01.021 · Cited by 43 · see citing works
Self-citationAccreditation of medical laboratories in the European Union
Huisman, W., Horvath, A. R., Burnett, D. A., Blaton, V. H., Czikkely, R., Jansen, R., Kallner, A., Kenny, D., Brguljan, P. M., Plebani, M., et al. (2007) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2007.037 · Cited by 42 · see citing works
Collecting factors related to the haemolysis of blood specimens
Fang, L., Fang, S., Chung, Y., et al. (2008) · Journal of Clinical Nursing · doi:10.1111/j.1365-2702.2006.02057.x · Cited by 37 · see citing works
Söderberg, J., Wallin, O., Grankvist, K., et al. (2010) · Journal of Evaluation in Clinical Practice · doi:10.1111/j.1365-2753.2009.01179.x · Cited by 33 · see citing works
Self-citationJansen, R., Kenny, D., Blaton, V., et al. (2000) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2000.158 · Cited by 30 · see citing works
Laboratory sampling. Does the process affect the outcome?
Becan-McBride, K. (2000) · PubMed · Cited by 28 · see citing works
Wallin, O., Söderberg, J., Guelpen, B. R. V., et al. (2007) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365510701370675 · Cited by 24 · see citing works
Bilić-Zulle, L., Šimundić, A.-M., Smolčić, V. Š., et al. (2010) · Biochemia Medica · doi:10.11613/bm.2010.008 · Cited by 24 · see citing works
Laboratory quality improvement by implementation of phlebotomy guidelines.
Lippi, G., Mattiuzzi, C., & Guidi, G. C. (2006) · PubMed · Cited by 22 · see citing works
Self-citationBölenius, K., Söderberg, J., Hultdin, J., Lindkvist, M., Brulin, C., & Grankvist, K. (2012) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2012-0159 · Cited by 22 · see citing works
Self-citationQuality indicators and specifications for strategic and support processes in laboratory medicine
Ricós, C., Biosca, C., Ibarz, M., et al. (2008) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2008.225 · Cited by 21 · see citing works
How can we make laboratory testing safer?
Boone, D. J. (2007) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2007.169 · Cited by 21 · see citing works
National survey on the pre-analytical variability in a representative cohort of Italian laboratories
Lippi, G., Montagnana, M., & Giavarina, D. (2006) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2006.274 · Cited by 20 · see citing works
Self-citationAccreditation in clinical laboratories
Zima, T. (2010) · Biochemia Medica · doi:10.11613/bm.2010.026 · Cited by 20 · see citing works
Ho, B., & Ho, E. (2012) · Biochemia Medica · doi:10.11613/bm.2012.027 · Cited by 18 · see citing works
Huisman, W. (2001) · Clinica Chimica Acta · doi:10.1016/s0009-8981(01)00505-8 · Cited by 8 · see citing works
Communication strategies for reducing hospital error and professional liability.
Amon, E. (2002) · PubMed · doi:10.1097/01.ogx.0000035709.94922.e7 · Cited by 6 · see citing works
Amon, E. (2002) · Obstetrical & Gynecological Survey · doi:10.1097/00006254-200211000-00001 · Cited by 5 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Joint EFLM-COLABIOCLI Recommendation for venous blood sampling
Simundic, A.-M., Bölenius, K., Cadamuro, J., Church, S., Cornes, M. P., van Dongen-Lases, E. C., Eker, P., Erdeljanovic, T., Grankvist, K., Guimaraes, J. T., et al. (2018) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2018-0602 · Cited by 240 · see citing works
Self-citationTheodorsson, E., Magnusson, B., & Leito, I. (2014) · Bioanalysis · doi:10.4155/bio.14.249 · Cited by 73 · see citing works
Lippi, G., Salvagno, G. L., Lima-Oliveira, G., Brocco, G., Danese, E., & Guidi, G. C. (2015) · Clinica Chimica Acta · doi:10.1016/j.cca.2014.11.024 · Cited by 57 · see citing works
Self-citationCornes, M. P., Church, S., van Dongen-Lases, E., Grankvist, K., Guimarães, J. T., Ibarz, M., Kovalevskaya, S., Kristensen, G. B., Lippi, G., Nybo, M., et al. (2016) · Annals of Clinical Biochemistry: International Journal of Laboratory Medicine · doi:10.1177/0004563216643969 · Cited by 43 · see citing works
Self-citation
- Authors
- Simundic, A.-M., Cornes, M., Grankvist, K., Lippi, G., Nybo, M., Kovalevskaya, S., Sprongl, L., Sumarac, Z., & Church, S.
- Affiliations
- EFLM WG-PA Chair, University Department of Chemistry, Sestre milosrdnice University Hospital Center, Vinogradska 29, 10 000 Zagreb, Croatia
- New Cross Hospital (GB)
- Umeå University (SE)
- Ospedale di Parma (IT)
- Odense University Hospital (DK)
- Labstory Company , Saint Petersburg , Russian Federation
- Central Laboratory, Sumperk, Czech Republic
- Center for Medical Biochemistry, Clinical Center of Serbia , Belgrade , Serbia
- BD Diagnostics, Preanalytical Systems, Oxford, UK
- Venue
- Clinical Chemistry and Laboratory Medicine (CCLM)
- Publisher
- Walter de Gruyter GmbH
- Published
- 2013-05-14
- Volume
- 51
- Issue
- 8
- Pages
- 1585-1593
- ISSN
- 1434-6621 (ISSN-L)
- 1437-4331
- OpenAlex
- W2127617996
- Semantic Scholar
- 099cb6e29f6a116f021fa3cf6436cc655464ad00
- Cited by
- 84
- References
- 45
- Citations by year
- 2026: 3 · 2025: 1 · 2024: 6 · 2023: 3 · 2022: 6 · 2021: 2 · 2020: 8 · 2019: 10 · 2018: 4 · 2017: 13 · 2016: 11 · 2015: 6 · 2014: 9 · 2013: 2
- FWCI
- 6.48
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Clinical Chemistry and Laboratory Medicine (CCLM) (publishedVersion)
- PubMed (publishedVersion)
- University of Southern Denmark Research Portal (University of Southern Denmark) (publishedVersion)
- KTH Publication Database DiVA (KTH Royal Institute of Technology) (submittedVersion, other-oa)
- Topics
- Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Health Professions › Health Information Management › Healthcare Regulation
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Innovations in Medical Education
- Keywords
- phlebotomy, pre-analytical phase, guideline adherence, nurses, laboratory technicians, external quality assessment, European countries
- MeSH terms
- Chemistry, Clinical, Data Collection, Educational Status, European Union, Humans, Professional Practice, Surveys and Questionnaires, Medical Laboratory Science, Guidelines as Topic, Phlebotomy
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review