High accuracy measurements of dry mole fractions of carbon dioxide and methane in humid air

Open access

Published version of a preprint

The DOI matches its Crossref record; 39 of the 39 references it cites were found in a registry or an index.

Abstract

Traditional techniques for measuring the mole fractions of greenhouse gases in the well-mixed atmosphere have required dry sample gas streams (dew point < −25 °C) to achieve the inter-laboratory compatibility goals set forth by the Global Atmosphere Watch programme of the World Meteorological Organisation (WMO/GAW) for carbon dioxide (±0.1 ppm in the Northern Hemisphere and ±0.05 ppm in the Southern Hemisphere) and methane (±2 ppb). Drying the sample gas to low levels of water vapour can be expensive, time-consuming, and/or problematic, especially at remote sites where access is difficult. Recent advances in optical measurement techniques, in particular cavity ring down spectroscopy, have led to the development of greenhouse gas analysers capable of simultaneous measurements of carbon dioxide, methane and water vapour. Unlike many older technologies, which can suffer from significant uncorrected interference from water vapour, these instruments permit accurate and precise greenhouse gas measurements that can meet the WMO/GAW inter-laboratory compatibility goals (WMO, 2011a) without drying the sample gas. In this paper, we present laboratory methodology for empirically deriving the water vapour correction factors, and we summarise a series of in-situ validation experiments comparing the measurements in humid gas streams to well-characterised dry-gas measurements. By using the manufacturer-supplied correction factors, the dry-mole fraction measurements have been demonstrated to be well within the GAW compatibility goals up to a water vapour concentration of at least 1%. By determining the correction factors for individual instruments once at the start of life, this water vapour concentration range can be extended to at least 2% over the life of the instrument, and if the correction factors are determined periodically over time, the evidence suggests that this range can be extended up to and even above 4% water vapour concentrations.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.5194/amt-6-837-2013

Cites

39 of 39 references verified0 citation errors · 0 not checkable

Statistics

Figures

Fig. 1 Schematic for the setup of Method #1, used by MPI-Jena, NOAA, and LSCE. Fig. 1. Schematic for the setup of Method #1, used by MPI-Jena, NOAA, and LSCE.
Fig. 1 Schematic for the setup of Method #1, used by MPI-Jena, NOAA, and LSCE. Fig. 1. Schematic for the setup of Method #1, used by MPI-Jena, NOAA, and LSCE.CC-BY · From the PDF
Fig. 2 Schematics of the setup for Method #2 (NOAA / MPI implementation). The water droplet is injected through a tee connector before the hydrophobic membrane filter. The components enclosed in a dashed rectangle are optional, and are used to verify the droplet method. Fig. 2. Schematics of the setup for Method #2 (NOAA/MPI implementation). The water droplet is injected through a tee connector before the hydrophobic membrane filter. The components enclosed in a dashed rectangle are optional, and are used to verify the droplet method.
Fig. 2 Schematics of the setup for Method #2 (NOAA / MPI implementation). The water droplet is injected through a tee connector before the hydrophobic membrane filter. The components enclosed in a dashed rectangle are optional, and are used to verify the droplet method. Fig. 2. Schematics of the setup for Method #2 (NOAA/MPI implementation). The water droplet is injected through a tee connector before the hydrophobic membrane filter. The components enclosed in a dashed rectangle are optional, and are used to verify the droplet method.CC-BY · From the PDF
Fig. 3 Schematic for the setup of Method #2 (Empa implementation) 2 3 Fig. 3. Schematic for the setup of Method #2 (Empa implementation).
Fig. 3 Schematic for the setup of Method #2 (Empa implementation) 2 3 Fig. 3. Schematic for the setup of Method #2 (Empa implementation).CC-BY · From the PDF
Image 4
CC-BY · From the PDF
Image 5
CC-BY · From the PDF
Image 6
CC-BY · From the PDF
Image 7
CC-BY · From the PDF
Image 8
CC-BY · From the PDF
Image 9
CC-BY · From the PDF
Fig. 9 Deviations between the correction factors determined on a single instrument over 20 days and 5 replications at LSCE using methods #1 and #2, and the values reported in Chen et al. (2010). Left panel: CO2 measured at 400 ppm. Right panel: CH4 measured at 1900 ppb. The WMO/GAW targets are indicated by the green dashed lines in both panels. Fig. 9. Deviations between the correction factors determined on a single instrument over 20 days and 5 replications at LSCE using methods #1 and #2, and the values reported in Chen et al. (2010). Left panel: CO 2 measured at 400 ppm. Right panel: CH 4 measured at 1900 ppb. The WMO/GAW targets are indicated by the green dashed lines in both panels. 2 3 4
Fig. 9 Deviations between the correction factors determined on a single instrument over 20 days and 5 replications at LSCE using methods #1 and #2, and the values reported in Chen et al. (2010). Left panel: CO2 measured at 400 ppm. Right panel: CH4 measured at 1900 ppb. The WMO/GAW targets are indicated by the green dashed lines in both panels. Fig. 9. Deviations between the correction factors determined on a single instrument over 20 days and 5 replications at LSCE using methods #1 and #2, and the values reported in Chen et al. (2010). Left panel: CO 2 measured at 400 ppm. Right panel: CH 4 measured at 1900 ppb. The WMO/GAW targets are indicated by the green dashed lines in both panels. 2 3 4CC-BY · From the PDF
Image 11
CC-BY · From the PDF
Fig. 11 Dry mole-fraction corrections for several instruments at Empa calculated for mole fractions of 400 ppm CO2 and 1900 ppb CH4, compared against the coefficients from Chen et al. (2010), indicating instrument transferability of the correction coefficients. Fig. 11. Dry mole-fraction corrections for several instruments at Empa calculated for mole fractions of 400 ppm CO 2 and 1900 ppb CH4, compared against the coefficients from Chen et al. (2010), indicating instrument transferability of the correction coefficients.
Fig. 11 Dry mole-fraction corrections for several instruments at Empa calculated for mole fractions of 400 ppm CO2 and 1900 ppb CH4, compared against the coefficients from Chen et al. (2010), indicating instrument transferability of the correction coefficients. Fig. 11. Dry mole-fraction corrections for several instruments at Empa calculated for mole fractions of 400 ppm CO 2 and 1900 ppb CH4, compared against the coefficients from Chen et al. (2010), indicating instrument transferability of the correction coefficients.CC-BY · From the PDF
Fig. 13 Distribution of empirically determined correction coefficients (red points) for CO2 (left panel) and CH4 (right panel) at LSCE, NOAA, MPI-Jena, and Empa. Also shown in both panels are the values from Chen et al. (2010) (yellow points), and the means of each distribution (blue crosses). The gray points represent the Monte Carlo simulation (N = 2000) of the coefficients as modeled by a bivariate Gaussian distribution (see text for more information). Fig. 13. Distribution of empirically determined correction coefficients (red points) for CO 2 (left panel) and CH 4 (right panel) at LSCE, NOAA, MPI-Jena, and Empa. Also shown in both panels are the values from Chen et al. (2010) (yellow points), and the means of each distribution (blue crosses). The gray points represent the Monte Carlo simulation ( N = 2000) of the coefficients as modelled by a bivariate Gaussian distribution (see text for more information). Fig. 14: Uncertainty analysis based on the Monte Carlo simulation ( N = 2000) of the correction coefficients. The grey area indicates the uncertainty (1σ ) introduced by using the Chen et al. (2010) coefficients on any given instrument. The purple hashed area indicates the uncertainty introduced by a single measurement of the correction coefficients when the instrument is deployed. Finally, the blue hashed area indicates the overall repeatability of the measurement methodologies described in this paper. See the text for more information.
Fig. 13 Distribution of empirically determined correction coefficients (red points) for CO2 (left panel) and CH4 (right panel) at LSCE, NOAA, MPI-Jena, and Empa. Also shown in both panels are the values from Chen et al. (2010) (yellow points), and the means of each distribution (blue crosses). The gray points represent the Monte Carlo simulation (N = 2000) of the coefficients as modeled by a bivariate Gaussian distribution (see text for more information). Fig. 13. Distribution of empirically determined correction coefficients (red points) for CO 2 (left panel) and CH 4 (right panel) at LSCE, NOAA, MPI-Jena, and Empa. Also shown in both panels are the values from Chen et al. (2010) (yellow points), and the means of each distribution (blue crosses). The gray points represent the Monte Carlo simulation ( N = 2000) of the coefficients as modelled by a bivariate Gaussian distribution (see text for more information). Fig. 14: Uncertainty analysis based on the Monte Carlo simulation ( N = 2000) of the correction coefficients. The grey area indicates the uncertainty (1σ ) introduced by using the Chen et al. (2010) coefficients on any given instrument. The purple hashed area indicates the uncertainty introduced by a single measurement of the correction coefficients when the instrument is deployed. Finally, the blue hashed area indicates the overall repeatability of the measurement methodologies described in this paper. See the text for more information.CC-BY · From the PDF
Fig. 14 Uncertainty analysis based on the Monte Carlo simulation (N = 2000) of the 2 correction coefficients. The gray area indicates the uncertainty (1-sigma) introduced by using 3 the Chen et al. (2010) coefficients on any given instrument. The purple hashed area indicates 4 Fig. 14. Uncertainty analysis based on the Monte Carlo simulation ( N = 2000) of the correction coefficients. The grey area indicates the uncertainty (1σ ) introduced by using the Chen et al. (2010) coefficients on any given instrument. The purple hashed area indicates the uncertainty introduced by a single measurement of the correction coefficients when the instrument is deployed. Finally, the blue hashed area indicates the overall repeatability of the measurement methodologies described in this paper. See the text for more information.
Fig. 14 Uncertainty analysis based on the Monte Carlo simulation (N = 2000) of the 2 correction coefficients. The gray area indicates the uncertainty (1-sigma) introduced by using 3 the Chen et al. (2010) coefficients on any given instrument. The purple hashed area indicates 4 Fig. 14. Uncertainty analysis based on the Monte Carlo simulation ( N = 2000) of the correction coefficients. The grey area indicates the uncertainty (1σ ) introduced by using the Chen et al. (2010) coefficients on any given instrument. The purple hashed area indicates the uncertainty introduced by a single measurement of the correction coefficients when the instrument is deployed. Finally, the blue hashed area indicates the overall repeatability of the measurement methodologies described in this paper. See the text for more information.CC-BY · From the PDF
Image 15
CC-BY · From the PDF
Image 16
CC-BY · From the PDF
Image 17
CC-BY · From the PDF
Fig. 17 Comparison between G1301 (wet) and G2301(dry) at MHD for CO2 (left panels) and 2 CH4 (right panels). The top panel for each species shows the hourly data for both instruments 3 (green and blue curves, left axes) as well as the difference (black points, right axes). The 4 bottom panels indicate the results for the target tank measurements for each instrument (green 5 and black curves, left axes), as well as the difference (black points, right axis). The right hand 6 Fig. 17. Comparison between G1301 (wet) and G2301(dry) at MHD for CO 2 (left panels) and CH 4 (right panels). The top panel for each species shows the hourly data for both instruments (green and blue curves, left axes) as well as the difference (black points, right axes). The bottom panels indicate the results for the target tank measurements for each instrument (green and black curves, left axes), as well as the difference (black points, right axis). The right-hand axis of each panel also shows a histogram of the differences, and the mean and standard deviations for these distributions are noted in each panel. The green dashed lines indicate the WMO/GAW targets for each gas. Note, the dramatically higher drift in the G2301 target tank, which is mirrored in the difference between the two instruments for both the target tank and ambient measurements. This drift is due to a failure of the G2301 instrument (see text).
Fig. 17 Comparison between G1301 (wet) and G2301(dry) at MHD for CO2 (left panels) and 2 CH4 (right panels). The top panel for each species shows the hourly data for both instruments 3 (green and blue curves, left axes) as well as the difference (black points, right axes). The 4 bottom panels indicate the results for the target tank measurements for each instrument (green 5 and black curves, left axes), as well as the difference (black points, right axis). The right hand 6 Fig. 17. Comparison between G1301 (wet) and G2301(dry) at MHD for CO 2 (left panels) and CH 4 (right panels). The top panel for each species shows the hourly data for both instruments (green and blue curves, left axes) as well as the difference (black points, right axes). The bottom panels indicate the results for the target tank measurements for each instrument (green and black curves, left axes), as well as the difference (black points, right axis). The right-hand axis of each panel also shows a histogram of the differences, and the mean and standard deviations for these distributions are noted in each panel. The green dashed lines indicate the WMO/GAW targets for each gas. Note, the dramatically higher drift in the G2301 target tank, which is mirrored in the difference between the two instruments for both the target tank and ambient measurements. This drift is due to a failure of the G2301 instrument (see text).CC-BY · From the PDF
Image 19
CC-BY · From the PDF
Image 20
CC-BY · From the PDF
1 Fig. 19: Dependence of the difference between the two instruments against H2O and the 2 Fig. 19. Dependence of the difference between the two instruments against H 2 O and the analyte gas, for CO 2 (left panels) and CH 4 (right panels).
1 Fig. 19: Dependence of the difference between the two instruments against H2O and the 2 Fig. 19. Dependence of the difference between the two instruments against H 2 O and the analyte gas, for CO 2 (left panels) and CH 4 (right panels).CC-BY · From the PDF
Fig. 20 Upper panels: 1-min average CO2 (left panels) and CH4 (right panels) mole fractions 3 measured at Pallas by the Pallas G2401 with sample drying and the WCC G2401 without 4 sample drying. The WCC instrument was alternating between its own inlet system and the 5 Pallas manifold. The right axes of the upper panels show H2O as measured by the WCC 6 Fig. 20. Upper panels: 1-min average CO 2 (left panels) and CH 4 (right panels) mole fractions measured at Pallas by the Pallas G2401 with sample drying and the WCC G2401 without sample drying. The WCC instrument was alternating between its own inlet system and the Pallas manifold. The right axes of the upper panels show H 2 O as measured by the WCC instrument. Lower panels: difference between the two analysers, along with a histogram of the difference; the mean and standard deviation of the distributions are noted in the lower panels. Dashed lines correspond to WMO/GAW targets. 1
Fig. 20 Upper panels: 1-min average CO2 (left panels) and CH4 (right panels) mole fractions 3 measured at Pallas by the Pallas G2401 with sample drying and the WCC G2401 without 4 sample drying. The WCC instrument was alternating between its own inlet system and the 5 Pallas manifold. The right axes of the upper panels show H2O as measured by the WCC 6 Fig. 20. Upper panels: 1-min average CO 2 (left panels) and CH 4 (right panels) mole fractions measured at Pallas by the Pallas G2401 with sample drying and the WCC G2401 without sample drying. The WCC instrument was alternating between its own inlet system and the Pallas manifold. The right axes of the upper panels show H 2 O as measured by the WCC instrument. Lower panels: difference between the two analysers, along with a histogram of the difference; the mean and standard deviation of the distributions are noted in the lower panels. Dashed lines correspond to WMO/GAW targets. 1CC-BY · From the PDF
Image 23
CC-BY · From the PDF
Image 24
CC-BY · From the PDF
Image 25
CC-BY · From the PDF
Image 26
CC-BY · From the PDF
62 63 Fig. 24: Difference in CO2 as measured between two CRDS systems in Indianapolis measuring atmospheric air from a height of 121 m above ground level, one with drying and one without, averaged on an hourly basis. Upper panel: CO2 mole fraction by the G1301 (wet) and the G2401 (dry). The right axis of the upper panel shows H2O as measured by the G1301. Lower panels: difference between the two analyzers, along with a histogram of the difference; the mean and standard deviation of the distributions are noted in the lower panels. Dashed lines correspond to WMO/GAW targets. Fig. 24. Difference in CO 2 as measured between two CRDS systems in Indianapolis measuring atmospheric air from a height of 121 m above ground level, one with drying and one without, averaged on an hourly basis. Upper panel: CO 2 mole fraction by the G1301 (wet) and the G2401 (dry). The right axis of the upper panel shows H 2 O as measured by the G1301. Lower panels: difference between the two analysers, along with a histogram of the difference; the mean and standard deviation of the distributions are noted in the lower panels. Dashed lines correspond to WMO/GAW targets.
62 63 Fig. 24: Difference in CO2 as measured between two CRDS systems in Indianapolis measuring atmospheric air from a height of 121 m above ground level, one with drying and one without, averaged on an hourly basis. Upper panel: CO2 mole fraction by the G1301 (wet) and the G2401 (dry). The right axis of the upper panel shows H2O as measured by the G1301. Lower panels: difference between the two analyzers, along with a histogram of the difference; the mean and standard deviation of the distributions are noted in the lower panels. Dashed lines correspond to WMO/GAW targets. Fig. 24. Difference in CO 2 as measured between two CRDS systems in Indianapolis measuring atmospheric air from a height of 121 m above ground level, one with drying and one without, averaged on an hourly basis. Upper panel: CO 2 mole fraction by the G1301 (wet) and the G2401 (dry). The right axis of the upper panel shows H 2 O as measured by the G1301. Lower panels: difference between the two analysers, along with a histogram of the difference; the mean and standard deviation of the distributions are noted in the lower panels. Dashed lines correspond to WMO/GAW targets.CC-BY · From the PDF
Fig. 25 Dependence of the difference between the two instruments in Indianapolis against H2O (top panel) and CO2 (bottom panel). Fig. 25. Dependence of the difference between the two instruments in Indianapolis against H 2 O (top panel) and CO 2 (bottom panel). 1 2
Fig. 25 Dependence of the difference between the two instruments in Indianapolis against H2O (top panel) and CO2 (bottom panel). Fig. 25. Dependence of the difference between the two instruments in Indianapolis against H 2 O (top panel) and CO 2 (bottom panel). 1 2CC-BY · From the PDF
Image 29
CC-BY · From the PDF
Image 30
CC-BY · From the PDF

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 39 references

Research works printed but not linked: 1

  1. IPCC: Summary for Policymakers, in: Climate Change 2007: The Physical Science Basis. Contribution of Working Group I to the Fourth Assessment Report of the Intergovernmental Panel on Climate Change, edited by: Solomon, S., Qin, D., Manning, M., Chen, Z., Marquis, M., Averyt, K. B., Tignor, M., and Miller, H. L., Cambridge University Press, Cambridge, United Kingdom and New York, NY, USA, 2007.

Cited by

See all 207

Metadata

Authors
Rella, C. W., Chen, H., Andrews, A. E., Filges, A., Gerbig, C., Hatakka, J., Karion, A., Miles, N. L., Richardson, S. J., Steinbacher, M., Sweeney, C., Wastine, B., & Zellweger, C.
Affiliations
  • Picarro (United States) (US)
  • National Oceanic and Atmospheric Administration (US)
  • Max Planck Institute for Biogeochemistry (DE)
  • Finnish Meteorological Institute (FI)
  • Pennsylvania State University (US)
  • Empa (CH)
  • Laboratoire des Sciences du Climat et de l'Environnement (FR)
Venue
Atmospheric Measurement Techniques
Published
2013-03-27
Volume
6
Issue
3
Pages
837-860
ISSN
  • 1867-1381 (ISSN-L)
  • 1867-8548
OpenAlex
W2153796351
Cited by
194
References
36
Citations by year
2026: 5 · 2025: 14 · 2024: 11 · 2023: 9 · 2022: 18 · 2021: 16 · 2020: 13 · 2019: 16 · 2018: 18 · 2017: 20 · 2016: 14 · 2015: 16 · 2014: 14 · 2013: 10
FWCI
8.09
Type
journal article
Language
en
Open access
Locations
  • Atmospheric measurement techniques (publishedVersion, cc-by)
  • Repository for Publications and Research Data (ETH Zurich) (acceptedVersion, cc-by)
  • DORA Empa (Swiss Federal Laboratories for Materials Science and Technology (Empa)) (submittedVersion, cc-by)
  • HAL (submittedVersion, other-oa)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
  • Max Planck Digital Library (submittedVersion, other-oa)
  • Oskar-Bordeaux (Universite de Bordeaux) (submittedVersion)
  • MPG.PuRe (Max Planck Society) (submittedVersion, cc-by)
  • ETH Zürich Research Collection (cc-by)
Funders
European Commission (212128)
Topics
  • Physical Sciences › Chemistry › Spectroscopy › Spectroscopy and Laser Applications
  • Physical Sciences › Environmental Science › Global and Planetary Change › Atmospheric and Environmental Gas Dynamics
  • Physical Sciences › Engineering › Aerospace Engineering › Calibration and Measurement Techniques
Keywords
carbon dioxide, methane, humid air, cavity ring-down spectroscopy, greenhouse gas measurement, in situ validation, water vapor interference

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: High accuracy measurements of dry mole fractions of carbon dioxide and methane in humid air. https://app.cochat.ai/papers/10.5194/amt-6-837-2013
The paper
Rella, C. W., Chen, H., Andrews, A. E., Filges, A., Gerbig, C., Hatakka, J., Karion, A., Miles, N. L., Richardson, S. J., Steinbacher, M., Sweeney, C., Wastine, B., & Zellweger, C. (2013). High accuracy measurements of dry mole fractions of carbon dioxide and methane in humid air. Atmospheric Measurement Techniques. https://doi.org/10.5194/amt-6-837-2013

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.