State of the Art in Trueness and Interlaboratory Harmonization for 10 Analytes in General Clinical Chemistry
The DOI matches its Crossref record; 26 of the 27 references it cites were found in a registry or an index.
Abstract
Context. —Harmonization and standardization of results among different clinical laboratories is necessary for clinical practice guidelines to be established. Objective. —To evaluate the state of the art in measuring 10 routine chemistry analytes. Design. —A specimen prepared as off-the-clot pooled sera and 4 conventionally prepared specimens were sent to participants in the College of American Pathologists Chemistry Survey. Analyte concentrations were assigned by reference measurement procedures. Participants. —Approximately 6000 clinical laboratories. Results. —For glucose, iron, potassium, and uric acid, more than 87.5% of peer groups meet the desirable bias goals based on biologic variability criteria. The remaining 6 analytes had less than 52% of peer groups that met the desirable bias criteria. Conclusions. —Routine measurement procedures for some analytes had acceptable traceability to reference systems. Conventionally prepared proficiency testing specimens were not adequately commutable with a fresh frozen specimen to be used to evaluate trueness of methods compared with a reference measurement procedure.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Comparison of results from approximately 6,000 clinical laboratories reveals that the mean glucose concentrations measured in serum samples by the hexokinase and glucose oxidase methods are essentially the same (86).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [86]
In the citing paper
“However, compared with a reference measurement procedure, significant (P < 0.001) bias (up to 13%) was observed for 40.6% of the peer groups (86).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [86]
Citations
Cites
THE COLORIMETRIC DETERMINATION OF PHOSPHORUS
Fiske, C. H., & Subbarow, Y. (1925) · Journal of Biological Chemistry · doi:10.1016/s0021-9258(18)84756-1 · Cited by 18,997
Clinical and Laboratory Standards Institute
(2019) · Springer Reference Medizin · doi:10.1007/978-3-662-48986-4_300416 · Cited by 5,091
Current databases on biological variation: pros, cons and progress
C. Ricós, V. A. (1999) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365519950185229 · Cited by 987
Hoelzel, W., Weykamp, C., Jeppsson, J.-O., Miedema, K., Barr, J. R., Goodall, I., Hoshino, T., John, W. G., Kobold, U., Little, R., et al. (2004) · Clinical Chemistry · doi:10.1373/clinchem.2003.024802 · Cited by 690
Self-citationBiological Variation: From Principles to Practice
Fraser, C. G. (2001) · Cited by 653
Clinical Chemistry : Theory, Analysis, Correlation
Kaplan, L. A., & Pesce, A. J. (1996) · Medical Entomology and Zoology · Cited by 602
Enzymatische Harnstoffbestimmung in Blut und Serum im optischen Test nachWarburg
Talke, H., & Schubert, G. E. (1965) · Klinische Wochenschrift · doi:10.1007/bf01484513 · Cited by 481
Creatinine Measurement: State of the Art in Accuracy and Interlaboratory Harmonization
Greg Miller, W., Myers, G. L., Ashwood, E. R., Killeen, A. A., Wang, E., Thienpont, L. M., & Siekmann, L. (2005) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2005-129-297-cmsota · Cited by 286
Self-citationDoumas, B. T., Kwok-Cheung, P. P., Perry, B. W., Jendrzejczak, B., McComb, R. B., Schaffer, R., & Hause, L. L. (1985) · Clinical Chemistry · doi:10.1093/clinchem/31.11.1779 · Cited by 207
Miller, W. G., Myers, G. L., & Rej, R. (2006) · Clinical Chemistry · doi:10.1373/clinchem.2005.063511 · Cited by 178
Self-citationIntroduction: Strategies to set global quality specifications in laboratory medicine
C. G. Fraser, A. K. (1999) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365519950185184 · Cited by 172
Sampson, E. J., Baird, M. A., Burtis, C. A., Smith, E. M., Witte, D. L., & Bayse, D. D. (1980) · Clinical Chemistry · doi:10.1093/clinchem/26.7.816 · Cited by 107
Miller, W. G. (2002) · Clinica Chimica Acta · doi:10.1016/s0009-8981(02)00370-4 · Cited by 104
Self-citationD. Kenny, C. G. F. (1999) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365519950185409 · Cited by 98
Ross, J. W., Miller, W. G., Gl, M., & Præstgaard, J. (1998) · PubMed · Cited by 85
Self-citationStöckl, D., & Reinauer, H. (1993) · Clinical Chemistry · doi:10.1093/clinchem/39.6.993 · Cited by 83
Siekmann, L. (1985) · PubMed · Cited by 68
Self-citationA candidate reference method for uric acid in serum. I. Optimization and evaluation.
Duncan, P. H., Gochman, N., Cooper, T., Smith, E., & Bayse, D. D. (1982) · Clinical Chemistry · doi:10.1093/clinchem/28.2.284 · Cited by 47
Cobbaert, C., Weykamp, C., Baadenhuijsen, H., Kuypers, A., Lindemans, J., & Jansen, R. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.9.1526 · Cited by 45
Libeer, J.-C., Baadenhuijsen, H., Fraser, C. G., Petersen, P. H., Ricós, C., Stöckl, D., & Thienpont, L. M. (1996) · PubMed · Cited by 45
Self-citationMeasurement of Urea in Human Serum by Isotope Dilution Mass Spectrometry: A Reference Procedure
Keßler, A., & Siekmann, L. (1999) · Clinical Chemistry · doi:10.1093/clinchem/45.9.1523 · Cited by 42
Self-citationLibeer, J.-C., Baadenhuijsen, H., Fraser, C. G., et al. (1996) · Radboud Repository (Radboud University) · Cited by 35
Clinical chemistry : theory, analysis, and correlation
Kaplan, L. A., & Pesce, A. J. (1984) · Virtual Defense Library (Ministerio de Defensa) · Cited by 34
Revisions of the 1963 semidine HC1 standard method for inorganic phosphorus.
Cc, G., & Miller, R. C. (1983) · Clinical Chemistry · doi:10.1093/clinchem/29.1.184 · Cited by 20
Miller, W. G., & Hw, K. (1993) · PubMed · Cited by 20
Self-citation
Research works printed but not linked: 1
- Fraser, C. E. Biological Variation: From Principles to Practice. Washington, DC: AACC Press; 2001.
Cited by
International Expert Committee report on the role of the A1c assay in the diagnosis of diabetes
The International Expert Committee (2009) · Diabetes Care · doi:10.2337/dc09-9033 · Cited by 3,185
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Kirkman, M. S., Lernmark, A., Metzger, B. E., & Nathan, D. M. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2010.161596 · Cited by 1,360
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Kirkman, M. S., Lernmark, A., Metzger, B. E., & Nathan, D. M. (2011) · Diabetes Care · doi:10.2337/dc11-9998 · Cited by 718
A1c versus glucose testing: a comparison
Sacks, D. B. (2011) · Diabetes Care · doi:10.2337/dc10-1546 · Cited by 461
The Pros and Cons of Diagnosing Diabetes With A1C
Bonora, E., & Tuomilehto, J. (2011) · Diabetes Care · doi:10.2337/dc11-s216 · Cited by 379
Metadata
- Authors
- Miller, W. G., Myers, G. L., Ashwood, E. R., Killeen, A. A., Wang, E., Ehlers, G. W., Hassemer, D., Lo, S. F., Seccombe, D., Siekmann, L., Thienpont, L. M., & Toth, A.
- Affiliations
- Centers for Disease Control and Prevention (US)
- Children's Hospital of Wisconsin (US)
- Venue
- Archives of Pathology & Laboratory Medicine
- Published
- 2008-05
- Volume
- 132
- Issue
- 5
- Pages
- 838-846
- ISSN
- 0003-9985 (ISSN-L)
- 1543-2165
- OpenAlex
- W2163860957
- Cited by
- 87
- References
- 27
- Citations by year
- 2026: 3 · 2025: 1 · 2024: 2 · 2023: 4 · 2022: 2 · 2021: 4 · 2020: 4 · 2019: 2 · 2018: 2 · 2017: 4 · 2016: 5 · 2015: 2 · 2014: 6 · 2013: 5 · 2012: 12
- FWCI
- 6.56
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Archives of Pathology & Laboratory Medicine (publishedVersion)
- PubMed (publishedVersion)
- Ghent University Academic Bibliography (Ghent University) (submittedVersion)
- Topics
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Physics and Astronomy › Instrumentation › Scientific Measurement and Uncertainty Evaluation
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Keywords
- trueness, clinical chemistry analytes, bias, proficiency testing, pooled sera, reference measurement procedure, traceability, glucose, iron, potassium, uric acid, commutability, biological variation
- MeSH terms
- Blood Chemical Analysis, Chemistry, Clinical, Humans, Laboratories, Pathology, Clinical, Quality Control, Reference Values, United States, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.