Effect of a Second-Generation Venous Catheter Impregnated with Chlorhexidine and Silver Sulfadiazine on Central Catheter–Related Infections: A Randomized, Controlled Trial
The DOI matches its Crossref record; 26 of the 26 references it cites were found in a registry or an index.
Abstract
BACKGROUND: Central venous catheter-related infections are a significant medical problem. Improved preventive measures are needed. OBJECTIVE: To ascertain 1) effectiveness of a second-generation antiseptic-coated catheter in the prevention of microbial colonization and infection; 2) safety and tolerability of this device; 3) microbiology of infected catheters; and 4) propensity for the development of antiseptic resistance. DESIGN: Multicenter, randomized, double-blind, controlled trial. SETTING: 9 university-affiliated medical centers. PATIENTS: 780 patients in intensive care units who required central venous catheterization. INTERVENTION: Patients received either a standard catheter or a catheter coated with chlorhexidine and silver sulfadiazine. MEASUREMENTS: The authors assessed catheter colonization and catheter-related infection, characterized microbes by molecular typing, and determined their susceptibility to antiseptics. Patient tolerance of the catheter was monitored. RESULTS: Patients with the 2 types of catheters had similar demographic features, clinical interventions, laboratory values, and risk factors for infection. Antiseptic catheters were less likely to be colonized at the time of removal compared with control catheters (13.3 vs. 24.1 colonized catheters per 1000 catheter-days; P < 0.01). The center-stratified Cox regression hazard ratio for colonization controlling for sampling design and potentially confounding variables was 0.45 (95% CI, 0.25 to 0.78). The rate of definitive catheter-related bloodstream infection was 1.24 per 1000 catheter-days (CI, 0.26 to 3.62 per 1000 catheter-days) for the control group versus 0.42 per 1000 catheter-days (CI, 0.01 to 2.34 per 1000 catheter-days) for the antiseptic catheter group (P = 0.6). Coagulase-negative staphylococci and other gram-positive organisms were the most frequent microbes to colonize catheters. Noninfectious adverse events were similar in both groups. Antiseptic susceptibility was similar for microbes recovered from either group. LIMITATIONS: The antiseptic catheter was not compared with an antibiotic-coated catheter, and no conclusion can be made regarding its effect on bloodstream infection. CONCLUSIONS: The second-generation chlorhexidine-silver sulfadiazine catheter is well tolerated. Antiseptic coating appears to reduce microbial colonization of the catheter compared with an uncoated catheter.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Guidelines for the Prevention of Intravascular Catheter-related Infections
O'Grady, N. P., Alexander, M., Burns, L. A., Dellinger, E. P., Garland, J., Heard, S. O., Lipsett, P. A., Masur, H., Mermel, L. A., Pearson, M. L., et al. (2011) · Clinical Infectious Diseases · doi:10.1093/cid/cir257 · Cited by 4,699
Self-citationPrevention of perinatal group B streptococcal disease--revised guidelines from CDC, 2010.
Schrag, S. D., Gorwitz, R. J., Fultz-Butts, K., & Schuchat, A. (2002) · PubMed · Cited by 3,387
System, A. R. F. T. N. (2003) · American Journal of Infection Control · doi:10.1016/j.ajic.2003.09.002 · Cited by 2,915
A Semiquantitative Culture Method for Identifying Intravenous-Catheter-Related Infection
Maki, D. G., Weise, C. E., & Sarafin, H. W. (1977) · New England Journal of Medicine · doi:10.1056/nejm197706092962301 · Cited by 2,027 · cited 4×
“Proximal and distal segments were cultured by using the roll-plate method (11), and similarly, proximal and distal segments were cultured by using a sonication technique (12).”
MethodInfluentialSelf-citationGötz, F. (2002) · Molecular Microbiology · doi:10.1046/j.1365-2958.2002.02827.x · Cited by 1,178 · cited 1×
“Once organisms (mainly coagulase-negative staphylococci) gain access to the device, infection is derived from their ability to adhere, proliferate, and elaborate biofilm (17, 18).”
BackgroundPittet, D. (1994) · JAMA · doi:10.1001/jama.1994.03510440058033 · Cited by 926
Prevention of Intravascular Catheter–Related Infections
Mermel, L. A. (2000) · Annals of Internal Medicine · doi:10.7326/0003-4819-132-5-200003070-00009 · Cited by 919
Self-citationPrevention of Intravascular Catheter-Related Infections
(2021) · Indian Journal of Continuing Nursing Education · doi:10.4103/2230-7354.320831 · Cited by 855
A Comparison of Two Antimicrobial-Impregnated Central Venous Catheters
Darouiche, R. O., Raad, I. I., Heard, S. O., Thornby, J. I., Wenker, O. C., Gabrielli, A., Berg, J., Khardori, N., Hanna, H., Hachem, R., et al. (1999) · New England Journal of Medicine · doi:10.1056/nejm199901073400101 · Cited by 804
Raad, I. I., Hohn, D. C., Gilbreath, B. J., Suleiman, N., Hill, L. A., Bruso, P. A., Marts, K., Mansfield, P. F., & Bodey, G. P. (1994) · Infection Control & Hospital Epidemiology · doi:10.1086/646902 · Cited by 774
Maki, D. G., Stolz, S. M., Wheeler, S., & Mermel, L. A. (1997) · Annals of Internal Medicine · doi:10.7326/0003-4819-127-4-199708150-00001 · Cited by 753 · cited 3×
“Previous studies indicate that approximately 50% of infections result from microbes gaining access to the catheters from the cutaneous surface, whereas the remaining 50% of infections result from contamination of the hub and infusate (9, 10).”
BackgroundSelf-citationRaad, I., Darouiche, R., Dupuis, J., Abi-Said, D., Gabrielli, A., Hachem, R., Wall, M., Harris, R., Jones, J., Buzaid, A., et al. (1997) · Annals of Internal Medicine · doi:10.7326/0003-4819-127-4-199708150-00002 · Cited by 610
Chaiyakunapruk, N., Veenstra, D. L., Lipsky, B. A., & Saint, S. (2002) · Annals of Internal Medicine · doi:10.7326/0003-4819-136-11-200206040-00007 · Cited by 610 · cited 3×
“CAD or CHF 196 (50) 159 (41) Hemodialysis 42 (11) 41 (11) Neoplasm 119 ( 30 ) 124 (32) Type 1 diabetes 62 (16) 62 (16) COPD 103 (26) 83 (22) Gastrointestinal bleeding 47 (12) 61 (16) Infection 248 (63) 222 (58) Malnutrition 214 (55) 212 (56)”
BackgroundInfluentialCoagulase-Negative Staphylococci: Pathogens Associated with Medical Progress
Rupp, M. E., & Archer, G. L. (1994) · Clinical Infectious Diseases · doi:10.1093/clinids/19.2.231 · Cited by 487 · cited 3×
“Neoplasm 78 (20) 68 (18) Infection 50 (13) 42 (11) Cardiac (MI, angina, VHD, arrhythmia, CHF, cardiac arrest) 49 (12) 37 (10) Postoperative need for mechanical ventilation 30 (8) 17 (4) Sepsis 28 (7) 27 (7) Trauma 21 (5) 19 (5) Bleeding or hypovolemia 14 (4) 17 (4)”
BackgroundSelf-citationSherertz, R. J., Raad, I. I., Belani, A., Koo, L. C., Rand, K. H., Pickett, D. L., Straub, S. A., & Fauerbach, L. L. (1990) · Journal of Clinical Microbiology · doi:10.1128/jcm.28.1.76-82.1990 · Cited by 381 · cited 5×
“Proximal and distal segments were cultured by using the roll-plate method (11), and similarly, proximal and distal segments were cultured by using a sonication technique (12).”
BackgroundMethodInfluentialSelf-citationEvaluation of Outcome of Intravenous Catheter-related Infections in Critically Ill Patients
RELLO, J., OCHAGAVIA, A., SABANES, E., ROQUE, M., MARISCAL, D., REYNAGA, E., & VALLES, J. (2000) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/ajrccm.162.3.9911093 · Cited by 354 · cited 3×
“Neoplasm 78 (20) 68 (18) Infection 50 (13) 42 (11) Cardiac (MI, angina, VHD, arrhythmia, CHF, cardiac arrest) 49 (12) 37 (10) Postoperative need for mechanical ventilation 30 (8) 17 (4) Sepsis 28 (7) 27 (7) Trauma 21 (5) 19 (5) Bleeding or hypovolemia 14 (4) 17 (4)”
BackgroundDimick, J. B. (2001) · Archives of Surgery · doi:10.1001/archsurg.136.2.229 · Cited by 341 · cited 1×
“Neoplasm 78 (20) 68 (18) Infection 50 (13) 42 (11) Cardiac (MI, angina, VHD, arrhythmia, CHF, cardiac arrest) 49 (12) 37 (10) Postoperative need for mechanical ventilation 30 (8) 17 (4) Sepsis 28 (7) 27 (7) Trauma 21 (5) 19 (5) Bleeding or hypovolemia 14 (4) 17 (4)”
Veenstra, D. L. (1999) · JAMA · doi:10.1001/jama.282.6.554 · Cited by 299 · cited 2×
“Adherence to guidelines that emphasize aseptic techniques seems to decrease the rate of catheter-related bloodstream infection to a level below the threshold of cost-effectiveness demonstrated in previous studies of the first-generation antiseptic catheter (2, 29).”
BackgroundMethodCrnich, C. J., & Maki, D. G. (2002) · Clinical Infectious Diseases · doi:10.1086/339863 · Cited by 264 · cited 2×
“The use of chlorhexidine for insertion-site preparation might further reduce catheter-associated infection ( 28 , 30).”
BackgroundSelf-citation(2000) · American Journal of Infection Control · doi:10.1067/mic.2000.110544 · Cited by 221
Tambe, S. M., Sampath, L., & Modak, S. M. (2001) · Journal of Antimicrobial Chemotherapy · doi:10.1093/jac/47.5.589 · Cited by 177 · cited 3×
“Therefore, resident flora may be exposed to the antiseptic agents and the emergence of resistance is possible (27).”
BackgroundRijnders, B. J. A., Wijngaerden, E. V., & Peetermans, W. E. (2002) · Clinical Infectious Diseases · doi:10.1086/342905 · Cited by 164 · cited 3×
“the pathogenesis of catheter-related infection is colonization of the device, which may serve as a surrogate marker for bloodstream infection (22).”
BackgroundDisinfection, sterilization and preservation (5th ed)
Ison, S., & Beattie, M. (2002) · Australian Infection Control · doi:10.1071/hi02074 · Cited by 147
Armstrong, C. W., Mayhall, C. G., Miller, K. B., Newsome, H. H., Sugerman, H. J., Dalton, H. P., Hall, G. O., & Gennings, C. (1986) · Journal of Infectious Diseases · doi:10.1093/infdis/154.5.808 · Cited by 136 · cited 1×
“This study supports the recommendation that a catheter should be promptly removed if it is inserted through guidewire exchange in the setting of a catheter-related infection or in emergent, nonaseptic conditions (2, 25).”
ResultNosocomial Bloodstream Infections in the Critically III
Heiselman, D. (1994) · JAMA · doi:10.1001/jama.1994.03520230029022 · Cited by 107 · cited 2×
“Catheter-related bloodstream infection in patients in intensive care units costs approximately $30 000 per episode (2, 5, 6, 28).”
BackgroundBassetti, S., Hu, J., D'Agostino, R. B., & Sherertz, R. J. (2001) · Antimicrobial Agents and Chemotherapy · doi:10.1128/aac.45.5.1535-1538.2001 · Cited by 96 · cited 2×
“The second-generation antiseptic catheter seemed to be more efficacious than the first-generation catheter in both in vitro and in vivo models (19).”
BackgroundSelf-citation
Cited by
Guidelines for the Prevention of Intravascular Catheter-related Infections
O'Grady, N. P., Alexander, M., Burns, L. A., Dellinger, E. P., Garland, J., Heard, S. O., Lipsett, P. A., Masur, H., Mermel, L. A., Pearson, M. L., et al. (2011) · Clinical Infectious Diseases · doi:10.1093/cid/cir257 · Cited by 4,699
Self-citationChurch, D., Elsayed, S., Reid, O., Winston, B., & Lindsay, R. (2006) · Clinical Microbiology Reviews · doi:10.1128/cmr.19.2.403-434.2006 · Cited by 1,876
Lebeaux, D., Ghigo, J.-M., & Beloin, C. (2014) · Microbiology and Molecular Biology Reviews · doi:10.1128/mmbr.00013-14 · Cited by 1,318
Guidelines for the prevention of intravascular catheter-related infections
O'Grady, N. P., Alexander, M., Burns, L. A., Dellinger, E. P., Garland, J., Heard, S. O., Lipsett, P. A., Masur, H., Mermel, L. A., Pearson, M. L., et al. (2011) · American Journal of Infection Control · doi:10.1016/j.ajic.2011.01.003 · Cited by 1,066
Self-citationKostakioti, M., Hadjifrangiskou, M., & Hultgren, S. J. (2013) · Cold Spring Harbor Perspectives in Medicine · doi:10.1101/cshperspect.a010306 · Cited by 999
Metadata
- Authors
- Rupp, M. E., Lisco, S. J., Lipsett, P. A., Perl, T. M., Keating, K., Civetta, J. M., Mermel, L. A., Lee, D., Dellinger, E. P., Donahoe, M., Giles, D., Pfaller, M. A., Maki, D. G., & Sherertz, R.
- Venue
- Annals of Internal Medicine
- Publisher
- American College of Physicians
- Published
- 2005-10-18
- Volume
- 143
- Issue
- 8
- Pages
- 570-580
- ISSN
- 0003-4819 (ISSN-L)
- 1539-3704
- OpenAlex
- W2154156472
- Cited by
- 233
- References
- 26
- FWCI
- 23.57
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.