Magnetic Resonance Imaging Versus Computed Tomography for Three‐Dimensional Bone Imaging of Musculoskeletal Pathologies: A Review
Open accessThe DOI matches its Crossref record; 73 of the 73 references it cites were found in a registry or an index.
Abstract
Magnetic resonance imaging (MRI) is increasingly utilized as a radiation‐free alternative to computed tomography (CT) for the diagnosis and treatment planning of musculoskeletal pathologies. MR imaging of hard tissues such as cortical bone remains challenging due to their low proton density and short transverse relaxation times, rendering bone tissues as nonspecific low signal structures on MR images obtained from most sequences. Developments in MR image acquisition and post‐processing have opened the path for enhanced MR‐based bone visualization aiming to provide a CT‐like contrast and, as such, ease clinical interpretation. The purpose of this review is to provide an overview of studies comparing MR and CT imaging for diagnostic and treatment planning purposes in orthopedic care, with a special focus on selective bone visualization, bone segmentation, and three‐dimensional (3D) modeling. This review discusses conventional gradient‐echo derived techniques as well as dedicated short echo time acquisition techniques and post‐processing techniques, including the generation of synthetic CT, in the context of 3D and specific bone visualization. Based on the reviewed literature, it may be concluded that the recent developments in MRI‐based bone visualization are promising. MRI alone provides valuable information on both bone and soft tissues for a broad range of applications including diagnostics, 3D modeling, and treatment planning in multiple anatomical regions, including the skull, spine, shoulder, pelvis, and long bones. Level of Evidence: 3 Technical Efficacy: Stage 3
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Lecture Notes in Computer Science 1205
(1999) · Industrial Robot: An International Journal · doi:10.1108/ir.1999.04926fae.001 · Cited by 38,662
Simple proton spectroscopic imaging.
Dixon, W. T. (1984) · Radiology · doi:10.1148/radiology.153.1.6089263 · Cited by 2,276 · cited 1×
“Alternatively, water-only images can be generated by acquiring a GRE sequence with specific echo times to perform a Dixon water–fat separation.(25,28,29,68,69)”
MethodWarner, D. O., Zaccariello, M. J., Katusic, S. K., Schroeder, D. R., Hanson, A. C., Schulte, P. J., Buenvenida, S. L., Gleich, S. J., Wilder, R. T., Sprung, J., et al. (2018) · Anesthesiology · doi:10.1097/aln.0000000000002232 · Cited by 598 · cited 1×
“The use of anesthesia is however not riskfree, especially on repeated occurrences.(113) In addition, uncontrolled motion was seen in the spine,(36) and the jaw,(53) and is common in clinical care of the shoulder, weakening the diagnostic power of MRI.”
BackgroundMandl, P., Navarro-Compán, V, Terslev, L., Aegerter, P., van der Heijde, D., D'Agostino, M. A., Baraliakos, X., Pedersen, S. J., Jurik, A. G., Naredo, E., et al. (2015) · Annals of the Rheumatic Diseases · doi:10.1136/annrheumdis-2014-206971 · Cited by 572
Abdominal MR Imaging with a Volumetric Interpolated Breath-hold Examination
Rofsky, N. M., Lee, V. S., Laub, G., Pollack, M. A., Krinsky, G. A., Thomasson, D., Ambrosino, M. M., & Weinreb, J. C. (1999) · Radiology · doi:10.1148/radiology.212.3.r99se34876 · Cited by 538
A review of substitute CT generation for MRI-only radiation therapy
Edmund, J. M., & Nyholm, T. (2017) · Radiation Oncology · doi:10.1186/s13014-016-0747-y · Cited by 378
Towards quantitative PET/MRI: a review of MR-based attenuation correction techniques
Hofmann, M., Pichler, B., Schölkopf, B., & Beyer, T. (2009) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-008-1007-7 · Cited by 343
Dual‐echo Dixon imaging with flexible choice of echo times
Eggers, H., Brendel, B., Duijndam, A., & Herigault, G. (2011) · Magnetic Resonance in Medicine · doi:10.1002/mrm.22578 · Cited by 320 · cited 1×
“Alternatively, water-only images can be generated by acquiring a GRE sequence with specific echo times to perform a Dixon water–fat separation.(25,28,29,68,69)”
MethodA review on medical imaging synthesis using deep learning and its clinical applications
Wang, T., Lei, Y., Fu, Y., Wynne, J. F., Curran, W. J., Liu, T., & Yang, X. (2021) · Journal of Applied Clinical Medical Physics · doi:10.1002/acm2.13121 · Cited by 279 · cited 1×
“The most promising sCT generation models are deep learning-based and rely on various network architectures, including UNet,(73) generative adversarial network,(74) and their derivatives.(75) The use of sCT images has already been reviewed multiple times for radiotherapy purposes and PET-MR but their use for orthopedic…”
BackgroundQualitative and quantitative ultrashort echo time (UTE) imaging of cortical bone
Du, J., Carl, M., Bydder, M., Takahashi, A., Chung, C. B., & Bydder, G. M. (2010) · Journal of Magnetic Resonance · doi:10.1016/j.jmr.2010.09.013 · Cited by 263 · cited 2×
“An image with a CT-like contrast containing signal mainly in short T2 components can then be obtained by suppressing the long T2 signal.(72) However, the fast gradient switching between TR’s and the acquisition during the gradient ramp up (Fig.”
BackgroundMethodLeynes, A. P., Yang, J., Wiesinger, F., Kaushik, S. S., Shanbhag, D. D., Seo, Y., Hope, T. A., & Larson, P. E. Z. (2018) · Journal of Nuclear Medicine · doi:10.2967/jnumed.117.198051 · Cited by 250
Zero TEMR bone imaging in the head
Wiesinger, F., Sacolick, L. I., Menini, A., Kaushik, S. S., Ahn, S., Veit‐Haibach, P., Delso, G., & Shanbhag, D. D. (2016) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25545 · Cited by 242 · cited 3×
“ZTE-MRI with good to excellent inter- and intraobserver variability,(53,57) benefitting from a good UTE/ZTE-to-CT voxelwise intensity correlation in healthy and diseased bones.(21,53) ZTE-MRI revealed flattening and osteophytes in the mandibular condyles with near-perfect agreement to cone beam CT(57) whereas intermod…”
BackgroundMethodLenchik, L., Rogers, L. F., Delmas, P. D., & Genant, H. K. (2004) · American Journal of Roentgenology · doi:10.2214/ajr.183.4.1830949 · Cited by 228
Deep learning based synthetic‐CT generation in radiotherapy and PET: A review
Spadea, M. F., Maspero, M., Zaffino, P., & Seco, J. (2021) · Medical Physics · doi:10.1002/mp.15150 · Cited by 201
Botser, I. B., Ozoude, G. C., Martin, D. E., Siddiqi, A. J., Kuppuswami, S., & Domb, B. G. (2012) · Arthroscopy The Journal of Arthroscopic and Related Surgery · doi:10.1016/j.arthro.2011.10.021 · Cited by 197
Jans, L. B. O., Chen, M., Elewaut, D., Van den Bosch, F., Carron, P., Jacques, P., Wittoek, R., Jaremko, J. L., & Herregods, N. (2021) · Radiology · doi:10.1148/radiol.2020201537 · Cited by 175 · cited 5×
“Standard MR images were also able to detect 92% of joint space alterations, and to a lesser extent sclerosis,(92) while sCT could diagnose sclerosis and ankylosis with accuracies higher than 90%.(78) Dedicated MR imaging was also reliable owing to a good to excellent interobserver agreement(6,78) and repeatable with g…”
BackgroundMethodResultInfluentialMoro‐oka, T., Hamai, S., Miura, H., SHIMOTO, T., HIGAKI, H., Fregly, B. J., Iwamoto, Y., & Banks, S. A. (2007) · Journal of Orthopaedic Research® · doi:10.1002/jor.20355 · Cited by 149 · cited 18×
“Sequences like VIBE, UTE, or multi-echo steady state (MESS) can provide bone structural information while providing complementary information on other tissue, including cartilage.118–121 This can promote joint biomechanical and kinematic modeling by limiting the need for registration.18,32,122 Some of the dedicated im…”
MethodInfluentialClinical Evaluation of Zero-Echo-Time MR Imaging for the Segmentation of the Skull
Delso, G., Wiesinger, F., Sacolick, L. I., Kaushik, S. S., Shanbhag, D. D., Hüllner, M., & Veit-Haibach, P. (2015) · Journal of Nuclear Medicine · doi:10.2967/jnumed.114.149997 · Cited by 137 · cited 2×
“Low-resolution images induce more partial volume effects that can mask(54) or, on the contrary, enlarge structures of interest, potentially resulting in the under- or over-segmentation of bone on MR images.(58) Furthermore, for 3D bone modeling, low resolutions result in high interpolation uncertainty and can cause st…”
BackgroundEley, K. A., Watt-Smith, S. R., Sheerin, F., & Golding, S. J. (2014) · European Radiology · doi:10.1007/s00330-014-3286-7 · Cited by 136 · cited 4×
“Black bone/skull was reprinted by permission from Springer Nature.(20) Original images were all cropped to only show the region of interest.”
BackgroundMethodInfluentialFlorkow, M. C., Zijlstra, F., Willemsen, K., Maspero, M., van den Berg, C. A. T., Kerkmeijer, L. G. W., Castelein, R. M., Weinans, H., Viergever, M. A., van Stralen, M., et al. (2019) · Magnetic Resonance in Medicine · doi:10.1002/mrm.28008 · Cited by 135 · cited 1×
“However, getting acquainted with the use of MR images for measurements and diagnosis might be easier and faster on images with a CT-like contrast, like ZTE or sCT, that have a high correspondence to CT images.(21,27,53,80) As of now, only a few studies compared multiple MR sequences with equivalent acquisition paramet…”
BackgroundSelf-citation“Black bone” MRI: a partial flip angle technique for radiation reduction in craniofacial imaging
Eley, K. A., McIntyre, A., Watt‐Smith, S. R., & Golding, S. J. (2012) · British Journal of Radiology · doi:10.1259/bjr/95110289 · Cited by 124
Maspero, M., Bentvelzen, L. G., Savenije, M. H. F., Guerreiro, F., Seravalli, E., Janssens, G. O. R. J., van den Berg, C. A. T., & Philippens, M. E. P. (2020) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2020.09.029 · Cited by 111 · cited 1×
“Transforming a CT-MR multimodal workflow into a simplified radiation-free MR-only workflow, as previously proposed in radiotherapy treatment planning,(9) could lead to less hospital visits, lower costs, allow for the fusion of soft tissue and bone information, and reduce the time under sedation for younger patients.(1…”
MethodA Combined Experimental and Computational Approach to Subject-Specific Analysis of Knee Joint Laxity
Harris, M. D., Cyr, A. J., Ali, A. A., Fitzpatrick, C. K., Rullkoetter, P. J., Maletsky, L. P., & Shelburne, K. B. (2016) · Journal of Biomechanical Engineering · doi:10.1115/1.4033882 · Cited by 107 · cited 1×
“This can promote joint biomechanical and kinematic modeling by limiting the need for registration.(18,32,122) Some of the dedicated images also revealed fractures and lesions, like cysts or edema, that can improve patients’ clinical management but which were not visible on CT.”
Background3D-MR vs. 3D-CT of the shoulder in patients with glenohumeral instability
Stillwater, L., Koenig, J., Maycher, B., & Davidson, M. (2016) · Skeletal Radiology · doi:10.1007/s00256-016-2559-4 · Cited by 106 · cited 17×
“Sequences like VIBE, UTE, or multi-echo steady state (MESS) can provide bone structural information while providing complementary information on other tissue, including cartilage.118–121 This can promote joint biomechanical and kinematic modeling by limiting the need for registration.18,32,122 Some of the dedicated im…”
BackgroundMethodInfluentialSegmentation accuracy of long bones
den Broeck, J. V., Vereecke, E., Wirix-Speetjens, R., & Sloten, J. V. (2014) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2014.03.016 · Cited by 105 · cited 18×
“Sequences like VIBE, UTE, or multi-echo steady state (MESS) can provide bone structural information while providing complementary information on other tissue, including cartilage.118–121 This can promote joint biomechanical and kinematic modeling by limiting the need for registration.18,32,122 Some of the dedicated im…”
BackgroundMethodInfluentialStaartjes, V. E., Seevinck, P. R., Vandertop, W. P., van Stralen, M., & Schröder, M. L. (2021) · Neurosurgical FOCUS · doi:10.3171/2020.10.focus20801 · Cited by 96 · cited 1×
“However, getting acquainted with the use of MR images for measurements and diagnosis might be easier and faster on images with a CT-like contrast, like ZTE or sCT, that have a high correspondence to CT images.(21,27,53,80) As of now, only a few studies compared multiple MR sequences with equivalent acquisition paramet…”
BackgroundSelf-citationSchwaiger, B. J., Schneider, C., Kronthaler, S., Gassert, F. T., Boehm, C., Pfeiffer, D., Baum, T. H., Kirschke, J. S., Karampinos, D. C., Makowski, M. R., et al. (2021) · European Radiology · doi:10.1007/s00330-020-07597-9 · Cited by 93 · cited 9×
“In general, awareness of the possible artifacts and MR image specificities, especially regarding air, ligaments, tendons, or water/fat interfaces, is required for all anatomical regions and MR sequences.(5,20,30,36,57,61,81) However, getting acquainted with the use of MR images for measurements and diagnosis might be…”
BackgroundMethodInfluentialSEGUE: A Speedy rEgion-Growing Algorithm for Unwrapping Estimated Phase
Karsa, A., & Shmueli, K. (2018) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2018.2884093 · Cited by 93 · cited 1×
“To facilitate the distinction between tissues and air, phase information could complement the magnitude images,(90) although the processing of phase images is complex and can be error prone.(90,91)”
MethodDeep learning-based synthetic-CT generation in radiotherapy and PET: A review
Spadea, M. F., Maspero, M., Zaffino, P., Seco, J. C., Imaging, C., & Cancer (2021) · Utrecht University Repository (Utrecht University) · Cited by 92
Diagnostic Accuracy of Zero-Echo Time MRI for the Evaluation of Cervical Neural Foraminal Stenosis
Argentieri, E. C., Koff, M. F., Breighner, R. E., Endo, Y., Shah, P. H., & Sneag, D. B. (2017) · Spine · doi:10.1097/brs.0000000000002462 · Cited by 90
Pediatric skull fracture diagnosis: should 3D CT reconstructions be added as routine imaging?
Orman, G., Wagner, M. W., Seeburg, D. P., Zamora, C. A., Oshmyansky, A., Tekes, A., Poretti, A., Jallo, G. I., Huisman, T. A. G. M., & Bosemani, T. (2015) · Journal of Neurosurgery Pediatrics · doi:10.3171/2015.3.peds1553 · Cited by 90 · cited 1×
“7 were stated to be valuable for diagnosis in most patients,(70) which is in line with results obtained in CT images.(83) When UTE-MR images were used, promising results were reported for the detection of fractures in FIGURE 6: Example of an ex vivo radius used for evaluating different modalities for osteotomy plannin…”
ResultEvaluation of Osseous Morphology of the Hip Using Zero Echo Time Magnetic Resonance Imaging
Breighner, R. E., Bogner, E. A., Lee, S. C., Koff, M. F., & Potter, H. G. (2019) · The American Journal of Sports Medicine · doi:10.1177/0363546519878170 · Cited by 89 · cited 3×
“on CT.(23) These differences might originate from the fact that pelvic tilt was not standardized in the ZTE study.”
BackgroundInfluentialFlorkow, M. C., Willemsen, K., Zijlstra, F., Foppen, W., van der Wal, B. C. H., van Zyp, J. R. N. V. D. V., Viergever, M. A., Castelein, R. M., Weinans, H., van Stralen, M., et al. (2021) · Journal of Orthopaedic Research® · doi:10.1002/jor.25127 · Cited by 81 · cited 4×
“Overall, the CT-to-MR intermodal agreement for the diagnosis of osseous pathologies and for the measurement of anatomical parameters was good to excellent with multiple reports of statistical equivalence.(16,45,65,77) In addition, MRI could provide 3D bone renderings, critical in the clinical care”
BackgroundMethodSelf-citationZheng, W., Kim, J. P., Kadbi, M., Movsas, B., Chetty, I. J., & Glide‐Hurst, C. K. (2015) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2015.07.001 · Cited by 80 · cited 1×
“To facilitate the distinction between tissues and air, phase information could complement the magnitude images,(90) although the processing of phase images is complex and can be error prone.(90,91)”
Methodde Mello, R. A. F., Ma, Y., Ashir, A., Jerban, S., Hoenecke, H. R., Carl, M., Du, J., & Chang, E. Y. (2020) · Arthroscopy The Journal of Arthroscopic and Related Surgery · doi:10.1016/j.arthro.2020.05.042 · Cited by 80 · cited 4×
“When MR-based 3D bone renderings were compared to CT, bone defects were also equivalently visible on the bone reconstructions(16) with no statistically significant differences between MR- and CT-based renderings,(7,29,86) good to excellent intermodal correlation,(7,86) excellent intermodal agreement,(19,26) and submil…”
BackgroundMethodInfluentialCT‐MR Image Fusion for the Management of Skull Base Lesions
Leong, J., Batra, P. S., & Citardi, M. J. (2006) · Otolaryngology · doi:10.1016/j.otohns.2005.11.015 · Cited by 77 · cited 1×
“Such fusions could be useful for the design of patient-specific implants by combining bone and joint capsule information(124) and have proven their potential for diagnostic and treatment purposes, by easing the diagnosis for junior readers,(125) and by facilitating treatment guidance(9,126) and surgical navigation.(49…”
BackgroundNeuroimaging of Traumatic Brain Injury
Douglas, D., Ro, T., Toffoli, T., Krawchuk, B., Muldermans, J., Gullo, J., Dulberger, A. R., Anderson, A., Douglas, P. K., & Wintermark, M. (2018) · Medical Sciences · doi:10.3390/medsci7010002 · Cited by 75
Huang, R., Liu, X., Li, H., & Zhou, P. (2020) · Dose-Response · doi:10.1177/1559325820923828 · Cited by 72 · cited 1×
“Such a multimodal workflow is logistically complex and induces an adverse radiation burden inherent to CT imaging, especially harmful in young population.(8)”
BackgroundGersing, A. S., Pfeiffer, D., Kopp, F. K., Schwaiger, B. J., Knebel, C., Haller, B., Noël, P. B., Settles, M., Rummeny, E. J., & Woertler, K. (2018) · European Radiology · doi:10.1007/s00330-018-5450-y · Cited by 71 · cited 1×
“Furthermore, MRI can be used to generate simulated radiographs with a diagnostic quality similar to CT for imaging bone tumors, while providing additional information on tumor architecture and soft tissue extension.(48) Lastly, MR sequences for imaging bone can also be combined with other sequences for specific imagin…”
MethodUTE bi-component analysis of T2* relaxation in articular cartilage
Shao, H., Chang, E. Y., Pauli, C., Zanganeh, S. A., Bae, W. C., Chung, C. B., Tang, G., & Du, J. (2015) · Osteoarthritis and Cartilage · doi:10.1016/j.joca.2015.08.017 · Cited by 68
Anatomical evaluation of CT-MRI combined femoral model
Lee, Y. S., Seon, J. K., Shin, V. I., Kim, G. H., & Jeon, M. (2008) · BioMedical Engineering OnLine · doi:10.1186/1475-925x-7-6 · Cited by 67 · cited 1×
“1 mm.(88) In ex vivo skulls, which can be harder to register and segment, BB-MRI segmentations deviated on average by 1.”
BackgroundLansdown, D. A., Cvetanovich, G. L., Verma, N. N., Cole, B. J., Bach, B. R., Nicholson, G. P., Romeo, A. A., Dawe, R. J., & Yanke, A. B. (2019) · Arthroscopy The Journal of Arthroscopic and Related Surgery · doi:10.1016/j.arthro.2018.10.119 · Cited by 66 · cited 3×
“Such a technique has been applied on standard GRE,(48) VS-GRE,(16) or Dixon water-only images.(26,29) More advanced processing has been searched to convert MR image intensities to CT Hounsfield units (HU), creating so-called synthetic CT (sCT).”
BackgroundMethodDremmen, M. H. G., Wagner, M. W., Bosemani, T., Tekes, A., Agostino, D., Day, E., Soares, B. P., & Huisman, T. A. G. M. (2017) · American Journal of Neuroradiology · doi:10.3174/ajnr.a5405 · Cited by 66 · cited 22×
“In general, awareness of the possible artifacts and MR image specificities, especially regarding air, ligaments, tendons, or water/fat interfaces, is required for all anatomical regions and MR sequences.(5,20,30,36,57,61,81) However, getting acquainted with the use of MR images for measurements and diagnosis might be…”
BackgroundInfluentialBae, W. C., Du, J., Bydder, G. M., & Chung, C. B. (2010) · Topics in Magnetic Resonance Imaging · doi:10.1097/rmr.0b013e31823ccebc · Cited by 66
“Black Bone” MRI: a novel imaging technique for 3D printing
Eley, K. A., Watt‐Smith, S. R., & Golding, S. J. (2017) · Dentomaxillofacial Radiology · doi:10.1259/dmfr.20160407 · Cited by 64
Saarikko, A. M., Mellanen, E. H., Kuusela, L., Leikola, J. P., Karppinen, A., Autti, T. H., Virtanen, P. L. J., & Brandstack, N. (2019) · Journal of Plastic Reconstructive & Aesthetic Surgery · doi:10.1016/j.bjps.2019.11.006 · Cited by 62 · cited 4×
“Nonetheless, motion artifacts can be reduced by using motion insensitive acquisition methods, including breath holds,(67) interleaved scanning, increased parallel imaging with higher signal averaging, or radial sampling of the k-space.(21,55,64,70)”
BackgroundMethodInfluentialSu, A. W. Y., Hillen, T. J., Eutsler, E. P., Bedi, A., Ross, J. R., Larson, C. M., Clohisy, J. C. F., & Nepple, J. J. (2019) · Arthroscopy The Journal of Arthroscopic and Related Surgery · doi:10.1016/j.arthro.2018.11.013 · Cited by 62
Kralik, S. F., Supakul, N., Wu, I. C., Delso, G., Radhakrishnan, R., Ho, C. Y., & Eley, K. A. (2018) · Neuroradiology · doi:10.1007/s00234-018-2127-9 · Cited by 61 · cited 7×
“In children under 2 years of age, sutures are wider(61,83) and harder to distinguish from trauma-induced fractures.(61,70) Lastly, cranial sutures tend to be less conspicuous in adults than children on BB-MRI(20) and UTE-MRI(54) as compared to CT because of a lower suture-to-skull contrast in adults.”
BackgroundMethodInfluentialFlorkow, M. C., Guerreiro, F., Zijlstra, F., Seravalli, E., Janssens, G. O. R. J., Maduro, J. H., Knopf, A. C., Castelein, R. M., van Stralen, M., Raaymakers, B. W., et al. (2020) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2020.09.056 · Cited by 59 · cited 1×
“Transforming a CT-MR multimodal workflow into a simplified radiation-free MR-only workflow, as previously proposed in radiotherapy treatment planning,(9) could lead to less hospital visits, lower costs, allow for the fusion of soft tissue and bone information, and reduce the time under sedation for younger patients.(1…”
MethodSelf-citationJerban, S., Lü, X., Jang, H., Ma, Y., Namiranian, B., Le, N., Li, Y., Chang, E. Y., & Du, J. (2019) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2019.06.016 · Cited by 48
Zeng, G., Schmaranzer, F., Degonda, C., Gerber, N., Gerber, K., Tannast, M., Burger, J., Siebenrock, K. A., Guoyan Zheng, A., & Lerch, T. D. (2020) · European Journal of Radiology Open · doi:10.1016/j.ejro.2020.100303 · Cited by 47 · cited 2×
“Using bone renderings, the measurement of local morphometric parameters, including the center-edge angle and acetabular version was similar between CT and MR, with average intermodal differences under 4 .(28,30,44) Hip range of FIGURE 10: Imaging examples.”
BackgroundMethodStecco, A., Guenzi, E., Cascone, T., Fabbiano, F., Fornara, P., Oronzo, P., Grassi, F. A., Cecchi, G., Caniggia, M., Brambilla, M., et al. (2013) · La radiologia medica · doi:10.1007/s11547-013-0927-x · Cited by 45
Comparative accuracy of lower limb bone geometry determined using MRI, CT, and direct bone 3D models
Stephen, J. M., Calder, J. D. F., Williams, A. M., & Daou, H. E. (2020) · Journal of Orthopaedic Research® · doi:10.1002/jor.24923 · Cited by 41 · cited 18×
“Sequences like VIBE, UTE, or multi-echo steady state (MESS) can provide bone structural information while providing complementary information on other tissue, including cartilage.118–121 This can promote joint biomechanical and kinematic modeling by limiting the need for registration.18,32,122 Some of the dedicated im…”
BackgroundMethodInfluentialHuber, F. A., Schumann, P., von Spiczak, J., Wurnig, M. C., Klarhöfer, M., Finkenstaedt, T., Bedogni, A., & Guggenberger, R. (2019) · Investigative Radiology · doi:10.1097/rli.0000000000000617 · Cited by 40 · cited 5×
“However, getting acquainted with the use of MR images for measurements and diagnosis might be easier and faster on images with a CT-like contrast, like ZTE or sCT, that have a high correspondence to CT images.(21,27,53,80) As of now, only a few studies compared multiple MR sequences with equivalent acquisition paramet…”
BackgroundInfluentialFoetal “black bone” MRI: utility in assessment of the foetal spine
Robinson, A. J., Blasér, S., Vladimirov, A., Drossman, D., Chitayat, D. A., & Ryan, G. (2014) · British Journal of Radiology · doi:10.1259/bjr.20140496 · Cited by 40
Finkenstaedt, T., Siriwanarangsun, P., Achar, S. A., Carl, M., Finkenstaedt, S., Abeydeera, N. A., Chung, C. B., & Bae, W. C. (2018) · Investigative Radiology · doi:10.1097/rli.0000000000000506 · Cited by 39 · cited 10×
“Additional MRI is sometimes acquired, mainly to rule out occult injuries and to identify spinal cord lesions.4 In this context, MRI can also aid in distinguishing acute from old fractures,3 and can help diagnose specific types of fractures, such as stress fractures.39 Diagnostic performance statistics for detecting ac…”
MethodInfluentialGoronzy, J., Blum, S. F. U., Hartmann, A., Plodeck, V., Franken, L., Günther, K., & Thielemann, F. (2018) · Acta Radiologica · doi:10.1177/0284185118795331 · Cited by 35 · cited 3×
“Overall, the CT-to-MR intermodal agreement for the diagnosis of osseous pathologies and for the measurement of anatomical parameters was good to excellent with multiple reports of statistical equivalence.(16,45,65,77) In addition, MRI could provide 3D bone renderings, critical in the clinical care”
BackgroundMethodSuchyta, M. A., Gibreel, W. O., Hunt, C. H., Gorny, K. R., Bernstein, M. A., & Mardini, S. (2018) · Plastic & Reconstructive Surgery · doi:10.1097/prs.0000000000004396 · Cited by 32 · cited 7×
“1 mm).(54,63,65) The T1w-SE and BB acquisitions also had high receiver bandwidth (>610 Hz/px— Table 1) to maintain geometrical integrity.”
BackgroundMethodInfluentialWang, Y.-H., Li, G., Ma, R., et al. (2020) · Clinical Oral Investigations · doi:10.1007/s00784-020-03497-w · Cited by 29
Malloy, P. J., Gasienica, J., Dawe, R. J., Orías, A. A. E., Nwachukwu, B. U., Inoue, N., Yanke, A. B., & Nho, S. J. (2020) · Journal of Orthopaedic Research® · doi:10.1002/jor.24596 · Cited by 28
Deininger‐Czermak, E., Villefort, C., Doeberitz, N. V. K., et al. (2020) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.27478 · Cited by 25 · cited 11×
“Additional MRI is sometimes acquired, mainly to rule out occult injuries and to identify spinal cord lesions.4 In this context, MRI can also aid in distinguishing acute from old fractures,3 and can help diagnose specific types of fractures, such as stress fractures.39 Diagnostic performance statistics for detecting ac…”
BackgroundMethodInfluentialZijlstra, F., Willemsen, K., Florkow, M. C., Sakkers, R. J. B., Weinans, H., van der Wal, B. C. H., van Stralen, M., & Seevinck, P. R. (2019) · SPIE Medical Imaging: Image Processing · doi:10.1117/12.2512857 · Cited by 23
Self-citationFritz, J. M. (2020) · Radiology · doi:10.1148/radiol.2020204045 · Cited by 20
Yan, K., Xi, Y., Sasiponganan, C., Zerr, J., Wells, J. E., & Chhabra, A. B. (2018) · British Journal of Radiology · doi:10.1259/bjr.20180474 · Cited by 19 · cited 3×
“phometric parameters is especially important in the hip joint for the diagnosis of hip dysplasia and femoroacetabular impingement and has been investigated on VS-GRE,(43,45) ZTE,(23) sCT,(77) or intermediate-weighted images.(43) All imaging techniques found good to excellent intermodal agreement for measuring the acet…”
BackgroundMethodBone-Selective MRI as a Nonradiative Alternative to CT for Craniofacial Imaging
Zhang, R., Lee, H., Zhao, X., Song, H. K., Vossough, A., Wehrli, F. W., & Bartlett, S. P. P. (2020) · Academic Radiology · doi:10.1016/j.acra.2020.03.001 · Cited by 19 · cited 6×
“1 mm).(54,63,65) The T1w-SE and BB acquisitions also had high receiver bandwidth (>610 Hz/px— Table 1) to maintain geometrical integrity.”
BackgroundInfluentialWu, H., Zhong, Y., Nie, Q., Chen, W., Guo, L., Yang, X., Zhang, H., Lin, Y., Xu, J., Dai, Y., et al. (2015) · European Radiology · doi:10.1007/s00330-015-3804-2 · Cited by 18 · cited 19×
“Sequences like VIBE, UTE, or multi-echo steady state (MESS) can provide bone structural information while providing complementary information on other tissue, including cartilage.118–121 This can promote joint biomechanical and kinematic modeling by limiting the need for registration.18,32,122 Some of the dedicated im…”
BackgroundInfluentialWillemsen, K., Ketel, M. H. M., Zijlstra, F., Florkow, M. C., Kuiper, R. J. A., van der Wal, B. C. H., Weinans, H., Pouran, B., Beekman, F. J., Seevinck, P. R., et al. (2021) · 3D Printing in Medicine · doi:10.1186/s41205-021-00103-x · Cited by 17 · cited 20×
“On average, CT segmentation overestimated the actual bone shape, whereas MR segmentation mostly underestimated it, although not consistently.(25,82) Nevertheless, surface distances between the MR-based segmentation and the cadaveric specimen were on average submillimeter,(17,25,31,33,82) with”
BackgroundMethodSelf-citationNoorda, Y. H., Bartels, L. W., Huisman, M., Nijenhuis, R. J., van den Bosch, M. A. A. J., & Pluim, J. P. W. (2014) · Physics in Medicine and Biology · doi:10.1088/0031-9155/59/15/4167 · Cited by 15
Campanelli, V., Howell, S. M., & Hull, M. L. (2019) · Computer Methods in Biomechanics and Biomedical Engineering Imaging & Visualization · doi:10.1080/21681163.2018.1559101 · Cited by 14 · cited 1×
“This can promote joint biomechanical and kinematic modeling by limiting the need for registration.(18,32,122) Some of the dedicated images also revealed fractures and lesions, like cysts or edema, that can improve patients’ clinical management but which were not visible on CT.”
BackgroundSchröder, F. F., Post, C. E., van Raak, S. M., Simonis, F. F. J., Wagenaar, F. B. M., Veld, R. M. H. A. H. I. ’., & Verdonschot, N. J. J. (2020) · Journal of Experimental Orthopaedics · doi:10.1186/s40634-020-00274-2 · Cited by 11 · cited 1×
“Other advantages of low-field MRI include its low cost (purchase and maintenance), and, when considering musculoskeletal radiology, the ability to scan in weight-bearing position.(115) This, however, comes at the expanse of SNR and resolution.”
BackgroundAutomated Segmentation of the Craniofacial Skeleton With “Black Bone” Magnetic Resonance Imaging
Eley, K. A., & Delso, G. (2020) · Journal of Craniofacial Surgery · doi:10.1097/scs.0000000000006552 · Cited by 11 · cited 1×
“Such interfaces can cause misdiagnoses(58,61) and complicate automated processes for segmentation.(54,87) To facilitate the distinction between tissues and air, phase information could complement the magnitude images,(90) although the processing of phase images is complex and can be error prone.”
BackgroundZijlstra, F., & Seevinck, P. R. (2021) · Magnetic Resonance in Medicine · doi:10.1002/mrm.28921 · Cited by 6
Self-citationAnatomical MR imaging of long bones: Comparative performance of MRI at 1.5 T and 3 T
Rathnayaka, K., Momot, K. I., Coulthard, A., Volp, A., Sahama, T. R., Schütz, M. A., & Schmutz, B. (2013) · Biomedical Spectroscopy and Imaging · doi:10.3233/bsi-120030 · Cited by 5
Metadata
- Authors
- Florkow, M. C., Willemsen, K., Mascarenhas, V. V., Oei, E. H. G., van Stralen, M., & Seevinck, P. R.
- Venue
- Journal of Magnetic Resonance Imaging
- Publisher
- Wiley
- Published
- 2022-01-19
- Volume
- 56
- Issue
- 1
- Pages
- 11-34
- ISSN
- 1053-1807 (ISSN-L)
- 1522-2586
- OpenAlex
- W4206781950
- Cited by
- 232
- References
- 73
- FWCI
- 21.50
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.