Enhanced Biocompatibility in Biostable Poly(carbonate)urethane
Abstract
Summary: In this work, we synthesized two MDI‐based polyurethanes, including a poly(ether)urethane (PEU) and a poly(carbonate)urethane (PCU), by using different soft segments, poly(tetramethylene oxide) and poly(hexyl, ethyl)carbonate diol ($\overline M$ ∼ 2 000). We demonstrated that, in addition to the enhanced biostability of PCU over PEU, the biological performances of PCU in vitro were also improved in general. These included, better cellular attachment and proliferation, less platelet activation, as well as reduced monocyte activation. The unusual wide‐ranging enhancement in biocompatibility for PCU was believed to be related to the larger micro‐phase separation in PCU (∼25 nm) that caused distinct protein adsorption on the surface. The total number of adherent monocytes (nonactivated and activated) on the bare sample surfaces, albumin pre‐adsorbed sample surfaces, and fibrinogen pre‐adsorbed sample surfaces. imageThe total number of adherent monocytes (nonactivated and activated) on the bare sample surfaces, albumin pre‐adsorbed sample surfaces, and fibrinogen pre‐adsorbed sample surfaces.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Polymer surfaces with segmented block co-polymers that form phase-separated structures exhibit less bacterial adhesion and cell attachment.
“Some studies have demonstrated that polymer surfaces with segmented block co-polymers (hard and soft domains) that result in phase-separated structures also exhibit less bacterial adhesion and cell attachment.61–63”
From Recent advances in thromboresistant and antimicrobial polymers for biomedical applications: just say yes to nitric oxide (NO) (2016) · cites this paper as [61]
Cites
Dalby, M. J., Yarwood, S. J., Riehle, M. O., Johnstone, H., Affrossman, S., & Curtis, A. S. G. (2002) · Experimental Cell Research · doi:10.1006/excr.2002.5498 · Cited by 354 · see citing works
Zhao, Q., Topham, N., Anderson, J. M., Hiltner, A., Lodoen, G., & Payet, C. R. (1991) · Journal of Biomedical Materials Research · doi:10.1002/jbm.820250205 · Cited by 318 · see citing works
Rapid fibroblast adhesion to 27 nm high polymer demixed nano-topography
Dalby, M. J., Giannaras, D., Riehle, M. O., Gadegaard, N., Affrossman, S., & Curtis, A. S. G. (2003) · Biomaterials · doi:10.1016/s0142-9612(03)00475-7 · Cited by 236 · see citing works
Human monocyte/macrophage activation and interleukin 1 generation by biomedical polymers
Miller, K. M., & Anderson, J. M. (1988) · Journal of Biomedical Materials Research · doi:10.1002/jbm.820220805 · Cited by 156 · see citing works
PDMS-based polyurethanes with MPEG grafts: synthesis, characterization and platelet adhesion study
Park, J. H., Park, K. D., & Bae, Y. H. (1999) · Biomaterials · doi:10.1016/s0142-9612(98)00250-6 · Cited by 145 · see citing works
Surface properties and blood compatibility of polyurethaneureas
Grasel, T. G., & Cooper, S. L. (1986) · Biomaterials · doi:10.1016/0142-9612(86)90002-5 · Cited by 122 · see citing works
Texturing of polymer surfaces at the cellular level
Schmidt, J. A., & von Recum, A. F. (1991) · Biomaterials · doi:10.1016/0142-9612(91)90006-v · Cited by 89 · see citing works
Effect of hard segment chemistry and strain on the stability of polyurethanes: in vivo biostability
Hergenrother, R. W., Wabers, H. D., & Cooper, S. L. (1993) · Biomaterials · doi:10.1016/0142-9612(93)90148-u · Cited by 76 · see citing works
Yuan, Y. L., Ai, F., Zhang, J., et al. (2002) · Journal of Biomaterials Science Polymer Edition · doi:10.1163/156856202320813800 · Cited by 73 · see citing works
Park, J. H., Cho, Y. W., Kwon, I. C., Jeong, S. Y., & Bae, Y. H. (2002) · Biomaterials · doi:10.1016/s0142-9612(02)00144-8 · Cited by 61 · see citing works
Microphase separated structure and protein adsorption of polyurethanes with butadiene soft segment
Huang, S., Chao, M., Ruaan, R., & Lai, J. (2000) · European Polymer Journal · doi:10.1016/s0014-3057(99)00063-4 · Cited by 56 · see citing works
Structure and Adsorption Properties of Fibrinogen
Feng, L., & Andrade, J. D. (1995) · ACS symposium series · doi:10.1021/bk-1995-0602.ch005 · Cited by 44 · see citing works
Hsu, S., Tseng, H., & Wu, M. (2000) · Artificial Organs · doi:10.1046/j.1525-1594.2000.06338.x · Cited by 40 · see citing works
Self-citationPark, J. H., Lee, K. B., Kwon, I. C., et al. (2001) · Journal of Biomaterials Science Polymer Edition · doi:10.1163/156856201316883458 · Cited by 38 · see citing works
In vitro activation of human monocytes by silicones
Naim, J. O., van Oss, C. J., Ippolito, K. M. L., Zhang, J.-W., Jin, L.-P., Fortuna, R., & Buehner, N. A. (1998) · Colloids and Surfaces B Biointerfaces · doi:10.1016/s0927-7765(98)00023-x · Cited by 32 · see citing works
Park, J. H., & Bae, Y. H. (2002) · Journal of Biomaterials Science Polymer Edition · doi:10.1163/15685620260178382 · Cited by 31 · see citing works
Reduced Platelet Adhesion on the Surface of Polyurethane Bearing Structure of Sulfobetaine
Yuan, J., Zhang, J., Zhu, J., et al. (2003) · Journal of Biomaterials Applications · doi:10.1177/088532803038100 · Cited by 25 · see citing works
On Matching Compliance Between Canine Carotid Arteries and Polyurethane Grafts
Hsu, S., & Karnbic, H. (1997) · Artificial Organs · doi:10.1111/j.1525-1594.1997.tb00485.x · Cited by 21 · see citing works
Self-citationHsu, S., & Chen, J.-Y. (1999) · Polymer Degradation and Stability · doi:10.1016/s0141-3910(98)00219-5 · Cited by 18 · see citing works
Self-citationFatigue Cut Growth of Thermoplastic Polyurethanes
Rodriguez, E. L., Basar, R., Kolycheck, E. G., et al. (1994) · Journal of Reinforced Plastics and Composites · doi:10.1177/073168449401300101 · Cited by 8 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Wo, Y., Brisbois, E. J., Bartlett, R. H., & Meyerhoff, M. E. (2016) · Biomaterials Science · doi:10.1039/c6bm00271d · Cited by 251 · see citing works
Polyurethane Shape‐Memory Polymers Demonstrate Functional Biocompatibility In Vitro
Cabanlit, M., Maitland, D., Wilson, T., Simon, S., Wun, T., Gershwin, M. E., & Van de Water, J. (2007) · Macromolecular Bioscience · doi:10.1002/mabi.200600177 · Cited by 69 · see citing works
Biocompatibility of Poly(carbonate urethane)s with Various Degrees of Nanophase Separation
Hsu, S., & Kao, Y. (2005) · Macromolecular Bioscience · doi:10.1002/mabi.200400163 · Cited by 39 · see citing works
Self-citation
- Authors
- Hsu, S., Kao, Y., & Lin, Z.
- Affiliations
- National Chung Hsing University (TW)
- Corresponding author
- Hsu, Shan‐hui
- ORCID
- Hsu, Shan‐hui
- Venue
- Macromolecular Bioscience
- Publisher
- Wiley
- Published
- 2004-04-13
- Volume
- 4
- Issue
- 4
- Pages
- 464-470
- ISSN
- 1616-5187 (ISSN-L)
- 1616-5195
- OpenAlex
- W2153893069
- Cited by
- 33
- References
- 20
- Citations by year
- 2026: 2 · 2024: 2 · 2022: 1 · 2018: 2 · 2017: 1 · 2016: 4 · 2014: 1 · 2013: 4 · 2012: 2
- FWCI
- 0.56
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Macromolecular Bioscience (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Materials Science › Surfaces, Coatings and Films › Polymer Surface Interaction Studies
- Physical Sciences › Materials Science › Biomaterials › biodegradable polymer synthesis and properties
- Physical Sciences › Engineering › Biomedical Engineering › Bone Tissue Engineering Materials
- Keywords
- polycarbonate urethane, polyether polyurethane, biocompatibility, biostability, poly(tetramethylene oxide), soft segment, microphase separation, protein adsorption, platelet activation, monocyte activation, albumin, fibrinogen
- MeSH terms
- Adsorption, Biocompatible Materials, Fibrinogen, Materials Testing, Microscopy, Electron, Scanning, Molecular Weight, Polymers, Polyurethanes, Serum Albumin, Surface Properties, Thermodynamics, Microscopy, Atomic Force
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review