Rapid high‐resolution T1 mapping by variable flip angles: Accurate and precise measurements in the presence of radiofrequency field inhomogeneity
Open accessAbstract
Rapid 3D mapping of T1 relaxation times is valuable in diverse clinical applications. Recently, the variable flip angle (VFA) spoiled gradient recalled echo approach was shown to be a practical alternative to conventional methods, providing better precision and speed. However, the method is known to be sensitive to transmit field (B1+) inhomogeneity and can result in significant systematic errors in T1 estimates, especially at high field strengths. The main challenge is to improve the accuracy of the VFA approach without sacrificing speed. In this article, the VFA method was optimized for both accuracy and precision by considering the influence of imperfect transmit fields, noise bias, and selection of flip angles. An analytic solution was developed for systematic B1+‐induced T1 errors and allows simple correction of T1 measurements acquired with any imaging parameters. A noise threshold was also identified and provided a guideline for avoiding T1 biases. Finally, it was shown that three flip angles were the most efficient for maintaining accuracy and high precision over large ranges of T1. A rapid B1+ mapping sequence was employed in all phantom experiments and high‐field in vivo brain scans. Experimental results confirmed the theory and validated the accuracy of the proposed method. Magn Reson Med, 2006. © 2006 Wiley‐Liss, Inc.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Data Reduction and Error Analysis for the Physical Sciences
Bevington, P. R., Robinson, D. K., Blair, J. M., Mallinckrodt, A. J., & McKay, S. (1993) · Computers in Physics · doi:10.1063/1.4823194 · Cited by 14,824
T1, T2 relaxation and magnetization transfer in tissue at 3T
Stanisz, G. J., Odrobina, E. E., Pun, J., Escaravage, M., Graham, S. J., Bronskill, M. J., & Henkelman, R. M. (2005) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20605 · Cited by 1,266
Measurement of signal intensities in the presence of noise in MR images
Henkelman, R. M. (1985) · Medical Physics · doi:10.1118/1.595711 · Cited by 1,017
Time Saving in Measurement of NMR and EPR Relaxation Times
Look, D. C., & Locker, D. R. (1970) · Review of Scientific Instruments · doi:10.1063/1.1684482 · Cited by 819
Rapid combined T1 and T2 mapping using gradient recalled acquisition in the steady state
Deoni, S. C. L., Rutt, B. K., & Peters, T. M. (2003) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10407 · Cited by 744
Snapshot flash mri. applications to t1, t2, and chemical‐shift imaging
Haase, A. (1990) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910130109 · Cited by 693
Deoni, S. C. L., Peters, T. M., & Rutt, B. K. (2005) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20314 · Cited by 478
Imaging of the active B1 field in vivo
Stollberger, R., & Wach, P. (1996) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910350217 · Cited by 357
Mapping of the Radiofrequency Field
Insko, E. K., & Bolinger, L. (1993) · Journal of Magnetic Resonance, Series A · doi:10.1006/jmra.1993.1133 · Cited by 323
Optimizing the precision in T1 relaxation estimation using limited flip angles
Wang, H. Z., Riederer, S. J., & Lee, J. N. (1987) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910050502 · Cited by 271
Cited by
Alexander, A. L., Hurley, S. A., Samsonov, A. A., Adluru, N., Hosseinbor, A. P., Mossahebi, P., Tromp, D. P. M., Zakszewski, E., & Field, A. S. (2011) · Brain Connectivity · doi:10.1089/brain.2011.0071 · Cited by 449
Jiang, Y., Ma, D., Seiberlich, N., Gulani, V., & Griswold, M. A. (2015) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25559 · Cited by 422
Using high-resolution quantitative mapping of R1 as an index of cortical myelination
Lutti, A., Dick, F., Sereno, M. I., & Weiskopf, N. (2014) · NeuroImage · doi:10.1016/j.neuroimage.2013.06.005 · Cited by 346
Models and methods for analyzing DCE‐MRI: A review
Khalifa, F., Soliman, A., El‐Baz, A., Abou El‐Ghar, M., El‐Diasty, T., Gimel'farb, G., Ouseph, R., & Dwyer, A. C. (2014) · Medical Physics · doi:10.1118/1.4898202 · Cited by 322
What are normal relaxation times of tissues at 3 T?
Bojorquez, J. Z., Bricq, S., Acquitter, C., Brunotte, F., Walker, P. M., & Lalande, A. (2017) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2016.08.021 · Cited by 286
Metadata
- Authors
- Cheng, H. M., & Wright, G. A.
- Venue
- Magnetic Resonance in Medicine
- Publisher
- Wiley
- Published
- 2006-01-31
- Volume
- 55
- Issue
- 3
- Pages
- 566-574
- ISSN
- 0740-3194 (ISSN-L)
- 1522-2594
- OpenAlex
- W2151682263
- Cited by
- 351
- References
- 23
- FWCI
- 8.31
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Rapid high‐resolution T1 mapping by variable flip angles: Accurate and precise measurements in the presence of radiofrequency field inhomogeneity. https://app.cochat.ai/papers/10.1002/mrm.20791
- The paper
- Cheng, H. M., & Wright, G. A. (2006). Rapid high‐resolution T1 mapping by variable flip angles: Accurate and precise measurements in the presence of radiofrequency field inhomogeneity. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.20791
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.