Cited by 119Non-Invasive Vital Sign MonitoringOptical Imaging and Spectroscopy TechniquesHemodynamic Monitoring and Therapy

A review of the effect of skin pigmentation on pulse oximeter accuracy

Open access
Al-Halawani, Raghda, Charlton, Peter H, Qassem, Meha, Kyriacou, Panayiotis A·2023·Physiological Measurement

Abstract

Objective . Pulse oximetry is a non-invasive optical technique used to measure arterial oxygen saturation (SpO 2 ) in a variety of clinical settings and scenarios. Despite being one the most significant technological advances in health monitoring over the last few decades, there have been reports on its various limitations. Recently due to the Covid-19 pandemic, questions about pulse oximeter technology and its accuracy when used in people with different skin pigmentation have resurfaced, and are to be addressed. Approach . This review presents an introduction to the technique of pulse oximetry including its basic principle of operation, technology, and limitations, with a more in depth focus on skin pigmentation. Relevant literature relating to the performance and accuracy of pulse oximeters in populations with different skin pigmentation are evaluated. Main Results . The majority of the evidence suggests that the accuracy of pulse oximetry differs in subjects of different skin pigmentations to a level that requires particular attention, with decreased accuracy in patients with dark skin. Significance . Some recommendations, both from the literature and contributions from the authors, suggest how future work could address these inaccuracies to potentially improve clinical outcomes. These include the objective quantification of skin pigmentation to replace currently used qualitative methods, and computational modelling for predicting calibration algorithms based on skin colour.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1088/1361-6579/acd51a

Cites

18 of 18 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • There have been systematic reviews on PPG skin pigmentation bias, but these have been limited to either consumer device PR or pulse oximeter SpO2.

    “There have been systematic reviews on PPG skin pigmentation bias, but these have been limited to either consumer device PR [35] or pulse oximeter SpO2 [36-38].”

    From Impact of skin pigmentation on pulse oximetry blood oxygenation and wearable pulse rate accuracy: systematic review and meta-analysis (2024) · cites this paper as [38]

    The ‘Skin pigmentation as a limitation of pulse oximetry’ section describes a review of studies comparing pulse-oximeter SpO2 with reference SaO2 across skin pigmentation groups, supporting the subject attributed to this source. But the ‘Statistical significance’ section explicitly calls it a narrative review, not a systematic review; the source therefore does not support that classification. (LLM, conf=0.87)

    Experimental — accuracy not yet measured

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Research works printed but not linked: 124

  1. Abay T Y Kyriacou P A 2021 Photoplethysmography in oxygenation and blood volume measurements Photoplethysmography: Technology, Signal Analysis, and Applications Kyriacou P Allen J Amsterdam Elsevier 1st edn
  2. Adler J N Hughes L A Vivilecchia R Camargo C A 1998 Effect of skin pigmentation on pulse oximetry accuracy in the emergency department Acad. Emerg. Med. 5 965 970 965–70 10.1111/j.1553-2712.1998.tb02772.x 9862586 · DOI · Paper page
  3. Agache F Fanian H Maibach I Humbert P 2017 Agache’s Measuring the Skin 2nd edn Cham, Switzerland Springer 549 562 pp 549–62
  4. Ancans J 2001 Melanosomal pH controls rate of melanogenesis, eumelanin/phaeomelanin ratio and melanosome maturation in melanocytes and melanoma cells Exp. Cell. Res. 268 26 35 26–35 10.1006/excr.2001.5251 11461115 · DOI
  5. Anliker U 2004 AMON: a wearable multiparameter medical monitoring and alert system IEEE Trans. Inf. Technol. Biomed. 8 415 427 415–27 10.1109/TITB.2004.837888 15615032 · DOI · Paper page
  6. Badgujar K C Badgujar A B Dhangar D V Badgujar V C 2020 Importance and use of pulse oximeter in COVID-19 pandemic: general factors affecting the sensitivity of pulse oximeter Indian Chem. Eng. 62 374 384 374–84 10.1080/00194506.2020.1845988 · DOI
  7. Baker K 2021 Performance of five pulse oximeters to detect hypoxaemia as an indicator of severe illness in children under five by frontline health workers in low resource settings - a prospective, multicentre, single-blinded, trial in Cambodia, Ethiopia, South Sudan, and Uganda E. Clin. Med. 38 10.1016/j.eclinm.2021.101040 PMC8326731 34368660 · DOI
  8. Baker S J Wilson C W 2022 Racial effects on Masimo pulse oximetry: a laboratory study J Clin Monit Comput 37 567 574 567–74 10.1007/s10877-022-00927-w 36370242 PMC9652601 · DOI · Paper page
  9. Balaraman K Rausch D Bellomy R Chong K Kameoka A Simmons Z Jenkins C Sumibcay T Liu E Wong K 2020 1167: effect of nail polish on oxygen saturation readings Crit. Care Med. 48 561 10.1097/01.ccm.0000643592.11619.5c · DOI
  10. Bangash M N 2022 Determining the impact of ethnicity on the accuracy of measurements of oxygen saturations: A retrospective cohort study eClinicalMedicine 48 10.1101/2021.11.21.21266662 PMC9096912 35706489 · DOI
  11. Banik P P Hossain S Kwon T-H Kim H Kim K-D 2020 Development of a wearable reflection-type pulse oximeter system to acquire clean PPG signals and measure pulse rate and SpO2 with and without finger motion Electronics 9 1905 10.3390/electronics9111905 · DOI
  12. Baquié M Kasraee B 2014 Discrimination between cutaneous pigmentation and erythema: comparison of the skin colorimeters dermacatch and mexameter Skin Res. Technol. 20 218 227 218–27 10.1111/srt.12109 24033842 · DOI
  13. Barker S J Curry J Redford D Morgan S 2006 Measurement of carboxyhemoglobin and methemoglobin by pulse oximetry: a human volunteer study Anesthesiology. 105 892 897 892–7 10.1097/00000542-200611000-00008 17065881 · DOI
  14. Bashkatov A N Genina E A 2000 Optical properties of melanin in the skin and skinlike phantoms Proceedings of the SPIE 4162 10.1117/12.405946 · DOI
  15. Bickler P E Feiner J R Severinghaus J W 2005 Effects of skin pigmentation on pulse oximeter accuracy at low saturation Anesthesiology. 102 715 719 715–9 10.1097/00000542-200504000-00004 15791098 · DOI · Paper page
  16. Bothma P A Joynt G M Lipman J Hon H Mathala B Scribante J Kromberg J 1996 Accuracy of pulse oximetry in pigmented patients S. Afr. Med. J. 86 594 596 594–6 8914569
  17. Brouqui P Amrane S Million M Cortaredona S Parola P Lagier J C Raoult D 2021 Asymptomatic hypoxia in COVID-19 is associated with poor outcome Int. J. Infect. Dis. 102 233 238 233–8 10.1016/j.ijid.2020.10.067 33130200 PMC7604151 · DOI
  18. Brown S Liyanage S Mikrou P Singh A Ewer A K 2020 Newborn pulse oximetry screening in the UK: a 2020 survey Lancet Amsterdam Elsevier vol 396 p 881 32950070
  19. Burnett G W Stannard B Wax D B Lin H M Pyram-Vincent C DeMaria S Levin M A 2022 Self-reported race/ethnicity and intraoperative occult hypoxemia: a retrospective cohort study Anesthesiology. 136 688 696 688–96 10.1097/ALN.0000000000004153 35231085 · DOI
  20. Cahan C Decker M J Hoekje P L Strohl K P 1990 Agreement between noninvasive oximetric values for oxygen saturation Chest. 97 814 819 814–9 10.1378/chest.97.4.814 2323252 · DOI
  21. Cajanding R 2021 Oxygen use and saturation targets in patients with COVID-19: are we giving too much or aiming too low? Nurs. Crit. Care 27 282 285 282–5 10.1111/nicc.12709 34476873 PMC8661992 · DOI
  22. Calvin J E 1992 Noninvasive blood gas monitoring: a review for use in the adult critical care unit. Technology Subcommittee of the Working Group on Critical Care, Ontario Ministry of Health CMAJ 146 703 712 703–12 1562943 PMC1488288
  23. Chacon P J Pu L da Costa T H Shin Y-H Ghomian T Shamkhalichenar H Wu H-C Irving B A Choi J-W 2018 A wearable pulse oximeter with wireless communication and motion artifact tailoring for continuous use IEEE Trans. Biomed. Eng. 66 1505 1513 1505–13 10.1109/TBME.2018.2874885 30307850 · DOI
  24. Chong A V Terosiet M Histace A Romain O 2017 Prototype of a multispectral and multiwavelength pulse oximeter based on a BQJ sensor 2017 15th IEEE Int. New Circuits and Systems Conf. (NEWCAS) 373 376 pp 373–6
  25. Çiçek H S Gümü S, S Deniz Ö Yildiz S Açikel C H Çakir E Tozkoparan E Uçar E Bilgiç H 2011 Effect of nail polish and henna on oxygen saturation determined by pulse oximetry in healthy young adult females Emerg. Med. J. 28 783 785 783–5 10.1136/emj.2010.096073 20923819 · DOI
  26. Coté C J Goldstein E A Fuchsman W H Hoaglin D C 1988 The effect of nail polish on pulse oximetry Anesth. Analg. 67 683 686 683–6 3382042
  27. Crooks C J 2022 Pulse oximeters’ measurements vary across ethnic groups: an observational study in patients with Covid-19 infection Eur. Respiratory J. 59 1 4 1–4 10.1183/13993003.03246-2021. PMC8796692 35086839 · DOI · Paper page
  28. Cysewska-Sobusiak A 2021 Problems with evaluating readings of pulse oximeters used at home during the ongoing COVID-19 pandemic Photonics Lett. Poland 13 19 21 19–21 10.4302/plp.v13i1.1093 · DOI
  29. Dumas C Wahr J A Tremper K K 1996 Clinical evaluation of a prototype motion artifact resistant pulse oximeter in the recovery room Anesthesia Analgesia 83 269 272 269–72 10.1213/00000539-199608000-00012 8694304 · DOI
  30. Dyer O 2020 Pulse oximetry may underestimate hypoxaemia in black patients, study finds Brit. Med. J. 371 m4926 10.1136/bmj.m4926 · DOI · Paper page
  31. Ebmeier S J 2018 A two centre observational study of simultaneous pulse oximetry and arterial oxygen saturation recordings in intensive care unit patients Anaesthesia Intensive Care 46 297 303 297–303 10.1177/0310057X1804600307 29716488 · DOI · Paper page
  32. Eichhorn J H 1993 Pulse oximetry as a standard of practice in anesthesia Anesthesiology. 78 423 426 423–6 8117343
  33. Falconer R J Robinson B J 1990 Comparison of pulse oximeters: accuracy at low arterial pressure in volunteers Br. J. Anaesth. 65 552 557 552–7 10.1093/bja/65.4.552 2248826 · DOI
  34. Fawzy A 2022 Racial and ethnic discrepancy in pulse oximetry and delayed identification of treatment eligibility among patients with COVID-19 JAMA Intern. Med. 21224 1 9 1–9 10.1001/jamainternmed.2022.1906 PMC9257583 35639368 · DOI · Paper page
  35. Feiner J R Rollins M D Sall J W Eilers H Au P Bickler P E 2013 Accuracy of carboxyhemoglobin detection by pulse CO-oximetry during hypoxemia Anesth. Analg. 117 847 858 847–58 10.1213/ANE.0b013e31828610a0 23477959 PMC4476500 · DOI
  36. Ferrari M Quaresima V Scholkmann F 2022 Pulse oximetry, racial bias and statistical bias: further improvements of pulse oximetry are necessary Ann. Intensive Care 12 12 10.1186/s13613-022-00992-z 35190907 PMC8860624 · DOI
  37. Fitzpatrick J E High W A Kyle W L 2017 Introduction to Clinical Dermatology Urgent care dermatology Symptom-Based Diagnosis. Fitzpatrick J E s.I. Elsevier 1 6 pp 1–6
  38. Foglia E E Whyte R K Chaudhary A Mott A Chen J Propert K J Schmidt B 2017 The effect of skin pigmentation on the accuracy of pulse oximetry in infants with hypoxemia J. Pediatr. 182 375 377 375–7 10.1016/j.jpeds.2016.11.043 27939107 PMC5328979 · DOI · Paper page
  39. Gabrielczyk M R Buist R J 1988 Pulse oximetry and postoperative hypothermia. An evaluation of the Nellcor N-100 in a cardiac surgical intensive care unit Anaesthesia. 43 402 404 402–4 10.1111/j.1365-2044.1988.tb09025.x. 3400853 · DOI · Paper page
  40. Goldman J M Petterson M T Kopotic R J Barker S J 2000 Masimo signal extraction pulse oximetry J. Clin. Monit. Comput. 16 475 483 475–83 10.1023/A:1011493521730 12580205 · DOI
  41. Gootenberg D B 2021 Developing a pulse oximetry home monitoring protocol for patients suspected with COVID-19 after emergency department discharge BMJ Health Care Inform. 28 e100330 10.1136/bmjhci-2021-100330 PMC8313308 34301725 · DOI
  42. Guber A Epstein S G Kohn S Shitrit D 2019 Wrist-sensor pulse oximeter enables prolonged patient monitoring in chronic lung diseases J. Med. Syst. 43 230 10.1007/s10916-019-1317-2 31201528 · DOI
  43. Greenhalgh T Knight M Inda-Kim M Fulop N J Leach J Vindrola-Padros C 2021 Remote management of covid-19 using home pulse oximetry and virtual ward support Brit. Med. J. 372 n677 10.1136/bmj.n677. 33766809 · DOI
  44. Hay 0 Y Cohen M Nitzan I Kasirer Y Shahroor-karni S Yitzhaky Y Engelberg S Nitzan M 2018 Pulse Oximetry with two infrared wavelengths without calibration in extracted arterial blood Sensors 18 3457 10.3390/s18103457 30326552 PMC6211094 · DOI
  45. Henry N R Hanson A C Schulte P J Warner N S Manento M N Weister T J Warner M A 2022 Disparities in hypoxemia detection by pulse oximetry across self-identified racial groups and associations with clinical outcomes Crit. Care Med. 50 204 211 204–11 10.1097/CCM.0000000000005394 35100193 PMC9070439 · DOI · Paper page
  46. Holder A L Wong A I 2022 The big consequences of small discrepancies: why racial differences in pulse oximetry errors matter Crit. Care Med. 50 335 337 335–7 10.1097/CCM.0000000000005447 35100196 PMC8887789 · DOI · Paper page
  47. Hoppe-Seyler F 1864 Ueber die chemischen und optischen Eigenschaften des Blutfarbstoffs Archiv. f. pathol. Anat. 29 233 235 233–5 10.1007/BF01937823 · DOI
  48. Hoskin R W Granger R 2001 Intraoperative decrease in pulse oximeter readings following injection of isosul-fan blue Can. J. Anaesth. 48 38 40 38–40 10.1007/BF03019812 11212047 · DOI
  49. Hunasikatti M 2021 Racial bias in accuracy of pulse oximetry and its impact on assessments of hypopnea and T90 in clinical studies J. Clin. Sleep Med. 17 1145 10.5664/jcsm.9178 33583491 PMC8320489 · DOI · Paper page
  50. Huong Ngu X 2014 The application of extended modified lambert beer model for measurement of blood carboxyhemoglobin and oxyhe-moglobin saturation J. Innov. Opt. Health Sci. 07 1450026 10.1142/S1793545814500266 · DOI
  51. Ishiyama T Kotoda M Asano N Ikemoto K Mitsui K Sato H Matsukawa T Sessler D I 2015 The effects of patent blue dye on peripheral and cerebral oxyhaemoglobin saturations Anaesthesia 70 429 433 429–33 10.1111/anae.12932 25389044 · DOI
  52. Johnson Jr B L Moy R L White G M 1998 Ethnic Skin: Medical and Surgical. St. Louis, Missouri Mosby 3 5 pp 3–5
  53. Jothishankar B Stein S L 2019 Impact of skin color and ethnicity Clin. Dermatol. 37 418 429 418–29 10.1016/j.clindermatol.2019.07.009 31896399 · DOI
  54. Jubran A 2015 Pulse oximetry Crit. Care 19 272 10.1186/s13054-015-0984-8 26179876 PMC4504215 · DOI
  55. Kanellis V G 2019 A review of melanin sensor devices Biophys. Rev. 11 6 843 849 843–9 10.1007/s12551-019-00581-8 31396804 PMC6874939 · DOI · Paper page
  56. Kelleher J F 1989 Pulse oximetry J. Clin. Monit. 5 37 62 37–62 10.1007/BF01618369 2647912 · DOI
  57. Knight M J Subbe C P Inada-Kim M 2022 Racial discrepancies in oximetry: where do we stand? Anaesthesia. 77 129 131 129–31 10.1111/anae.15635 34844284 · DOI
  58. Kyriacou P A Charlton P H Al-Halawani R Shelley K H 2022 Inaccuracy of pulse oximetry with dark skin pigmentation: clinical implications and need for improvement Br J. Anaesth. 130 e33 10.1016/j.bja.2022.03.011 35430087 · DOI
  59. Lancet E A 2021 Prehospital hypoxemia, measured by pulse oximetry, predicts hospital outcomes during the New York City COVID-19 pandemic J. Am. Coll. Emerg. Physicians Open. 2 e12407 10.1002/emp2.12407 33748809 PMC7967703 · DOI
  60. Lawless S T 1994 Crying wolf: false alarms in a pediatric intensive care unit Crit. Care Med. 22 981 985 981–5 8205831
  61. Lee K H Hui K P Tan W C Lim T K 1993 Factors influencing pulse oximetry as compared to functional arterial saturation in multi-ethnic Singapore Singap. Med. J. 34 385 387 385–7 8153680
  62. Lemay M 2021 Chapter 18 -Applications of optical cardiovascular monitoring Edward Sazonov, Wearable Sensors E Sazonov New York Academic 487 517 pp 487–517
  63. Levitan R M 2020 Pulse oximetry as a biomarker for early identification and hospitalization of COVID-19 pneumonia Acad Emerg Med. 27 785 786 785–6 10.1111/acem.14052 32779867 PMC7323007 · DOI
  64. Lipnick M S Feiner J R Au P Bernstein M Bickler P E 2016 The accuracy of 6 Inexpensive pulse oximeters not cleared by the food and drug administration: the possible global public health implications Anesth. Analg. 123 338 345 338–45 10.1213/ANE.0000000000001300 27089002 · DOI
  65. Luks A M Swenson E R 2020 Pulse oximetry for monitoring patients with COVID-19 at home. potential pitfalls and practical guidance Ann. Am. Thorac. Soc. 17 1040 1046 1040–6 10.1513/AnnalsATS.202005-418FR 32521167 PMC7462317 · DOI
  66. Ly B C K 2020 Research techniques made simple: cutaneous colorimetry: a reliable technique for objective skin color measurement J. Investigative Dermatology 140 3 12 3–12 10.1016/j.jid.2019.11.003 31864431 · DOI · Paper page
  67. Mahle W T 2009 American heart association congenital heart defects committee of the council on cardiovascular disease in the young, council on cardiovascular nursing, and interdisciplinary council on quality of care and outcomes research; american academy of pediatrics section on cardiology and cardiac surgery, and committee on fetus and newborn. Role of pulse oximetry in examining newborns for congenital heart disease: a scientific statement from the american heart association and american academy of pediatrics Circulation 120 447 458 447–58 10.1161/CIRCULATIONAHA.109.192576 19581492 · DOI
  68. Mannheimer P D Casciani J R Fein M E Nierlich S L 1997a Wavelength selection for low-saturation pulse oximetry IEEE Trans. Biomed. Eng. 44 148 158 148–58 10.1109/10.554761 9216128 · DOI
  69. Mannheimer P D Fein M E Casciani J R 1997b Physio-optical considerations in the design of fetal pulse oximetry sensors Eur. J. Obstet. Gynecol. Reprod. Biol. 72 9 19 9–19 10.1016/s0301-2115(97)02713-9 9134408 · DOI
  70. Mathes A M Kreuer S Schneider S O Ziegeler S Grundmann U 2008 The performance of six pulse oximeters in the environment of neuronavigation Anesth. Analg. 107 541 544 541–4 10.1213/ane.0b013e31817e6778 18633032 · DOI
  71. Michard F Shelley K L’Her E 2021 COVID-19: pulse oximeters in the spotlight J. Clin. Monit. Comput. 35 11 14 11–4 10.1007/s10877-020-00550-7 32578070 PMC7308445 · DOI · Paper page
  72. Millikan G A 1942 The oximeter, an instrument for measuring continuously the oxygen saturation of arterial blood in man Rev. Sci. Instrum. 13 434 444 434–44 10.1063/1.1769941 · DOI
  73. Morris R W Nairn M Torda T A 1989 A comparison of fifteen pulse oximeters: I. A clinical comparison; II. A test of performance under conditions of poor perfusion Anaesth. Intensive Care 17 62 73 62–73 10.1177/0310057X8901700113 2712278 · DOI
  74. Murphy S M Omar S 2018 The clinical utility of noninvasive pulse co-oximetry hemoglobin measurements in dark-skinned critically ill patients Anesth. Analg. 126 1519 1526 1519–26 10.1213/ANE.0000000000002721 29239951 · DOI
  75. Nafiu O O Mpody C Kim S S Uffman J C Tobias J D 2020 Race, postoperative complications, and death in apparently healthy children Pediatrics 146 e20194113 10.1542/peds.2019-4113 32690804 · DOI
  76. Naik P P Farrukh S N 2022 Influence of ethnicities and skin color variations in different populations: a review Skin Pharmacol Physiol. 35 65 76 65–76 10.1159/000518826 34515229 · DOI
  77. Ngiam J N Chew N W Sia C H Kong W K Poh K K 2021 Silent hypoxia: pulse oximetry and its relation to COVID-19 in Singapore Singap. Med. J. 64 222 225 222–5 10.11622/smedj.2021134 PMC10144454 34581542 · DOI
  78. Nicolai L 1932 Uber Sichtbarmachung, Verlauf und chemische Kinetik der Oxyhemoglobinreduktion im ebenden Gewebe, besonders in der menschlichen Arch. Ges. Physiol. 229 372 389 372–89 10.1007/BF01754476 · DOI
  79. Nitzan M Nitzan I Arieli Y 2020 The various oximetric techniques used for the evaluation of blood oxygenation Sensors (Basel). 20 4844 10.3390/s20174844 32867184 PMC7506757 · DOI
  80. Perkins G D McAuley D F Giles S Routledge H Gao F 2003 Do changes in pulse oximeter oxygen saturation predict equivalent changes in arterial oxygen saturation? Crit. Care 7 R67 10.1186/cc2339 12930558 PMC270702 · DOI · Paper page
  81. Philip K E J 2020 Working accuracy of pulse oximetry in COVID-19 patients stepping down from intensive care: a clinical evaluation BMJ Open Respir. Res. 7 e000778 10.1136/bmjresp-2020-000778 PMC7759755 33361436 · DOI
  82. Philip K E J Tidswell R McFadyen C 2021 Racial bias in pulse oximetry: more statistical detail may help tackle the problem Brit. Med. J. 372 n298 10.1136/bmjresp-2020-000778 33531354 · DOI
  83. Quaresima V Ferrari M 2020 COVID-19: efficacy of prehospital pulse oximetry for early detection of silent hypoxemia Crit. Care 24 501 10.1186/s13054-020-03185-x 32791979 PMC7424128 · DOI
  84. Ralston A C Webb R K Runciman W B 1991 Potential errors in pulse oximetry: III. Effects of interferences, dyes, dyshaemoglobins and other pigments Anaesthesia 46 291 295 291–5 10.1111/j.1365-2044.1991.tb11501.x 2024749 · DOI
  85. Rawlings A V 2006 Ethnic skin types: are there differences in skin structure and function? Int. J. Cosmet. Sci. 28 79 93 79–93 10.1111/j.1467-2494.2006.00302.x 18492142 · DOI · Paper page
  86. Reuss J L 2005 Multilayer modeling of reflectance pulse oximetry IEEE Trans. Biomed. Eng. 52 153 159 153–9 10.1109/TBME.2004.840188 15709652 · DOI
  87. Ries A L Prewitt L M Johnson J J 1989 Skin color and ear oximetry Chest 96 287 290 287–90 10.1378/chest.96.2.287 2752811 · DOI · Paper page
  88. Rodden A M Spicer L Diaz V A Steyer T E 2007 Does fingernail polish affect pulse oximeter readings? Intensive Crit. Care Nurs. 23 51 55 51–5 10.1016/j.iccn.2006.08.006 17064901 · DOI
  89. Rodriguez C 2020 Using pulse oximetry to monitor high-risk patients with COVID-19 at home Nursing 50 15 16 15–6 10.1097/01.NURSE.0000718376.94916.eb 33048856 · DOI
  90. Rosychuk R J Hudson-Mason A Eklund D Lacaze-Masmonteil T 2012 Discrepancies between arterial oxygen saturation and functional oxygen saturation measured with pulse oximetry in very preterm infants Neonatology. 101 14 19 14–9 10.1159/000326797 21791935 · DOI
  91. Saito J Kitayama M Kato R Hirota K 2017 Interference with pulse oximetry by the stealth station image guidance system J. Clin. Rep. 3 6 10.1186/s40981-017-0076-7 PMC5813676 29492445 · DOI
  92. Schulte T E Ohnoutka J R Agrawal A 2012 BrainLAB interference with pulse oximetry during stereotactic brain biopsy J. Clin. Anesth. 24 675 10.1016/j.jclinane.2012.03.010 23164645 · DOI
  93. Shah N Ragaswamy H B Govindugari K Estanol L 2012 Performance of three new-generation pulse oximeters during motion and low perfusion in volunteers J. Clin. Anesth. 24 385 391 385–91 10.1016/j.jclinane.2011.10.012 22626683 · DOI
  94. Shah S Majmudar K Stein A Gupta N Suppes S Karamanis M Capannari J Sethi S Patte C 2020 Novel use of home pulse oximetry monitoring in COVID-19 patients discharged from the emergency department identifies need for hospitalization Acad. Emerg. Med. 27 681 692 681–92 10.1111/acem.14053 32779828 PMC7323027 · DOI
  95. Sjoding M Iwashyna T J Valley T S 2021 More on racial bias in pulse oximetry measurement Reply. N. Engl. J. Med. 384 1278 10.1056/NEJMc2101321 33789021 · DOI
  96. Smith R N Hofmeyr R 2019 Perioperative comparison of the agreement between a portable fingertip pulse oximeter v. a conventional bedside pulse oximeter in adult patients (COMFORT trial) S. Afr. Med. J. 109 154 158 154–8 10.7196/SAMJ.2019.v109i3.13633 30834870 · DOI
  97. So-Ling C Li L 2001 A multi-layered reflection model of natural human skin Computer Graphics International 249 256 pp 249–56 10.1109/CGI.2001.934681 · DOI
  98. Sriganesh K Vinay B Bhadrinarayan V 2013 Indocyanine green dye administration can cause oxygen desaturation J. Clin. Monit. Comput. 27 371 10.1007/s10877-012-9423-4 23250655 · DOI
  99. Teo J 2020 Early detection of silent hypoxia in Covid-19 pneumonia using smartphone pulse oximetry J. Med. Syst. 44 134 32562006 PMC7305055
  100. Tobin M J Jubran A 2022 Pulse oximetry, racial bias and statistical bias Ann. Intensive Care. 12 2 10.1186/s13613-021-00974-7 34981267 PMC8723900 · DOI
  101. Tobin M J Laghi F Jubran A 2020 Why COVID-19 silent hypoxemia is baffling to physicians Am. J. Respir. Crit. Care. Med. 202 356 360 356–60 10.1164/rccm.202006-2157CP 32539537 PMC7397783 · DOI
  102. Toda K 1972 Alteration of racial differences in melanosome distribution in human epidermis after exposure to ultraviolet light Nat. New Biol. 236 143 145 143–5 10.1038/newbio236143a0 4502818 · DOI
  103. Todd B 2021 Pulse oximetry may be inaccurate in patients with darker skin AJN Am. J. Nurs. 121 16 33755610
  104. Tremper K K Barker S J 1989 Pulse oximetry Anesthesiology 70 98 108 98–108 10.1097/00000542-198901000-00019 2643368 · DOI
  105. Tytler J A Seeley H F 1986 The nellcor N-101 pulse oximeter Anaesthesia 41 302 305 302–5 10.1111/j.1365-2044.1986.tb12793.x 3963334 · DOI
  106. Valbuena V S M 2021 Racial bias in pulse oximetry measurement among patients about to undergo extracorporeal membrane oxygenation in 2019-2020: a retrospective cohort study Chest 161 971 978 971–8 10.1016/j.chest.2021.09.025 34592317 PMC9005857 · DOI
  107. Verdugo-Naranjo I 2022 Radiodermatitis grade estimation by RGB color imaging Artif. Life Robot. 27 58 63 58–63 10.1007/s10015-022-00739-w · DOI
  108. Vesoulis Z 2021 Racial discrepancy in pulse oximeter accuracy in preterm infants J. Perinatol. 42 79 85 79–85 10.1038/s41372-021-01230-3 34642469 PMC8508473 · DOI · Paper page
  109. Visscher M O 2017 Skin color and pigmentation in ethnic skin Facial Plast. Surg. Clin. North Am. 25 119 125 119–25 10.1016/j.fsc.2016.08.011 27888889 · DOI
  110. Volgyesi G A Spahr-Schopfer I 1991 Does skin pigmentation affect the accuracy of pulse oximetry? an in vitro study Anesthesiology 75 A406 10.1097/00000542-199109001-00406 · DOI
  111. Wahr J A Tremper K K Diab M 1995 Pulse oximetry Respir. Care Clin. N Am. 1 77 105 77–105 9390852
  112. Wang Y T Poh S C 1985 Noninvasive oximetry in pigmented patients Ann. Acad. Med. 14 427 429 427–9 4073807
  113. Watabe H 2004 Regulation of tyrosinase processing and trafficking by organellar pH and by proteasome activity J. Biol. Chem. 279 7971 7981 7971–81 10.1074/jbc.M309714200 14634018 · DOI
  114. 1997 Design of Pulse Oximeters. Webster J G Boca Raton, FL CRC Press 1st edn 10.1201/9780367802592 · DOI
  115. Westhorpe R N Ball C 2008 The pulse oximeter Anaesthesia Intensive Care 36 767 10.1177/0310057X0803600602 19115641 · DOI
  116. Whitehead C T 2021 More on racial bias in pulse oximetry measurement N. Engl. J. Med. 384 1278 10.1056/NEJMc2101321 33789020 · DOI
  117. Wiles M D 2022 Effect of patient ethnicity on the accuracy of peripheral pulse oximetry in patients with COVID-19 pneumonitis requiring mechanical ventilation Anaesthesia 77 489 491 489–91 10.1111/anae.15656 35001371 · DOI
  118. Wilkes M Wright C Y du Plessis J L Reeder A 2015 Fitzpatrick skin type, individual typology angle, and melanin index in an african population: steps toward universally applicable skin photosensitivity assessments JAMA Dermatol. 151 902 903 902–3 10.1001/jamadermatol.2015.0351 25923982 · DOI
  119. Wilkins C J Moores M Hanning C D 1989 Comparison of pulse oximeters: effects of vasoconstriction and venous engorgement Br. J. Anaesth. 62 439 444 439–44 2706182
  120. Wilson-Baig N McDonnell T Bentley A 2021 Discrepancy between SpO2 and SaO2 in patients with COVID-19 Anaesthesia 76 6 7 6–7 10.1111/anae.15228 PMC7436453 32738055 · DOI
  121. Witting M D Scharf S M 2008 Diagnostic room-air pulse oximetry: effects of smoking, race, and sex Am. J. Emerg. Med. 26 131 136 131–6 10.1016/j.ajem.2007.04.002 18272090 · DOI
  122. Wood E H 1950 A single scale absolute reading ear oximeter Proc Staff MeetMayo Clin. 25 384 391 384–91 15430456
  123. Yamamoto L G Yamamoto J A Yamamoto J B Yamamoto B E Yamamoto P P 2008 Nail polish does not significantly affect pulse oximetry measurements in mildly hypoxic subjects Respir. Care 53 1470 1474 1470–4 18957149
  124. Yönt G H Korhan E A Dizer B 2014 The effect of nail polish on pulse oximetry readings Intensive Crit. Care Nurs. 30 111 115 111–5 10.1016/j.iccn.2013.08.003 24054173 · DOI

Cited by

Sorted by citations

Authors
Al-Halawani, R., Charlton, P. H., Qassem, M., & Kyriacou, P. A.
Affiliations
University of London (GB)
Corresponding author
Al-Halawani, Raghda
Venue
Physiological Measurement
Published
2023-05-01
Volume
44
Issue
5
Pages
05TR01
ISSN
  • 0967-3334 (ISSN-L)
  • 1361-6579
OpenAlex
W4376272121
Cited by
119
References
18
Citations by year
2026: 34 · 2025: 51 · 2024: 29 · 2023: 5
FWCI
2.67
Type
journal article
Language
en
Open access
Locations
  • Physiological Measurement (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • City Research Online (City University London) (submittedVersion)
  • Apollo (University of Cambridge) (submittedVersion, cc-by)
Funders
British Heart Foundation (FS/20/20/34626)
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
  • Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
Keywords
pulse oximetry, skin pigmentation, arterial oxygen saturation, measurement accuracy, dark skin, skin colour, calibration algorithm, COVID-19 pandemic
MeSH terms
COVID-19, Humans, Oximetry, Oxygen, Skin Pigmentation, Pandemics

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review