Characterization of the oscillometric method for measuring indirect blood pressure
The DOI matches its Crossref record; 11 of the 11 references it cites were found in a registry or an index.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Among BPMDs using fixed-ratio coefficients, there is systematic bias towards greater underestimation of systolic BP as systolic BP increases, because the optimal fixed-ratio coefficient for accurate systolic BP estimation becomes progressively lower as systolic BP increases (e.g., the optimal fixed-ratio approximates 0.57 at 100 mmHg but this falls to 0.45 at 190 mmHg) [12].”
From Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure (2022) · cites this paper as [12]
In the citing paper
“However, the range of optimal coefficients for accurate systolic and diastolic BP measurements varies greatly between devices [12, 45, 50, 54].”
From Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure (2022) · cites this paper as [12]
In the citing paper
“specifically designed to provide BP values equivalent to the systolic and diastolic BP values measured with a mercury sphygmomanometer [12].”
From Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure (2022) · cites this paper as [12]
Citations
Cites
Noninvasive automatic determination of mean arterial pressure
Ramsey, M. (1979) · Medical and Biological Engineering and Computing · doi:10.1007/bf02440948 · Cited by 287
PHYSIOLOGIC RELATIONSHIPS BETWEEN INTRATHORACIC, INTRASPINAL AND ARTERIAL PRESSURES
HAMILTON, W. F. (1936) · Journal of the American Medical Association · doi:10.1001/jama.1936.02770370017005 · Cited by 277
Recommendations for Human Blood Pressure Determinations by Sphygmomanometers
BORDLEY, J., CONNOR, C. A. R., HAMILTON, W. F., KERR, W. J., & WIGGERS, C. J. (1951) · Circulation · doi:10.1161/01.cir.4.4.503 · Cited by 262
Mauck, G. W., Smith, C. R., Geddes, L. A., & Bourland, J. D. (1980) · Journal of Biomechanical Engineering · doi:10.1115/1.3138195 · Cited by 222
Self-citationIndirect Measurement of Mean Blood Pressure in the Anesthetized Patient
Yelderman, M., & Ream, A. K. (1979) · Anesthesiology · doi:10.1097/00000542-197903000-00022 · Cited by 142
The error in indirect blood pressure measurement with the incorrect size of cuff
Geddes, L. A., & Whistler, S. J. (1978) · American Heart Journal · doi:10.1016/0002-8703(78)90119-9 · Cited by 117
Self-citation(1969) · PubMed · Cited by 96
Self-citationIndirect Mean Blood Pressure in the Anesthetized Pony
Geddes, L. A., Chaffee, V., Whistler, S. J., BOURLAND, J. D., & Tacker, W. A. (1977) · American Journal of Veterinary Research · doi:10.2460/ajvr.1977.38.12.2055 · Cited by 45
Self-citationIndirect Measurement of Mean Blood Pressure in the Normotensive and Hypotensive Horse
Latshaw, H. S., FESSLER, J. F., Whistler, S. J., & Geddes, L. A. (1979) · Equine Veterinary Journal · doi:10.1111/j.2042-3306.1979.tb01340.x · Cited by 44
Self-citationGraphic recording of the korotkoff sounds
Geddes, L. A., Spencer, W. A., & HOFF, H. E. (1959) · American Heart Journal · doi:10.1016/0002-8703(59)90317-5 · Cited by 26
Self-citation
Research works printed but not linked: 2
- Geddes, L.A., W. Combs, W. Denton, S.J. Whistler, and J.D. Bourland. Indirect mean blood pressure in the dog.Am. J. Physiol. (Heart & Circulation) 7:H663-H666, 1980.
- Steele, J.M. Comparison of simultaneous indirect (auscultatory) and direct (intra-arterial) measurements of arterial pressure in man.J. Mt. Sinai Hosp. 8:1042–1050, 1941.
Cited by
Items 201–300 of 339
Dhamotharan, V., Cheng, H., Sung, S., Chen, C., Landry, C., Freithaler, M., Mahajan, A., Shroff, S. G., Hahn, J., & Mukkamala, R. (2025) · Frontiers in Physiology · doi:10.3389/fphys.2025.1611096 · Cited by 4
Non-Invasive Blood Pressure Measurement Using A Mobile Phone Camera
Zhang, Y., Zhang, X., & Du, J. (2022) · IEEE Conference on Computational Intelligence in Bioinformatics and Computational Biology (CIBCB) · doi:10.1109/cibcb55180.2022.9863015 · Cited by 5
Celler, B. G., Yong, A. S. C., Rubenis, I., Butlin, M., Argha, A., Rehan, R., & Avolio, A. P. (2024) · Journal of Hypertension · doi:10.1097/hjh.0000000000003651 · Cited by 5
Monitoring Personalized Trait Using Oscillometric Arterial Blood Pressure Measurements
Shin, Y.-S. (2012) · Journal of Applied Mathematics · doi:10.1155/2012/591252 · Cited by 4
Peng, Z., Zhang, M., Sun, Y., Bai, J., Gu, H., Liu, P., Li, M., & Cai, M. (2016) · Pediatric Anesthesia · doi:10.1111/pan.12989 · Cited by 4
Han, W.-R., Chuang, C.-Y., Li, P.-C., Kk, C., & Young, S.-T. (2009) · International Journal of Impotence Research · doi:10.1038/ijir.2009.30 · Cited by 4
Jilek, J., & Štork, M. (2016) · International Conference on Applied Electronics · doi:10.1109/ae.2016.7577254 · Cited by 4
Moderate morning rise in blood pressure has lowest risk of stroke but only in women
Head, G. A., Sata, Y., Imai, Y., Kikuya, M., Ohkubo, T., Reid, C. M., McGrath, B. M., & Lukoshkova, E. V. (2019) · Journal of Hypertension · doi:10.1097/hjh.0000000000002027 · Cited by 4
Cremer, J., da Cunha, A. F., Paul, L. J., Liu, C., & Acierno, M. J. (2019) · American Journal of Veterinary Research · doi:10.2460/ajvr.80.12.1067 · Cited by 4
Blood pressure measurement using finger cuff
Lee, J. Y., Choi, E.-O., Jeong, H. J., Kim, K. H., & Park, J. C. (2005) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2005.1617252 · Cited by 4
A New Design of Blood Pressure Measurement for Family Health Care
Mei, Z., Wang, Q., Zhou, P., Shao, L. N., & Liu, Z. W. (2009) · International Conference on Biomedical Engineering and Informatics (BMEI) · doi:10.1109/bmei.2009.5304905 · Cited by 4
Asymmetrical oscillometric pulse waveform envelopes in normotensive and hypertensive subjects
Zheng, D., Giovannini, R., & Murray, A. (2010) · Pure (Coventry University) · doi:10.25405/ncl.34171152 · Cited by 7
The Indirect Measurement of Blood Pressure
Geddes, L. A. (1991) · Humana Press eBooks · doi:10.1007/978-1-4684-7170-0_2 · Cited by 4
Self-citationMeasurements of arterial pressure and flow in vivo
Sharman, J. E., Segers, P., & Chirinos, J. A. (2022) · Textbook of Arterial Stiffness and Pulsatile Hemodynamics in Health and Disease · doi:10.1016/b978-0-323-91391-1.00002-9 · Cited by 3
Kario, K., Saito, K., Sato, K., Hamasaki, H., Suwa, H., Okura, A., & Hoshide, S. (2018) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000347 · Cited by 3
Performance Analysis of Oscillometric Blood Pressure Estimation Techniques in Cardiac Patients
Sidhu, E., Yoshida, M., Groza, V. Z., Dajani, H. R., & Bolić, M. (2019) · IEEE International Symposium on Medical Measurements and Applications (MeMeA) · doi:10.1109/memea.2019.8802200 · Cited by 5
Wrist cuff emulates auscultation in dual-cuff noninvasive blood pressure system
Jílek, J. O., & Štork, M. (2012) · International Conference on Applied Electronics · Cited by 3
Blood Pressure Measurement in Noise Intensive Environments Using Adaptive Interference Cancellation
Pinto, L., Dhanantwari, A. C., Wong, W., Stergiopoulos, S., & Maris, M. (2002) · Annals of Biomedical Engineering · doi:10.1114/1.1481051 · Cited by 3
Wrist cuff method determines mean arterial pressure in dual-cuff blood pressure system
Jilek, J., & Štork, M. (2014) · International Conference on Applied Electronics · doi:10.1109/ae.2014.7011687 · Cited by 3
Control System for DC Motor Based Micro Air Pump to Simulate Oscillograph of Blood Pressure
Santoso, N. A., & Ughi, F. (2017) · International Conference on Instrumentation, Communications, Information Technology, and Biomedical Engineering · doi:10.1109/icici-bme.2017.8537768 · Cited by 3
Jones, R., Sun, M., & Rithalia, S. (1999) · CRC Press eBooks · doi:10.1201/9780415876179.ch75 · Cited by 3
Nichtinvasives Monitoring von Druck und Strömung im Kreislauf
Pasch, T. (1985) · Springer eBooks · doi:10.1007/978-3-642-69893-4_22 · Cited by 3
Novelty detection with extreme value theory in vital-sign monitoring
Hugueny, S., & Hugueny, S. (2013) · Oxford University Research Archive (ORA) (University of Oxford) · doi:10.5287/ora-o5qeqd2d0 · Cited by 5
Blood Pressure Prediction Based on Single Photoplethysmography
Akuthota, U. C., Hassan, M. M., & Bhargava, L. (2022) · IEEE International Symposium on Smart Electronic Systems (iSES) · doi:10.1109/ises54909.2022.00075 · Cited by 2
Advances in principles and technologies of non-mechanical blood pressure monitoring
Zheng, Z., Hao, H., Huang, Y., Lyu, R., Ma, H., Zhang, Y., Liu, J., Shao, C., & Zhou, A. (2026) · npj Cardiovascular Health · doi:10.1038/s44325-025-00102-5 · Cited by 3
A new methodology of blood pressure measurement through the oscillometric method
Chabane, L., Omari, T., Benlatreche, M. S., & Bereksi-Reguig, F. (2021) · Measurement Science and Technology · doi:10.1088/1361-6501/ac0b14 · Cited by 2
Kim, Y., Park, S. H., Shin, J., Sung, K., Ihm, S., Kim, S. H., Kim, D., Park, S. M., Cho, I., Oh, K. W., et al. (2024) · Journal of Clinical Hypertension · doi:10.1111/jch.14805 · Cited by 2
Celler, B. G., Yong, A. S. C., Rubenis, I., Butlin, M., Argha, A., Rehan, R., & Avolio, A. P. (2024) · Journal of Hypertension · doi:10.1097/hjh.0000000000003719 · Cited by 3
First true blood pressure measurement with micropulse detection of arterial opening achieved
Murray, A., Zheng, D., Griffiths, C. J., Liu, C., Graham, D. J., Neasham, J. A., & Cossor, A. J. (2025) · Scientific Reports · doi:10.1038/s41598-025-87855-2 · Cited by 2
A popular validated home monitor uses the maximum oscillogram amplitude to compute blood pressure
Dhamotharan, V., Jazini, M., Kumar, R., Daher, H., Mahajan, A., Hahn, J., Shroff, S. G., & Mukkamala, R. (2025) · Scientific Reports · doi:10.1038/s41598-025-18850-w · Cited by 3
Evaluation of pulse features for oscillometry
Colak, S., & Isik, C. (2004) · IEEE Annual Northeast Bioengineering Conference (NEBEC) · doi:10.1109/nebc.2004.1300009 · Cited by 2
Jung, W.-M., Sim, M.-H., Jung, S.-O., Kim, M.-Y., Yoon, C.-S., Jung, I.-C., & Yoon, H.-R. (2011) · The Transactions of The Korean Institute of Electrical Engineers · doi:10.5370/kiee.2011.60.8.1606 · Cited by 2
A developed algorithm for oscillometric blood pressure measurement
Chen, M., Wang, Q., Mei, Z., & Liu, Z. (2013) · IEEE Region 10 Conference (TENCON) · doi:10.1109/tencon.2013.6718980 · Cited by 2
A Matlab demo software for NIBP measurement based on hybrid oscillometric method
Huang, P. C., Fu, J., & Zhao, Y. (2011) · International Congress on Image and Signal Processing (CISP) · doi:10.1109/cisp.2011.6100763 · Cited by 2
Naqvi, K. R., Jávorfí, T., & Vellani, C. W. (2008) · Journal of Biomedical Science and Engineering · doi:10.4236/jbise.2008.12012 · Cited by 2
Kusakabe, T. (2014) · Angiology Open Access · doi:10.4172/2329-9495.1000123 · Cited by 2
Lee, S. (2014) · Mathematical Problems in Engineering · doi:10.1155/2014/231925 · Cited by 2
Nguyen, T. H. M., Dinh, A., Bui, F. M., & Luu, L. (2015) · IEEE Annual Information Technology, Electronics and Mobile Communication Conference (IEMCON) · doi:10.1109/iemcon.2015.7344515 · Cited by 2
The Effect of Occlusion with the Cuff
Jobbágy, Á., & Nagy, P. (2017) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-981-10-5122-7_3 · Cited by 3
Inspection and Testing of Noninvasive Blood Pressure Measuring Devices
Lacković, I. (2017) · Series in biomedical engineering · doi:10.1007/978-981-10-6650-4_5 · Cited by 2
Development of a Public Database of Raw Data From a Blood Pressure Cuff
Jilek, J., & Štork, M. (2018) · International Conference on Applied Electronics · doi:10.23919/ae.2018.8501420 · Cited by 2
Designing a Biosignal Extraction Device for Measuring a Blood Pressure
Al-Masri, S., Safi, A., EL-Farra, A. A. H., & Shehada, M. (2018) · International Conference on Promising Electronic Technologies (ICPET) · doi:10.1109/icpet.2018.00026 · Cited by 2
Cuff Design for Home Blood Pressure Monitors
Palatini, P., Asmar, R., Bilo, G., & Parati, G. (2020) · Updates in Hypertension and Cardiovascular Protection · doi:10.1007/978-3-030-23065-4_2 · Cited by 2
Non-invasive methods for studying the dynamics of blood pressure
Ivanov, S. Y., & Бондаренко, Б. Б. (2019) · Arterial’naya Gipertenziya (Arterial Hypertension) · doi:10.18705/1607-419x-2018-24-6-637-645 · Cited by 3
Jilek, J., & Štork, M. (2020) · International Conference on Applied Electronics · doi:10.23919/ae49394.2020.9232699 · Cited by 2
Random Forest Based Blood Pressure Prediction Model from ECG And PPG Signal
Wang, H. (2022) · International Conference on Bioscience, Biochemistry and Bioinformatics (ICBBB) · doi:10.1145/3510427.3510428 · Cited by 2
Mieke, S., & Murray, A. (2023) · Biomedical Physics & Engineering Express · doi:10.1088/2057-1976/acf5f4 · Cited by 2
Vascular Mechanics in Artificial and Human Arteries
Whitt, M. D., & Drzewiecki, G. M. (2009) · Series on bioengineering and biomedical engineering · doi:10.1142/9789812779793_0007 · Cited by 2
Sidhu, E., & Groza, V. F. (2018) · International Conference on Intelligent Systems · Cited by 2
Improvements on cuff measurement of arterial pressure
Avolio, A. P., & Parati, G. (2013) · Journal of Hypertension · doi:10.1097/hjh.0b013e32835d2a22 · Cited by 2
Colak, S., & Isik, C. (2004) · Annual Meeting of the North American Fuzzy Information Processing Society (NAFIPS) · doi:10.1109/nafips.2004.1337413 · Cited by 2
COMPARISON OF BLOOD PRESSURE IN THE ARM AND ANKLE IN PATIENTS IN THE EMERGENCY DEPARTMENT
Goldstein, L. N. (2013) · University of the Witwatersrand, Johannesburg Institutional Repository on DSpace (University of the Witwatersrand, Johannesburg) · Cited by 2
Ratio-Independent Arterial Stiffness-Based Blood Pressure Estimation
Baktash, S. (2014) · uO Research (University of Ottawa) · doi:10.20381/ruor-5024 · Cited by 3
Methods for Non-invasive Trustworthy Estimation of Arterial Blood Pressure
Koohi, I. (2017) · uO Research (University of Ottawa) · doi:10.20381/ruor-20113 · Cited by 2
Saldin, T. K. (2019) · doi:10.15368/theses.2019.4 · Cited by 2
Simulations of Cuff-Tissue-Bone-Artery Systems for Blood Pressure Measurements
Xu, J. X. J., & Hu, D. H. D. (2022) · East Asian Journal on Applied Mathematics · doi:10.4208/eajam.020322.050422 · Cited by 1
Wearable Ring-Based System for Measuring Hemodynamic Parameters
Lindström, N., Sirkiä, J., Panula, T., & Kaisti, M. (2025) · IEEE Sensors Letters · doi:10.1109/lsens.2025.3592756 · Cited by 1
Oscillometric blood pressure estimation using machine learning-based mapping of waveform features
Ezeddin, M., Chowdhury, M. H., Khandakar, A. A., Faisal, M. A. A., Gonzales, A., Hossain, M. S. A., Murugappan, M., Naik, G. R., & Chowdhury, M. E. H. (2025) · Biomedical Engineering Letters · doi:10.1007/s13534-025-00496-2 · Cited by 2
A Center-of-Pressure Guided Finger-Press Sensor for Cuffless Blood Pressure Estimation
Gul, F. A. I., & Tudose, D. (2026) · Sensors · doi:10.3390/s26175563
Amplitude envelope slopes of oscillometric blood pressure waveforms as defined by amplitude ratios
Jílek, J. O., & Štork, M. (2009) · International Conference on Applied Electronics · Cited by 1
Metoda oscylometryczna pomiaru ciśnienia tętniczego krwi - modelowanie układu
Juroszek, B. (2008) · PRZEGLĄD ELEKTROTECHNICZNY · Cited by 1
Blood Pressure Simulation using an Arterial Pressure-volume Model
Yoon, S.-H., Kim, J.-H., Ye, S.-Y., Kim, C.-H., & Jeon, G.-R. (2008) · Transactions on Electrical and Electronic Materials · doi:10.4313/teem.2008.9.1.038 · Cited by 1
Lee, S., Jeon, G., & Kang, S. (2015) · BioMed Research International · doi:10.1155/2015/920206 · Cited by 1
Wrist cuff method determines diastolic pressure in dual cuff blood pressure system
Jílek, J. O., & Štork, M. (2015) · Digital Library (University of West Bohemia) · Cited by 1
Estimation of Blood Pressure Measurements for Hypertension Diagnosis Using Oscillometric Method
Shin, Y.-S. (2011) · IEICE Transactions on Fundamentals of Electronics Communications and Computer Sciences · doi:10.1587/transfun.e94.a.806 · Cited by 1
Ro, J. H., Bae, J.-W., Ye, S.-Y., Shin, B.-J., & Jeon, G.-R. (2009) · Journal of the Korea Academia-Industrial cooperation Society · doi:10.5762/kais.2009.10.8.1926 · Cited by 1
Validation of a blood pressure simulator that regenerates oscillometric cuff pressure waveforms
Zheng, D., Liu, C., Amoore, J., Mieke, S., & Murray, A. (2015) · Newcastle University ePrints (Newcastle Univesity) · doi:10.25405/ncl.34304589 · Cited by 1
Avolio, A. P., & Butlin, M. (2015) · Journal of Hypertension · doi:10.1097/hjh.0000000000000687 · Cited by 1
Mesure de pression non invasive, des valeurs à vérifier
Pernel, H. (2017) · Oxymag · doi:10.1016/j.oxy.2017.03.015 · Cited by 1
Development of equations for converting random-zero to automated oscillometric blood pressure values
Yan, L., Wen, X., Dyer, A. R., Chen, H., Zhou, L., Elliott, P., Wu, Y., Chan, Q. M., & Zhao, L. (2019) · Wellcome Open Research · doi:10.12688/wellcomeopenres.15407.1
Radhakrishnan, S., Lorenzutti, A. M., Araos, J. D., & Martin‐Flores, M. (2021) · PubMed Central · Cited by 1
Abdel-Galeel, A., Fawzy, N. N., Ali, W. A., & Fouad, D. A. (2023) · BMC Cardiovascular Disorders · doi:10.1186/s12872-023-03292-x · Cited by 2
Grove, M., Paliwal, M., Shearin, A., Kaiser, J., Koo, E. S., Howey, D., Galati, M., Czekalski, B., Dumawal, J., DeCarvalho, B., et al. (2023) · Stroke Vascular and Interventional Neurology · doi:10.1161/svin.122.000711 · Cited by 1
Development of an Innovative Algorithm for Accurate Measurement of Human Blood Pressure
Soni, A., Kumar, S., & Kumar, A. (2023) · MAPAN · doi:10.1007/s12647-023-00686-2 · Cited by 1
Physical Noninvasive Attacks on Photoplethysmogram by Computer Controlled Blood Pressure Cuff
Yoshida, K., Sawano, R., Okamoto, M., Murao, K., Tsuchida, S., & Terada, T. (2023) · Sensors · doi:10.3390/s23249764 · Cited by 1
Yu, J., Huang, Z., Shan, Y., Liu, Y., & Wang, X. (2023) · IEEE International Conference on Medical Artificial Intelligence (MedAI) · doi:10.1109/medai59581.2023.00018 · Cited by 1
a-STEP: Oscillometry-Augmented Auscultation Method for Improved Blood Pressure Measurement
Krishna, A. R., Kiran, V. R., Nabeel, P. M., & Joseph, J. (2024) · IEEE International Symposium on Medical Measurements and Applications (MeMeA) · doi:10.1109/memea60663.2024.10596731 · Cited by 1
Streamlining pediatric vital sign assessment: innovations and insights
Goo, S., Jang, W., Kim, Y. S., Ji, S., Park, T., Park, J. D., & Lee, B. (2024) · Scientific Reports · doi:10.1038/s41598-024-73148-7 · Cited by 1
Kim, G.-R., Ye, S.-Y., Kim, J., & Jeon, G.-R. (2008) · Journal of Sensor Science and Technology · doi:10.5369/jsst.2008.17.2.087 · Cited by 1
Raamat, R., Jagomägi, K., Talts, J., & Kivastik, J. (2010) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-642-13039-7_36 · Cited by 1
On the need to estimate the adequacy of measurement results
Leca, R., & Groza, V. F. (2009) · IEEE International Workshop on Advanced Methods for Uncertainty Estimation in Measurement (AMUEM) · doi:10.1109/amuem.2009.5207604 · Cited by 1
She, J., Guan, X., Liu, Y., & Xiang, H. (2016) · Journal of Clinical Hypertension · doi:10.1111/jch.12859 · Cited by 1
Hendrayana, Y. H., Riyadi, M. A., & Darjat, D. (2016) · Transmisi: Jurnal Ilmiah Teknik Elektro · doi:10.12777/transmisi.18.1.38-42 · Cited by 4
Hwang, Y. J., Kim, G. H., Yun, S. U., & Nam, K. W. (2019) · BioMedical Engineering OnLine · doi:10.1186/s12938-019-0719-1 · Cited by 1
Evaluation of Devices for Measurement of Blood Pressure
Meidert, A. S., & Saugel, B. (2021) · Springer eBooks · doi:10.1007/978-3-030-73387-2_19 · Cited by 1
The Effects and Consequences of Blood Pressure Cuff Pulse Wave Filtering
Jilek, J., & Štork, M. (2021) · International Conference on Applied Electronics · doi:10.23919/ae51540.2021.9542906 · Cited by 1
Home Blood Pressure Monitors For 21st Century
Jilek, J., & Štork, M. (2022) · International Conference on Applied Electronics · doi:10.1109/ae54730.2022.9919887 · Cited by 2
Monitoring of the blood pressure in the perioperative period
Pasch, T. (1989) · Annales Françaises d Anesthésie et de Réanimation · doi:10.1016/s0750-7658(89)80036-x · Cited by 1
Herbst, T., Anwar, Y. A., Giacco, S., & White, W. B. (2002) · Blood Pressure Monitoring · doi:10.1097/00126097-200210000-00007 · Cited by 1
Mesure de la pression artérielle en période opératoire
Pasch, T. (1989) · Annales Françaises d Anesthésie et de Réanimation · doi:10.1016/s0750-7658(89)80025-5 · Cited by 1
Interference resistance in automated oscillometric sphygmomanometers
Lazarashvili, L. T. (1993) · Biomedical Engineering · doi:10.1007/bf00556824 · Cited by 1
Tovey, F. I., Karanjia, N. D., Plant, G. R., & Rees, M. (1990) · Practical Diabetes · doi:10.1002/pdi.1960070505 · Cited by 1
New algorithm of cuff-tissue-artery system modeled as the space axisymmetric problemPreprint
Xu, J., & Hu, D. (2020) · arXiv (Cornell University) · doi:10.48550/arxiv.2007.06322 · Cited by 1
Mean Arterial Pressure in Acute Ischemic Stroke Study
Grove, M. (2020) · doi:10.21007/etd.cghs.2020.0511 · Cited by 1
NonInvasive Monitoring of Vital Signs and Traumatic Brain Injuries
Stergiopoulos, S., Freibert, A., Zhang, J., & Hatzinakos, D. (2017) · CRC Press eBooks · doi:10.4324/9781315219042-16
Método robusto de medición de la presión sanguínea arterial en niños
Arias, R., & de Jesus, S. (2009) · Cited by 1
Evaluation the Accuracy of Oscillometric Blood Pressure Measurement According to the AAMI SP10
Xiang, H. Y., Liu, Y. Y., Guo, T., Qin, Y., & Yu, M. (2015) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-319-19387-8_243
A new arterial blood pressure Holter based on the oscillometric method
Vallascas, R. (2015) · IEEE International Symposium on Medical Measurements and Applications (MeMeA) · doi:10.1109/memea.2015.7145203
Hemodynamic Measurements and Clinical Monitoring
Li, J. K. ‐. (2000) · Humana Press eBooks · doi:10.1007/978-1-59259-034-6_5
PERSONALIZED TRAITS MONITORING USING A NEURAL NETWORK BASED ON OSCILLOMETRIC MEASUREMENTS
Shin, Y.-S. (2013) · Biomedical Engineering Applications Basis and Communications · doi:10.4015/s101623721350021x
Metadata
- Authors
- Geddes, L. A., Voelz, M., Combs, C., Reiner, D., & Babbs, C. F.
- Affiliations
- Purdue University West Lafayette (US)
- Venue
- Annals of Biomedical Engineering
- Published
- 1982-11
- Volume
- 10
- Issue
- 6
- Pages
- 271-280
- ISSN
- 0090-6964 (ISSN-L)
- 1573-9686
- OpenAlex
- W2003899531
- Semantic Scholar
- 11130bb58509d11506466b456e8e31460a7b40c6
- Cited by
- 340
- References
- 11
- Citations by year
- 2026: 4 · 2025: 9 · 2024: 12 · 2023: 13 · 2022: 14 · 2021: 17 · 2020: 11 · 2019: 22 · 2018: 16 · 2017: 11 · 2016: 15 · 2015: 20 · 2014: 12 · 2013: 8 · 2012: 18
- FWCI
- 0.55
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.springer.com/tdm (tdm)
- Locations
- Annals of Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- Purdue e-Pubs (Purdue University System) (submittedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Keywords
- oscillometric blood pressure measurement, systolic blood pressure, diastolic blood pressure, cuff deflation, auscultatory method
- MeSH terms
- Adolescent, Adult, Animals, Biomedical Engineering, Blood Pressure Determination, Diastole, Dogs, Evaluation Studies as Topic, Humans, Middle Aged, Oscillometry, Systole
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.