Characterization of the oscillometric method for measuring indirect blood pressure
The DOI matches its Crossref record; 11 of the 11 references it cites were found in a registry or an index.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Among BPMDs using fixed-ratio coefficients, there is systematic bias towards greater underestimation of systolic BP as systolic BP increases, because the optimal fixed-ratio coefficient for accurate systolic BP estimation becomes progressively lower as systolic BP increases (e.g., the optimal fixed-ratio approximates 0.57 at 100 mmHg but this falls to 0.45 at 190 mmHg) [12].”
From Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure (2022) · cites this paper as [12]
In the citing paper
“However, the range of optimal coefficients for accurate systolic and diastolic BP measurements varies greatly between devices [12, 45, 50, 54].”
From Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure (2022) · cites this paper as [12]
In the citing paper
“specifically designed to provide BP values equivalent to the systolic and diastolic BP values measured with a mercury sphygmomanometer [12].”
From Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure (2022) · cites this paper as [12]
Citations
Cites
Noninvasive automatic determination of mean arterial pressure
Ramsey, M. (1979) · Medical and Biological Engineering and Computing · doi:10.1007/bf02440948 · Cited by 287
PHYSIOLOGIC RELATIONSHIPS BETWEEN INTRATHORACIC, INTRASPINAL AND ARTERIAL PRESSURES
HAMILTON, W. F. (1936) · Journal of the American Medical Association · doi:10.1001/jama.1936.02770370017005 · Cited by 277
Recommendations for Human Blood Pressure Determinations by Sphygmomanometers
BORDLEY, J., CONNOR, C. A. R., HAMILTON, W. F., KERR, W. J., & WIGGERS, C. J. (1951) · Circulation · doi:10.1161/01.cir.4.4.503 · Cited by 262
Mauck, G. W., Smith, C. R., Geddes, L. A., & Bourland, J. D. (1980) · Journal of Biomechanical Engineering · doi:10.1115/1.3138195 · Cited by 222
Self-citationIndirect Measurement of Mean Blood Pressure in the Anesthetized Patient
Yelderman, M., & Ream, A. K. (1979) · Anesthesiology · doi:10.1097/00000542-197903000-00022 · Cited by 142
The error in indirect blood pressure measurement with the incorrect size of cuff
Geddes, L. A., & Whistler, S. J. (1978) · American Heart Journal · doi:10.1016/0002-8703(78)90119-9 · Cited by 117
Self-citation(1969) · PubMed · Cited by 96
Self-citationIndirect Mean Blood Pressure in the Anesthetized Pony
Geddes, L. A., Chaffee, V., Whistler, S. J., BOURLAND, J. D., & Tacker, W. A. (1977) · American Journal of Veterinary Research · doi:10.2460/ajvr.1977.38.12.2055 · Cited by 45
Self-citationIndirect Measurement of Mean Blood Pressure in the Normotensive and Hypotensive Horse
Latshaw, H. S., FESSLER, J. F., Whistler, S. J., & Geddes, L. A. (1979) · Equine Veterinary Journal · doi:10.1111/j.2042-3306.1979.tb01340.x · Cited by 44
Self-citationGraphic recording of the korotkoff sounds
Geddes, L. A., Spencer, W. A., & HOFF, H. E. (1959) · American Heart Journal · doi:10.1016/0002-8703(59)90317-5 · Cited by 26
Self-citation
Research works printed but not linked: 2
- Geddes, L.A., W. Combs, W. Denton, S.J. Whistler, and J.D. Bourland. Indirect mean blood pressure in the dog.Am. J. Physiol. (Heart & Circulation) 7:H663-H666, 1980.
- Steele, J.M. Comparison of simultaneous indirect (auscultatory) and direct (intra-arterial) measurements of arterial pressure in man.J. Mt. Sinai Hosp. 8:1042–1050, 1941.
Cited by
Validation of the Omron HEM-907 device for blood pressure measurement
El Assaad, M. A., Topouchian, J. A., Darné, B. M., & Asmar, R. G. (2002) · Blood Pressure Monitoring · doi:10.1097/00126097-200208000-00006 · Cited by 279
Theory of the oscillometric maximum and the systolic and diastolic detection ratios
Drzewiecki, G., Hood, R., & Apple, H. (1994) · Annals of Biomedical Engineering · doi:10.1007/bf02368225 · Cited by 251
Human Pulse Diagnosis for Medical Assessments Using a Wearable Piezoelectret Sensing System
Chu, Y., Zhong, J., Liu, H., Ma, Y., Liu, N., Song, Y., Liang, J., Shao, Z., Sun, Y., Dong, Y., et al. (2018) · Advanced Functional Materials · doi:10.1002/adfm.201803413 · Cited by 249
Methods of Blood Pressure Measurement in the ICU*
Lehman, L.-W. H., Saeed, M., Talmor, D., Mark, R., & Malhotra, A. (2013) · Critical Care Medicine · doi:10.1097/ccm.0b013e318265ea46 · Cited by 224
El Assaad, M. A., Topouchian, J. A., & Asmar, R. G. (2003) · Blood Pressure Monitoring · doi:10.1097/00126097-200306000-00006 · Cited by 224
Metadata
- Authors
- Geddes, L. A., Voelz, M., Combs, C., Reiner, D., & Babbs, C. F.
- Affiliations
- Purdue University West Lafayette (US)
- Venue
- Annals of Biomedical Engineering
- Published
- 1982-11
- Volume
- 10
- Issue
- 6
- Pages
- 271-280
- ISSN
- 0090-6964 (ISSN-L)
- 1573-9686
- OpenAlex
- W2003899531
- Semantic Scholar
- 11130bb58509d11506466b456e8e31460a7b40c6
- Cited by
- 340
- References
- 11
- Citations by year
- 2026: 4 · 2025: 9 · 2024: 12 · 2023: 13 · 2022: 14 · 2021: 17 · 2020: 11 · 2019: 22 · 2018: 16 · 2017: 11 · 2016: 15 · 2015: 20 · 2014: 12 · 2013: 8 · 2012: 18
- FWCI
- 0.55
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.springer.com/tdm (tdm)
- Locations
- Annals of Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- Purdue e-Pubs (Purdue University System) (submittedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Keywords
- oscillometric blood pressure measurement, systolic blood pressure, diastolic blood pressure, cuff deflation, auscultatory method
- MeSH terms
- Adolescent, Adult, Animals, Biomedical Engineering, Blood Pressure Determination, Diastole, Dogs, Evaluation Studies as Topic, Humans, Middle Aged, Oscillometry, Systole
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.