A sub-millimetric, 0.25 mN resolution fully integrated fiber-optic force-sensing tool for retinal microsurgery
Open accessAbstract
PURPOSE: Retinal microsurgery requires extremely delicate manipulation of retinal tissue where tool-to-tissue interaction forces are usually below the threshold of human perception. Creating a force-sensing surgical instrument that measures the forces directly at the tool tip poses great challenges due to the interactions between the tool shaft and the sclerotomy opening. METHODS: We present the design and analysis of a force measurement device that senses distal forces interior to the sclera using 1-cm long, 160 microm diameter Fiber Bragg Grating (FBG) strain sensors embedded in a 0.5 mm diameter tool shaft. Additionally, we provide an algorithm developed to cancel the influence of environmental temperature fluctuations. RESULTS: The force-sensing prototype measures forces with a resolution of 0.25 mN in 2 DOF while being insensitive to temperature. CONCLUSION: Sub-millinewton resolution force sensors integrated into microsurgical instruments are feasible and have potential applications in both robotic and freehand microsurgery.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“Micron is a handheld actively stabilized 3-DOF micromanipulator that is mainly designed to remove unintentional motion and enhance positional accuracy [16].”
From Preliminary evaluation of a micro-force sensing handheld robot for vitreoretinal surgery (2012) · cites this paper as [16]
Experimental — accuracy not yet measuredThe check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“This micromanipulator robot can sense its own motion, estimate the undesired hand tremor component in real time, and moves the tool tip relative to the handle to counteract this component.”
From Preliminary evaluation of a micro-force sensing handheld robot for vitreoretinal surgery (2012) · cites this paper as [16]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Fiber Bragg grating technology fundamentals and overview
Hill, K. O., & Meltz, G. (1997) · Journal of Lightwave Technology · doi:10.1109/50.618320 · Cited by 3,071
Methods for haptic feedback in teleoperated robot‐assisted surgery
Okamura, A. M. (2004) · Industrial Robot the international journal of robotics research and application · doi:10.1108/01439910410566362 · Cited by 494
Development and application of a new steady-hand manipulator for retinal surgery
Mitchell, B., Koo, J., Iordachita, I., Kazanzides, P., Kapoor, A., Handa, J., Hager, G., & Taylor, R. (2007) · Proceedings 2007 IEEE International Conference on Robotics and Automation · doi:10.1109/robot.2007.363056 · Cited by 174
Self-citationBerkelman, P. J., Whitcomb, L. L., Taylor, R. H., & Jensen, P. (2003) · IEEE Transactions on Robotics and Automation · doi:10.1109/tra.2003.817526 · Cited by 172
Self-citationSurgical Forces and Tactile Perception During Retinal Microsurgery
Gupta, P. K., Jensen, P. S., & de Juan, E. (1999) · Lecture notes in computer science · doi:10.1007/10704282_132 · Cited by 163
Fusing force and vision feedback for micromanipulation
Zhou, Y., Nelson, B. J., & Vikramaditya, B. (2002) · Proceedings - IEEE International Conference on Robotics and Automation/Proceedings · doi:10.1109/robot.1998.677265 · Cited by 90
ERGONOMICS APPLIED TO THE PRACTICE OF MICROSURGERY1
Patkin, M. (1977) · Australian and New Zealand Journal of Surgery · doi:10.1111/j.1445-2197.1977.tb04297.x · Cited by 80
Applied force during vitreoretinal microsurgery with handheld instruments
Jagtap, A. D., & Riviere, C. N. (2005) · The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2004.1403792 · Cited by 79
Force feedback-based microinstrument for measuring tissue properties and pulse in microsurgery
Menciassi, A., Eisinberg, A., Scalari, G., Anticoli, C., Carrozza, M. C., & Dario, P. (2002) · Proceedings - IEEE International Conference on Robotics and Automation/Proceedings · doi:10.1109/robot.2001.932620 · Cited by 72
Calibration of Multi-Axis MEMS Force Sensors Using the Shape-From-Motion Method
Kim, K., Sun, Y., Voyles, R. M., & Nelson, B. J. (2007) · IEEE Sensors Journal · doi:10.1109/jsen.2006.890141 · Cited by 56
Flagged first, then by citations
Research works printed but not linked: 1
- Iordachita I et al (2006) Steady-hand manipulator for retinal surgery. In: Proceedings of the MICCAI medical robotics workshop
Cited by
Sorted by citations
Items 101–181 of 181
Optical Millinewton Force Sensors Based on GaN Devices Integrated With Bionic-Structured PMDS Films
Chen, T., Zhu, L., Chen, J., Tang, X., & Li, K. H. (2023) · IEEE Transactions on Electron Devices · doi:10.1109/ted.2023.3279301 · Cited by 16
Abushagur, A. A. G., Bakar, A. A. A., Zan, M. S. D., & Arsad, N. (2016) · IEEE Sensors Journal · doi:10.1109/jsen.2016.2602239 · Cited by 16
Toward Medical Devices With Integrated Mechanisms, Sensors, and Actuators Via Printed-Circuit MEMS
Gafford, J. B., Ranzani, T., Russo, S., Değirmenci, A., Kesner, S. B., Howe, R. D., Wood, R. J., & Walsh, C. J. (2016) · Journal of Medical Devices · doi:10.1115/1.4035375 · Cited by 16
An Optimized Tilt Mechanism for a New Steady-Hand Eye Robot
Wu, J., Li, G., Urias, M. G., Patel, N. A., Liu, Y., Gehlbach, P. L., Taylor, R. H., & Iordachita, I. I. (2020) · IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) · doi:10.1109/iros45743.2020.9340741 · Cited by 15
Self-citationGao, Q., Ji, G., Sun, M., Xiao, Y., Rao, H., & Sun, Z. (2025) · IEEE Transactions on Robotics · doi:10.1109/tro.2025.3577011 · Cited by 14
Self-citationMeasuring the Forces of Middle Ear Surgery; Evaluating a Novel Force-Detection Instrument
Bergin, M. J., Sheedy, M., Ross, P. D., Wylie, G., & Bird, P. (2014) · Otology & Neurotology · doi:10.1097/mao.0000000000000173 · Cited by 14
Enabling Technology for Safe Robot-Assisted Retinal Surgery: Early Warning for Unsafe Scleral Force
He, C., Patel, N. A., Iordachita, I. I., & Kobilarov, M. (2019) · Proceedings - IEEE International Conference on Robotics and Automation/Proceedings · doi:10.1109/icra.2019.8794427 · Cited by 14
Self-citationLeslie, O., Bulens, D. C., Ulloa, P. M., & Redmond, S. J. (2023) · IEEE Sensors Journal · doi:10.1109/jsen.2023.3293967 · Cited by 13
Vision-based retinal membrane peeling with a handheld robot
Becker, B. C., MacLachlan, R. A., Lobes, L. A., & Riviere, C. N. (2012) · Proceedings - IEEE International Conference on Robotics and Automation/Proceedings · doi:10.1109/icra.2012.6224844 · Cited by 13
Robotic Retinal Surgery Impacts on Scleral Forces: In Vivo Study
Urias, M. G., Patel, N. A., Ebrahimi, A., Iordachita, I. I., & Gehlbach, P. L. (2020) · Translational Vision Science & Technology · doi:10.1167/tvst.9.10.2 · Cited by 13
Self-citationNovel force sensing module for a concentric tube-based vitreoretinal surgical robot
Xu, B., & Ko, S. Y. (2020) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2020.112395 · Cited by 13
Evaluation of the Hand Motion and Peeling Force in Inner Limiting Membrane Peeling
Zheng, Y. H., Yang, X., Mo, B., Qi, Y. A., Yang, Y., Lin, C., Han, S., Wang, N., Guang, C., & Liu, W. (2023) · Translational Vision Science & Technology · doi:10.1167/tvst.12.3.32 · Cited by 12
Evaluation of a Force-Sensing Handheld Robot for Assisted Retinal Vein Cannulation
Gonenc, B., Patel, N. A., & Iordachita, I. I. (2018) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2018.8513304 · Cited by 12
Self-citationHand-Held Force Magnifier for Surgical Instruments: Evolution toward a Clinical Device
Lee, R., Wu, B., Klatzky, R. L., Shivaprabhu, V., Galeotti, J., Horvath, S., Siegel, M., Schuman, J. S., Hollis, R. L., & Stetten, G. (2013) · Lecture notes in computer science · doi:10.1007/978-3-642-38085-3_9 · Cited by 12
Ranjbar-Naeini, O. R., Latifi, H., Zibaii, M. I., & Mousavian, A. (2018) · Optics Letters · doi:10.1364/ol.43.005210 · Cited by 11
Sabieleish, M., Heryan, K., Boese, A., Hansen, C., Friebe, M., & Manríquez, A. I. (2021) · International Journal of Computer Assisted Radiology and Surgery · doi:10.1007/s11548-021-02479-x · Cited by 11
Fiber Optical Sensors in Biomechanics
Roriz, P., & Ribeiro, A. B. L. (2018) · Elsevier eBooks · doi:10.1016/b978-0-12-803131-5.00010-6 · Cited by 11
Zaltieri, M., Presti, D. L., Massaroni, C., Schena, E., D’Abbraccio, J., Massari, L., Oddo, C. M., Formica, D., Caponero, M., Bravi, M., et al. (2020) · IEEE International Instrumentation and Measurement Technology Conference · doi:10.1109/i2mtc43012.2020.9128974 · Cited by 12
Effects of micro-vibratory modulation during robot-assisted membrane peeling
Gonenc, B., Gehlbach, P. L., Taylor, R. H., & Iordachita, I. I. (2015) · IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) · doi:10.1109/iros.2015.7353912 · Cited by 11
Self-citationHangai, S., Nozaki, T., Soma, T., Miyashita, H., Asoda, S., Yazawa, M., Sato, K., Kawana, H., Ohnishi, K., & Kobayashi, E. (2021) · The International Journal of Medical Robotics and Computer Assisted Surgery · doi:10.1002/rcs.2205 · Cited by 10
Real-time measurement of tool-tissue interaction forces in neurosurgery: Quantification and analysis
Maddahi, Y., Huang, J. J., Huang, J., Gan, L. S., Hoshyarmanesh, H., Zareinia, K., & Sutherland, G. R. (2016) · IEEE International Conference on Advanced Intelligent Mechatronics (AIM) · doi:10.1109/aim.2016.7576967 · Cited by 10
Xiao, J., Wu, Q., Sun, D., He, C., & Chen, Y. (2019) · International Conference on Intelligent Transportation, Big Data & Smart City (ICITBS) · doi:10.1109/icitbs.2019.00122 · Cited by 10
Development of a smart handheld surgical tool with tactile feedback
Lee, C., Kim, U., Lee, D., Nguyen, C. T., Nguyen, D. T., Phung, H., Park, J., Jung, H., & Choi, H. R. (2017) · Intelligent Service Robotics · doi:10.1007/s11370-016-0214-5 · Cited by 10
Wang, H., Du, C., Wang, G., Xia, Z., Men, Y., Du, Z., & Wang, W. (2023) · IEEE Sensors Journal · doi:10.1109/jsen.2023.3283559 · Cited by 9
A Robotic Microforceps for Retinal Microsurgery With Adaptive Clamping Method
Zhang, H., Yi, H., Wang, C., Yang, J., Jin, T., & Zhao, J. (2024) · IEEE/ASME Transactions on Mechatronics · doi:10.1109/tmech.2024.3378275 · Cited by 9
Toward hybrid position/force control for an active handheld micromanipulator
Wells, T. S., MacLachlan, R. A., & Riviere, C. N. (2014) · Proceedings - IEEE International Conference on Robotics and Automation/Proceedings · doi:10.1109/icra.2014.6906942 · Cited by 9
Sub-millinewton force sensor for Vitreoretinal microsurgery using linear chirp fiber Bragg grating
Abushagur, A. A. G., Bakar, A. A. A., Arsad, N., & Shaari, S. B. (2014) · IEEE International Conference on Photonics · doi:10.1109/icp.2014.7002305 · Cited by 9
A Multifunction Sensing Forceps Based on Fiber Bragg Gratings for Membrane Peeling
Liu, J., Yang, Y., Yu, J., Zheng, Y., Lin, C., Wang, Z., & Guang, C. (2024) · IEEE Sensors Journal · doi:10.1109/jsen.2024.3350068 · Cited by 8
Patel, N. A., Urias, M. G., Ebrahimi, A., Taylor, R. H., Gehlbach, P. L., & Iordachita, I. I. (2022) · IEEE Transactions on Medical Robotics and Bionics · doi:10.1109/tmrb.2022.3191441 · Cited by 8
Self-citationQuantifying efficacy of the fiber bragg grating sensors in medical applications: a survey
Prathap, P. B., & Saara, K. (2023) · Journal of Optics · doi:10.1007/s12596-023-01480-8 · Cited by 9
Qin, W., Yi, H., Fan, Z., & Zhao, J. (2025) · Sensors · doi:10.3390/s25020405 · Cited by 7
Callaghan, D., Rajan, G., McGrath, M. M., Coyle, E., Semenova, Y., & Farrell, G. (2011) · Smart Materials and Structures · doi:10.1088/0964-1726/20/10/105004 · Cited by 7
Abushagur, A. A. G., Arsad, N., Elgaud, M. M., & Bakar, A. A. A. (2019) · IEEE Access · doi:10.1109/access.2019.2954914 · Cited by 7
Zheng, Y. H., Yang, Y., Han, S., & Ma, K. (2021) · International Journal of Mechanical Sciences · doi:10.1016/j.ijmecsci.2021.107025 · Cited by 7
A hand‐held device with 3‐DOF haptic feedback mechanism for microsurgery
Wang, Z., Wang, S., & Zuo, S. (2019) · International Journal of Medical Robotics and Computer Assisted Surgery · doi:10.1002/rcs.2025 · Cited by 7
Taniguchi, Y., Sugiura, H., Yamanaka, T., Watanabe, S., Omata, S., Harada, K., Mitsuishi, M., Shiraya, T., Sugimoto, K., Ueta, T., et al. (2022) · Microsystems & Nanoengineering · doi:10.1038/s41378-022-00417-8 · Cited by 6
Force feedback without sensor: A preliminary study on haptic modeling
Sun, Z., Wang, Z., & Phee, S. J. (2012) · IEEE/RAS-EMBS International Conference on Biomedical Robotics and Biomechatronics (BioRob) · doi:10.1109/biorob.2012.6290279 · Cited by 6
Self-citationFeasibility Check: Can Audio Be a Simple Alternative to Force-Based Feedback for Needle Guidance?
Manríquez, A. I., Boese, A., Friebe, M., & Hansen, C. (2020) · Lecture notes in computer science · doi:10.1007/978-3-030-59716-0_3 · Cited by 6
Schwarz, A., Mehrfard, A., Amirkhani, G., Phalen, H., Ma, J. H., Grupp, R. B., Gomez, A. M., & Armand, M. (2024) · Proceedings - IEEE International Conference on Robotics and Automation/Proceedings · doi:10.1109/icra57147.2024.10610024 · Cited by 5
Ren, H., & Meng, M. Q. ‐. (2011) · Journal of Mechanics in Medicine and Biology · doi:10.1142/s0219519411004289 · Cited by 5
Real Time Prediction of Sclera Force with LSTM Neural Networks in Robot-Assisted Retinal Surgery
He, C. Y., Patel, N. A., Kobilarov, M., & Iordachita, I. I. (2020) · Applied Mechanics and Materials · doi:10.4028/www.scientific.net/amm.896.183 · Cited by 5
Self-citationOperation behaviours of surgical forceps in continuous curvilinear capsulorhexis
Zheng, Y. H., Lin, C., Guang, C., Han, S., Ma, K., & Yang, Y. (2022) · International Journal of Medical Robotics and Computer Assisted Surgery · doi:10.1002/rcs.2424 · Cited by 5
Calibration of a needle tracking device with fiber Bragg grating sensors
Mandal, K., Parent, F., Martel, S. M., Kashyap, R., & Kadoury, S. (2015) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.2081198 · Cited by 5
Xu, B., & Ko, S. Y. (2023) · Mechatronics · doi:10.1016/j.mechatronics.2023.103106 · Cited by 4
Joža, A. V., Bajić, J. S., Manojlović, L. M., Milosavljević, V. A., Batinić, B., Lakovic, N. M., & Živanov, M. B. (2017) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2017.04.042 · Cited by 4
Chen, Y., Zhu, L., Gao, W.-Q., Yu, J., Wang, J., Deng, X., Tao, J., Qu, J., & Shen, L.-J. (2022) · International Journal of Ophthalmology · doi:10.18240/ijo.2022.02.10 · Cited by 4
Shape detection of thin diameter flexible sensors with non-uniform gratings
Chen, X., Qian, J., Yu, Y., Zhang, Y., & Shen, T. (2024) · Measurement Science and Technology · doi:10.1088/1361-6501/ad1e4e · Cited by 4
Comparing FBG and PCF Force Sensors in a Laparoscopic Smart Surgical Scissor Instrument
Callaghan, D., Rajan, G., McGrath, M. M., Coyle, E. D., Semenova, Y., & Farrell, G. (2011) · Arrow@dit (Dublin Institute of Technology) · Cited by 4
Development of a surgical instrument for measuring forces applied to the ossicles of the middle ear
Sheedy, M., Bergin, M. H., Wylie, G., Ross, P. D., Dove, R., & Bird, P. (2012) · Australasian Physical & Engineering Sciences in Medicine · doi:10.1007/s13246-012-0172-y · Cited by 3
High Sensitive Microsurgical Force Sensor Using Spectral-Width of Tapered Fiber Bragg Gratings
Abushagur, A. A. G., Arsad, N., Mokhtar, M. H. H., & Bakar, A. A. A. (2019) · doi:10.1109/biophotonics.2019.8896754 · Cited by 3
Ranjbar-Naeini, O. R., جعفری, ف., Zarafshani, P., Zibaii, M. I., & Latifi, H. (2017) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.2272013 · Cited by 3
Modelling and Characterization of FBG Sensors
Joža, A. V., Lakovic, N. M., Korcok, N., Kokotovic, V., Batinić, B., & Bajić, J. S. (2019) · Zooming Innovation in Consumer Technologies Conference (ZINC) · doi:10.1109/zinc.2019.8769416 · Cited by 3
Luo, B., Deng, Y., Yang, T., Dai, S., & Song, G. (2019) · Chinese Automation Congress · doi:10.1109/cac48633.2019.8996395 · Cited by 3
Karaçalı, Ö., & Çetiner, D. (2014) · Acta Physica Polonica A · doi:10.12693/aphyspola.125.180 · Cited by 2
Preliminary Estimation of the Friction between Force-sensing Forceps and Cornea
Zheng, Y. H., Yang, Y., Lin, C., Guang, C., Zong, J.-J., & Ma, K. (2022) · IEEE/RAS-EMBS International Conference on Biomedical Robotics and Biomechatronics (BioRob) · doi:10.1109/biorob52689.2022.9925455 · Cited by 2
Teleoperated MRI-guided prostate needle placement
Seifabadi, R. (2013) · Cited by 2
Force Sensing Surgical Scissor Blades using Fibre Bragg Grating Sensors
Callaghan, D. (2013) · Arrow - TU Dublin (Technological University Dublin) · doi:10.21427/d77g87 · Cited by 2
Artificial Intelligence in Intraocular Robotic Microsurgery
Iordachita, I. I., Navab, N., Gehlbach, P. L., Kobilarov, M., & Nasseri, M. A. (2025) · Springer eBooks · doi:10.1007/978-981-96-6635-5_11 · Cited by 1
Self-citationFBG-based Actuation and Data Driven Contact Detection for Smart Steerable Instruments
Mehdi, S. Z., Janssens, W., Gielen, M., Vanderschueren, E., Ourak, M., Verslype, C., Laleman, W., & Poorten, E. V. (2025) · IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) · doi:10.1109/iros60139.2025.11247263 · Cited by 1
Esfandiari, M., Kim, J. W., Zhang, P., Heng, J. S., Gehlbach, P., Taylor, R. H., & Iordachita, I. I. (2025) · IEEE Transactions on Medical Robotics and Bionics · doi:10.1109/tmrb.2025.3617962 · Cited by 1
Self-citationFast fiber shape sensing with high accuracy by a two-layer LSTM network
Zhou, T., Zheng, X., Yunlu, F. A. N., Wang, X., Wu, J., Zhu, R., Yang, T., Zhou, R., yuming, D., & Liu, H. (2026) · Optics Letters · doi:10.1364/ol.586239 · Cited by 1
Abushagur, A. A. G., Bakar, A. A. A., Zan, M. S. D., Arsad, N., & Shaari, S. B. (2015) · International Conference on Sensing Technology · doi:10.1109/icsenst.2015.7438410 · Cited by 1
Djinović, Z., Tomić, M., & Plank, N. (2021) · International Convention on Information and Communication Technology, Electronics and Microelectronics · doi:10.23919/mipro52101.2021.9596875 · Cited by 1
Batch fabrication of force sensors for robotic surgery using optical interference
Kashimura, Y., Tsuroka, N., Arata, J., & Haga, Y. (2023) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2023.114672 · Cited by 1
Development of a Back Bone Shape Allay Force Sensor using Optical Fiber
Arata, J., Terakawa, S., & FUJIMOTO, H. (2012) · Journal of Japan Society of Computer Aided Surgery · doi:10.5759/jscas.14.455 · Cited by 1
Innovative Vitreoretinal Surgery Technologies
Chen, A., & Weng, C. Y. (2018) · International Ophthalmology Clinics · doi:10.1097/iio.0000000000000247 · Cited by 1
Optical fiber sensors applications for human health
Junior, A. G. L., & Frizera, A. (2022) · Elsevier eBooks · doi:10.1016/b978-0-32-385952-3.00023-8 · Cited by 1
Inner Ear Effects of Middle Ear Surgery
Bergin, M. J. (2012) · Otago University Research Archive (University of Otago) · Cited by 1
ASTRO: A Novel Robotic Tool for Laser Surgery of the Prostate
Russo, S., Dario, P., & Menciassi, A. (2013) · CINECA IRIS Institutional Research Information System (Sant'Anna School of Advanced Studies)
Robotics in ophthalmic surgery
Clarkson, D. (2017) · Optician · doi:10.12968/opti.2017.10.6779
Force Sensing in Arthroscopic Instruments using Fiber Bragg Gratings
Yurkewich, D. S. (2015) · Scholarship@Western (Western University)
Collaborative Handheld Robots Near the Limits of Human Dexterity—A Pilot Study
Ullah, Y. F., Keefe, D. F., & Kowalewski, T. M. (2025) · International Symposium on Medical Robotics · doi:10.1109/ismr67322.2025.11025975
Li, H., Men, Y., Li, X., Fang, H., Wang, W., & Du, Z. (2025) · IEEE International Conference on Manipulation, Manufacturing and Measurement on the Nanoscale (3M-NANO) · doi:10.1109/3m-nano65639.2025.11260986
Simultaneous ranging and force sensing on a single-mode fiber for minimally invasive interventions
Liu, C., Zhang, W., Zhang, H., Lan, Q., & Yang, J. (2025) · Optics Letters · doi:10.1364/ol.583811
FBG-based 3D micro-force sensor integrated into ophthalmic microsurgical robotic forceps
Wang, N., Zhang, X., Li, M., Stilli, A., Bano, S., & Stoyanov, D. V. (2026) · Measurement · doi:10.1016/j.measurement.2026.120909
Mühlenbeck, B., Opie, J., Fernandez, C. S., Jaufuraully, S., Parris, D., Desjardins, A. E., David, A. L., Siassakos, D. M., Tiwari, M. K., & BLANDFORD, A. E. (2026) · JMIR Formative Research · doi:10.2196/78845
Compliant Glass Mechanism Instrumented with a Bragg Grating to Measure Indentation Force
Marchandise, M., Chafaï, A., Caucheteur, C., & Lambert, P. (2026) · Micromachines · doi:10.3390/mi17050572
Manan, A., Azam, M., Ahmad, N., Ali, W., Sabah, F., & Sarwar, R. (2026) · Research Square · doi:10.21203/rs.3.rs-9731703/v1
Metadata
- Authors
- Iordachita, I., Sun, Z., Balicki, M., Kang, J. U., Phee, S. J., Handa, J., Gehlbach, P., & Taylor, R.
- Affiliations
- Johns Hopkins University (US)
- Nanyang Technological University (SG)
- Corresponding author
- Iordachita, Iulian
- ORCID
- Venue
- International Journal of Computer Assisted Radiology and Surgery
- Published
- 2009-04-15
- Volume
- 4
- Issue
- 4
- Pages
- 383-390
- ISSN
- 1861-6410 (ISSN-L)
- 1861-6429
- PMC
- PMC2801926
- OpenAlex
- W1974070762
- Cited by
- 181
- References
- 11
- Citations by year
- 2026: 5 · 2025: 9 · 2024: 5 · 2023: 13 · 2022: 11 · 2021: 10 · 2020: 10 · 2019: 14 · 2018: 12 · 2017: 10 · 2016: 13 · 2015: 8 · 2014: 15 · 2013: 18 · 2012: 13
- FWCI
- 7.14
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- Locations
- International Journal of Computer Assisted Radiology and Surgery (publishedVersion, cc-by-nc)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- National Science Foundation
- Johns Hopkins University
- National Institutes of Health
- National Institute of Biomedical Imaging and Bioengineering (R01EB007969)
- Topics
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Advanced Fiber Optic Sensors
- Health Sciences › Medicine › Ophthalmology › Retinal and Macular Surgery
- Health Sciences › Medicine › Ophthalmology › Intraocular Surgery and Lenses
- Keywords
- retinal microsurgery, fiber Bragg grating strain sensors, sclerotomy, temperature compensation, microsurgical instruments, robotic microsurgery
- MeSH terms
- Calibration, Equipment Design, Fiber Optic Technology, Humans, Microsurgery, Retinal Diseases, Sclera, Surgery, Computer-Assisted
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review